Emerson Rosemount 396P Quick Start Manual
Emerson Rosemount 396P Quick Start Manual

Emerson Rosemount 396P Quick Start Manual

Ph/orp sensors
Show thumbs Also See for Rosemount 396P:
Quick Start Guide
00825-0100-3096, Rev AA
November 2020
Rosemount
396P/396PVP
pH/ORP Sensors
Table of Contents
loading

Summary of Contents for Emerson Rosemount 396P

  • Page 1 Quick Start Guide 00825-0100-3096, Rev AA November 2020 ™ Rosemount 396P/396PVP pH/ORP Sensors...
  • Page 2 • Read all instructions prior to installing, operating, and servicing the product. • If you do not understand any of the instructions, contact your Emerson representative for clarification. • Follow all warnings, cautions, and instructions marked on and supplied with the product.
  • Page 3: Table Of Contents

    November 2020 Quick Start Guide WARNING Physical access Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could be intentional or unintentional and needs to be protected against. Physical security is an important part of any security program and fundamental to protecting your system.
  • Page 4 Quick Start Guide November 2020 Emerson.com/Rosemount...
  • Page 5: Install

    Mount sensor. Mount sensor Emerson has designed the sensor to be located in industrial process environments. Do not exceed temperature and pressure limitations at any time. A caution label regarding this matter is attached to the sensor. Please do not remove the label.
  • Page 6 A. 1-in. (25.4 mm) male national pipe thread (MNPT), two places B. 1-in. (25.4 mm) wrench opening C. Reference junction D. End tip option -41 E. Solution ground F. Electrode G. Sensor cable (or Variopol connector - not shown) Emerson.com/Rosemount...
  • Page 7 1.2.1 Flow through and insertion mounting Rosemount 396P and 396PVP Sensors have a 1-in. (25.4 mm) male national pipe thread (MNPT) process connection at the front of the sensor for mounting into a 1½-in. (38.1 mm) tee or the process pipes.
  • Page 8 B. O-ring must be in place prior to use C. Process connection threads - two places D. 2-in. (50.8 mm) sched. 80 T E. Sensor cable Ordering option (xx) Process connection threads ¾ in. (19.1 mm) 1 in. (25.4 mm) 1½ in. (38.1 mm) Emerson.com/Rosemount...
  • Page 9 November 2020 Quick Start Guide Figure 1-3: Flow-Through and Insertion Installations A. 1½-in. x 1-in. (38.1 mm x 25.4 mm) reducing bushing B. 1½-in. (38.1 mm) pipe tee PN 2002011 C. 1½-in. (38.1 mm) pipe Y D. Flow Quick Start Guide...
  • Page 10 Quick Start Guide November 2020 Figure 1-4: Rosemount 396P with Insertion Mounting Adapter (PN 23242-02) Not for use with Rosemount 396PVP. Mounting adapter allows for sensor removal without twisting or disconnecting interconnecting cable for ease of maintenance. A. Cable or Variopol connector (not shown) B.
  • Page 11 November 2020 Quick Start Guide Figure 1-5: Low Flow Cell (PN 24091-00) A. O-ring PN 9550298 B. Adapter with one national pipe thread (NPT) thread C. ¼-in. (6.4 mm) O.D. tubing outlet D. Supplied hardware: 10-32 x 2.5-in. (63.5 mm) stainless steel (two) 10-32 hex nut, stainless steel (two) #10 flat washer, stainless steel (two) 10-32 cap nut, nylon (two)
  • Page 12 Quick Start Guide November 2020 3. This flow cell assembly is used for pressurized applications. See limitations. Unless otherwise specified. Emerson.com/Rosemount...
  • Page 13 B. Sensor cable C. 1½-in. (38.1 mm) PVC pipe schedule 80 D. 1½-in. (38.1 mm) pipe clamp, three places E. Sensor model Rosemount 396P. Regularly check to make sure connections are water tight. F. Sweep pipe with 1-in. (25.4 mm) FNPT adapter G.
  • Page 14 November 2020 Figure 1-7: Junction Box A. Junction box B. 2-in. (50.8 mm) pipe mounting bracket (PN 2002565) C. Flexible conduit if required D. 1-in. (25.4 mm) pipe by others E. 1-in. (25.4 mm) FNPT CPVC union (PN 9320057) Emerson.com/Rosemount...
  • Page 15 November 2020 Quick Start Guide Figure 1-8: Jet Spray Cleaner (PN 12707-00) A. Cleaning solution by others B. Solenoid valve or manual valve (supplied by others) C. Corrosion resistant tubing (supplied by others) D. Polypropylene ¼-in. (6.4 mm) compression fitting E.
  • Page 16 1.2.2 Submersion mounting Rosemount 396P and 396PVP sensors also have a 1-in. (25.4 mm) male national pipe thread (MNPT) process connection at the back of the sensor. Using a standard 1-in. (25.4 mm) union, you can mount the sensor to a 1-in.
  • Page 17: Wire

    Run signal cable in a dedicated conduit (preferably an earth grounded metallic conduit) and keep it away from AC power lines. For your convenience, Emerson has furnished a wire nut kit (in a plastic bag wrapped around the cable). CAUTION...
  • Page 18 Note Remove electrical tape or shrink sleeve away from gray reference wire before connecting wire to terminal. Note For additional information on this product, including sensor combinations not shown here, please refer to our website Wiring Diagrams. Emerson.com/Rosemount...
  • Page 19 If there is a shrink sleeve on the gray reference wire, remove it before connecting the wire to the terminal. For Rosemount 1056, terminate inner drain with a wire nut. Table 2-1: Wiring for Rosemount 396P (Gray Cable) and Rosemount 1056, 56, or 1057 Terminal number...
  • Page 20 If there is a shrink sleeve on the gray reference wire, remove it before connecting the wire to the terminal. For Rosemount 1056, terminate inner drain with a wire nut. Table 2-2: Wiring for Rosemount 396P (Blue Cable) and Rosemount 1056, 56, or 1057 Terminal number...
  • Page 21 If there is a shrink sleeve on the gray reference wire, remove it before connecting the wire to the terminal. For Rosemount 1056, terminate inner drain with a wire nut. Table 2-3: Wiring for Rosemount 396P and Rosemount 1056, 56, or 1057 Terminal number Wire color...
  • Page 22 If there is a shrink sleeve on the gray reference wire, remove it before connecting the wire to the terminal. For Rosemount 1056, terminate inner drain with a wire nut. Table 2-4: Wiring for Rosemount 396P (Blue Cable) and Rosemount 1056, 56, or 1057 Terminal number...
  • Page 23 November 2020 Quick Start Guide Figure 2-5: Wiring for Rosemount 396PVP without Integral Preamplifier and Rosemount 1056, 56, or 1057 If there is a shrink sleeve on the gray reference wire, remove it before connecting the wire to the terminal. For Rosemount 1056, terminate inner drain with a wire nut.
  • Page 24 Terminal number Wire color Connects to White RTD return White/red Sense RTD in Blue Solution ground Inner drain +5 Vdc White/gray -5 Vdc Clear pH shield Orange pH in Reference shield Gray Reference Green No connection or earth ground Emerson.com/Rosemount...
  • Page 25 If ground lead is present, terminate it to green ground screw on inner enclosure. • TB5, TB7, and TB8 not used for pH/ORP sensor wiring. Table 2-7: Wiring for Rosemount 396P (Gray Cable) and Rosemount 1066 Terminal number Wire color Connects to...
  • Page 26 If ground lead is present, terminate it to green ground screw on inner enclosure. • TB5, TB7, and TB8 not used for pH/ORP sensor wiring. Table 2-8: Wiring for Rosemount 396P (Blue Cable) and Rosemount 1066 Terminal number Wire color Connects to...
  • Page 27 If ground lead is present, terminate it to green ground screw on inner enclosure. • TB5, TB7, and TB8 not used for pH/ORP sensor wiring. Table 2-9: Wiring for Rosemount 396P (Gray Cable) and Rosemount 1066 Terminal number Wire color Connects to...
  • Page 28 If ground lead is present, terminate it to green ground screw on inner enclosure. • TB5, TB7, and TB8 not used for pH/ORP sensor wiring. Table 2-10: Wiring for Rosemount 396P (Blue Cable) and Rosemount 1066 Terminal number Wire color Connects to...
  • Page 29 November 2020 Quick Start Guide Figure 2-11: Wiring for Rosemount 396PVP and Rosemount 1066 Note • If ground lead is present, terminate it to green ground screw on inner enclosure. • TB5, TB7, and TB8 not used for pH/ORP sensor wiring. Table 2-11: Wiring for Rosemount 396PVP and Rosemount 1066 Terminal block Wire color...
  • Page 30 Table 2-12: Wiring for Rosemount 396PVP-70 and Rosemount 1066 Terminal block Wire color Connects to White RTD return White/red RTD sense RTD return in Inner drain Preamplifier: +volts White/gray Preamplifier: -volts Shield pH shield Orange pH in Blue Solution ground Gray Reference in Emerson.com/Rosemount...
  • Page 31 Figure 2-13: Wiring for Rosemount 396P (Gray Cable) and Rosemount 5081-P-HT If there is a shrink sleeve on the gray reference wire, remove it before connecting the wire to the terminal. Table 2-13: Wiring for Rosemount 396P (Gray Cable) and Rosemount 5081-P-HT Terminal Wire color Connects to...
  • Page 32 Figure 2-14: Wiring for Rosemount 396P (Blue Cable) and Rosemount 5081-P-HT If there is a shrink sleeve on the gray reference wire, remove it before connecting the wire to the terminal. Table 2-14: Wiring for Rosemount 396P (Blue Cable) and Rosemount 5081-P-HT Terminal Wire color Connects to...
  • Page 33 Figure 2-15: Wiring for Rosemount 396P (Gray Cable) and Rosemount 5081-P-HT If there is a shrink sleeve on the gray reference wire, remove it before connecting the wire to the terminal. Table 2-15: Wiring for Rosemount 396P (Gray Cable) and Rosemount 5081-P-HT Terminal Wire color Connects to...
  • Page 34 Figure 2-16: Wiring for Rosemount 396P (Blue Cable) and Rosemount 5081-P-HT If there is a shrink sleeve on the gray reference wire, remove it before connecting the wire to the terminal. Table 2-16: Wiring for Rosemount 396P (Blue Cable) and Rosemount 5081-P-HT Terminal Wire color Connects to...
  • Page 35 November 2020 Quick Start Guide Figure 2-17: Wiring for Rosemount 396PVP and Rosemount 5081-P-HT If there is a shrink sleeve on the gray reference wire, remove it before connecting the wire to the terminal. Table 2-17: Wiring for Rosemount 396PVP and Rosemount 5081-P-HT Terminal Wire color Connects...
  • Page 36 Wire color Connects to number number Green Clear Drain Orange mV in White/gray -5 Vdc White RTD return Inner drain +5 Vdc White/red RTD sense Anode RTD in Cathode Drain -24 Vdc Gray Reference +24 Vdc Blue Solution ground Emerson.com/Rosemount...
  • Page 37: Start Up And Calibrate

    November 2020 Quick Start Guide Start up and calibrate Calibrate Rosemount 396P and 396PVP pH sensors 3.1.1 Prepare sensor Procedure 1. Shake down the sensor to remove any air bubbles that may be present at the tip of the pH glass bulb.
  • Page 38 1. While obtaining a process solution sample, record the pH value that is shown on the transmitter display. Emerson recommends taking the sample close to the sensor. 2. Measure and record the pH of the process solution sample with another temperature compensated, calibrated pH instrument.
  • Page 39 November 2020 Quick Start Guide Quinhydrone is only slightly soluble; therefore, use only a few crystals. 3. Immerse the sensor in the standard solution. Allow one to two minutes for the ORP sensor to stabilize. 4. Adjust the standardized control of the transmitter to the solution value shown in Table 3-1.
  • Page 40: Maintenance

    Values for Temperature Compensation Elements Temperature element PT-100 107.7 0.00385 Table 4-2: Temperature vs. Resistance of Automatic Temperature Compensation Elements Temperature °F (°C) Resistance (Ohms) ±1% PT-100 32 (0) 100.0 50 (10) 103.8 68 (20) 107.7 77 (25) 109.6 86 (30) 111.5 Emerson.com/Rosemount...
  • Page 41 November 2020 Quick Start Guide Table 4-2: Temperature vs. Resistance of Automatic Temperature Compensation Elements (continued) Temperature °F (°C) Resistance (Ohms) ±1% PT-100 104 (40) 115.4 122 (50) 119.2 140 (60) 123.1 158 (70) 126.9 176 (80) 130.8 194 (90) 134.6 212 (100) 138.5...
  • Page 42 (ORP) electrode/transmitter system. Type 1: One type of commonly used ORP standard solution is the saturated quinhydrone solution (PN R508-8OZ). Refer to Calibrate Rosemount 396P and 396PVP oxidation reduction potential (ORP) sensors. WARNING The solution used during the following check is an acid.
  • Page 43 November 2020 Quick Start Guide measuring this strongly acidic and concentrated solution. However, if you keep measuring electrodes clean and in good operating condition, you can carry out consistently repeatable calibrations using this standard solution. Clean platinum electrode To restore the electrode to normal operation, clean the platinum electrode with baking soda.
  • Page 44: Diagnostics And Troubleshooting

    Make sure electrode measuring tip is fully submerged in the process liquid. 5.1.2 Cracked glass failure GLASS fAIL Potential cause Broken or cracked glass. Recommended action Replace sensor. 5.1.3 High reference impede rEF fAIL or rEF WArn Potential cause Liquid junction coated. Emerson.com/Rosemount...
  • Page 45 November 2020 Quick Start Guide Recommended action Clean sensor; replace if necessary. Potential cause Reference cell gel depleted. Recommended action Replace sensor. Potential cause Sensor not fully immersed. Recommended action Make sure electrode tip is fully immersed in the process solution. 5.1.4 Input voltage high or input voltage low Potential cause...
  • Page 46 Check wiring. Potential cause Bad preamplifier. Recommended action Replace preamplifier (code -01 only). 5.1.9 Temp error high or temp error low tEMP HI or tEMP LO Potential cause Open or shorted resistance temperature device (RTD). Recommended action Replace the sensor. Emerson.com/Rosemount...
  • Page 47 November 2020 Quick Start Guide Potential cause Temperature out of range. Recommended action Check process temperature. Troubleshooting without advanced diagnostics The sections below list common problems, causes, and remedies typically encountered in process measurement. 5.2.1 Reading is off scale Display reads overrange. Potential cause Defective preamplifier.
  • Page 48 Sluggish or slow meter indication for real changes in pH level Potential cause Electrode coated. Recommended actions 1. Clean sensor as recommended in Clean electrode Clean platinum electrode. 2. Replace sensor if cracked. Potential cause Electrode defective. Recommended action Replace the sensor. Emerson.com/Rosemount...
  • Page 49 November 2020 Quick Start Guide 5.2.6 Transmitter cannot be standardized Potential cause Electrode coated. Recommended actions 1. Clean sensor as recommended in Clean electrode Clean platinum electrode. 2. Replace sensor if cracked. Potential cause Defective preamplifier. Recommended action Replace preamplifier. 5.2.7 Transmitter short spans between two different buffer values Potential cause...
  • Page 50: Return Of Material

    Quick Start Guide November 2020 Return of material For all repair or warranty inquiries, please contact our Customer Care department at 800-999-9307. Emerson.com/Rosemount...
  • Page 51: Rosemount Ph/Orp Sensor(S) Product Certifications

    A copy of the EU Declaration of Conformity can be found at the end of the Quick Start Guide. The most recent revision of the EU Declaration of Conformity can be found at Emerson.com/Rosemount. Ordinary location certification As standard, the transmitter has been examined and tested to determine...
  • Page 52 Model 396VP-a-b. Submersion/insertion pH sensor g. Model 396P-a-b-c-d-e. Submersion/insertion pH/ORP sensor h. Model 396PVP-a-b-c-d. Submersion/insertion pH/ORP sensor i. Model 396R-a-b-c-d-e. Retraction/submersion/insertion pH/ORP sensor j. Model 396RVP-a-b-c-d. Retraction/submersion/insertion pH/ORP sensor k. Model 397-a-b-c-d-e. pH sensor l. Model 398-a-b-c-d-e. pH/ORP sensor Emerson.com/Rosemount...
  • Page 53 November 2020 Quick Start Guide m. Model 398VP-a-b-c. pH/ORP sensor n. Model 398R-a-b-c-d-e-f. pH/ORP sensor o. Model 398RVP-a-b-c-d-e-f. pH/ORP sensor p. Model 3200HP-00. High purity water pH sensor q. Model 3300HT-a-b-c-d. High performance pH and ORP sensor r. Model 3300HTVP-a-b-c-d. High performance pH and ORP sensor s.
  • Page 54 –20 °C to +60 °C: (simple apparatus) Europe 7.6.1 ATEX Intrinsic Safety Certificate Baseefa10ATEX0156 Standards EN 60079-0: 2012+A11: 2013, EN 60079-11: 2012 Markings pH/ORP sensors with no preamplifier fitted II 1 G Ex ia IIC T4 Ga (–20 °C to +60 °C) Emerson.com/Rosemount...
  • Page 55 November 2020 Quick Start Guide pH sensors with integral smart preamplifier fitted II 1 G Ex ia IIC T4 Ga (–20 °C to +60 °C) ORP sensors with integral standard preamplifier fitted II 1 G Ex ia IIC T4 Ga (–20 °C to +80 °C) Ex ia IIC T5 Ga (–20 °C to +40 °C) pH sensors with integral standard preamplifier fitted II 1 G Ex ia IIC T4 Ga (–20 °C to +80 °C)
  • Page 56 ORP sensors with integral standard preamplifier fitted Ex ia IIC T4 Ga (–20 °C to +80 °C) Ex ia IIC T5 Ga (–20 °C to +40 °C) pH sensors with integral standard preamplifier fitted Ex ia IIC T4 Ga (–20 °C to +80 °C) Emerson.com/Rosemount...
  • Page 57 November 2020 Quick Start Guide Ex ia IIC T5 Ga (–20 °C to +40 °C) Specific Condition for Safe Use (X): See certificate for special conditions. Quick Start Guide...
  • Page 58: Declaration Of Conformity

    Quick Start Guide November 2020 Declaration of Conformity Emerson.com/Rosemount...
  • Page 59 November 2020 Quick Start Guide Quick Start Guide...
  • Page 60: China Rohs Table

    Quick Start Guide November 2020 China RoHS table Emerson.com/Rosemount...
  • Page 61: Intrinsically Safe Sensor Installation Drawing - Fm

    November 2020 Quick Start Guide Intrinsically safe sensor installation drawing - Quick Start Guide...
  • Page 62 Quick Start Guide November 2020 Emerson.com/Rosemount...
  • Page 63 November 2020 Quick Start Guide Quick Start Guide...
  • Page 64 LinkedIn.com/company/Emerson- Automation-Solutions The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a Twitter.com/rosemount_news mark of one of the Emerson family of companies. Facebook.com/Rosemount All other marks are the property of their Youtube.com/RosemountMeasurement respective owners.