Thermo Scientific Heraeus Cryofuge 5500i Instruction Manual 20057781-e • 07 / 2020 Visit us online to register your warranty: thermofisher.com/labwarranty...
Page 2
WEEE Conformity This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked by the following symbol:...
Rotor Installation ..................... 4-3 Entering Parameters ....................4-4 Centrifugation ......................4-8 Temperature Adaptation during Standstill ............... 4-9 Short-term Centrifugation ..................4-9 Removing the Rotor ....................4-9 Aligning the Centrifuge ..................4-10 Audible Alarm ....................... 4-10 Key Switch ......................4-11 Thermo Scientific Cryofuge 5500i...
Page 4
Disinfection ......................5-3 Decontamination ..................... 5-4 Service of Thermo Fisher Scientific ................5-5 Chapter 6 Troubleshooting ........................6-1 Mechanical Emergency Door Release ..............6-2 Troubleshooting by User ..................6-3 When to contact a Service Technician ..............6-5 Cryofuge 5500i Thermo Scientific...
CAUTION means that material damage could occur. WARNING means that injuries or material damage or contamination could occur. This symbol refers to biological hazards. Observe the information contained in the instruction manual to keep yourself and your environment safe. Thermo Scientific Cryofuge 5500i...
Maximum sample density at maximum speed: 1.2 ---------- - Accident Prevention Prerequisite for the safe operation of the Heraeus Cryofuge 5500i is a work environment in compliance with standards, directives and trade association safety regulations and proper instruction of the user.
Page 7
• Rotor installation: Make sure the rotor is locked properly into place before operating the centrifuge. • Especially when working with corrosive samples (salt solutions, acids, bases), the accessory parts and vessel have to be cleaned carefully. • Always balance the samples. Thermo Scientific Cryofuge 5500i...
Page 8
• Highly corrosive substances which can cause material damage and impair the mechanical stability of the rotor, should only be centrifuged in corresponding protective tubes. WARNING If a hazardous situation occurs, turn off the power supply to the centrifuge and leave the area immediately. Cryofuge 5500i Thermo Scientific...
• The integrated rotor detection systems ensure that no inadmissible speed settings can be preselected. • Electronic imbalance detection, SMARTspin . • Door emergency release: For emergencies only, e.g. during power failures (see “Mechanical Emergency Door Release” page 6-2). Cryofuge 5500i Thermo Scientific...
-9 °C to +40 °C Dimensions Height 960 mm 960 mm 960 mm Width 680 mm 680 mm 680 mm Depth 720 mm 720 mm 720 mm Weight without Rotor 286 kg 286 kg 286 kg Thermo Scientific Cryofuge 5500i...
RCF-Selection, Temperature Regulation during Standstill, Pre-Temp, and PULSE Acceleration / Braking Profiles 9 Acceleration and 9 Braking Curves Rotor Recognition Automatic Imbalance Detection SMARTspin Electronic, contingent on Rotor and Speed Door Lock Motor assisted Door Looking Cryofuge 5500i Thermo Scientific...
14.5 / 16 A 75004479 200 V 2.7 kW 16 AT 20 AT 60 Hz 14.5 A 75004479 208 V 2.7 kW 16 AT 20 AT 60 Hz 75004479 230 V 14.5 A 2.7 kW 16 AT 20 AT Thermo Scientific Cryofuge 5500i...
1.43 t 75004479 Contains fluorinated greenhouse gases in a hermetically sealed system. Rotor Selection The Cryofuge 5500i is supplied without a rotor. Various Thermo Scientific rotors are available to choose from. LH-4000 75006475 with round Buckets 1000 75006477 Double rectangular Bucket DoubleSpin...
1. Place the pallet with the centrifuge so that you have atleast 2 m room in front of the pallet. 2. Release the rails from the pallet. 3. Screw the rails onto the pallet in front of the centrifuge. Screws Rails Cryofuge 5500i Thermo Scientific...
Page 17
9. Screw the threaded nuts down. 10. Place the locking stabilizers centrally under the threaded rods and align the centrifuge. All the locking stabilizers should carry uniformly (check by a tilt test). The casters are some 5-15 mm off the ground. Thermo Scientific Cryofuge 5500i...
People and hazardous substances must be kept out of the safety zone while centrifuging. • The supporting structure must be stable and free of resonance. • The supporting structure must be suitable for horizontal setup of the centrifuge. Cryofuge 5500i Thermo Scientific...
3. Check whether the cable complies with the safety standards of your country. 4. Make sure that the voltage and frequency correspond to the figures on the rating plate. 5. Establish the connection to the power supply with the connecting cable. Thermo Scientific Cryofuge 5500i...
• Before storing the centrifuge and the accessories it must be cleaned and if necessary, disinfected and decontaminated. • Store the centrifuge in a clean, dust-free location. • Be sure to place the centrifuge on its feet. • Avoid direct sunlight. Cryofuge 5500i Thermo Scientific...
By pressing the key briefly, you increase or decrease the speed in one step. When pressing both arrows the cursor moves to the left. Cryofuge 5500i Thermo Scientific...
WARNING Do not reach into the crack between the door and the housing. The door is drawn shut automatically. Use the emergency release only for malfunctions and power failures (see “Mechanical Emergency Door Release” page 6-2). CAUTION Do not slam the door. Cryofuge 5500i Thermo Scientific...
6. If available close the rotor with the rotor lid. Be sure to check all sealings before starting any aerosol-tight applications. See the information in the rotor instruction manual. 7. Close the centrifuge door. Thermo Scientific Cryofuge 5500i...
10 rpm. The RCF can be set in steps of 1. 4. The value is stored when the value stops flashing. Note If an extremely low RCF value has been selected, it will be corrected automatically, if the resulting speed is less than 300 rpm. Cryofuge 5500i Thermo Scientific...
Page 27
Run Time Preselection 1. Press the key below the field running time. 2. Enter the desired value by pressing up or down. Up to 9 minutes 59 seconds the duration is given exactly to the second. Thermo Scientific Cryofuge 5500i...
Page 28
1. Press the key below the field temperature. 2. Enter the desired value by pressing up or down. The temperature can be set in steps of 1. The value is stored when the value stops flashing. Cryofuge 5500i Thermo Scientific...
Page 29
The value that is being changed flashes. 3. Press the key again when all displays are flashing again. The display with the number of the program memory flashes and all settings are stored. Thermo Scientific Cryofuge 5500i...
The display shows the following message: The run will terminate. Check the loading and start the centrifuge once again. See the information on proper loading in the rotor instruction manual. “Troubleshooting by User” page 6-3). For information on troubleshooting, see section Cryofuge 5500i Thermo Scientific...
Note The centrifuge accelerates to maximum speed, regardless of which rotor was installed. Check carefully whether you have to maintain a certain speed for your application. During the acceleration process, time is counted forwards in seconds. The reading stays displayed until the centrifuge door is opened. Thermo Scientific Cryofuge 5500i...
Note If the message rotor appears in the display, the selected speed or RCF value are too high. If you press the key the centrifuge accelerates to the shown maximum value. 4-10 Cryofuge 5500i Thermo Scientific...
All set parameters are blocked, but the actual program can be operated as often as needed with the control keys ( All set values can be seen with the arrow keys. The selected program cannot be changed. PULSE and the pre-temp function are locked. Thermo Scientific Cryofuge 5500i 4-11...
Clean centrifuge and accessories as follow: 1. Open the centrifuge. 2. Turn off the centrifuge. 3. Pull out the power supply plug. 4. Release the rotor. 5. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle. Cryofuge 5500i Thermo Scientific...
Keep in mind the risk of infection when touching the rotor and take all necessary precautions. In case of contamination, make sure that others are not put at risk. Decontaminate the affected parts immediately. Take other precautions if need be. Thermo Scientific Cryofuge 5500i...
For general radioactive decontamination use a solution of equal parts of 70 % ethanol, 10 % SDS and water. 1. Open the centrifuge. 2. Turn off the centrifuge. 3. Pull out the power supply plug. Cryofuge 5500i Thermo Scientific...
Thermo Fisher Scientific offers inspection and service contracts for this work. Any necessary repairs are performed for free during the warranty period and afterwards for a charge. This is only valid, if the centrifuge has only been maintained by a Thermo Fisher Scientific service technician. Thermo Scientific Cryofuge 5500i...
4. Push the release cords back into the centrifuge and mount the plug. 5. Reconnect the centrifuge once the power has been restored. Switch on the centrifuge. Press the key to have the door locks operative again. Cryofuge 5500i Thermo Scientific...
After 15 sec. the rotor will be decelerated. Always wait until the rotor has come to a stop without braking. rotor By opening and closing the door you reset the message After entering a permissible speed, you can start again. Thermo Scientific Cryofuge 5500i...
Page 44
Restart the centrifuge. If an error message appears again, inform a service technician. E-10 Self-test after switching on the centrifuge. NV-RAM; error in program memory Restart the centrifuge. If an error message appears again, inform a service technician. Cryofuge 5500i Thermo Scientific...
1. Turn on the power switch on the right side of the device. The centrifuge performs a self-check of its software. Subsequently, the following readings will be displayed. The order is: Software version key panel Software version NV-RAM-Version 1 Thermo Scientific Cryofuge 5500i...
Page 46
When to contact a Service Technician NV-RAM-Version 2 Cycle counter Note The values given are only examples. 2. Afterwards the different programs are tested as well. Program test 3. Communicate the software version to the service technician. Cryofuge 5500i Thermo Scientific...
Page 47
U S S M S S U S S S S S S S S S U S S S S S S S S S S Boric Acid Cesium Acetate S S S - S S S S - S S - S S S S S S M S S S Thermo Scientific Cryofuge 5500i...
Page 48
S M U S S S S M S U U U - Hydrofluoric Acid (50 %) U U U U - U U U U S U U U S S M M S U U U - Cryofuge 5500i Thermo Scientific...
Page 49
S S S S - S S - S S S S S S M S S S Lactic Acid (100 %) M S U - S S S M S S - M S M S S - Thermo Scientific Cryofuge 5500i...
Page 50
S S S S S S M S S S Rubidium Chloride Sucrose M S S - S S S S S S S S S S S S S S S S S S S S S S S Cryofuge 5500i Thermo Scientific...
Page 51
S S S S S S S S S S S S S S S S S S S S S S S Citric Acid (10 %) M S S M S S M S S S S S S S S S M S S S S S S S S S S Polyethyleneterephthalate Thermo Scientific Cryofuge 5500i...
Page 52
Performance unknown; suggest testing, using sample to avoid loss of valuable material. Chemical resistance data is included only as a guide to product use. No organized chemical resistance data exists for materials under the stress of centrifugation. When in doubt, we recommend pretesting sample lots. Cryofuge 5500i Thermo Scientific...
Page 53
Explanation of RCF Value ......4-5 Run Time Preselection ........4-5 Features ............1-4 Safety features ..........1-2 Safety zone ............2-4 Select Program ..........4-7 Imbalance Indicator ........4-8 Service Technician ........5-5, 6-5 Short-term Centrifugation .......4-9 Starting Centrifuge Program ......4-8 Thermo Scientific Cryofuge 5500i...
Page 54
Index Stopping the Centrifugation Program ..... 4-9 Storage ............2-6 Switch on Centrifuge ........4-2 Technical Data ..........1-3 Temperature Adaptation .........4-9 Transporting the Centrifuge ......2-4 Trouble Shooting ..........6-1 Cryofuge 5500i Thermo Scientific...