Extech Instruments GX900 User Manual

True rms industrial multimeter
Hide thumbs Also See for GX900:
Table of Contents

Quick Links

TRUE RMS INDUSTRIAL MULTIMETER 
MODEL GX900 
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
USER GUIDE 
Table of Contents
loading

Summary of Contents for Extech Instruments GX900

  • Page 1 USER GUIDE  TRUE RMS INDUSTRIAL MULTIMETER  MODEL GX900  99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentDepot.com...
  • Page 2: Introduction

    Extech  Instruments  is  an  ISO‐ 9001 certified company.  Please visit the Extech Instruments website (www.extech.com) for  the latest versions of Manuals, Software, Product updates, and  other current information.  KEY FEATURES   Measures AC/DC Voltage and Current (including remote Clamp current), Resistance, Capacitance, Frequency (electrical and electronic), RPM, Duty Cycle, dBm/dBV, Diode, Continuity, and Temperature  Built‐in datalogger and trend features store measurement data with convenient recall functionality  Maximum reading (MAX) and Minimum Reading (MIN) memory features  Reading HOLD feature for freezing displayed information (with Auto Hold)  Intuitive TFT color LCD display and convenient keypad push‐button control  On‐screen HELP utility for user hints and application information  Quick A/D conversion sample rates with low pass AC filtering  Internal real‐time calendar clock for time stamping  Industrial strength meter housing GX900-EU-EN V1.6 4/14...
  • Page 3: Table Of Contents

      LCD DISPLAY DESCRIPTION ..............  1 2   QUICK START GUIDE ................  1 3   MEASUREMENTS  ...................  1 4   DC VOLTAGE MEASUREMENTS ..............  1 4   AC VOLTAGE MEASUREMENTS ..............  1 5   dBm/dBV AC MEASUREMENTS  ..............  1 5   LOW PASS FILTER FOR ACV MEASUREMENTS ........  1 6   DC/AC MILLIVOLT MEASUREMENTS ............  1 6   GX900-EU-EN V1.6 4/14...
  • Page 4 AC AND DC SIGNALS ................  2 6   MINIMUM (MIN) AND MAXIMUM (MAX) READINGS ......  2 7   CAPTURING PEAK VALUES ..............  2 8   RELATIVE MODE ..................  2 8   DATA HOLD AND AUTO‐HOLD MODE ............  2 9   STORING INDIVIDUAL SCREEN‐SHOTS ...........  2 9   VIEWING MEMORY DATA  ..............  2 9   VIEW TREND DATA ................  3 0   RECORDING MEASUREMENTS (DATALOGGING) ........  3 1   GX900-EU-EN V1.6 4/14  ...
  • Page 5 SETTING THE EVENT THRESHOLD VALUE ..........  3 3   CHANGE CALIBRATION PASSWORD ............  3 3   CALIBRATION ACCESS  ................  3 4   BLUETOOTH  CAPABILITY ..............  3 4   GX900 SOFTWARE .................  3 4   SETTING THE DATE AND TIME ...............  3 5   AUTOMATIC POWER OFF ..............  3 5   TIME AND DATE FORMATS ..............  3 6   NUMERIC FORMAT  ................  3 6  ...
  • Page 6: Safety

    Equipment  of  OVERVOLTAGE  CATEGORY  I  is  equipment  for  connection  to  circuits  in  which  measures  are  taken  to  limit  the  transient  over‐voltages  to  an  appropriate  low  level.  Examples  include protected electronic circuits.  OVERVOLTAGE CATEGORY II  Equipment of OVERVOLTAGE CATEGORY II is energy‐consuming equipment to be supplied from  the fixed installation. Examples include household, office, and laboratory appliances.  OVERVOLTAGE CATEGORY III  Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.    Note – Examples include switches in the fixed installation and some equipment for industrial use  with permanent connection to the fixed installation.  OVERVOLTAGE CATEGORY IV  Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation. Examples  include electricity meters and primary over‐current protection equipment.  GX900-EU-EN V1.6 4/14  ...
  • Page 7: Cautions

     Always remove the test leads before replacing the battery or fuses.   Inspect the condition of the test leads and the meter itself for any damage before  operating the meter.    Use great care when making measurements if the voltages are greater than 25VAC  rms or 35VDC. These voltages are considered a shock hazard.   Warning! This is a class A equipment. This equipment can cause interference in  residential areas; in these cases the operator may be required to take steps to  avoid or minimize the interference.   Always discharge capacitors and remove power from the device under test before  performing Diode, Resistance, or Continuity tests.   Voltage checks on electrical outlets can be difficult and misleading because of the  uncertainty of connection to the recessed electrical contacts. Other means should  be used to ensure that the terminals are not "live".   If the equipment is used in a manner not specified by the manufacturer, the  protection provided by the equipment may be impaired.   This device is not a toy and must not reach children’s hands. It contains hazardous  objects as well as small parts that the children could swallow.   Do not leave batteries and packing material unattended; they can be dangerous in  the hands of children.   In the event that the device is going to be unused for an extended period of time,  remove the batteries.   Expired or damaged batteries can cause cauterization on contact with the skin.  Always, therefore, use suitable gloves in such cases.   Ensure that the batteries are not short‐circuited. Never incinerate batteries.  GX900-EU-EN V1.6 4/14  ...
  • Page 8: Safety Instructions - Input Protection Limits

    Function  Maximum Input  V DC or V AC  1000V DC/AC rms  mA AC/DC  800mA 1000V fast acting fuse  A AC/DC  10A 1000V fast acting fuse  Frequency, Resistance,  Capacitance, Duty Cycle, Diode  1000VDC/AC rms  Test, Continuity  Temperature  1000VDC/AC rms  2. USE EXTREME CAUTION when working with high voltages.  3. DO NOT measure voltage if the voltage on the "COM" input jack exceeds 1000V  above earth ground.  4. NEVER connect the meter leads across a voltage source while the function switch is  in the current, resistance, or diode mode. Doing so can damage the meter.  5. ALWAYS discharge filter capacitors in power supplies and disconnect the power  when making resistance or diode tests.  6. ALWAYS turn off the power and disconnect the test leads before opening the covers  to replace the fuse or batteries.  7. NEVER operate the meter unless the back cover and the battery and fuse covers are  in place and fastened securely.  8. If the equipment is used in a manner not specified by the manufacturer, the  protection provided by the equipment may be impaired.      GX900-EU-EN V1.6 4/14  ...
  • Page 9: Fcc Compliance

    Extech Instruments  FCC COMPLIANCE  This device complies with part 15 of the FCC Rules. Operation is subject to the following  two conditions:    1. This device may not cause harmful interference.  2. This device must accept any interference received, including interference that may  cause undesired operation.    This equipment has been tested and found to comply with the limits for a Class B digital  device, pursuant to part 15 of the FCC Rules. These limits are designed to provide  reasonable protection against harmful interference in a residential installation. This  equipment generates, uses, and can radiate radio frequency energy and, if not installed  and used in accordance with the instructions, may cause harmful interference to radio  communications. However, there is no guarantee that interference will not occur in a  particular installation.    If this equipment does cause harmful interference to radio or television reception, which  can be determined by turning the equipment off and on, the user is encouraged to try to  correct the interference by one or more of the following measures:    • Reorient or relocate the receiving antenna.  • Increase the separation between the equipment and receiver.  • Connect the equipment into an outlet on a circuit different from that to which the  receiver is connected.  • Consult the dealer or an experienced radio/TV technician for help.    GX900-EU-EN V1.6 4/14  ...
  • Page 10: Descriptions

    All  functions,  excepting  current,  use  the  V  and  COM  input  terminals.  Separate  terminals are provided for current above 500mA and for current below 500mA.  Input for 0 A to 10.00 A current Input for 0 A to 500 mA current measurements Return terminal for all measurements Input for voltage, continuity, resistance, diode test, conductance, capacitance GX900-EU-EN V1.6 4/14  ...
  • Page 11: Keypad Description

      MAX/MIN button  Starts and stops the Maximum (MAX) and Minimum (MIN) reading memory    OK button  Confirms data entry    Rotary Switch  Selects the primary measurement function; for each, the meter presents a standard  display for that function (range, measurement units, and modifiers). See table below:  V~  AC voltage measurements  V–  DC and AC+DC voltage measurements    mV  DC(AC) millivolts, ac + dc millivolt, Temperature, Clamp Current measurements  Resistance, Diode test, capacitance, continuity measurements    Hz %  Frequency and duty cycle measurements  RPM Counter  RPM, Counter measurements  A  AC, dc and ac + dc amps measurements  mA  AC, dc and ac + dc milliamps measurements  uA  AC, dc and ac + dc microampere measurements up to 5,000 μA  INFO button  Opens the information window  Backlight button   Selects backlight intensity from low to high. Also used to return from an auto‐power  off condition  GX900-EU-EN V1.6 4/14  ...
  • Page 12: Lcd Display Description

    3. Primary measurement display area* (minus sign indicates a negative reading).  4. Range indicates the meter’s present range and the ranging mode (auto or manual).  5. Battery level indicates the battery charge level.  6. Time indicates the time set for the internal clock.  7. Mode annunciators indicate the Meter's mode.   8. Mini‐measurement display shows the input value when the primary and secondary  displays are busy with a menu screen or pop‐up message. Also shows the lightning  bolt icon (when necessary) in this area.  9. Main display provides measurement information regarding the input signal.  10. Date indicates the date set for the internal clock.  11. Beeper indicates the Meter’s audible feedback is enabled (this is not associated with  the continuity beeper).  12. Unit indicates the units of measure. Auxiliary Units indicates measurements that use  factors instead of units (such as Crest Factor).  13. Blue tooth indicates communication link activity.  14. Relative indicates that the displayed value is relative to a stored reference value.  15. Secondary display shows secondary measurement information concerning the input  signal.    *The Primary display area indicates the most important value of the selected function.  The secondary display contains the bar graph and other values that may be measured in  addition to the primary function (for example, frequency measurements along with the  primary AC voltage measurement).  GX900-EU-EN V1.6 4/14  ...
  • Page 13: Quick Start Guide

    Extech Instruments  QUICK START GUIDE  Step One: Ensure that batteries are installed and are fresh    Step Two: Carefully read the Safety section of this Instruction Manual      Step Three: Check the Table of Contents to find the section of this Instruction  Manual related to the desired test    Step Four: Connect the test leads to the meter and to the device under test as  described in this Instruction Manual’s section related to the test in question    Step Five: Safely perform the test as described in this Instruction Manual    Step Six: Read the measurement value on the LCD display    Step Seven: Carefully disconnected the meter’s test leads from the circuit or  device under test    Step Eight: Read the section of this Instruction Manual entitled Features and  Functions to learn more about the multitude of meter uses and features     Step Nine: Disconnect the test leads from the circuit under test and from the  meter and switch OFF the meter before storing.  GX900-EU-EN V1.6 4/14  ...
  • Page 14: Measurements

    COM  jack.  Insert  the  red  test  lead  banana  plug  into  the  positive V jack.  3. Touch  the  black  test  probe  tip  to  the  negative  side  of  the  circuit.  4. Touch  the  red  test  probe  tip  to  the  positive  side  of  the  circuit.  5. Read the voltage in the display.   GX900-EU-EN V1.6 4/14  ...
  • Page 15: Ac Voltage Measurements

    CAUTION:  Do not measure AC voltages if a motor on the circuit is being switched ON or OFF. Large voltage  surges may occur that can damage the meter.  1. Set the function switch to the V~ position.   2. Insert the black test lead banana plug into the negative  COM jack. Insert red test lead banana plug into the  positive V jack.  3. Touch the black test probe tip to the neutral side of the  circuit. Touch the red test probe tip to the “hot” side of  the circuit.  4. Read the voltage in the display.   dBm/dBV AC MEASUREMENTS  The meter is capable of displaying AC voltage as a dB value;  relative to 1 milli‐watt (dBm), relative to 1V (dBV), or to a  custom reference value.  1. Set the function switch to the V~ position   2. Press the F1 key (MENU) on the LCD and then  navigate to the dBm/dBV field using the arrow keys  and then press OK.  3. Navigate to the menu item labeled dBm and then  press OK.  4. Insert the black test lead banana plug into the  negative COM jack. Insert red test lead banana plug into the positive V jack.  5. Touch the black test probe tip to the neutral side of the circuit. Touch the red test  probe tip to the “hot” side of the circuit.  6. Read the dBm in the primary display area and read the Voltage in the secondary  display area.  GX900-EU-EN V1.6 4/14  ...
  • Page 16: Low Pass Filter For Acv Measurements

    1. Set the rotary switch to the V~ position  2. Press the F1 key (MENU)  3. Move the menu selector to the LOW PASS field and then  press OK  4. Select LPF (Low Pass Filter) and then press OK  5. Press the F3 key (RANGE) and then press the F2 key (MANUAL)  6. Move the menu selector to the 500V field and press the OK key  DC/AC MILLIVOLT MEASUREMENTS   CAUTION:   Do not measure DC/AC voltages if a motor on the circuit is  being switched  ON or OFF. Large  voltage surges may occur that can damage the meter.  1. Set the function switch to the mV position.   2. Press the F1 key (MENU), move the menu selector to the menu item labeled mVDC  or mVAC, and then press the OK key  3. Select mVDC or mVAC from the key options and press the OK key.  4. Insert the black test lead banana plug into the negative COM jack.  5. Insert the red test lead banana plug into the positive V jack.  6. Touch the black test probe tip to the negative side of the circuit.  7. Touch the red test probe tip to the positive side of the circuit.  8. Read the voltage in the display.   GX900-EU-EN V1.6 4/14  ...
  • Page 17: Current Clamp Measurements

    Extech Instruments  CURRENT CLAMP MEASUREMENTS  1. This meter offers remote current clamp measurement functionality.  2. Set the rotary switch to the mV position.  3. Press the F1 key (MENU), move the menu  selector to the item labeled CLAMP DC/AC,  and then press the OK key.  4. Select the item labeled RANGE  (0.1~100mV/A) and then press the OK key.  5. Connect the remote clamp to the input  terminal jacks ensuring proper polarity.  Insert the clamp’s negative lead into the negative COM jack and connect the  clamp’s positive lead into the meters positive V jack.  6. Read the measurement in the display.   GX900-EU-EN V1.6 4/14  ...
  • Page 18: Dc Current Measurements

    Extech Instruments  DC CURRENT MEASUREMENTS  CAUTION:  Do not make 20A current measurements for longer than 30 seconds. Exceeding 30 seconds may  cause damage to the meter and/or the test leads.  1. Insert black test lead banana plug into the negative COM jack.   2. For current measurements up to 5000µA DC, set the  function switch to the µA position and insert the red test  lead banana plug into the µA/mA jack.  Press the F1 key  (MENU) and select uADC, Press the OK key.  3. For current measurements up to 500mA DC, set the function  switch to the mA position and insert the red test lead  banana plug into the µA/mA jack.  Press the F1 key (MENU)  and select mADC, Press the OK key.  4. For current measurements up to 10A DC, set the function  switch to the 10A position and insert the red test lead  banana plug into the 10A jack.  Press the F1 key (MENU) and  select ADC, Press the OK key.  5. Remove power from the circuit under test, then open up the  circuit at the point where current is to be measured.  6. Touch the black test probe tip to the negative side of the  circuit.  7. Touch the red test probe tip to the positive side of the  circuit.  8. Apply power to the circuit.  9. Read the current in the display.   GX900-EU-EN V1.6 4/14  ...
  • Page 19: Ac Current Measurements

    Extech Instruments  AC CURRENT MEASUREMENTS  CAUTION:  Do not make 20A current measurements for longer than 30 seconds. Exceeding 30 seconds may  cause damage to the meter and/or the test leads.  1. Insert the black test lead banana plug into the negative  COM jack.   2. For current measurements up to 5000µA AC, set the  function switch to the µA position and insert the red test  lead banana plug into the µA/mA jack.  Press the F1 key  (MENU) and select uAAC, Press the OK key.  For current measurements up to 500mA AC, set the  function switch to the mA position and insert the red test  lead banana plug into the µA/mA jack.  Press the F1 key  (MENU) and select mAAC, Press the OK key.  For current measurements up to 10A AC, set the function  switch to the 10A position and insert the red test lead  banana plug into the 10A jack.  Press the F1 key (MENU)  and select AAC, Press the OK key.  5. Press the key labeled MENU, move the menu selector to  the item labeled AC, and then press the OK key  6. Remove power from the circuit under test, then open up  the circuit at the point where current is to be measured.  7. Touch the black test probe tip to the neutral side of the circuit. Touch the red test  probe tip to the “hot” side of the circuit.  8. Apply power to the circuit.  9. Read the current in the display.  GX900-EU-EN V1.6 4/14  ...
  • Page 20: Resistance Measurements

    Extech Instruments  RESISTANCE MEASUREMENTS  WARNING:  To  avoid  electric  shock,  disconnect  power  to  the  unit  under  test  and  discharge  all  capacitors  before taking any resistance measurements. Remove the batteries and unplug the line cords.  1. Set the function switch to the            Ω CAP position.  2. Insert the black test lead banana plug into the negative  COM jack.  3. Insert the red test lead banana plug into the positive Ω jack  4. Touch the test probe tips across the circuit or part under  test. It is best to disconnect one side of the part under test  so the rest of the circuit will not interfere with the  resistance reading.  5. Read the resistance in the display.  GX900-EU-EN V1.6 4/14  ...
  • Page 21: Continuity Check

    2. Press the F1 key (MENU), move the menu selector to the  RESISTANCE, and then press the OK key  3. Move the menu selector to the item labeled CONTINUITY  and press the OK key  4. Insert the black lead banana plug into the negative COM  jack.  5. Insert the red test lead banana plug into the positive          jack.  6. Touch the test probe tips to the circuit or wire under test.   7. If the resistance is less than approximately 35 Ω, the audible  signal will sound. If the circuit is open, the display will  indicate OL.  DIODE TEST  1. Set the function switch to the        Ω CAP position  2. Press the F1 key (MENU), move the menu selector to the  DIODE field, and then press the OK key  3. Move the menu selector to the item labeled DIODE and  press the OK key  4. Insert the black test lead banana plug into the negative  COM jack and the red test lead banana plug into the  positive V jack.  5. Touch the test probes to the diode under test. Forward  voltage will typically indicate 0.400 to 3.200V. Reverse  voltage will indicate OL. Shorted devices will indicate near  0V and an open device will indicate OL in both polarities.  GX900-EU-EN V1.6 4/14  ...
  • Page 22: Temperature Measurements

    Extech Instruments  TEMPERATURE MEASUREMENTS    1. Set the function switch to the TEMP position.   2. Press the F1 key (MENU), move the menu selector to the  item labeled TEMP, and then press the OK key  3. Move the menu selector to menu item labeled  FAHRENHEIT, Celsius, or Kelvin and then press the OK key  4. Insert the Temperature Probe into the input jacks ensuring  correct polarity.  5. Touch the Temperature Probe head to the device under  test. Maintain contact with the device until the reading  stabilizes (about 30 seconds).  6. Read the temperature in the display.  To input a temperature offset value, press the F3 key  (OFFSET).  A message box will appear showing the currently  selected offset value. Use the right/left arrow keys to position  the cursor over one of the digits (or the polarity sign). Then  use the up/down arrows to edit the selected digit or polarity  sign. With the desired value displayed, press the F1 key (OK) to  confirm the temperature offset.  Note: The temperature probe is fitted with a type K mini connector.   A mini connector to banana connector adaptor is supplied for connection to the input  banana jacks.  The adapter connects to COM (‐) and V (+).    Note: The temperature range of the supplied thermocouple probe is ‐20 to 250°C (‐4  to 482°F)  GX900-EU-EN V1.6 4/14  ...
  • Page 23: Capacitance Measurements

    WARNING:  To  avoid  electric  shock,  disconnect  power  to  the  unit  under  test  and  discharge  all  capacitors  before taking any capacitance measurements. Remove the batteries and unplug the line cords.  1. Set the rotary function switch to the CAP position.   2. Press the F1 key (MENU), move the menu selector to the  menu item labeled CAPACITANCE,  and then press the OK  key.  3. Select CAPACITANCE and then press the OK key.  4. Insert the black test lead banana plug into the negative COM  jack.  5. Insert the red test lead banana plug into the positive V jack.  6. Touch the test leads to the capacitor to be tested.  7. Read the capacitance value in the display  FREQUENCY, DUTY CYCLE, PULSE WIDTH, and PERIOD  MEASUREMENTS (ELECTRONIC)  1. Set the rotary function switch to the Hz/% position.   2. Press the F1 key (MENU), move the menu selector to the  menu item of your choice, either FREQUENCY(Hz), DUTY  CYCLE, PULSE WIDTH, or PERIOD, and then press the OK key.  3. Insert the black lead banana plug into the negative COM jack  and the red test lead banana plug into the positive Hz jack.  4. Touch the test probe tips to the circuit under test.  5. Read the measurement on the display.  GX900-EU-EN V1.6 4/14  ...
  • Page 24: Rpm Measurements

    RPM MEASUREMENTS  1. Set the rotary function switch to the RPM position.  2. Insert the black lead banana plug into the negative COM jack and the red test lead  banana plug into the positive V jack.  3. Touch the test probe tips to the circuit under test.  4. Read the RPM value on the display.  COUNTER MEASUREMENTS  1. Set the rotary function switch to the RPM position.  2. Press the F1 key (MENU), move the menu selector to the item labeled COUNTER,  and then press the OK key  3. Move the menu selector to the item labeled START and press the OK key  4. Insert the black test lead into the negative COM jack and the red test lead into the  positive V jack.  5. Touch the test probe tips to the circuit under test.  6. Read the Period value on the display.  7. To set the Count Threshold press the F1 key (MENU), move the menu selector to  the item labeled COUNTER, and press the OK key  8. Move the menu selector to the menu item labeled THRESHOLD and press the OK  key  9. Enter 1 or 2 or 3 using the arrow keys and then press the F1 key to save.  Press the  F1 key to return to the measure screen.      GX900-EU-EN V1.6 4/14  ...
  • Page 25: Functions And Features

    Use  the  keys  labeled  NEXT  and  PREV  to  move  from  topic  to  topic.  Use  the  key  labeled  MORE,  UP,  and  DOWN  to  scroll  through the information a full screen at a time.  GX900-EU-EN V1.6 4/14  ...
  • Page 26: Automatic And Manual Ranging

    Select  one  of  the  above  combinations  using  the  FUNCTION  and  MODE  menu  as  described in the steps below:  1. With the rotary switch set to V, mV, A, mA, or uA press the key labeled MENU  2. Move the menu selector to the menu item labeled MATH  3. Press OK  4. The three labels (AC+DC, AC/DC, and DC/AC) will appear  5. Move the menu selector to the desired menu label  6. Press OK  While  in  any  of  the  three  modes  described  above  Peak  Measurements,  Frequency,  Duty Cycle, MAX‐MIN, Relative %, and Period Measurements are not available for use.   GX900-EU-EN V1.6 4/14  ...
  • Page 27: Minimum (Min) And Maximum (Max) Readings

      To save the MIN‐MAX screen data, the MIN‐MAX session must be ended by pressing  the key labeled STOP. After pressing STOP, press the key labeled SAVE. A dialogue box  will  open  prompting  for  a  filename  (a  name  is  automatically  proposed  which  can  be  accepted  or  changed by the  user).  Press  the  key  labeled  SAVE.  Press  the  key  labeled  RESTART while MIN‐MAX is running stops the MIN‐MAX session, discards all MIN‐MAX  data, and starts a new MIN‐MAX recording session.  GX900-EU-EN V1.6 4/14  ...
  • Page 28: Capturing Peak Values

    Move  the  menu  selector  to  the  menu  item  labeled  PEAK/CREST  F  (Crest  Factor)  and  then  press  the  OK  key.  Move the menu selector to the PEAK or the CREST F label  and then press OK to start the PEAK recording session.   RELATIVE MODE  To activate the Relative mode, press the F1 key (MENU) and move the menu selector  next to the menu item labeled REL/REL% and then press the OK key.  Move the menu selector to the REL or the REL% label and press OK to start the Relative  mode.  If  the  meter  is  currently  in  the  Relative  mode,  pressing  REL%  causes  the  meter  to  switch OFF the Relative mode and display Relative percent.  GX900-EU-EN V1.6 4/14  ...
  • Page 29: Data Hold And Auto-Hold Mode

    (Auto‐Hold threshold) for at least one second. The meter filters open lead conditions  so that the meter leads can be moved between test points without triggering a display  update.  STORING INDIVIDUAL SCREEN‐SHOTS  For  all  measurement  functions,  a  snapshot  of  the  screen  data is saved by pressing the key labeled SAVE.         VIEWING MEMORY DATA  Viewing data stored in the meter’s memory is performed through the SAVE menu as  described below:  1. Press the F2 key (SAVE).  2. Position the menu selector on MEMORY and press the OK key.  3. Position the menu selector on VIEW and press the OK key.  4. Position the menu selector on LOG and press the OK key.  5. Press F1 (PREV) or F2 (NEXT) to view your saved data sets.  GX900-EU-EN V1.6 4/14  ...
  • Page 30: View Trend Data

        If there are previously stored records, press the F1 key   (PREV) to page through them or press the F2 key (NEXT) to  scroll in the opposite direction.   Press  F4  (CLOSE)  to  return  the  meter  to  the  normal  operating mode.   Press the F3 key (TREND) to display the recorded data in a  trend‐plot view. Press the F2 key (DELETE) to clear recorded  data.    Press  the  F3  key  (SUMMARY)  to  get  back  to  the  data  summary  and  Press  F4  key  (CLOSE) to return to normal operating mode.  GX900-EU-EN V1.6 4/14  ...
  • Page 31: Recording Measurements (Datalogging)

      7. Press the key labeled START to start recording.  8. The recording session will continue until the allocated memory is consumed, the  batteries  expire,  the  rotary  switch  is  moved,  or  the  session  is  terminated  by  pressing the key labeled STOP.    GX900-EU-EN V1.6 4/14  ...
  • Page 32: Zooming On Trend Data

    Meter  information  such  as  serial  number,  model,  etc.  is  also available in the Setup mode.  Access to the setup options is available by pressing the F4  key (SETUP).   RESET THE METER SETUP OPTIONS  The meter’s setup options can be reset to factory default conditions through the Setup  menu.  1. Open the setup menu by pressing the F4 key (SETUP).  2. Position the menu selector next to the item labeled RESET and then press the  OK key  3. Position the menu selector next to the item labeled SETUP and then press the  OK key, press the OK key again.  4. A message will appear asking “Reset settings to Factory Defaults?” to confirm  the  Reset  action.  Press  the  F1  key  (OK)  or  the  F4  key  (Cancel)  to  cancel  the  reset.    GX900-EU-EN V1.6 4/14  ...
  • Page 33: Meter Information Screen

    The  Calibration  selection  allows  a  qualified  calibration  technician  to  enter  a password that grants access to the meter for calibration purposes.  1. Press the F4 key (SETUP) to access the Setup menu.  2. Using  the  cursor  buttons,  move  the  menu  selector  next  to  the  item  labeled  Calibration and press OK.   3. Select PASSWORD and press OK.  4. Select EDIT and press OK.  5. Using the menu arrow keys enter a password and Click F1 key (OK) of F4 key  (CANCEL).  Note: default password is 1234  GX900-EU-EN V1.6 4/14  ...
  • Page 34: Calibration Access

    5. Input  the  current  password  using  the  menu  keys    and  press  F1  key  (OK)  or  press the F4 key (CANCEL).  Calibration instructions are only available to qualified certified technicians. Contact  Extech Instruments for further information concerning calibration instructions    BLUETOOTH  CAPABILITY  A Bluetooth communication link can be used to connect to the GX900 software.  It  will also allow you to monitor the meter from an Android device using ExView™.  1. Open the setup menu by pressing the F4 key (SETUP).  2. Position the menu selector next to the INSTRUMENT label and press OK  3. Position the menu selector next to the COMMUNICATE label and press OK  4. Press the F1 key (START) to turn on the Bluetooth™ signal.  5. Repeat steps 1~3 and then press the F2 key (STOP) to turn off the Bluetooth™  signal.  Note: when the meter is shut off and then back on, the Bluetooth signal will have  to be turned back on manually.    GX900 SOFTWARE  The GX900 meter comes with software that will allow you to capture measured data  in Real Time and download logged data from the GX900 to the software.  Refer to  the software’s help file for instructions for use.  GX900-EU-EN V1.6 4/14  ...
  • Page 35: Setting The Date And Time

    5. Use the up and down arrows to change the selected date or time values.  6. Press the F1 key (OK) to complete and confirm the actions.    AUTOMATIC POWER OFF  1. Open the Setup menu by pressing the F4 Key (SETUP).  2. Position the menu selector next to INSTRUMENT and press OK.  3. Position the menu selector next to SETTINGS and press OK.  4. Position the menu selector next to the item labeled POWER OFF and press OK.  5. To set the Auto Power OFF time interval, use the up and down arrows to select  the time in minutes (05 to 60).  6. Set the time to zero (00) to disable the Auto Power OFF feature.  7. Press the F1 key (OK) to confirm the selection.  8. Press  the  F4  key  (CANCEL)  to  return  to  the  normal  operating  mode  without  saving the edit.  GX900-EU-EN V1.6 4/14  ...
  • Page 36: Time And Date Formats

    4. Position  the  menu  selector  next  to  the  item  labeled  TIME  FORMAT  or  DATE  FORMAT and press OK to begin editing  5. TIME FORMAT – select F1 (24 HOUR) or F2 (12 HOUR).  DATE FORMAT – select F1(MM/DD/YY) or F2 (DD/MM/YY).  Press F3 (CANCEL) to exit without changing the format.  NUMERIC FORMAT  1. Press the F4 key (SETUP).  2. Position the menu selector next to the item labeled DISPLAY and press OK.  3. Position the menu selector next to the item labeled FORMAT and press OK  4. Position the menu selector next to the item labeled NUMERIC and press OK.  5. Select F1 (0.000) or F2 (0,000) or F4 (CANCEL).   LANGAUGE FORMAT  1. Press the F4 key (SETUP).  2. Position the menu selector next to the item labeled DISPLAY and press OK.  3. Position the menu selector next to the item labeled FORMAT and press OK  4. Position the menu selector next to the item labeled LANGAUGE and press OK.  5. Select F1 (ENG) or F4 (CANCEL).       GX900-EU-EN V1.6 4/14  ...
  • Page 37: Battery And Fuse Replacement

    Never  dispose  of  used  batteries  or  rechargeable  batteries  in  household  waste.  As  consumers,  users  are  legally  required  to  take  used  batteries  to  appropriate  collection  sites,  the  retail  store  where  the  batteries  were  purchased, or wherever batteries are sold.  Disposal: Do not dispose of this instrument in household waste. The user is obligated  to take end‐of‐life devices to a designated collection point for the disposal of electrical  and electronic equipment.   GX900-EU-EN V1.6 4/14  ...
  • Page 38: Fuse Replacement

    Extech Instruments  FUSE REPLACEMENT  1. Switch the meter off and remove the test leads from the meter’s input jacks  2. Remove the battery compartment cover.  3. Remove  the  fuse  from  the  battery  compartment  by  first  carefully  prying  one  end loose and then sliding the fuse from its bracket  4. Install only specified replacement fuses  5. Replace the battery compartment cover      800mA/1000V ceramic ‐ 6.3mm x 32mm (SIBA 7017240.0,8)      10A/1000V ceramic ‐ 10mm x 38mm (SIBA 5019906.10)  GX900-EU-EN V1.6 4/14  ...
  • Page 39: Specifications

    ANSI / NEDA- 5004LC,IEC-CR2032; Normal Voltage: 3.0 Volts; Built-in lithium Typical Capacity: 240 mAh ; Storage 5 Year Chemical type: Lithium polymer, Standard: GB/T Dimensions 235 x 108 x 63.5mm (9.25 x 4.25 x 2.5”) Weight 839g (1.85 lbs.) GX900-EU-EN V1.6 4/14  ...
  • Page 40: Electrical Specifications

    0.01µA 50/60Hz (0.6% + 25) <1KHz (1.5% + 25) 5000µA 0.1µA <10KHz (3% + 25) 50mA 0.001mA 500mA 0.01mA 0.001A Accuracy for all AC current ranges are specified from 5% of range to 100% of range GX900-EU-EN V1.6 4/14  ...
  • Page 41 ±(2% + 40) 50nF[¹] 0.01nF 500nF 0.1nF ±(2% + 40 digits) 5µF 0.001µF 50µF 0.01µF 500µF 0.1µF ±(5% +40 digits) 10mF 0.01mF [1] With a film capacitor or better, use relative mode (REL ∆ ) to zero residual. GX900-EU-EN V1.6 4/14...
  • Page 42 Resolution Accuracy Temperature -50 to 1000°C 0.1°C ±(1.0% reading (type-K) + 2.5°C) -58 to 1832°F 0.1°F ±(1.0% reading + 4.5°F) (Probe accuracy not included) GX900-EU-EN V1.6 4/14 Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com...