2.11.4
Min/Max Tracking diagnostic
Minimum and Maximum Temperature Tracking (Min/Max Tracking) when enabled records minimum and
maximum temperatures with date and time stamps on Rosemount 644 HART Head mount Temperature
Transmitters. This feature records values for Sensor 1, Sensor 2, Differential, Average, First Good and
Terminal temperatures. Min/Max Tracking only records temperature maxima and minima obtained since
the last reset, and is not a logging function.
To track maximum and minimum temperatures, Min/Max Tracking must be enabled using a Field
Communicator, AMS Device Manager, Local Operator Interface, or other communicator. While enabled,
this feature allows for a reset of information at any time, and all variables can be reset simultaneously.
Additionally, each of the individual parameter's minimum and maximum values may be reset
individually. Once a particular field has been reset, the previous values are overwritten.
Using a Field Communicator
From the HOME screen, enter the Fast Key sequence
Device Dashboard Fast Keys
Using AMS Device Manager
1. Right click on the device and select Configure.
2. In the left navigation window select Manual Setup.
3. On the Diagnostics tab, there is a group box labeled Sensor and Process Diagnostics; choose the
button for Configure Min/Max Tracking.
4. Follow the screen prompts to Enable and configure the settings for the diagnostic.
Using Local Operator Interface
Reference the figure below to find the path to the Thermocouple Diagnostic in the LOI menu.
Figure 2-23. Configuring Min/Max Tracking with LOI
VIEW CONFIG
SENSOR CONFIG
UNITS
RERANGE
LOOP TEST
DISPLAY
EXTENDED MENU
EXTENDED MENU
EXIT MENU
* Available only if option code (S) is ordered.
38
CALIBRAT
MIN-MAX MODE
DAMPING
PARAM CONFIG
VARIABLE MAP
VIEW VALUES
TAG
RESET VALUES
ALM SAT VALUES
BACK TO MENU
PASSWORD
EXIT MENU
SIMULATE
HART REV
HOT BACK CONFIG*
DRIFT ALERT*
TC DIAG CONFIG
MIN MAX TRACK
MIN MAX TRACK
BACK TO MENU
EXIT MENU