Rf Auto Shutoff - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents
Unit Settings - Alarms
Control
Function
Non-
Set latched/non-latched status
arrhythmia
for SpO
Yellow
arrhythmia yellow alarms.
Alarms
Leads Off
Adjust the severity level of this
technical alarm (INOP).
Replace
Adjust the severity level of this
Battery
technical alarm (INOP).
RF Auto
Shutoff
, ST, and other non-
2
When RF Auto Shutoff is enabled (default), the transceiver stops broadcasting a
radio signal in order to prevent interference with other transceivers in use. This
condition occurs if there is no ECG signal for 10 minutes and the SpO
cable is not inserted in the SpO
followed by the "Transmitter Off" INOP will be displayed at the Information
Center. Remove the batteries to conserve battery charge. However, if there is no
ECG signal for 10 minutes but the SpO
not stop broadcasting the signal. This feature can be disabled in Unit Settings.
Telemetry Functions at the Information Center & TeleMon
Telemetry Functions at the Information Center
Settings
Latched
Non-latched
Low
Medium
High
Low
Medium
High
sensor port. The technical alarm, "No Signal",
2
cable is inserted, the transceiver does
2
Factory
Default
Latched
Low
Low
sensor
2
9-15