Important information on wall thickness measurement using the DMS Go+

Probes

Please note that only probes specified in chapter 2 of
this manual are accepted for the DMS Go+.
V-path correction
The following probes have only one transducer element
and therefore require no V-path correction:
● K 1 SC
● K-PEN
● G 2 N
● CLF 5
● CA 215
● CA 214
● CA 211 A
● A DFRP
● A2 DFR
● A2 10 M
DMS Go+
All probes equipped with a transmitter and a receiver el-
ement require a V-path correction due to the inclined po-
sition of their transducer elements (with wall thickness-
es < 60 mm in steel).
The V-path correction is stored in the DMS Go+ for the
following probes:
● Alpha2 D
● DA 303
● G5 KB
● DA 501 EN
● CLF 4
● DA 512
● CA 211
● HT 400 A
● A2 DFR
● OSS 10
● 5 M 5 DFR
Edition 4 (05/2014)
● CA 211A
● DA 301
● DA 312
● DA 312 B1
● DA 503 EN
● DA 507
● D 590
● FH 2 E
● KBA 525
● KBA 560
● TC 560
● D 790