Quick Links

Table of Contents
loading

Summary of Contents for Love Plasma Love Plasma

  • Page 2 The proof of purchase and the delivery note must be enclosed with the return shipment to the dealer or manufacturer Love Plasma Ltd assumes no transport costs and no liability for transport damages.
  • Page 3 Customer service For questions about your device and further information, please contact the customer service of Love Plasma Ltd Love Plasma Ltd 46 Silver Street Salisbury SP1 2NE United Kingdom www.loveplasma.com Mobile: 07830699360 Landline: 01722237127...
  • Page 4: Table Of Contents

    User manual Love Plasma device Table of contents 1 Safety and responsibility ................4 1.1 Intended use ..................4 1.2 Hygienic requirements ................ 4 1.3 Qualification of the user ..............4 1.4 Professional user training ..............5 1.5 Contraindications ................5 1.6 Side effects and residual risks ............
  • Page 5: Safety And Responsibility

    When the device is sold, the instructions for use must be included. Intended use The plasma device “Love Plasma” for cosmetic treatment is intended exclusively for commercial use by qualified person in clinical environment that meets the hygiene requirements.
  • Page 6: Professional User Training

    Professional user training Training includes the following topics. ■ Theory ■ Special application techniques. ■ Treatment of sensitive areas, e.g. eye and mouth area. ■ Further treatment options, e.g. upper and lower lid tightening. ■ Exercises on models. ■ Effective aftercare to support the treatment. Contraindications Only use the device on healthy persons who qualify for treatment.
  • Page 7: Storage

    Electrical and Electronic Equipment (WEEE), device must not be disposed of with conventional household waste. Inquire at the waste offices of the municipalities about collection points or return the device to Love Plasma Ltd Company. Our WEEE registration number: WEE/AX4054VE...
  • Page 8: Device Overview

    = Power bank (6) = Cable for handpiece = Charging cable for power bank = Love Plasma device Connecting the device Connect the cable for handpiece to the USB-A input of the power bank and connect the other end of the cable to the plasma device (right-angle connector).
  • Page 9: Load The Powerbank

    Load the power bank Connect the supplied Micro USB cable (3) to the Micro USB input of the power bank and connect the other end of the USB cable to a standard USB port of a computer or USB power supply. The charging process and the current charging status are indicated by the LED status lights on the power bank.
  • Page 10 reduction effect...
  • Page 11: Treatment

    Treatment Treatment options WARNING Skin injuries due to improper use The treatment requires special knowledge. Treatment is only permitted after prior user training. eyebrows lift upper eyelid lift crow's feet nasal folds l ower eyelid lift ear wrinkles nasolabial folds mouth wrinkles neck wrinkles Preliminary discussion...
  • Page 12: Pretreatment

    ■ Indication of downtime after treatment. ■ Information on the declaration of consent A written declaration of consent is required to perform the treatment. It is recommended to keep a treatment book to document the areas of application and progress of the treatment. Pre-treatment Carry out the following steps before treatment: ►...
  • Page 13 Danger due to infections Wear hygienic disposable gloves and mask. Treatment steps: ► Put the magnetic applicator into the device. ► Press the release button and guide the needle tip 0,5 mm to 2 mm to the skin. This produces a plasma flash that must not last longer than 1 second.
  • Page 14: Cleaning Parts Of The Unit

    ATTENTION Applicators must be disposed properly after treatment. ► Clean all parts of the unit. Cleaning parts of the unit ATTENTION Improper cleaning will damage the surfaces of the unit and cause malfunctions. ■ Unplug the rear connector before starting cleaning. ■...
  • Page 15: Troubleshooting

    WARNING Danger to life from electricity Never open the device or power bank. Repairs may only be carried out by the customer service of Love Plasma Ltd Company to ensure continued compliance with the safety requirements and the safety of persons, see page 2.
  • Page 16: Technical Specifications

    Technical specifications Device and power bank Technical Specifications power supply for power bank 5 VDC power supply for the device 5 VDC, 0.5 A max DC connection 2-pin cable, hollow plug 3.6 mm approx. 130 mm (with needle 150 Unit dimensions, total (L x H) mm) x 25 mm Weight approx.
  • Page 17: Ec Declaration Of Conformity

    Salisbury SP1 2NE United Kingdom The manufacturer declares conformity to the guidelines for the plasma device “Love Plasma” described in this manual: ■ Low voltage directive: 2014/35/EU ■ EMC directive: 2014/30/EU ■ RoHS directive 2011/65/EU ■ Waste electrical and electronic equipment 2012/19/EU and compliance with the following harmonised standards: ■...
  • Page 18 Instruction sheet for aftercare Customer (m/f): __________________________ Attending person: __________________________ Place, date: __________________________ Safety notice ATTENTION Scar formation on removal of the crust After approx. 1 to 2 days a protective crust forms on the treated area, which falls off after approx. 5 to 8 days. The skin is pink and very sensitive. ■...
  • Page 19 Recommended care products (to be completed by the attending person) Product application frequency Post-treatment care A lack of hygiene can lead to inflammation. Twice a day, morning and evening: ► Place curd soap in a lukewarm water bath and gently dab the treated areas with a cotton pad.

Table of Contents