Table of Contents Introduction The DPI 832 Electrical Loop Calibrator is part of the Introduction ................. 1 Druck DPI 800 series of hand held instruments. Safety ..................1 The DPI 800 series uses Intelligent Digital Output Sensor Safety - Marks and symbols on the instrument ....2 (IDOS) technology to give instant plug and play To start ..................
Safety (Continued) To start - Items on the display Before you start an operation or procedure in this Item Description publication, make sure that you have the necessary skills Task indication for the switch test. (if necessary, with qualifications from an approved = switch closed = switch open training establishment).
To start - Power on or off To start - Select a task (Measure and/or supply) To turn the instrument on or off, press (A1 - item [1]). The When the instrument is set up (Table 1), use the Select Task instrument does a self test and then shows the applicable menu to select the applicable task.
To start - Set up the settings Table 4: (Part of table) Menu options - Settings (Input) When the task is set up (Table 2/3), use the Settings menu Options Description to adjust the input and/or output operation. (If applicable) To select and set up the filter values to give a Display: Task Settings selection...
Page 7
Table 5: Menu options - Settings (Output) Table 6: Additional data for Settings (Output): Options Description Item Value To select and set up a value for the “Nudge” Span Check output. Example: 1.000mA increments. Low (0%) Set the 0% value. Additional data: Select Settings ( High (100%) Set the 100% value.
Operation Operation - Change the output values When the output operation is set up (Table 5), use one of This section gives examples of how to connect and use the these procedures to change the output values: instrument. Before you start: •...
Operation - Supply mV or Volts To supply mV or Volts: 1. Connect the instrument (Figure 2) and, if necessary, Maximum: 30V adjust the Set Up (Table 1). 2. Select the Volts output task from Select Task (Table 2/3) and select the necessary range (mV or Volts). 3.
Operation - Switch test Operation - UPM Pressure measurements To do tests on a switch: Read all the instructions supplied with the UPM and then use the specified procedures to connect it (Figure 6/7). 1. Connect the instrument (Figure 5) and, if necessary, adjust the Set Up (Table 1).
All the other configuration options stay in memory. Calibration Note: GE can provide a calibration service that is traceable to international standards. We recommend that you return the instrument to the manufacturer or an approved service agent for calibration.
Selection sequence: 6. Make sure that the error is in the specified limits (Table 10 or Table 11). Select Task (Table 2) Set Up (Table 1) Calibration Display: Enter Menu: Table 10: mV input error limits Calibration PIN Select Channel Applied Calibrator Permitted...
Calibration - Procedures (mV or Volts output) Calibration - Procedures (mA output) 1. Connect the instrument to the calibration equipment 1. Connect the instrument to the calibration equipment (Figure 2). (Figure 3). 2. Let the equipment get to a stable temperature 2.
Specification data All accuracy statements are for one year. Specification - General Languages English [Default] BS EN 61326-1:1998 + A2:2001 Operating 14 … 122°F (-10 … 50°C) Safety Electrical - BS EN 61010:2001; CE Marked temperature Size (L: W: H) 7.1 x 3.3 x 2.0 in Storage -4 …...