7 Repair procedures
7.9.8 Replace gas inlet
filters
7-36
Each of the gas inlets has a sinter filter to protect the pneumatic assembly for
foreign material. If the filter's resistance is increased due to debris, the system will
alarm with a message during system test to clean the filter element. To replace the
filter:
1. Remove the gas inlet connector according to section 7.9.7.
2. Remove and replace the gas filter from the gas connector.
Note
Note that the filter for AIR and O2 are identical.
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a