Measurements - HP 54620A User's And Service Manual

Logic analyzer
Hide thumbs Also See for 54620A:
Table of Contents
The HP 54620A/C at a Glance
Display shows the current input signals
and much more
Up to sixteen (16) channels in normal
or delayed mode
Indicators for glitch or normal
acquisition, time base, channel
activity, trigger and acquisition status
Softkey labels
Measurement results
Color Palettes
Channel Controls select, position, and
label inputs
Turn channels on or off individually
or in groups of 8
Rearrange order of channels to
group related signals
Create and display labels to identify
channels, and set channel colors
General Controls measure, save and
restore results, and configure the
analyzer
Measurement keys provide
automatic single or dual-channel
time and frequency measurements,
or use cursors to make manual

measurements

Save or recall measurement
configurations or previous results
Autoscale performs simple
one-button setup of the analyzer
Horizontal Controls select sweep speed
and delay parameters
Sweep speeds from 5 ns/div to 1 s/div
Delay control moves waveform
display to point of interest
Delayed mode and delay allow
zooming in to show a portion of
waveform in detail (split screen)
2
Storage Keys begin and end data
acquisition
Run/Stop starts and stops continuous
acquisitions
Single performs one acquisition
Autostore accumulates and displays
the results of multiple acquisitions
(like persistence on an oscilloscope)
Erase clears the current acquisition
and all accumulated data
Trigger Keys define what data will be
captured by the analyzer
Edge mode allows triggering on a
positive or negative edge (or both) of
any single channel
Pattern mode allows triggering on a
pattern of channels either high, low,
or don't care, with a single edge
qualifier
Advanced mode allows sequential
triggers, combined patterns, patterns
and edges, pattern durations, and
edge occurrence counts
Softkeys extend the functionality of
command keys
Select measurement types, operating
modes, trigger specifications, label data,
and more
Trigger Input/Output
Trigger the HP 54620A/C from other
instruments or use the HP54620A/C to
trigger other instruments
Channel Inputs through a flexible
probing system
Sixteen channels through a dual
8-channel cable with micro-grabbers
Set logiclevels as TTL, CMOS, ECL, or
to a user-definable voltage