Honeywell MDA Scientific MIDAS Quick Start Manual
Honeywell MDA Scientific MIDAS Quick Start Manual

Honeywell MDA Scientific MIDAS Quick Start Manual

Gas detector sensor cartridge installation

Quick Links

Quick Start Guide
��������� �������
�������� ���������
��� ������
���������
���
������
MDA Scientific MIDAS
®
Sensor Cartridge Installation
�����
������
�����
2. Sensor cartridge fitting/
replacement
�������� �����
Diagram 1.
��� ��������� �� ������
Removing detector cover
�������
�������
�������� �����
��������
�������
�������
�����
��������
�������
�������
��� ����� �� ������
�������� ������
�������
��������
�������
�������� ������
��� ��
��� �����
� ������
���
���� ���
� ����
��� ����
�� � ���
���
���� ����
�� � ����
��� ��
����
�����
� ����
��
�����
���� ��
����
���� �����
��������� ������
1. Introduction
®
The MIDAS
sensor cartridge is supplied separately
and needs to be fitted to the detector's main chassis.
Before installing a sensor cartridge check that the
number on the sensor cartridge label matches the
gas type and range for the function required. All
sensor cartridges are supplied calibrated.
�������� �����
�������
Some MIDAS
sensor cartridges are capable of monitoring for more
®
than one type of gas. For details of how to set the
different gas types for a multi gas sensor cartridge refer to the
MIDAS
operating manual MIDAS-A-001.
®
��������
�������
NOTE: To avoid potential contamination of the gas sensing
surface, avoid handling the sensor cone. Hold sensor cartridge
by the locking tabs. Avoid exposure to dust, grease or any other
chemical contaminants.
�������� ������
��������
�������
��������� �������
�������� ���������
��� ������
Diagram 2.
Removing sensor cartridge
���������
������
��� ��
��� �����
� ������
���
���� ���
� ����
��������� �������
��� ����
���
�� � ���
�� � ����
���� ����
����
��� ��
��
�������� ���������
�����
�����
� ����
���� ��
����
���� �����
��� ������
��������� ������� ��
������ �� � ����
���������
������
���������
���
������
������
Diagram 3.
������
Fitting/replacing sensor cartridge
��� ��������� �� ������
�����
�������
������� ������ ������ �����
��� ��������� �� ������
�������
���������
�������
��� ��
��� �����
�����
� ������
���
���� ���
� ����
��� ����
�� � ���
���
�� � ����
���� ����
��� ��
����
� ����
��
�����
���� ��
����
���� �����
���������
��������� �������
�������� ���������
��� ������
������
�����
������
��� ��������� �� ������
�������
Diagram 4.
Refitting detector cover
���
��� ����� �� ������
�����
��������
�������
��� �����
���� ���
�� � ���
�� � ����
��� ��
�����
����
��� ��
��� �����
� ������
���
���� ���
� ����
���
��� ����
�� � ���
���� ����
�� � ����
��� ��
����
� ����
��
�����
����
���� ��
���� �����
���������
���������
For further details refer to the MIDAS
MIDAS-A-001.
2. Sensor cartridge fitting/
1. Unscrew the captive thumbscrew located on the front panel
and remove the cover by pulling it forwards off the main chassis
(see diagram 1). Switch power off using the power switch
on the backplane of the unit.
2. Ensure the power switch on the terminal module is in the
'off' position.
3. Remove the old sensor cartridge from the unit (if fitted) by
unclipping the two sensor cartridge locking tabs located either
side of the sensor cartridge and using them to firmly pull
the sensor cartridge out (see diagram 2).
���������
���
4. Fit the new sensor cartridge by aligning the pins at the top
of the sensor cartridge with the socket in the sensor cartridge
chamber.
5. Carefully push the sensor cartridge into the sensor cartridge
chamber until fully home and lock in place using the tabs
either side of the sensor cartridge (see diagram 3).
�����
6. Switch the power switch located on the terminal board
to the 'on' position.
7. Refit the detector's cover by aligning the slots either side
with the locating tabs on the mounting bracket assembly
and
pushing the cover horizontally until home.
�������
8. Tighten the thumbscrew located on the front panel
��� ��
��� �����
(see diagram 4).
� ������
���
���� ���
� ����
��� ����
�� � ���
���
�� � ����
���� ����
��� ��
����
� ����
��
�����
���� ��
����
���� �����
���������
Find out more
www.honeywellanalytics.com
Customer business centre
Europe and the rest of the world
Honeywell Analytics AG
Wilstrasse 11-U11
CH-8610 Uster
Switzerland
Tel: +41 (0)44 943 4300
Fax: +41 (0)44 943 4398
�������
��� ��
� ������
���
� ����
���
��� ����
���� ����
����
��
�����
Customer business center
� ����
����
���� ��
���� �����
A mericas
��������� ������
Honeywell Analytics Distribution, Inc.
400 Sawgrass Corporate Pkwy
Suite 100
Sunrise, FL 33325
USA
Tel: +1 954 514 2700
Toll free: +1 800 538 0363
Fax: +1 954 514 2784
www.honeywell.com
operating manual
®
replacement
Issue1.1 12/2005
H_MIDAS-A-021_V1
© 2005 Honeywell Analytics
loading