Emerson Rosemount 1595 Reference Manual
Emerson Rosemount 1595 Reference Manual

Emerson Rosemount 1595 Reference Manual

Conditioning orifice plate
Hide thumbs Also See for Rosemount 1595:

Quick Links

Reference Manual
00809-0100-4828, Rev AA
October 2003
Rosemount 1595 Conditioning Orifice Plate
www.rosemount.com
Table of Contents
loading

Summary of Contents for Emerson Rosemount 1595

  • Page 1 Reference Manual 00809-0100-4828, Rev AA October 2003 Rosemount 1595 Conditioning Orifice Plate www.rosemount.com...
  • Page 3: Table Of Contents

    Installation ..........2-6 Rosemount 1595 Types: ....... . 2-6 Rosemount 1496 Types: .
  • Page 4 Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003 TOC-2...
  • Page 5 Using non-nuclear qualified products in applications that require nuclear-qualified hardware or products may cause inaccurate readings. For information on Rosemount nuclear-qualified products, contact your local Emerson Process Management Representative. This device is intended for use in temperature monitoring applications and should not be used in control and safety applications.
  • Page 7: Using This Manual

    Returning the Product ......page 1-1 USING THIS MANUAL This product manual provides installation, configuration, calibration, troubleshooting, and maintenance instructions for the Rosemount 1595 Conditioning Orifice Plate Section 2: Installation...
  • Page 8 Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003...
  • Page 9: Safety Messages

    Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003 Section 2 Installation Safety Messages ....... . . page 2-1 Location and Orientation .
  • Page 10: Location And Orientation

    Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003 LOCATION AND The orifice plate electronics must be installed in the proper orientation relative to the pipe and the fluid measured. ORIENTATION Horizontal Pipe NOTE The maximum acceptable temperature is –100 to 700 °F (-73 to 370 °C).
  • Page 11: Vertical Pipe Installation

    Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003 Figure 2-3. Liquid in Horizontal Pipes Recommended Zone 45 degrees 90 degrees Recommended Zone 45 degrees Vertical Pipe Installation Gas in Vertical Pipes Mount the electronics above the pipe with the instrument lines sloping down.
  • Page 12 Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003 Liquid or Steam in Vertical Pipes Mount the transmitter below the pipe with the instrument lines sloping up. Figure 2-5. Liquid in Vertical Pipes LO Instrument Line LO Transmitter Connection HI Instrument Line...
  • Page 13: 1595 Straight Pipe Requirements

    Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003 1595 Straight Pipe Use the appropriate lengths of straight pipe upstream and downstream of the 1595 to minimize the effects of moderate flow disturbances in the pipe. Requirements Beta 0.40 0.65 Single 90°...
  • Page 14: Installation

    Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003 INSTALLATION Rosemount 1595 Types: 1595 P 1595 G 1595 U Rosemount 1496 Types: 1496 WN 1496 SO Step 1: Determine the Proper Orientation 1496 RJ See “Location and Orientation” on page 2-2.
  • Page 15: Rosemount 1497 Types

    Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003 Rosemount 1497 Types: 1497 WNP 1497 RJ Step 1: Determine the Proper Orientation See “Location and Orientation” on page 2-2. Step 2: Install the Meter Section Follow these steps to weld the orifice flanges to the pipe.
  • Page 16 Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003...
  • Page 17: Specification

    Liquid, gas or vapor turbulent flow, for pipe Reynold’s Numbers greater than 2,000. Pipe Sizes 2 to 24-in. (50 to 600 mm). Contact Emerson Process Management for other pipe sizes. Operating Temperature Limits -100 to 700 °F (-73 to 370 °C) (Contact Emerson Process Management for temperatures above 700 °F)
  • Page 18: Performance

    Perform a flow calculation using the Instrument Toolkit software package. Alternatively, contact a Emerson Process Management sales representative or Emerson Process Management Customer Central at 1-800-999-9307 for assistance. A Configuration Data Sheet is required prior to order for application verification.
  • Page 19: Dimensional Drawings

    Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003 DIMENSIONAL DRAWINGS Rosemount 1595 Conditioning Orifice Plate (Paddle, Square edged, Concentric) Paddle Width .75-in. Paddle Length Diameter Diameter for Paddle Type Paddle Paddle Line Size 150# 300# 600# 900# 1500# 2500#...
  • Page 20: Ordering Information

    Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003 ORDERING INFORMATION Rosemount 1595 Orifice Plate Ordering Table Model Product Description 1595 Conditioning Orifice Plate Code Plate Type Paddle, Square Edged Code Line Size 2-in. (50 mm) 3-in. (76 mm) 4-in. (100 mm) 6-in.
  • Page 21: Standard Dimensional Drawings For Rosemount 1496

    Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003 Appendix B Installation Drawings STANDARD DIMENSIONAL DRAWINGS FOR ROSEMOUNT 1496 www.rosemount.com...
  • Page 22 Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003...
  • Page 23 Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003...
  • Page 24 Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003...
  • Page 25: Standard Dimensional Drawings For Rosemount 1497

    Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003 STANDARD DIMENSIONAL DRAWINGS FOR ROSEMOUNT 1497...
  • Page 26 Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003...
  • Page 27 Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003...
  • Page 28 Reference Manual 00809-0100-4828, Rev AA Rosemount 1595 October 2003...
  • Page 30 October 2003 Rosemount and the Rosemount logotype are registered trademarks of Rosemount Inc. PlantWeb is a registered trademark of one of the Emerson Process Management group of companies. All other marks are the property of their respective owners. Emerson Process Management...

Table of Contents