Midas gas detector high-temperature pyrolyzer option

Quick Links

Quick Start Guide
Midas
Gas Detector
®
High-Temperature Pyrolyzer Option
(MIDAS-T-HTP)
3. Reassembling the Midas® detector
1. Align the PCB at the top rear of the main chassis
with the connector at the top of the mounting
bracket.
2. At the same time, align the two tubes at the
bottom rear of the main chassis with the two
tubes located on the bottom of the mounting
bracket.
3. Slide the chassis backwards on the mounting
bracket assembly so that the PCB, connector
and tubes engage fully by firmly pushing
the main chassis horizontally backwards on the
mounting bracket.
WARNING:
DO NOT PUSH ON THE LCD AS THIS MAY
CAUSE DAMAGE.
4. Align the two fixing screws located at the
bottom of the chassis with the screw threads on
the mounting bracket.
1. Introduction
The MIDAS-T-HTP pyrolyzer option is installed
underneath the standard Midas
®
gas detector. The air
sample is drawn through the pyrolyzer to the sensor
cartridge. The pyrolyzer converts perfluorocompounds
(PFCs) present in the air sample into hydrogen
fluoride (HF) by means of pyrolysis. The HF can then
be measured by the sensor and the concentration
displayed as the equivalent reading in ppm.
NOTE: To maintain stated sensor accuracy when
using the pyrolyzer, ensure ambient temperature of
installation point does not exceed 30°C (86°F). It is
recommended that the ventilated Midas
®
top cover
part number MIDAS-A-039 is used in all pyrolyzer
applications. Usage above this temperature may
require more frequent bump testing or calibration to
confirm working specification.
(Contact Honeywell Analytics for the availablity of
other detectable gases using this pyrolyzer module)
5. Tighten the screws to secure the chassis to the
mounting bracket.
6. Fit the MIDAS-S-CFX sensor cartridge into the
sensor cartridge chamber (See Diagram 4, and
refer to the Sensor Cartridge Installation Quick
Start Guide MIDAS-A-021).
7. Set the power switch to the "on" position.
8. Refit optional Midas
®
ventilated top by aligning the
slots on either side with locating tabs on the
mounting bracket assembly and push the cover
horizontally until seated.
9. Tighten thumbscrew on the front panel
(See Diagram 5)
2. Fitting the pyrolyzer module
See reverse for all diagrams
1. Isolate power to the detector.
2. Unscrew the captive thumbscrew located on the
front panel.
3. Remove the cover by pulling it forward off the main
chassis (See Diagram 1).
4. Unscrew the two captive retaining screws located
at the bottom front of the chassis.
5. Pull the main chassis forward to disconnect it from
the mounting bracket assembly (See Diagram 2).
6. Thread the connector and wire harness from
the pyrolyzer through the rectangular access in the
bottom of the mounting bracket and secure wires
with retention clip (if available).
7. Plug the connector into the socket (con5) at the
bottom left of the terminal board.
4. Configuration
1. After completion of the startup sequence, press
the 's' button for a few seconds to select the
setup menu.
2. Enter the passcode (if set).
3. Use the 's' or 't' buttons to select the set up
menu '
' icon and press the '3' to accept.
4. Use the 's' or 't' buttons to select the set alarms
' ALm' submenu and press '3' to accept.
5. The flashing gas ID code is displayed along with
the gas cylinder and alarms icon '
'.
6. Use the 's' or 't' buttons to change the gas ID
number to to the appropriate PFC gas. See the
manual for further details.
7. Continue to accept or change the rest of the alarm
settings. *
8. Press '3' to update all changes (UPdt displayed).
9. Press 'X' twice to return to normal operation.
* For further details regrading the Midas
pyrolyzer
®
option, refer to the Midas
®
operating manual
MIDAS-A-001.
8. Offer the pyrolyzer module up underneath the
mounting bracket ensuring that the fitting at the
top rear of the pyrolyzer mates with the sample
inlet port at the bottom of the mounting bracket.
9. Align the three screw mounting bosses on the
top of the pyrolyzer with the three screw holes in
the mounting bracket.
10. Insert and tighten the three screws provided
(See Diagram 3).
Find out more
www.honeywellanalytics.com
Contact Honeywell Analytics:
Europe, Middle East, Africa, India
Life Safety Distribution AG
Wilstrasse 11-U31
CH-8610 Uster
Switzerland
Tel: +41 (0)44 943 4300
Fax: +41 (0)44 943 4398
Americas
Honeywell Analytics Inc.
405 Barclay Blvd.
Lincolnshire, IL 60069
USA
Tel: +1 847 955 8200
Toll free: +1 800 538 0363
Fax: +1 847 955 8208
Please Note:
While every effort has been made to ensure accuracy in this
publication, no responsibility can be accepted for errors or
Asia Pacific
omissions. Data may change, as well as legislation, and you are
strongly advised to obtain copies of the most recently issued
Honeywell Analytics Asia Pacific
regulations, standards, and guidelines.This publication is not
#508, Kolon Science Valley (I)
intended to form the basis of a contract.
187-10 Guro-Dong, Guro-Gu
Seoul, 152-050,
Korea
Tel: +82 (0)2 2025 0300
Fax: +82 (0)2 2025 0329
Technical Services
Issue 1.1 03.06
Part No. MIDAS-A-038
10/08
www.honeywell.com
© 2008 Honeywell Analytics
loading

Summary of Contents for Honeywell MIDAS-T-HTP

  • Page 1 Quick Start Guide 1. Introduction 2. Fitting the pyrolyzer module The MIDAS-T-HTP pyrolyzer option is installed See reverse for all diagrams 8. Offer the pyrolyzer module up underneath the underneath the standard Midas ® gas detector. The air mounting bracket ensuring that the fitting at the...
  • Page 2 Diagram 2: Diagram 1: Removing the main chassis Removing the Midas cover ® chassis loosen cover mounting bracket mounting bracket Diagram 3. Fitting the pyrolzer module Terminal module Power Switch Diagram 4: Diagram 5: Socket Fitting sensor cartridge Fitting cover (con5) chassis Rectangular access...