Maintenance

To keep the sampler in good operating condition, perform the
following basic maintenance as required:
Keep an operations log of all maintenance and fault
events.
Clean the exterior with a soft damp cloth. Do not use
solvents, soaps or polishes.
Do not immerse the sampler in liquids.
Confirm that the filter is free of dirt and replace it if
necessary.
Confirm that the water trap is empty.
optimumstores
www.
Shop for BW products online at:

Replacing the Batteries

a Warning
To avoid personal injury:
⇒ Replace the batteries as soon as the sampler
indicates a Low Battery Alarm.
⇒ Use only batteries recommended by
BW Technologies to prevent damage or personal
injury and maintain intrinsic safety.
⇒ Use only approved AA alkaline or charged AA NiMH
batteries, properly installed in the sampler. See the
Specifications section for approved batteries.
⇒ Charge batteries using only a recommended
charger. Do not use any other charger. Failure to
observe this precaution could lead to fire or
explosion.
⇒ Do not change or charge batteries in a hazardous
location. Doing so will impair the intrinsic safety of
the unit, and may lead to fire or explosion.