Quick Links

MEASURING CABLE REPLACEMENT
FOR MTA SERIES
For Models
MTA-3, MTA-5
Celesco Transducer Products 7800 Deering Avenue, Canoga Park, CA 91309 Tel: 800-884-
6860, Fax 818-34-1175
Created 2/22/99 C:MTA
Page 1 of 4
loading

Summary of Contents for celesco MTA Series

  • Page 1 MEASURING CABLE REPLACEMENT FOR MTA SERIES For Models MTA-3, MTA-5 Celesco Transducer Products 7800 Deering Avenue, Canoga Park, CA 91309 Tel: 800-884- 6860, Fax 818-34-1175 Created 2/22/99 C:MTA Page 1 of 4...
  • Page 2 MEASURING CABLE REPLACEMENT If measurement cable is still intact, cut cable end off and discard. Mark the end plate pot side (side with electrical wires) relative to the housing with a pen/pencil. A simple line starting on the end bracket and continuing to the edge of the housing will do.
  • Page 3 Celesco crimping pliers part number 9610035-0000. Squeeze the pliers until the crimping pliers release to insure a strong crimp. Secure the crimp to the spool using Loctite 401, with the flat portion of the crimp adhered to the spool.
  • Page 4 Remove all set screws one at a time, apply Loctite primer 7649 and loctite adhesive 222 on the threads, and replace. Page 4 of 4...

This manual is also suitable for:

Mta-3Mta-5