Electrical loop calibrator; k393
Hide thumbs Also See for Druck DPI 832:

Quick Links

GE Infrastructure
Sensing
Druck DPI 832
Electrical loop calibrator
User manual - K393
99 Washington Street
Melrose, MA 02176
Fax 781-665-0780
TestEquipmentDepot.com
Table of Contents
loading

Summary of Contents for GE Druck DPI 832

  • Page 1 GE Infrastructure Sensing 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com Druck DPI 832 Electrical loop calibrator User manual - K393...
  • Page 2 Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 K393 Issue 1 FAX 781.665.0780 - TestEquipmentDepot.com...
  • Page 3: Table Of Contents

    Table of Contents Introduction The DPI 832 Electrical Loop Calibrator is part of the Introduction ................. 1 Druck DPI 800 series of hand held instruments. Safety ..................1 The DPI 800 series uses Intelligent Digital Output Sensor Safety - Marks and symbols on the instrument ....2 (IDOS) technology to give instant plug and play To start ..................
  • Page 4: Safety - Marks And Symbols On The Instrument

    Safety (Continued) To start - Items on the display Before you start an operation or procedure in this Item Description publication, make sure that you have the necessary skills Task indication for the switch test. (if necessary, with qualifications from an approved = switch closed = switch open training establishment).
  • Page 5: To Start - Power On Or Off

    To start - Power on or off To start - Select a task (Measure and/or supply) To turn the instrument on or off, press (A1 - item [1]). The When the instrument is set up (Table 1), use the Select Task instrument does a self test and then shows the applicable menu to select the applicable task.
  • Page 6: To Start - Set Up The Settings

    To start - Set up the settings Table 4: (Part of table) Menu options - Settings (Input) When the task is set up (Table 2/3), use the Settings menu Options Description to adjust the input and/or output operation. (If applicable) To select and set up the filter values to give a Display: Task Settings selection...
  • Page 7 Table 5: Menu options - Settings (Output) Table 6: Additional data for Settings (Output): Options Description Item Value To select and set up a value for the “Nudge” Span Check output. Example: 1.000mA increments. Low (0%) Set the 0% value. Additional data: Select Settings ( High (100%) Set the 100% value.
  • Page 8: Operation

    Operation Operation - Change the output values When the output operation is set up (Table 5), use one of This section gives examples of how to connect and use the these procedures to change the output values: instrument. Before you start: •...
  • Page 9: Operation - Supply Mv Or Volts

    Operation - Supply mV or Volts To supply mV or Volts: 1. Connect the instrument (Figure 2) and, if necessary, Maximum: 30V adjust the Set Up (Table 1). 2. Select the Volts output task from Select Task (Table 2/3) and select the necessary range (mV or Volts). 3.
  • Page 10: Operation - Switch Test

    Operation - Switch test Operation - UPM Pressure measurements To do tests on a switch: Read all the instructions supplied with the UPM and then use the specified procedures to connect it (Figure 6/7). 1. Connect the instrument (Figure 5) and, if necessary, adjust the Set Up (Table 1).
  • Page 11: Operation - Error Indications

    All the other configuration options stay in memory. Calibration Note: GE can provide a calibration service that is traceable to international standards. We recommend that you return the instrument to the manufacturer or an approved service agent for calibration.
  • Page 12: Calibration - Procedures (Mv Or Volts Input)

    Selection sequence: 6. Make sure that the error is in the specified limits (Table 10 or Table 11). Select Task (Table 2) Set Up (Table 1) Calibration Display: Enter Menu: Table 10: mV input error limits Calibration PIN Select Channel Applied Calibrator Permitted...
  • Page 13: Calibration - Procedures (Mv Or Volts Output)

    Calibration - Procedures (mV or Volts output) Calibration - Procedures (mA output) 1. Connect the instrument to the calibration equipment 1. Connect the instrument to the calibration equipment (Figure 2). (Figure 3). 2. Let the equipment get to a stable temperature 2.
  • Page 14: Specification Data

    Specification data All accuracy statements are for one year. Specification - General Languages English [Default] BS EN 61326-1:1998 + A2:2001 Operating 14 … 122°F (-10 … 50°C) Safety Electrical - BS EN 61010:2001; CE Marked temperature Size (L: W: H) 7.1 x 3.3 x 2.0 in Storage -4 …...