Disassembly
a) For disassembly place the punch on a solid worktop. Unscrew the hex head screw (pos.1)
by turning it clockwise using the supplied screwdriver TXX-0X, and remove.
b) Pull the anterior handle element (pos.2) downward to separate it from the punch body
(pos. 4). To ease the unlocking the slider (Pos. 3) should be pushed toward punch foot at
the same time.
c) Now the slider (pos.3) can be easily removed from the punch body.
d) Always clean the four components punch body, slider, anterior handle element and screw
in the same container.
2
Assembly (with the aid of the fitting device TXW-7X)
M 11
ASSEMBLY INSTRUCTIONS
1-01/10
2
5
3
1
4
6
1
loading

Summary of Contents for Fehling Instruments CERAMO CONCEPT X PUNCHES

  • Page 1 FEHLING M 11 INSTRUMENTS ASSEMBLY INSTRUCTIONS 1-01/10 ® CERAMO CONCEPT X PUNCHES Disassembly a) For disassembly place the punch on a solid worktop. Unscrew the hex head screw (pos.1) by turning it clockwise using the supplied screwdriver TXX-0X, and remove. b) Pull the anterior handle element (pos.2) downward to separate it from the punch body (pos.
  • Page 2 FEHLING M 11 INSTRUMENTS ASSEMBLY INSTRUCTIONS 1-01/10 Only components belonging to the same unit may be assembled, with the exception of the screw which fits all punches. Component compatibility can be verified by means of the identi- fication number (pos.5) that is marked on all components (except the screw). NOTE: Before assembling the contact areas (pos.
  • Page 3 FEHLING M 11 INSTRUMENTS ASSEMBLY INSTRUCTIONS 1-01/10 The screwdriver TXX-0X and the fitting device TXW-6X can be cleaned with warm water and a mild detergent. Hot water and steam sterilisation at 134°C can destroy screwdriver and fitting device. FEHLING Hanauer Landstr. 7A · 63791 Karlstein/Germany · www.fehling-instruments.de ...