Chapter 3. OPERATION
c.
The ADTS 405F file system directories (under folder fs\etfs) are visible in the right hand
pane as shown in figure 3-14 File system directories screen below:
The example ETP (Test Programs) directory can be seen to contain two default test programs,
"altosc" and "asposc", used for oscillation of altitude and airspeed. New prepared programs may be
added from the PC left panes by drag and drop to the ETP directory. They will then become available
for stand-alone use under the front panel ETP key. Unwanted programs may be deleted from the
ETP directory as required.
Similarly, any reports previously generated using the "PRINT" key will be available under the
"SPOOL" directory. This may be transfer to the PC window for archive or hard copy print as required.
d.
The access session can now be closed using the Win SCP "Session" tab > Disconnect.

3.7.3 Connection of the IEEE 488 Facility

CAUTION
THE OPTION CONNECTOR COVER MUST ALWAYS BE FITTED WHEN THE
EQUIPMENT IS USED OUTDOORS OR IN AN ENVIRONMENT WHERE MOISTURE MAY ENTER
THE UNIT THROUGH THE OPTION CONNECTOR PANEL.
To enable and disable this IEEE 488 facility access the configuration menu. When enabled,
this facility will be available from power-up of the ADTS 405 with the settings held in non-
volatile memory.
The IEEE 488 connector is located behind the (optional) communications module cover plate and is
configured as detailed in Table 3-2 IEEE 488 facility below.
48 | ADTS 405 Mk2 Instruction Manual-English
Figure 3-14: ADTS File System Directories Screen
© 2015 General Electric Company – All rights reserved.