Instructions for Use Operating / Service Manual IMPORTANT: Read all instructions before assembling or using this device. Employee-Owned...
Page 2
JEDMED is committed to delivering reliable and cost-effective solutions to medical professionals everywhere. The employee-owners of JEDMED would like to take this opportunity to thank you for all your support. We are pleased and proud to serve the ever-changing medical community.
Classic CSC Operating / Service Manual PRECAUTIONS The following conventions are used throughout this manual: WARNING Alerts user to the possibility of PERSONAL INJURY. CAUTION Alerts user to the possibility of DAMAGE to the EQUIPMENT. Ø NOTE Alerts user to PERTINENT FACTS or CONDITIONS. Ø...
Operating / Service Manual DESCRIPTION The JEDMED Classic CSC ENT Cabinet is a cost-effective workstation for the otolaryngic specialist. Its proven design has remained the industry standard for decades. The contemporary look of the Classic CSC will enhance the appearance of any exam room.
Classic CSC Operating / Service Manual PARTS IDENTIFICATION Main Cabinet (right side) Bottle shelf Work surface Tubing hanger Pressure connection Suction tubing Pressure tubing Tubing hanger Main Cabinet (left side) Instrument drawer Storage drawer Pump compartment & trap canister Control unit Cord cover Rear View...
Page 7
Classic CSC Operating / Service Manual PARTS IDENTIFICATION Control Unit (top view) Power input Main power switch Power indicator Auxiliary indicator Pressure indicator Suction indicator Control Unit (bottom view) Unswitched outlet Switched 1 outlet Switched 2 outlet Auxiliary outlet Pressure outlet Suction outlet Remote power strip connection Footswitch connection...
Page 8
Classic CSC Operating / Service Manual PARTS IDENTIFICATION Pump Compartment Components Pressure pump Suction pump Trap canister Pump compartment located in base of cabinet Suction Trap System Trap canister & lid Patient tubing Suction tubing (to pump) Hydrophobic filter Elbow fitting Trap canister configuration Suction Pump Detail Suction tubing...
Page 9
Classic CSC Operating / Service Manual PARTS IDENTIFICATION Pressure Pump Detail Power switch Power cord Intake filter Pressure tubing Hydrophobic filter (pressure) Pressure pump Securing strap (right side) Foam pad Accessories Nebulizer Powder Blower Atomizer 3” Cotton 4” Gauze #180 #175 #286 Container...
Page 10
Classic CSC Operating / Service Manual PARTS IDENTIFICATION - Options Standing Riser (10”) with drawer Ø NOTE Not available on stainless steel version. Riser Drawer Cabinet on Riser Flip-Up Shelf Ø NOTE Not available on stainless steel version. Retracted Extended Remote Footswitch Footswitch Option Plugs into rear of...
Page 11
Automatic Switching Tubing Hanger (pressure & suction) Pressure or suction pumps start automatically when tubing is lifted out of hanger bracket. Automatic switching option JEDMED K2 Recharger ® Otoscope Recharger Unit (Welch Allyn ® Works with traditional Welch Allyn 2.5V Rechargeable handle unit for or 3.5V rechargeable handles.
Classic CSC Operating / Service Manual ASSEMBLY INSTRUCTIONS INSTALL TRAP CANISTER 1. Place lid on canister and press to seal securely. 2. Slip canister with lid into metal support ring. 3. Attach tapered elbow fitting (on end of pump tubing) to top of canister.
Classic CSC Operating / Service Manual DEVICE OPERATION Controls Warning Possible explosion hazard if used in the presence of flammable anesthetics. Top view 1. Power Switch (main) Shuts off all AC power to the cabinet except “unswitched” receptacle on bottom of control box. "Unswitched" receptacle may be used for rechargeable otoscope unit or other device which requires power even if control box is shut off.
Page 14
Classic CSC Operating / Service Manual DEVICE OPERATION SPRAYING (Pressure) pressure handle spray bottle Begin spraying Connect bottle (Partially cover bleeder Turn on pressure hole with thumb) (Tapered Luer fitting) Warning Always exercise care when introducing pressurized medications onto the site. Use only enough pressure to accomplish the task.
Page 15
Classic CSC Operating / Service Manual DEVICE OPERATION REPLACEMENT OF TRAP CANISTER & PATIENT TUBING Disconnect Pump Tubing Securely Cap Pump Port Disinfect Patient Tubing on Canister Securely Cap Patient Tubing Remove & Discard Properly Carefully Pass Tubing Through Cabinet Trap canister and suction tubing are single-use, disposable items.
Classic CSC Operating / Service Manual CLEANING - DISINFECTION - MAINTENANCE WARNING Always unplug unit from main power outlet before cleaning, disinfecting, maintenance or service. Cleaning System may be wiped clean using a soft cloth moistened with water and a mild detergent. Do not use excessive liquids or allow liquids to enter control box, pumps, cable connections or other components.
Page 17
Classic CSC Operating / Service Manual FILTERS About Hydrophobic Filters Filter Wetting - A filter can become “wetted” if the filter Both the pressure and suction pumps are fitted with in-line membrane comes in contact with alcohol or any other hydrophobic filters.
Refer to the assembly drawings located at the back of this manual for replacement parts. Contact Information For additional information or to order replacement parts contact your Regional Sales Representative or JEDMED Technical Service. Refer to inside of back cover.
Page 19
Classic CSC Operating / Service Manual Blank...
Classic CSC Operating / Service Manual SPECIFICATIONS Overall Cabinet Dimensions See below Weight In use Shipping Laminate version 175 lbs. (80 kg) 200 lbs. (91 kg) Stainless steel version 190 lbs. (87 kg) 215 lbs. (98 kg) Control Box Input 115Vac, 60Hz Output 115Vac (Controls up to 10 Amps)
Page 24
Contact JEDMED for further information. WARNING Always unplug unit from main power outlet before cleaning, disinfecting, maintenance or service. Contact Information For additional information or to order replacement parts contact your Regional Sales Representative or JEDMED Technical Service. Refer to inside of back cover.
Corresponding pressure or suction indicator does illuminate. Blown fuse (corresponding pump). Replace fuse (Contact JEDMED) Pumps plugged into incorrect AC outlet Plug each pump into proper outlet. Pressure pump operates when on bottom of control box.