1 Technical data KERN EW 150-3M EW 600-2M EW 1500-2M Readout (d) 0,001 g 0,01 g 0,01 g Verification value (e) 0,01 g 0,1 g 0,1 g Weighing range (Max) 150 g 600 g 1.500 g Min. load (Min) 0,02 g...
Page 3
KERN EW 3000-2M EW 6000-1M Readout (d) 0,01 g 0,1 g Verification value (e) 0,1 g Weighing range (Max) 3.000 g 6.000 g Min. load (Min) 0,5 g Taring range 3.000 g 6.000 g Reproducibility 0,01 g 0,1 g Linearity ±...
2 Unpacking and standard accessories Take the balance carefully out of the packing and store the plastic wrap for a possible future transport. The standard accessories consist of: x Balance x Weighing pan x Weighing pan support x External mains adapter x Operating instructions 3 Setting up the balance x Place the pan base packed with the weighing pan on the balance.
Display panel 5 Installation Please note that your KERN balance is a precision weighing instrument. Even if the instrument is made for robust and continuous use, please handle it carefully and do avoid unnecessary use. The balance requires a warm-up time. The optimum weighing results are achieved when starting to used the balance only 5 minutes after turning it on.
This declaration ceases its validity at each changing of the above mentioned apparatus which is not in accordance with KERN. Date: 20 January 2000 Signature: Gottl. KERN & Sohn GmbH Managing Director Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72322 Balingen-Frommern, Tel. +49-07433/9933-0,Fax +49-07433/9933-149 EW-BA-defsi-0620...
After the successful adjusting it is necessary to make the adjusting possibility impossible on KERN EW balances. The switch for this step is located on the inside of the balance (beneath the housing, above the display) (only with models KERN EW).
Page 9
(see table), but not ideal from a metrological point of view. Model Alternative adjusting weight EW 150-3M 50 g EW 600-2M 200 g EW 1500-2M 500 g EW 3000-2M 1.000 g EW 6000-1M 2.000 g Information concerning the adjusting weights is available at: http://www.kern-sohn.com...
9 Operating the balance 9.1 Warm-up time Optimum weighing results are being achieved when starting to operate the balance only 5 minutes after turning the balance on. 9.2 Power display If the symbol (*) is visible, the balance is provided with power via the mains adapter. By pressing the ON/OFF key the balance is in the weighing mode.
9.3 Balance zero display By pressing the Tare key the balance is set to zero. On the display appears 0,00 g. Additionally, the symbol for the balance zero display (ZERO) is displayed. 9.4 Tare weighing (taring) x Place the tare cup on the weighing pan x Press the tare key x Balance display turns to 0,00 g x The weight of the cup has been stored now internally.
Page 12
10 General program settings The balance leaves the factory with the factory settings being set. This is the standard configuration of the balance. It is marked by a *. It can be altered as follows: Press the F key for about 4 seconds until Func appears on the display. While releasing the key, 1.
Page 13
Baud rate 7. b.L. 1200 Baud 2400 Baud 4800 Baud Unit commutation 8. Set. g/ct not documented not documented Not documented 0. GLP Always use this setting Not documented A. PrF. Always use this setting 11 Important information These electronic balances are precision instruments. It is designed for use in a self-controlled electromagnetic environment.
The adjusting switch (right) is not in the “WEIGH“ position. If other error messages occur, turn the balance off and on again. The weighing result is obviously wrong. If the error message still appears, please contact KERN. EW-BA-defsi-0620...