Gassing The Sensor; Equipment Required; Calibration Gassing Procedure - Honeywell 705 Series Operating Instructions Manual

H2s sensor and transmitter for electrochemical cells ul certified
Hide thumbs Also See for 705 Series:
Table of Contents
MAN0362.P65 Issue 03, Mar 01
00705M5001

6.1 EQUIPMENT REQUIRED

To test and calibrate the gas detection system, the following
items are required:
a.
b.
Note:

6.2 CALIBRATION GASSING PROCEDURE

1.
2.
3.
To calibrate a H
(1)

6. GASSING THE SENSOR

A suitable Calibration Gas Test Kit and a bottle of the
appropriate test gas.
Where the sensor accessories are not fitted, a Sample
Flow Housing will be required for accurate sensor
calibration.
Purpose built gassing equipment is available on
request from Zellweger Analytics Limited.
The calibration of H
care and accurate gas mixtures. Electrochemical
sensors are better left factory calibrated and only an
approximate calibration check made.
If applying H
as on the Weather Protection Housing, the control
system may indicate up to 20% high. This is normal
and the system does not need recalibrating.
Should recalibration of a H
Sample Flow Housing should be used to ensure
accuracy.
S sensor, proceed as follows:
2
Connect a suitable calibration gas test kit to the sensor
gassing nozzle or to the permanently connected
pipeline, if fitted.
CAUTIONS
S sensors requires extreme
2
S test gas via a gassing nozzle, such
2
S sensor be required, a
2
21
705 Sensor