Times Required To Perform Prime, Washes, And Cell Blank; Daily Operation; Using Workorders; Using Host Computer Workorders - Siemens ADVIA 1800 Operator's Manual

Chemistry system
Table of Contents

Times required to perform prime, washes, and cell blank

Typical times
Operation
PRIME1
WASH1
WASH2
WASH3
Cell blank
Times for combined operations during startup and shutdown
Startup
PRIME1 + WASH3 = 28 minutes
PRIME1 + WASH3 + Cell blank = 42 minutes
Shutdown
WASH2 + System End + Power OFF = 39 minutes
Automatic startup
PRIME1 + WASH3 = 28 minutes
PRIME1 +WASH3 + Cell blank (wait for CB Temp) = 72 minutes
PRIME1 + WASH3 + Cell blank (no wait for CB TEMP) = 42 minutes

Daily Operation

Using workorders

Using host computer workorders

1. Create the workorders at your host computer.
Each patient sample must have a workorder that contains a sample number and at least one
test request.
NOTE:
computer before operation. Instead, use the Automatic item select feature in the
Automatic transfer area on the Online Settings window.
2. Download workorders to the
a. At the Menu Panel, select
b. Select
c. In the Entry format area, select the means for identifying the first workorder in
d. In the Last no. entry format area, select the means for identifying the last
e. In the Start no. box, identify the first workorder you want downloaded.
78
Cycles
Time (minutes)
5
1:40
1
9:10
2
38
1
26
NA
14
Stop here if you do not want to download workorders manually from the host
.
Host request
step 2e.
workorder in step 2f.
hemistry system:
c
, then select
Request
Order Entry
.
ADVIA 1800 Operator's Guide