Intended Use
This device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with the measurement result.
Please read this instruction manual thoroughly before using the device. Please keep for future reference. For specific information about your own blood pressure, CONSULT YOUR PHYSICIAN.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
General Usage
DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician.
Only a physician is qualified to diagnose and treat High Blood Pressure.
This device is not intended to be a diagnostic device.
Consult your physician before using the device for any of the following conditions: common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, age, pregnancy, pre-eclampsia, renal diseases.
Note that PATIENT motion, trembling, shivering may affect the measurement reading.
Do not use the device on the injured arm or the arm under medical treatment.
Do not apply the arm cuff on the arm while being on an intravenous drip or blood transfusion.
Consult your physician before using the device on the arm with an arterio-venous (A-V) shunt.
Do not use the device with other medical electrical (ME) equipment simultaneously.
Do not use the device in the area the HF surgical equipment, MRI, or CT scanner exists, or in the oxygen rich environment.
The air tube or the AC adapter cable may cause accidental strangulation in infants.
Contains small parts that may cause a choking hazard if swallowed by infants.
Keep the battery out of reach of children.
Do not use this product on aircraft or in hospitals. Please remove the battery from the unit. This product emits radio frequencies (RF) in the 2.4 GHz band, use of this product in locations where RF is restricted is not recommended.
The use of RF in this product is licensed for use by the FCC, for further information on any potential restrictions refer to documentation on Bluetooth® usage by the FCC.
AC Adapter (optional accessory)
Do not use the AC adapter if the device or the power cord is damaged.
Turn off the power and unplug the power cord immediately.
Plug the AC adapter into the appropriate voltage outlet. Do not use in a multi-outlet plug.
Never plug in or unplug the power cord from the electric outlet with wet hands.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.
General Usage
Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous.
Consult your physician before using the device for any of the following conditions:
Do not take measurements more than necessary. It may cause bruising due to blood flow interference.
Remove the arm cuff if it does not start deflating during the measurement. Do not use this device on infants or persons who cannot express their intentions.
Do not use the device for any purpose other than measuring blood pressure.
Use only the approved arm cuff for this device. Use of other arm cuffs may result in incorrect measurement results.
Do not use a mobile phone or other devices that emit electromagnetic fields, near the device. This may result in incorrect operation of the device.
Do not disassemble the monitor or arm cuff. This may cause an inaccurate reading.
Do not use in a location with moisture, or a location where water may splash on the device. This may damage the device.
Do not use the device in a moving vehicle (car, airplane).
Read "What to do if your systolic pressure is more than 210 mmHg" of this instruction manual, if your systolic pressure is known to be more than 210 mmHg. Inflating to a higher pressure than necessary may result in bruising where the arm cuff is applied.
AC Adapter (optional accessory)
Fully insert the power plug into the outlet.
When disconnecting the power plug from the outlet, do not pull the power cord. Be sure to pull from the power plug safely.
When handling the power cord, take care not to do the following:
Do not damage.
Do not break it.
Do not tamper with it.
Do not forcibly bend or pull.
Do not twist.
Do not bundle during use.
Do not pinch.
Do not place under heavy objects.
Wipe the dust off from the power plug.
Unplug monitor when not in use.
Disconnect the power plug before cleaning.
Use only an OMRON AC adapter designed for this device. Use of unsupported adapters may damage and/or may be hazardous to the device.
Battery Usage
Do not insert the batteries with their polarities incorrectly aligned.
Use only 4 "AA" alkaline or manganese batteries with this device. Do not use other types of batteries. Do not use new and used batteries together.
Remove the batteries if the device will not be used for three months or more.
Contents:
Monitor, arm cuff, 4 "AA" batteries, instruction manual, quick start guide
Monitor:
Arm cuff: Arm circumference 9" - 17" (22 - 42 cm)
Display:
Irregular Heartbeat Symbol ()
When the monitor detects an irregular rhythm two or more times during the measurement, the irregular heartbeat symbol will appear on the display with the measurement values. An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure. If the irregular heartbeat symbol displays with your measurement results, we recommend you consult your physician. Follow the directions of your physician.
Movement Error Symbol ()
The movement error symbol is displayed if you move your body during the measurement. Please remove the arm cuff, and wait 2 - 3 minutes. Take another measurement, remain still during measurement.
SYNC Symbol ()
The SYNC symbol () is displayed if device is not connected to Telehealth service receiver or when data is not transmitted successfully. Please see "Connecting failure. Data is not being transmitted." in Troubleshooting section.
Before use, thoroughly read the instruction manual included with the Telehealth service receiver (in some cases, receiver maybe your smart phone) being used with this blood pressure monitor for instruction about your monitor to Telehealth service receiver, and receiver transmission range. It will be necessary for the blood pressure monitor to be within the receiver's transmission range to successfully transfer data. This Omron blood pressure monitor is designed to connect to specific Bluetooth® Smart receivers, and is not guaranteed to connect to all Bluetooth® Smart compatible devices and Telehealth service receiver. "Omron Healthcare, Inc. shall not be liable for loss of use or any other special, incidental, consequential or indirect costs, expenses or damages due to improper usage of this monitor."
To help ensure an accurate reading, follow these directions:
Below procedures are ONLY done at inserting batteries.
Notes:
Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm.
Don't place the arm cuff over thick clothes.
To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature.
Notes:
What to do if your systolic pressure is more than 210 mmHg
After the arm cuff starts to inflate, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
Notes:
Note: Wait 2-3 minutes before taking another measurement. Waiting between measurements allows the arteries to return to the condition prior to taking a measurement.
DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure.
This device is not intended to be a diagnostic device.
Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous.
Inflating to a higher pressure than necessary may result in bruising where the arm cuff is applied.
Error Display | Cause | Solution |
Irregular heartbeats are detected. | Remove the arm cuff. Wait 2 - 3 minutes and then take another measurement. Repeat the steps in Taking a Measurement section. If this error continues to appear, contact your physician. | |
Movement during measurement. | Carefully read and repeat the steps in Taking a Measurement section. | |
Connecting failure. Data is not being transmitted. | Refer "Connecting failure. Data is not being transmitted." in Troubleshooting section. | |
The batteries are low. | Recommend to replace the batteries with new ones ahead of time. Refer to Preparation section. | |
The batteries are depleted. | Replace 4 batteries with new ones at once. Refer to Preparation section. | |
Air plug disconnected. | Insert the plug securely. Refer to Applying the Arm Cuff section. | |
Arm cuff is applied too loosely. | Apply the arm cuff tighter. Refer to Applying the Arm Cuff section. | |
Air is leaking from the arm cuff. | Replace the arm cuff with a new one. Refer to Optional Accessories section. | |
Movement during measurement and the arm cuff has not been inflated sufficiently. | Repeat measurement. Remain still and do not talk during measurement. Refer to Taking a Measurement section. | |
If "E2" appears repeatedly, inflate the arm cuff manually until it is 30 to 40 mmHg above your previous measurement result. Refer to Taking a Measurement section. | ||
The arm cuff was inflated exceeding the maximum allowable pressure, and then deflated automatically when inflating the arm cuff manually. | Do not touch the arm cuff and/or bend the air tube while taking a measurement. Do not inflate the arm cuff more than necessary. Refer to Taking a Measurement section. | |
Movement during measurement. | Repeat measurement. Remain still and do not talk during measurement. Refer to Taking a Measurement section. | |
Clothing is interfering with the arm cuff. | Remove any clothing interfering with the arm cuff. Refer to Applying the Arm Cuff section. | |
Device error. | Contact your Telehealth service provider. |
Problem | Cause and Solution | |
No power. No display appears on the monitor. | Replace all batteries with new ones. Check the battery installation for proper placement of the battery polarities. Refer to Preparation section. | |
Measurement values appear too high or too low. | Blood pressure varies constantly. Many factors including stress, time of day, and how you wrap the cuff, may affect your blood pressure. Review the Before Taking a Measurement section and Taking a Measurement section. | |
Connecting failure. Data is not being transmitted. | The blood pressure monitor might not be properly placed within the receiver's transmission range and is too far from the receiver. If there are no causes of data transmission interference found near the blood pressure monitor, move the blood pressure monitor within 16 ft. (5 m) of the receiver and try again. | |
The Bluetooth® feature on the receiver is turned off. Turn the Bluetooth® feature on the receiver ON, and try to transmit the readings again. | ||
The blood pressure monitor did not pair successfully to the receiver. Try to pair the devices again. Refer to Preparation section. | ||
The application on the receiver or destination device is not ready. Check the application then try to transmit the readings again. Refer to Preparation section. If the ERR symbol is on the screen after checking the application, contact your Telehealth service provider. |
To protect your device from damage, please follow the directions below:
Do not store the device in the following situations:
To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first, and then remove the AC adapter plug from the monitor.
Model | HEM-9200T |
Display | LCD digital display |
Measurement range | Pressure: 0 to 299 mmHg Pulse: 40 to 180 beats / min. |
Accuracy | Pressure: ±3 mmHg or 2% of reading Pulse: ± 5% of display reading |
Inflation | Fuzzy-logic controlled by electric pump |
Deflation | Automatic pressure release valve |
Measurement method | Oscillometric method |
IP classification | IP 20 |
Power source | 4 "AA" batteries 1.5V or optional AC adapter (INPUT AC100-240V 50/60Hz 0.12A) |
Battery life | Approximately 1000 measurements (using new alkaline batteries) |
Operating temperature / humidity | 50°F to 104°F (10°C to 40°C) / 15 to 90% RH |
Storage temperature / humidity / air pressure | -4°F to 140°F (-20°C to 60°C) / 10 to 95% RH / 700 to 1060 hPa |
Weight | Monitor : Approximately 10 oz. (290 g) not including batteries Arm cuff: Approximately 6 oz. (170 g) |
Dimensions | Monitor : Approximately 4 1/4" (w) × 3 1/8" (h) × 5 1/2" (l) (107 mm × 79 mm × 141 mm) Arm cuff: Approximately 5 3/4" × 23 1/2" (air tube: 29 1/2") (145 mm × 594 mm (air tube: 750 mm)) |
Cuff circumference | 9" to 17" (220 to 420 mm) |
Contents | Monitor, arm cuff, 4 "AA" batteries, instruction manual, quick start guide |
Applied part | = Type BF |
Protection against electric shock | Internally powered ME equipment (When using only the batteries) = Class II ME equipment (Optional AC adapter) |
Notes:
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.