Inspection Before Use; Inspection For Appropriate Connection; Inspection For Material Surface - Hitachi EUP-V53W Instruction Manual

Table of Contents

2. Inspection before Use

Prior to use, the probe and accessories must be carefully inspected
that they are appropriate for use. If you find any damage, do not use
them and contact a service support immediately.

2.1. Inspection for Appropriate Connection

1) Confirm that the system is correctly operating. Refer to the
instruction manual for the Main unit.
2) Do not attach or connect unauthorized devices or instruments on the
probe, such as unauthorized biopsy attachments.
3) Confirm that the Sterile Puncture Adapter and software version and
then settings of the scanner are appropriate for the probe. Attach
the Sterile Puncture Adapter on the probe. Set the main unit to
display the "Needle guide line". (Refer to the operation manual for
the main unit.) Keep the probe head in the water and insert a puncture
needle in the Sterile Puncture Adapter. Then, confirm that the needle
is inserted smoothly and the echo of the needle is displayed on the
dot line "Needle guide line" on the monitor.

2.2. Inspection for Material Surface

1) Visually inspect the surface of the probe head, housing and cable
for any crack, scratch or denaturalization.
2) Visually inspect the envelope of the Sterile Puncture Adapter for
any break, deformation, crack or denaturalization. If you find any
damage, do not use the Sterile Puncture Adapter.
3) Visually inspect the Magnetic sensor attachment and the Spacer for
EZU-RV2S for any crack, deformation or denaturalization.
The Spacer for EZU-RV2S is small, please do not lose the Spacer.
CAUTION
-5-
Q1E-EP0325