Converting A Detector From Fixed To Portable; In-Line Filters - Honeywell SPM Flex User Manual

Single point monitor gas detector
Hide thumbs Also See for SPM Flex:
Table of Contents

Converting a detector from fixed to portable

A fixed-installation SPM Flex gas detector can be converted for portable
applications following this procedure:
1. Remove the detector from its mounting bracket (see
on page 29
or
"Backward-compatible mounting bracket" on page
Thumbscrew
Hex screw (2)
Figure 24. Handle removal
2. Slide the handle into place as shown above.
3. Secure the snap-in bracket on the bracket pin with one of the 1/8-inch hex screws,
as shown in the detail illustration.
4. Repeat on the other side of the detector.
5. The handle is hinged to allow access to the Chemcassette cartridge compartment.
Lower the handle and secure it with the thumbscrew.
6. If the optional shoulder strap is to be used, attach its spring clips to each of the
rings on the detector's handle. The strap must be laundered before use in clean-
room applications.

In-line filters

For most gases, it is imperative to use an external filter to protect the tubing and
the detector from contamination� Use particulate filter part number 780248 for
non-corrosive gases� Use filter 1991-0147 for corrosive gases� Refer to
Gases" on page 62
for specific gases� Replace the filters after every 3 to 6
months of operation, depending on the cleanliness of the installation environment�
Note that filters are not used with diisocyanates, hydrazine, ozone, or hydrogen
peroxide so regular maintenance cleaning is especially important for detectors
exposed to those gases�
SPM Flex Gas Detector
"Standard mounting bracket"
Handle
Shoulder
strap
ring (2)
30).
Detail
Snap-in
bracket
Hex
screw
Bracket
pin
"Detectable
33