Inlet Connections - Teledyne D series Installation And Operation Manual

Table of Contents
D Series Syringe Pumps
Section 4 Gradient Pumping for Pressure, Flow, and Concentration Modes

4.2.1 Inlet Connections

Figure 4-1 Pump inlet connections
4-4
Kit components and connections are shown in Figure 4-1.
1. Mount the inlet valve on the pump housing with the spacer
block and screws provided.
2. Connect the pre-bent SST tubing from one port of the valve
to the pump inlet. Use the nut and ferrule to connect the
tubing at the inlet and the valve fittings to connect the tub-
ing at the valve.
3. Connect the PTFE refill tubing (with the filter) to the other
port of the valve, using the nuts and ferrules supplied.
Note
When connecting to pressurized sources in supercritical fluid
applications, use the stainless steel tubing without a filter. An
in-line filter is contained in the CO
to technical bulletin
Notes
at http://www.isco.com/sp_applications/#techbulletins).
pre-bent
SST tubing
PTFE tubing
stainless steel tubing
connection package (refer
2
TB08 CO
Applications and Technical
2
nut, ferrule
or
filter
to
system
spacer block
valve