Service Modes - GE DINAMAP Compact TS Service Manual

Vital signs monitor
Table of Contents

3.2 Service Modes

There are three different service modes, each offering different levels of access to the monitor' s set-up and
calibration. These modes can be accessed using the rotor control and display panel. The modes are:
1. The " Clinician Menu" mode, allowing:
? ?
setting of temperature units,
? ?
checking NIBP and temperature calibration dates,
? ?
setting the alarms into " Permanent Silence" mode.
(This mode is described in the DINAMAP
2. The " Main Service Mode" provides the features allowed from the clinician menu (described
above) plus the following:
? ?
Checking NIBP calibration,
? ?
NIBP re-calibration,
? ?
Pop Off (overpressure detection) calibration,
? ?
Changing temperature algorithm,
? ?
Set temperature calibration date,
? ?
Configuration for use with Central Station,
? ?
Changing language of operation,
? ?
Monitor " Set-up" - specifying whether SpO
3. The " Super Service Mode" is used to allow the calibration of the Predictive Temperature circuitry and
Neonatal Overpressure checking procedure. Other menu options within this mode are for
manufacturing test purposes only.
Note : Certain menu functions may not be present or appearance may be slightly different than depicted in
the following section if unit is fitted with software earlier than 8613-V4.1.
To enter the service mode from the Main Menu, use the rotor control and LCD.
First select More...
then Service
Compact Monitor Operating Manual).
, printer, and temperature are fitted.
2
29
Table of Contents
loading

Table of Contents