CARESCAPE ONE User Manual DRAFT 26 October 2017
Pulse oximetry display data update
period
Pulse oximetry sweep speed options
Pulse oximetry waveform scale options
Wavelength information for SpO
LEDs
Pulse oximetry parameters monitored
Pulse oximetry saturation measurement
value and display range
Pulse oximetry peripheral pulse rate
range
Pulse oximetry saturation measurement
value accuracy
2094480-001
< 30 s
6.25, 12.5, 25, and 50 mm/s
TruSignal technology: AUTO, 50, 20, 10, 5, 2
Masimo and Nellcor technologies: 1x, 2x, 4x, and 8x
probe
This information can be especially useful to clinicians such as
2
those performing photodynamic therapy.
GE TruSignal pulse oximetry sensor LED peak wavelengths are
within 600 to 1000 nm and the maximum optical output power for
each LED is less than 15 mW.
Nellcor OxiMax pulse oximetry sensors contain LEDs that emit red
light at a wavelength of approximately 660 nm and infrared light
at a wavelength of approximately 900 nm. The total optical output
power of the sensor LEDs is less than 15 mW.
Masimo SET pulse oximetry sensors contain LEDs that emit red
light at a wavelength of approximately 660 nm and infrared light
at a wavelength of approximately 905 nm for LNOP and LNCS,
approximately 663 nm and 880 nm for LNOP and LNCS tip clips,
and approximately 660 nm and 880 nm for LNOP and LNCS
transflectance forehead sensors. The total optical output power
of the LEDs is less than 15 nW.
Arterial oxygen saturation (SpO
GE TruSignal: 0 to 100%
Masimo: 0 to 100%
Nellcor: 1 to 100%
GE TruSignal: 30 to 300 bpm
Masimo: 25 to 240 bpm
Nellcor: 20 to 300 bpm
The specified accuracy for each technology is the
root-mean-square (RMS) difference between the measured values
and the reference values. Because pulse oximetry equipment
measurements are statistically distributed, only about two-thirds
of the pulse oximetry equipment measurements can be expected
to fall within the ±1 A
Actual accuracy depends on sensor. Please refer to the sensor
instructions for use for more detailed information.
GE TruSignal accuracy:
Without motion:
SpO
(70 to 100%): ±2 Adult, ±2 Pediatric, ±3 Neonatal
2
SpO
(<70%): Unspecified
2
With motion:
SpO
(70 to 100%): ±3 Adult, ±3 Pediatric, ±3 Neonatal
2
SpO
(<70%): Unspecified
2
Low perfusion:
SpO
(70 to 100%): ±2 Adult, ±2 Pediatric, ±3 Neonatal
2
SpO
(<70%): Unspecified
2
CARESCAPE ONE
Parameter specifications
) and pulse rate
2
of the value measured by a CO-oximeter.
rms