Technical Data; Specifications - Siemens XRS-5 Compact Operating Instructions

Ultrasonic transducers for sitrans
Hide thumbs Also See for XRS-5:
Table of Contents

Technical data

Note
Device specifications
Siemens makes every attempt to ensure the accuracy of these specifications but reserves
the right to change them at any time.
For a complete listing, including Approvals, see the full operating instructions.
5.1

Specifications

Specification
Process Application
Application
Max. range
Min. range
Vessel Pres-
sure
Operation
Beam Angle
Frequency
Temperature
Sensor
Supply Source
Environmental
Location
Max. altitude
Max. tempera-
ture
Min. tempera-
ture
Pollution de-
gree
Construction
Transducers
Compact Operating Instructions, 01/2018, A5E43390688_AA
XRS-5/5C
ST-H
Liquids
8 m (26 ft)
10 m (33 ft)
1)
0.3 m (1 ft)
Vented to atmosphere
10°
12°
44 kHz
Internal
Transducer shall only be supplied by a Siemens certified controller
Indoor/outdoor
2000 m
65 °C (149 °F)
ATEX:
60 °C (140 °F)
-20 °C (-4 °F)
-20 °C (-5 °F)
4
XPS-10
XPS-15/15F
Liquids and Solids
15 m (50 ft)
Standard:
0.3 m (1 ft)
XPS-15F:
0.45 m (1.5 ft)
95 °C (203 °F)
-40 °C (-40
Standard:
°F)
-40 °C (-40 °F)
XPS-15F:
-20 °C (-4 °F)
5
XPS-30
30 m (100 ft)
0.6 m (2 ft)
30 kHz
-40 °C (-40 °F)
37
Table of Contents
loading

This manual is also suitable for:

Xrs-5cXps-10Xps-15fXps-15St-hXps-30

Table of Contents