Common Trigger
Configurations
In the previous paragraphs, you learned about the basic building
blocks allowed in a SCPI trigger system. Generally, an instrument
implements only a portion of the trigger features available. These
paragraphs discuss the simplest configurations: INIT and TRIG.

The INIT Configuration

The 1NIT configuration is the simplest possible trigger configuration.
It uses no event detection states, and requires only two subsystems
for programming,
implement these two subsystems.
Idle
Initiate
Sequence
Figure l-40. The INIT Trigger Configuration
Command
Example commands using the INIT trigger configuration:
:INIT:IMM
:INIT:CONT ON
: INIT: CONT OFF
All SCPI
and
.
Parameters Parameter Type
Boolean
abort operations, go to idle
execute one sequence operation
execute sequence operations continuously
stop sequence operations after the current one is
complete
Getting Started Programming l-1 15
instruments
Actions