YOKOGAWA ST401G User Manual

YOKOGAWA ST401G User Manual

Sampling system
Hide thumbs Also See for ST401G:

Quick Links

User's
Manual
Model ST401G
Sampling System
IM 12A0V2-E
IM 12A0V2-E
8th Edition
Table of Contents
loading

Summary of Contents for YOKOGAWA ST401G

  • Page 1 User’s Manual Model ST401G Sampling System IM 12A0V2-E IM 12A0V2-E 8th Edition...
  • Page 2: Introduction

    The product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor responsibility to any person or entity with respect to any direct or indirect loss or damage arising from using the product or any defect of the product that YOKOGAWA can not predict in advance. Media No. IM 12A0V2-E 8th Edition :Jul.
  • Page 3 Notes on Hardware Check the following when you receive the product: Appearance and Accessories Contact our sales representative or your local distributor if the product’s coating has come off or it has been damaged. l Model and Suffix Codes The name plate on the product contains the model and suffix codes. Compare them with those in the general specification to make sure the product is the correct one.
  • Page 4: After-Sales Warranty

    If we replace the product with a new one, we won’t provide you with a repair report. l Yokogawa warrants the product for the period stated in the pre-purchase quotation Yokogawa shall conduct defined warranty service based on its standard. When the customer site is located outside of the service area, a fee for dispatching the maintenance engineer will be charged to the customer.
  • Page 5 Blank Page...
  • Page 6: Table Of Contents

    Model ST401G Sampling System IM 12A0V2-E 8th Edition CONTENTS INTRODUCTION .......................i After-sales Warranty ...................iii OVERVIEW ....................1-1 1.1 System configuration ..................1-2 1.2 ST401G-FC4 ....................... 1-5 ST401G-PH6 ...................... 1-5 1.4 ST401G-PF6 ....................... 1-5 1.5 ST401G-FC5 ....................... 1-5 ST401G-PH7 ...................... 1-6 1.7 ST401G-PF7 ....................... 1-6 1.8 ST401G-PF8 .......................
  • Page 7 Blank Page...
  • Page 8: Overview

    < 1. Overview > OVERVIEW Model ST401G is a compact sampling system which can be used with the pH converter system and the model FC400G non-reagent type free available chlorine analyzer. The measurement can be started after only simple installation because the ST401G is provided with a pressure-reducing valve and ball valve, and shipped with customer ordered devices.
  • Page 9: System Configuration

    Model ST401G refers only to a sampling system and does not include a free available chlorine analyzer and/or a pH meter(sensor,holder,converter and ultrasonic oscillator). (Instrument should be purchased separately.) Note that the ST401G Sampling System accepts the following models of note 1 or note 2 only, and depending on system configuration. Note 1: Non-reagent type free available chlorine analyzer (refer to GS 12F5A1-E) When ordering, option code “/ST”...
  • Page 10 < 1. Overview > Dimensions, wiring and piping diagram for the ST401G are shown on following pages. For figure numbers to be referred corresponding each models and codes are shown in the Table 1.2 below. Table 1.2 uses following signs.
  • Page 11 < 1. Overview > Model and codes Dimensions Wiring Piping Figure 3.27 Figure 3.57 Figure 4.4 ST401G-PF7-¨-A ST401G-PF7-¨-A /TT3 Figure 3.28 Yes Figure 3.29 ST401G-PF7-¨-A /R ST401G-PF7-¨-A /R /TT3 Yes Figure 3.30 ST401G-PF7-¨-A /AZC Figure 3.31 Figure 3.58 Figure 4.5 Figure 3.32 ST401G-PF7-¨-A /AZC /TT3 ST401G-PF7-¨-A /R /AZC Yes Figure 3.33 Yes Figure 3.34...
  • Page 12: St401G-Fc4

    12F5A1-01E “Model FC400G Free Available Chlorine Analyzer (Non-reagent type)” ST401G-PH6 ST401G-PH6 is composed of the FLXA402, PH8EFP pH sensor, PH8HF flow-through type pH holder, and sampling system. Moreover, PUS400G ultrasonic cleaner can be built in (optional). Refer to Figure 3.9 to Figure 3.12 for the configuration.
  • Page 13: St401G-Ph7

    < 1. Overview > ST401G-PH7 ST401G-PH7 is composed of the FLXA402, PH8EFP pH sensor, PH8HF flow-through type pH holder, and sampling system. Moreover, PG400 pulse generator for clean unit can be built in (optional). Refer to Figure 3.23 to Figure 3.26 for the configuration.
  • Page 14: Specifications

    Cleaning system: Chlorine analyzer: Not provided pH analyzer: Ultrasonic cleaning Sample (process) conditions: Pressure: 100 to 750 kPa (Under 100 kPa, consult Yokogawa) Flow rate: Chlorine analyzer: 0.1 to 10 L/min pH analyzer: 3 to 10 L/min Temperature: 0 to 50 °C when using FC800D -F1 or -C1 0 to 40 °C when using FC800D -F2 or -F3...
  • Page 15 < 2. Specifications > When option code “/RR” specified, two measurement range set in the FC400G can be switched remotely. Remote range switching contact input: Applied voltage when contact input is opened: 5 V DC (Range 1) Applied current when contact input is closed: 20 μA (Range 2) Remote range switching contact output: Contact type: dry contact output...
  • Page 16: Installation, Piping, And Wiring

    < 3. INSTALLATION, PIPING AND WIRING > INSTALLATION, PIPING, AND WIRING Unpacking The ST401G sampling system is sufficiently packed for shipment to prevent damage during transportation. Unpack carefully near the place of installation. Installation place Install the ST401G sampling system at a location as below:...
  • Page 17 (Cable O.D. 9 to 12 mm) Sample inlet Drain outlet VP40 pipe VP16 pipe PR1 : Pressure reducing valve (Note) Option code: /ARS (with arrester) NV1 : Needle valve BV1 : Ball valve Weight : Approx. 35kg Figure 3.1 ST401G-FC4-N-A(/ARS) With free available chlorine analyzer, (arrester), (bottom piping) IM 12A0V2-E...
  • Page 18 (Cable O.D. 9 to 12 mm) Drain outlet VP40 pipe Sample inlet VP16 pipe PR1 : Pressure reducing valve (Note) Option code: /ARS (with arrester) NV1 : Needle valve BV1 : Ball valve Weight : Approx. 35kg Figure 3.2 ST401G-FC4-N-A/R(/ARS) With free available chlorine analyzer, (arrester), (back piping) IM 12A0V2-E...
  • Page 19 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space 4 - Ø15 holes Approx. 200 Approx. 200 For anchor bolts Maintenance View A Front Space Free available chlorine converter FC400G Free available chlorine detector FC400G Terminal box (option) (Note) Power line/output signal (Cable O.D.
  • Page 20 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space 4 - Ø15 holes Approx. 200 Approx. 200 For anchor bolts Maintenance Front View A Space Free available chlorine converter FC400G Free available chlorine detector FC400G Terminal box (option) (Note) Power line/output signal (Cable O.D.
  • Page 21 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance 4 - Ø15 holes Space Space Approx. 200 For anchor Approx. 200 bolts Maintenance View A Front Space Free available chlorine converter FC400G Termimal box for Free available "with arrester" and chlorine detector "remote range setting"...
  • Page 22 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space Approx. 200 Approx. 200 4 - Ø15 holes For anchor bolts Maintenance Front Space View A Free available chlorine converter FC400G Free available Termimal box for chlorine detector "with arrester"...
  • Page 23 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space 4 - Ø15 holes Approx. 200 Approx. 200 For anchor bolts Maintenance View A Front Space Free available chlorine converter FC400G Free available Termimal box for chlorine detector "with arrester"...
  • Page 24 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space 4 - Ø15 holes Approx. 200 Approx. 200 For anchor bolts Maintenance Front Space View A Free available chlorine converter FC400G Free available Termimal box for chlorine detector "with arrester"...
  • Page 25 VP40 pipe VP16 pipe (Note1) Suffix code: -U (with ultrasonic oscillator) PR1: Pressure reducing valve (Note2) Suffix code: -U (with ultrasonic oscillator) or BV1: Ball valve Option code: /ARS (with arrester) Weight: Approx. 35 kg Figure 3.9 ST401G-PH6-¨-A (/ARS) With pH converter, (ultrasonic oscillator), (arrester), (bottom piping) IM 12A0V2-E...
  • Page 26 VP40 pipe VP16 pipe (Note1) Suffix code: -U (with ultrasonic oscillator) PR1: Pressure reducing valve (Note2) Suffix code: -U (with ultrasonic oscillator) or BV1: Ball valve Option code: /ARS (with arrester) Weight: Approx. 35 kg Figure 3.10 ST401G-PH6-¨-A/TT3(/ARS) With pH converter, (ultrasonic oscillator), 500 ml KCl tank, (arrester), (bottom piping) IM 12A0V2-E...
  • Page 27 VP16 pipe PR1: Pressure reducing valve (Note 1) Suffix code: -U (with ultrasonic oscillator) BV1: Ball valve (Note 2) Suffix code: -U (with ultrasonic oscillator) or Option code: /ARS (with arrester) Weight: Approx. 35 kg Figure 3.11 ST401G-PH6-¨-A/R (/ARS) With pH converter, (ultrasonic oscillator), (arrester), (back piping) IM 12A0V2-E...
  • Page 28 VP16 pipe (Note 1) Suffix code: -U (with ultrasonic oscillator) (Note 2) Suffix code: -U (with ultrasonic oscillator) or PR1: Pressure reducing valve BV1: Ball valve Option code: /ARS (with arrester) Weight: Approx. 35 kg Figure 3.12 ST401G-PH6-¨-A/R/TT3 (/ARS) With pH converter, (ultrasonic oscillator), 500 ml KCl tank, (arrester), (back piping) IM 12A0V2-E...
  • Page 29 3-14 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space Approx. 200 Approx. 200 4-Ø15 holes For anchor bolts Maintenance Space Front (20) View A Free available pH ultrasonic oscillator chlorine converter pH converter PUS400G (Note) FC400G FLXA402 KCl tank...
  • Page 30 3-15 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space Approx. 200 Approx. 200 4 - Ø15 holes For anchor bolts Maintenance (20) Front Space View A pH converter Free available chlorine FLXA402 pH ultrasonic oscillator converter PUS400G (Note) FC400G...
  • Page 31 3-16 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space Approx. 200 Approx. 200 4-Ø15 holes For anchor bolts (20) Maintenance Front Space View A Free available chlorine pH ultrasonic oscillator converter pH converter PUS400G (Note) FC400G FLXA402 KCl tank...
  • Page 32 3-17 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space Approx. 200 Approx. 200 4-Ø15 holes For anchor bolts (20) Maintenance Front Space View A Free available pH converter chlorine FLXA402 pH ultrasonic oscillator converter PUS400G (Note) FC400G KCl tank Terminal box...
  • Page 33 3-18 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space 4-Ø15 holes Approx. 200 Approx. 200 For anchor bolts (20) Maintenance Front Space View A Free available chlorine pH ultrasonic oscillator converter pH converter PUS400G (Note) FC400G FLXA402 KCl tank...
  • Page 34 3-19 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space Approx. 200 Approx. 200 4-Ø15 holes For anchor bolts (20) Maintenance Front Space View A Free available chlorine pH converter pH ultrasonic oscillator converter FLXA402 PUS400G (Note) FC400G KCl tank Terminal box...
  • Page 35 3-20 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance 4-Ø15 holes Space Space For anchor bolts Approx. 200 Approx. 200 (20) Maintenance Front Space View A Free available chlorine pH ultrasonic oscillator converter pH converter PUS400G (Note) FC400G FLXA402 KCl tank...
  • Page 36 3-21 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space Approx. 200 Approx. 200 4-Ø15 holes For anchor bolts (20) Maintenance Front Space View A Free available chlorine pH converter pH ultrasonic oscillator converter FLXA402 PUS400G (Note) FC400G KCl tank 500 mL...
  • Page 37 BV1 to BV3 : Ball valve cable mating connection Ø6.5 to Ø12.5 ※ When only /AZC is specified, the box shown above is attached to the dotted area. Weight : Approx. 40 kg Figure 3.21 S T401G-FC5-N-A (/ARS) With free available chlorine analyzer, (arrester) ST401G-FC5-N-A /AZC With free available chlorine analyzer, auto zero calibration ST401G-FC5-N-A /AZC /ARS With free available chlorine analyzer, auto zero calibration, arrester IM 12A0V2-E...
  • Page 38 PR1 : Pressure reducing valve NV1 : Needle valve BV1 to BV3 : Ball valve ※ When only /AZC is specified, the box shown above is attached to the dotted area. 40kg Weight : Approx. Figure 3.22 S T401G-FC5-N-A /R (/ARS) With free available chlorine analyzer, back piping, (arrester) ST401G-FC5-N-A /R /AZC back piping, auto zero calibration ST401G-FC5-N-A /R /AZC /ARS back piping, arrester, auto zero calibration IM 12A0V2-E...
  • Page 39 -P is specified. pH holder PH8HF Terminal box cable mating connection Ø6.5 to Ø12.5 Drain outlet Sample inlet VP40 pipe VP16 pipe PR1 : Pressure reducing valve BV1 : Ball valve Weight : Approx. 35 kg Figure 3.23 S T401G-PH7-N-A (/ARS) With free available chlorine analyzer, (arrester) ST401G-PH7-P-A With free available chlorine analyzer, ultrasonic oscillator IM 12A0V2-E...
  • Page 40 Terminal box PH8HF cable mating connection Ø6.5 to Ø12.5 Sample inlet Drain outlet VP16 pipe VP40 pipe PR1 : Pressure reducing valve BV1 : Ball valve Weight : Approx. 35 kg Figure 3.24 S T401G-PH7-N-A /TT3 (/ARS) With free available chlorine analyzer, 500 ml KCl tank, (arrester) ST401G-PH7-P-A /TT3 With free available chlorine analyzer, ultrasonic oscillator, 500 ml KCl tank IM 12A0V2-E...
  • Page 41 Terminal box pH holder cable mating PH8HF connection Ø6.5 to Ø12.5 Drain outlet VP40 pipe Sample inlet VP16 pipe PR1 : Pressure reducing valve BV1 : Ball valve Weight : Approx. 35 kg Figure 3.25 S T401G-PH7-N-A /R (/ARS) With free available chlorine analyzer, back piping, (arrester) ST401G-PH7-P-A /R With free available chlorine analyzer, ultrasonic oscillator, back piping IM 12A0V2-E...
  • Page 42 Terminal box PH8HF cable mating connection Ø6.5 to Ø12.5 Drain outlet VP40 pipe Sample inlet VP16 pipe Weight : Approx. 35 kg PR1 : Pressure reducing valve BV1 : Ball valve Figure 3.26 S T401G-PH7-N-A /R /TT3 (/ARS) With free available chlorine analyzer, back piping, 500 ml KCl tank, (arrester) ST401G-PH7-P-A /R /TT3 With free available chlorine analyzer, ultrasonic oscillator, back piping, 500 ml KCl tank IM 12A0V2-E...
  • Page 43 PH8EFP Ø6.5 to Ø12.5 pH holder PH8HF Sample inlet VP16 pipe Drain outlet VP40 pipe PR1 : Pressure reducing valve NV1 : Needle valve BV1, BV2 : Ball valve Weight : Approx. 45 kg Figure 3.27 ST401G-PF7- -A With free available chlorine analyzer, (ultrasonic oscillator) ¨ IM 12A0V2-E...
  • Page 44 3-29 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance space space Approx. 200 Approx. 200 4 - Ø15 holes For anchor bolts Maintenance (20) Front space View A Liquid Analyzer for *1: Attached when -P Turbidity and Chlorine pH converter is specified.
  • Page 45 3-30 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance space space 4 - Ø15 holes Approx. 200 Approx. 200 For anchor bolts (20) Maintenance Front space View A Liquid Analyzer for KCI tank *1: Attached when -P Turbidity and Chlorine pH converter is specified.
  • Page 46 3-31 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance space space Approx. 200 Approx. 200 4 - Ø15 holes For anchor bolts Maintenance Front (20) space View A Liquid Analyzer for pH converter Turbidity and Chlorine *1: Attached when -P 500 mL FLXA402 FLXA402T...
  • Page 47 3-32 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance 4 - Ø15 holes space space For anchor bolts Approx. 200 Approx. 200 (20) Maintenance Front space View A Terminal box Liquid Analyzer for pH converter cable mating connection Turbidity and Chlorine KCI tank FLXA402...
  • Page 48 3-33 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance space space Approx. 200 Approx. 200 4 - Ø15 holes For anchor bolts (20) Maintenance Front space View A Terminal box Liquid Analyzer for cable mating connection pH converter KCI tank Turbidity and Chlorine Ø6.5 to Ø12.5...
  • Page 49 3-34 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance 4 - Ø15 holes Maintenance space space For anchor bolts Approx. 200 Approx. 200 (20) Maintenance Front space View A Terminal box Liquid Analyzer for KCI tank pH converter cable mating connection Turbidity and Chlorine FLXA402...
  • Page 50 3-35 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance space space Approx. 200 Approx. 200 4 - Ø15 holes For anchor bolts (20) Maintenance Front space View A Terminal box Liquid Analyzer for KCI tank cable mating connection pH converter Turbidity and Chlorine Ø6.5 to Ø12.5...
  • Page 51 3-36 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space 4 - Ø15 holes Approx. 200 Approx. 200 For anchor bolts Maintenance (20) Front Space View A Liquid Analyzer for *1: Attached when -P KCI tank Turbidity and Chlorine is specified.
  • Page 52 3-37 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space 4 - Ø15 holes Approx. 200 Approx. 200 For anchor bolts Maintenance Front (20) Space View A Liquid Analyzer for Turbidity and Chlorine *1: Attached when -P KCI tank FLXA402T is specified.
  • Page 53 3-38 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space Approx. 200 Approx. 200 4 - Ø15 holes For anchor bolts Maintenance Front Space (20) View A *1: Attached when -P KCI tank Liquid Analyzer for is specified.
  • Page 54 3-39 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space Approx. 200 Approx. 200 4 - Ø15 holes For anchor bolts Maintenance Front Space (20) View A Liquid Analyzer for Turbidity and Chlorine 500 mL *1: Attached when -P FLXA402T is specified.
  • Page 55 3-40 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space 4 - Ø15 holes Approx. 200 Approx. 200 For anchor bolts Maintenance Front Space (20) View A Terminal box Liquid Analyzer for KCI tank cable mating connection Turbidity and Chlorine Ø6.5 to Ø12.5 FLXA402T...
  • Page 56 3-41 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space 4 - Ø15 holes Approx. 200 Approx. 200 For anchor bolts Maintenance Front Space (20) View A Terminal box Liquid Analyzer for cable mating connection Turbidity and Chlorine KCI tank Ø6.5 to Ø12.5 FLXA402T...
  • Page 57 3-42 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space 4 - Ø15 holes For anchor bolts Approx. 200 Approx. 200 Maintenance Front (20) Space View A Terminal box Liquid Analyzer for KCI tank cable mating connection Turbidity and Chlorine Ø6.5 to Ø12.5 FLXA402T...
  • Page 58 3-43 < 3. INSTALLATION, PIPING AND WIRING > Unit: mm Maintenance Maintenance Space Space 4 - Ø15 holes Approx. 200 Approx. 200 For anchor bolts Maintenance Front Space (20) View A Liquid Analyzer for Terminal box Turbidity and Chlorine cable mating connection KCI tank FLXA402T Ø6.5 to Ø12.5...
  • Page 59: Piping

    3-44 < 3. INSTALLATION, PIPING AND WIRING > Piping Pressure range of measuring water is from 100 to 750 kPa (1 to 7.5 kg/cm ) and the flow rate is from 0.1 to 10 L/min. Measuring water pipe should be hard PVC and VP16 (22 mm O.D.). Drain pipe should be hard PVC and VP40 (48 mm O.D.).
  • Page 60 *2: Use cable with 9 to 12 mm OD. *3: Use two-core shielded cable with 9 to 12 mm OD. To avoid two-point ground, connect shield to ground at one end only. Figure 3.43 ST401G-FC4-N-A (/AZC¨) With free available chlorine analyzer, (auto zero calibration) Terminal box Analog output Free available...
  • Page 61 3-46 < 3. INSTALLATION, PIPING AND WIRING > Free available Free available chlorine converter chlorine detector Terminal box FC400G FC400G AR2x2 MAINT FAIL (Customer wiring) Dedicated cable Solenoid valve (SV1) Option for "/AZC□" (with auto zero calibration) AR1,2 : arresters CB1: Circuit breaker (Note) * Dotted wiring is external wiring and should be prepared by customer.
  • Page 62 * Use two-core shielded cable with 9 to 12 mm OD for output wiring. CB1: Circuit breaker Avoid two-point ground. Ground the G terminal of power wiring. Power, Ground Figure 3.47 ST401G-PH6-U-A With pH converter (FLXA402), ultrasonic oscillator IM 12A0V2-E...
  • Page 63 CB1: Circuit breaker * Use two-core shielded cable with 9 to 12 mm OD for output wiring. Avoid two-point ground. Ground the G terminal of power wiring. Power Output Figure 3.48 ST401G-PH6-¨-A/ARS With pH converter (FLXA402), (ultrasonic oscillator), arrester IM 12A0V2-E...
  • Page 64 * Use two-core shielded cable with 9 to 12 mm OD for output wiring. CB1: Circuit breaker Avoid two-point ground. Ground the G terminal of power wiring. Power, Ground Figure 3.49 ST401G-PF6-¨-A/AZC¨ With free available chlorine analyzer, (auto zero calibration), pH converter (FLXA402) , (ultrasonic oscillator) IM 12A0V2-E...
  • Page 65 Contact input (22) (-N2, -N4) (32)NC (24) (31)C (33)NO (87) Analog input (88) (-N4) (42)NC (41)C SHLD(89) (43)NO Contact output (-WR) Termination(95) (52)NC (51)C A+(91) (53)NO RS485 (-R) B-(92) (72)NC GND(93) (71)C SHLD(94) (73)NO Ethernet (-E) Figure 3.50 ST401G-FC5-N-A (/R) With free available chlorine analyzer, (back piping) IM 12A0V2-E...
  • Page 66 3-51 < 3. INSTALLATION, PIPING AND WIRING > Terminal box for connecting solenoid valve Fuse L(101) (11) Solenoid N(102) valve (12) Power (SV1) Supply G(001) L(201) N(202) G(002) (301) (302) FC800D Non-reagent Type Free Available Chlorine Sensor Unit FLXA402T Disuse Liquid Analyzer for Turbidity and Chlorine L(101) (201)L...
  • Page 67 (22) (-N2, -N4) (32)NC (24) (31)C (33)NO (87) + Analog input (88) - (-N4) (42)NC (41)C SHLD(89) (43)NO Contact output (-WR) Termination(95) (52)NC (51)C A+(91) (53)NO RS485 (-R) B-(92) (72)NC GND(93) (71)C SHLD(94) (73)NO Ethernet (-E) Figure 3.52 ST401G-FC5-N-A (/R) /ARS With free available chlorine analyzer, (back piping), arrester IM 12A0V2-E...
  • Page 68 3-53 < 3. INSTALLATION, PIPING AND WIRING > Terminal box for connecting solenoid valve L(101) (11) Solenoid valve Power N(102) (12) (SV1) Supply G(001) L(201) (301) N(202) (302) G(002) FC800D Non-reagent Type Free Available Chlorine Sensor Unit FLXA402T Disuse Liquid Analyzer for Turbidity and Chlorine (201)L L(101) (61) +...
  • Page 69 3-54 < 3. INSTALLATION, PIPING AND WIRING > FLXA402 pH converter Analog output Power (61) + Power (-N2, -N4) (62) Supply Analog output (65) + (66) (-N2, -N4) (81) + Analog output (82) (-N4) 1st input Analog output (85) + pH Sensor (86) (-N4)
  • Page 70 3-55 < 3. INSTALLATION, PIPING AND WIRING > FLXA402 Terminal box for power supply pH converter L(101) L(201) Analog output N(102) (61) + Power N(202) Power (-N2, -N4) (62) Supply G(001) G(003) Analog output (65) + (66) (-N2, -N4) (81) + Analog output (82) (-N4)
  • Page 71 (87) (88) (-N4) (42)NC (41)C SHLD(89) (43)NO Contact output (-WR) Termination(95) (52)NC (51)C A+(91) (53)NO S485 (-R) B-(92) (72)NC GND(93) (71)C SHLD(94) (73)NO Etherne t (-E) PG400 Pulse Generator for Clean Unit Power Oscillator Figure 3.57 S T401G-PF7-N-A (/R) (/TT3) ST401G-PF7-P-A (/R) (/TT3) With free available chlorine analyzer, (back piping), (500 ml KCl tank) IM 12A0V2-E...
  • Page 72 (-N4) (42)NC (41)C SHLD(89) (43)NO Contact output (-WR) Termination(95) (52)NC (51)C A+(91) (53)NO B-(92) S485 (-R) (72)NC GND(93) (71)C SHLD(94) (73)NO Etherne t (-E) PG400 Pulse Generator for Clean Unit Power Oscillator Figure 3.58 S T401G-PF7-P-A /AZC (/R) (/TT3) With free available chlorine analyzer, auto zero calibration, (back piping), (500 ml KCl tank) ST401G-PF7-N-A /AZC (/R) (/TT3) With free available chlorine analyzer, ultrasonic oscillator, auto zero calibration, (back piping), (500 ml KCl tank) IM 12A0V2-E...
  • Page 73 (88)- (42)NC (-N4) (41)C SHLD(89) (43)NO Termination(95) (52)NC (51)C A+(91) (53)NO RS485(-R) B-(92) pH Sensor (72)NC GND(93) (71)C SHLD(94) (73)NO Ethernet(-E) Flow-Through type holder Pulse Generator for Clean Unit PG400 Power Oscillator Figure 3.59 S T401G-PF8-N-A (/R) (/TT3) With free available chlorine analyzer and pH analyzer (back piping), (500 ml KCl tank) ST401G-PF8-P-A (/R) (/TT3) With free available chlorine analyzer and pH analyzer, ultrasonic oscillator, (back piping), (500 ml KCl tank) IM 12A0V2-E...
  • Page 74 (41)C SHLD(89) (43)NO Termination(95) (52)NC (51)C A+(91) (53)NO B-(92) RS485(-R) (72)NC GND(93) Contact output (71)C (-WR) SHLD(94) (73)NO pH Sensor Ethernet(-E) Flow-Through type holder PG400 Pulse Generator for Clean Unit Power Oscillator Figure 3.60 S T401G-PF8-N-A /AZC (/R) (/TT3) With free available chlorine analyzer and pH analyzer, auto zero calibration, (back piping), (500 ml KCl tank) ST401G-PF8-P-A /AZC (/R) (/TT3) With free available chlorine analyzer and pH analyzer, ultrasonic oscillator, auto zero calibration, (back piping), (500 ml KCl tank) IM 12A0V2-E...
  • Page 75 Blank Page...
  • Page 76: Operation

    VP16 Pipe VP40 Pipe PR1 : Pressure-reducing valve FCC : Free available chlorine converter FCD : Free available chlorine detector NV1 : Needle valve BV1 : Ball valve Measuring water (VP16) PVC tray Drain (VP40) Figure 4.1 P iping diagram of ST401G-FC4-N-A, ST401G-FC5-N-A (With free available chlorine analyzer) IM 12A0V2-E...
  • Page 77: St401G-Fc4-N-A/Azc¨, St401G-Fc5-N-A/Azc

    : Ball valve : Pressure-reducing valve : Free available chlorine converter : Free available chlorine detector : Needle valve : Activated charcoal filter : 3-way solenoid valve Measuring water (VP16) PVC tray Drain (VP40) Figure 4.2 P iping diagram of ST401G-FC4-N-A/AZC¨, ST401G-FC5-N-A/AZC¨ (With free available chlorine analyzer, auto zero calibration) IM 12A0V2-E...
  • Page 78: St401G-Ph6-¨-A, St401G-Ph7-¨-A

    : Ball valve : pH converter : Flow-through type holder : pH sensor : KCl tank : Ultrasonic oscillator (Suffix code: -U) Measuring water (VP16) BV1 PR1 PVC tray Drain (VP40) Figure 4.3 Piping diagram of ST401G-PH6-¨-A, ST401G-PH7¨-A (With pH converter, (ultrasonic oscillator)) IM 12A0V2-E...
  • Page 79: St401G-Pf6-¨-A, St401G-Pf7-¨-A, St401G-Pf8-¨-A

    BV1, 2 : Ball valve : pH converter : Flow-through type holder : pH sensor : KCl tank : Ultrasonic oscillator (Suffix code: -U) Measuring water (VP16) BV1 PR1 PVC tray Drain (VP40) Figure 4.4 P iping diagram of ST401G-PF6-¨-A, ST401G-PF7-¨-A (With free available chlorine analyzer and pH converter, (ultrasonic oscillator)), ST401G-PF8-¨-A (With free available chlorine analyzer and pH analyzer, (ultrasonic oscillator)) IM 12A0V2-E...
  • Page 80: St401G-Pf6-¨-A/Azc¨, St401G-Pf7-¨-A/Azc¨, St401G-Pf8-¨-A/Azc

    : 3-way solenoid valve : pH converter : Flow-through type holder : pH sensor : KCl tank : Ultrasonic oscillator (Suffix code: -U) Measuring water (VP16) PVC tray Drain (VP40) Figure 4.5 P iping diagram of ST401G-PF6-¨-A/AZC¨, ST401G-PF7-¨-A/AZC¨ (With free available chlorine analyzer and pH converter, (ultrasonic oscillator), auto zero calibration), ST401G-PF8-¨-A/AZC¨ (With free available chlorine analyzer and pH analyzer, (ultrasonic oscillator), auto zero calibration) IM 12A0V2-E...
  • Page 81 Blank Page...
  • Page 82: Maintenance

    < 5. MAINTENANCE > MAINTENANCE 5.1 Confirmation of flow rates of sample Check the flow rate of the sample regularly and adjust into necessary amounts. 5.2 Leakage check Check the leakage from pipes and fittings. At this time, check the appearance of pipes and fittings. Repair or exchange them when any crack or defect is observed. 5.3 Maintenance of each built-in equipment Maintain and inspect pH converter system and free available chlorine analyzer regularly...
  • Page 83 Blank Page...
  • Page 84: Customer Maintenance Parts List

    L9901CK Ø12 / Ø9 P.E. Tube L9831MH Elbow Union L9831JE Half Union L9831NE Elbow Union K9146BB Sensor Stand CMPL 12A00V02-00E All Rights Reserved, Copyright © 2002, Yokogawa Electric Corporation. Subject to change without notice. 5th Edition : May. 2013 (YK)
  • Page 85 When suffix code of "- PH¨" Item Part No. Description K9326VG Regulator K9326HT Ball Valve Assembly L9901CK Ø12 / Ø9 P.E. Tube L9831MH Elbow Union K9146BB Sensor Stand CMPL 12A00V02-00E 5th Edition : May. 2013 (YK)
  • Page 86 When suffix code of "- FC¨" Item Part No. Description L9852CB Needle Valve K9326VG Regulator K9326HT Ball Valve Assembly L9901CA Ø6 / Ø4 P.E. Tube L9831JE Half Union L9831NE Elbow Union L9901CA Ø6 / Ø4 P.E. Tube L9831NE Elbow Union 5th Edition : May.
  • Page 87 When option code of "/AZC" Item Part No. Description ─ Solenoid Valve A1029MV For 100 V AC K9460ET For 110 V AC K9726KR For 200 V AC K9726KS For 220 V AC K9726EG Filter Assembly L9862AY Filter Element CMPL 12A00V02-00E 5th Edition : May.
  • Page 88: Revision Information

    Sec.4 Operation some changed of Auto-zero Calibration; CMPL revised to 3rd edition, some parts added. Dec. 1996/2nd Edition Some page changed. May. 1996/1st Edition Newly published. n If you want to have more information about Yokogawa products, you can visit Yokogawa’s home page at the following web site. Home page: http://www.yokogawa.com/an IM 12A0V2-E...
  • Page 89 Blank Page...