Thank you for purchasing Omron Digital Thermometer.
The thermometer you purchased offers a safe, accurate and quick temperature reading. You can measure your temperature through oral, rectal or underarm. This thermometer eliminates any worries about broken glass or mercury hazards. Note that your temperature is affected by smoking, eating or drinking. Thoroughly read this instruction manual before using your OMRON MC-341 Digital Thermometer. We recommend that you practice taking your temperature with the thermometer. In this way you will gain confidence in your technique when using the thermometer. Contact your doctor if you have any questions regarding specific temperatures.
To assure the correct use of the product basic safety, measures should always be followed including the precautions listed below.
GENERAL SAFETY PRECAUTIONS
Press the button for 7 seconds until a beep is emitted.
The display will change to the other measurement scale at the time when the beep sounds.
All of the display symbols appear.
The thermometer will automatically switch itself off.
If possible, a full probe cover covering all sections should be used. Replace the probe cover with a new one after each use.
Probe covers are included with the thermometer, place a probe cover carefully over the tip.
BUZZER NOTIFICATION TIMING:
Oral : approx. 20 seconds
Rectal : approx. 10 seconds
Underarm : approx. 25 seconds
* 25 seconds is the earliest estimated temperature indication. To complete the measurement, it is required to measure the temperature in the underarm between 2 - 5 minutes (depending on the ambient temperature and condition).
NOTES:
NOTES:
Press the button to display the temperature that was measured most recently. For details, please refer to the "TAKING THE TEMPERATURE" section.
Measurement accuracy cannot be assured when the method used to measure the temperature is incorrect.
ORAL USE (for ages 4 and older):
The mouth should remain closed up to 5 minutes before attempting a reading.
RECTAL USE:
Always use a probe cover when taking rectal measurements.
UNDERARM (AXILLARY) USE:
* In the case of infants and very young children, gently hold the arm to prevent movement.
INCORRECT MEASUREMENTS
Only applies when taking underarm temperatures - The following may result in incorrect measurements. (See "UNDERARM (AXILLARY) USE")
|
Battery: Alkaline-Manganese Button Battery LR41 (available commercially)
Battery Replacement Indicator ( )
Replace the battery when the battery replacement indicator appears when the thermometer is switched on.
NOTES:
INDICATOR | CAUSE | REMEDY |
Displayed when the temperature of the sensing section is less than 32°C (89.6°F). | This is not an error so it may be ignored and the temperature taken. | |
Displayed when the temperature of the sensing section is greater than 42°C (107.6°F). | Cool the sensing section and then take the temperature. | |
Temperature Displayed | Displayed when the surrounding temperature exceeds 32°C (89.6°F). | This is not an error, so the temperature may be taken as is. |
PROBLEM | CAUSE | CORRECTION |
Wide differences in readings | | Check to see that the thermometer is being placed at the location as described in the "CORRECT MEASUREMENT" section. |
Nothing appears in the display section when the power switch is pressed. | The polarity is reversed. | Take out the battery and replace using the correct polarity. |
The battery is depleted. | Check to see if appears in the display section. Replace with a new battery (LR41). | |
[37.0°C (98.6°F)] is not displayed during the measurement preparation period. | A temperature from [36.9°C (98.4°F)] to [37.1°C (98.8°F)] is displayed. | The thermometer may be used as it is. |
A temperature other than that in the range from [36.9°C (98.4°F)] to [37.1°C (98.8°F)] is displayed, even after turning the thermometer on and off several times. | Please contact your nearest authorized OMRON dealer. |
Product Name: | DIGITAL THERMOMETER MC-341 |
Power Consumption: | 0.1mW |
Sensing Unit: | Thermistor |
Temperature Display: | 3-digit, +°C (°F) display in 0.1 degree increments |
Measurement Accuracy: ± 0.1°C (32.0 to 42.0°C) | ± 0.2°F (89.6 to 107.6°F) (when measured at a standard room temperature of 23°C (73.4°F) in a test tank held at constant temperature) |
Power Supply: | 1.5V DC, 1 LR41 Alkaline-Manganese Button Battery |
Battery Life: | Approx. 2 years or more (3 times per day) |
Measurement Method: | Actual measurement |
Applied Part: | =Type B |
Protection Against Electric Shock: | Internally powered ME equipment |
Measurement Range: | 32.0 to 42.0°C (89.6 to 107.6°F) |
Usage Environment Temp and Humidity: | Surrounding temperature +10 to +40°C (+50 to +104°F), Relative Humidity 30-85%RH |
Storage Environment Temp and Humidity: | Surrounding temperature -20 to +60°C (–4 to +140°F), Relative Humidity 10-95%RH |
Weight: | Approx. 12g (with battery installed) |
External Dimensions: | 19.4 mm (w) × 132.5 mm (l) × 10.0 mm (d) |
Accessories: | Test Battery (Alkaline-Manganese Button Battery LR41), Protective Case, Instruction Sheet, 5 Probe Covers. |
The specifications may be changed without prior notice.
This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co., Ltd., Japan.
OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto 617-0002 JAPAN | |
OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR Hoofddorp THE NETHERLANDS | |
Production Facility | OMRON (DALIAN) CO., LTD. Dalian, China |
Asia Pacific HQ | OMRON HEALTHCARE SINGAPORE PTE LTD. 438A Alexandra Road, #05-05/08, Alexandra Technopark Singapore 119967 www.omron-healthcare.com.sg |
Read the instruction in this manual carefully.
This device fulfills the provisions of the EC directive
93/42/EEC (Medical Device Directive) and the European Standard EN12470-3:2000, Clinical thermometers - Part 3: Performance of compact electrical thermometers (nonpredictive and predictive) with maximum device.
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.