2555
? I2C Internal Error; Block
Table full
2556
? I2C Not Enough Memory
2557
? I2C Internal Error; Block
Too Large on %4
2558
? I2C Internal Error; Block
Not Found on %4
2559
? I2C Internal Error; Invalid
Block Size on %4
2560
!
I2C Driver OS and Applica‐
tion version mismatch
2562
? I2C Internal Error; Bad
state on %4
2563
!
I2C FPGA bad version
2564
!
I2C Temperature Block
Version Invalid
2565
? I2C Opcode not expected
in current state on %4
2566
!
I2C Modules were reset fol‐
lowing multiple errors
2567
!
I2C Header Index Mis‐
match; Message Lost
2568
? I2C Internal Error; Invalid
Block Index on %4
2569
? I2C FPGA Queue Full on
%4
2570
? I2C FPGA Packet Too Big
on %4
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001
4.6 Alarm Codes, Descriptions, and Suggested Actions
Description
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: SNECON firmware
internal error.
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: The version num‐
ber of the SNECON OS is incompatible
with the SNECON APP version.
SNECON I2C driver: SNECON firmware
internal error.
SNECON I2C driver: The SNECON I2C
FPGA is incompatible or broken.
SNECON I2C driver: The temperature
controller PID parameter data block sup‐
plied by the SYSCON is incompatible.
SNECON I2C driver: Unexpected I2C
message opcode received by the SNE‐
CON from a module.I2C communication
error.
SNECON I2C driver: A fatal error or mul‐
tiple consecutive retries have forced the
I2C controller to reset.
System error
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: The SNECON I2C
FPGA queue is full, the PIC is no longer
processing I2C messages.
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
General Maintenance and Troubleshooting
Action
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reload the SNECON OS with the new‐
est version.
Replace the SNECON.
Reload the SNECON OS and APP soft‐
ware to the corresponding SYSCON
version.
Contact Customer Support.
Contact Customer Support.
Contact Customer Support.
Reset the analyzer.
Reload the SNECON OS software.
Reset the analyzer.
Reload the SNECON OS software.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.