Invasive Blood Pressure - Datex-Ohmeda Cardiocap/5 User Manual

For anesthesia related to software s-xane 01
Hide thumbs Also See for Cardiocap/5:
Table of Contents

INVASIVE BLOOD PRESSURE

You need
(1) Heparinized fluid bag with pressure infuser
(2) Flushing set
(3) Transducer
(4) Adapter cable for using disposable transducers
1
Disposable
Disposable
4
Reusable
NOTE: Patient connections made according to the picture above
using D-O specified accessories are defibrillator proof.
WARNING: Use only defibrillator proof transducers and cables.
You can monitor up to two pressure channels.
CAUTION: Mechanical shock to pressure transducer may change zero
balance and calibration
2
3
Starting
• When doing the setup, prepare the transducer kit according
to the manufacturer's instructions.
• Ensure there is no air in the line. Refer to transducer
manufacturer's instructions on how to remove trapped air
from the transducer.
• Zero the transducer; open the transducer to air:
Press the Invasive Pressures key or press the ComWheel and
select Parameters and Invasive Press. Select Zero ALL.
• Open the line to the patient.
Combining pressures
You can display two invasive pressure waveforms one on top of
the other, using an area of one normal waveform, or both
waveforms combined in the same field with the same zero line.
1. Press the ComWheel and select Monitor Setup.
2. Select Screen Setup.
3. Select Waveform Fields.
4. Select Combine Pressures and YES.
WARNING: Ensure proper contact of the return electrode of the
electrosurgery unit to your patient to avoid possible burns at sensor
sites.
WARNING: Make sure that no part of the patient connections
touches any electrically conductive material including earth.
WARNING: Use only defibrillator proof transducers and cables.
– 31 –