Remove Power Module - Emerson Rosemount 648 Reference Manual

Wireless temperature transmitter with rosemount x-well technology
Hide thumbs Also See for Rosemount 648:
Table of Contents
Reference Manual
00809-0200-4648
Figure 2-18: Example 2: Falling Alert
Field Communicator
To configure the process alerts with a Field Communicator, perform the following
procedure:
Procedure
1. From the HOME screen, follow the Fast Key sequence, 2 Configure → 1 Guided
2. Select 2 for Hi-Hi Alarm, or Select 3 for Hi Alarm, or Select 4 for LO Alarm, or
3. If the alarm is disabled, select 1 Enable and press Enter. If the alarm was previously
4. Enter the alarm limit and press Enter.
5. Enter the alarm deadband and press Enter.
2.10

Remove power module

After the sensor and network have been configured, remove the Power Module and
replace the transmitter cover. The Power Module should be inserted only when the device
is ready to be commissioned. Use caution when handling the Power Module. The Power
Module may be damaged if dropped from heights in excess of 20 feet.
Rosemount 648 Wireless
Alert "OFF"
Alert "ON"
Deadband
Setup → 1 Guided Setup.
Select 5 for LO-LO Alarm and press Enter.
enabled, select 2 Leave Enabled and press Enter.
Alert "OFF"
Time
Configuration
May 2020
Assigned Value
Alert Set Point
31