Abaxis VetScan HM5 Quick Reference Manual
Hide thumbs Also See for VetScan HM5:
QUICK REFERENCE GUIDE
Easy operation — fast results
Closer to Care.
Table of Contents
loading

Summary of Contents for Abaxis VetScan HM5

  • Page 1 QUICK REFERENCE GUIDE Easy operation — fast results Closer to Care.
  • Page 2: Unpacking And Setup

    Unpacking and Setup • Unpack the VetScan HM5 analyzer from its shipping container and put out the accesory box. Note: Store the shipping box, accessory box, and packaging materials in case you need to use them later. • Place the analyzer on a level surface.
  • Page 3 • Open the accessory box and use this list to make sure you received all components of the Vet-Scan HM5 system. • VetScan HM5 Analyzer. • VetScan HM5 Operator’s Manual. • External power supply. • Mini-keyboard • Four sample tube adaptors.
  • Page 4 Connect the reagent pack as follows. • Open the plastic bag located in the accessory box, and take out the six color-coded reagent tubes and caps with drop-down tubes. • Attach the supplied connector • Insert the six male connectors into tubing kit to the instrument, the free ends of the color-coded as shown.
  • Page 5 Touchscreen Icons (list of icons and their meanings) Test: analyze samples and Database: search, retrieve, view run blanks and manage up to 5,000 records Maintenance: clean, calibrate, Settings: manage printer, run quality controls and manage testing, export and date/time reagent status and other analyzer settings Exit: select shutdown options and Printer: Manage printer settings...
  • Page 6: Running A Sample

    Running a Sample • Touch the Measure icon. • Enter the Patient ID on the “New Sample Screen”. • Select animal Type. • Mix the sample well by slowly inverting the tube 10-15 times. Insert the well-mixed EDTA preserved whole blood sample into the appropriate tube adapter, on the sample rotor, and remove the cap.
  • Page 7: Interpreting Results

    Interpreting Results Analysis is complete in less than four minutes. All results and histograms are automatically saved in the analyzer’s database. • When analysis is complete, the analyzer displays all measured and calculated parameters, as well as the WBC, EOS, RBC, and PLT histograms: If the analyzer is not set to print results automatically, press Print to print the results.
  • Page 8 Running a Blank A blank must be run every 12 hours. Each blank measurement remains valid for 12 hours of continuous operation, after which the analyzer displays BLANK needed and a new blank is required. In addition, a new blank must be run each time the analyzer has been powered on and off again.
  • Page 9: Running Controls

    Running Controls Controls are run to assess if the analyzer is maintaining calibration. • Touch the Maintenance icon. • Touch the Quality control button. Touch the QC level (e.g. QC Normal) that will • be used. Touch References to load the assay and gap values •...
  • Page 10: Changing The Reagent Pack

    Changing the Reagent Pack • In the Maintenance menu, select Reagent status. Select Replace and then Change pack. • Select Prime All to prime the reagent tubing.
  • Page 11: Calibrating The Analyzer

    Calibrating the Analyzer Calibration is rarely necessary on the HM5, but should be performed when the controls are running out of range. • Warm up HM5 Normal Control to room temperature. Do not use other level controls for calibration. • Touch Maintenance and select Calibration. Touch Measure.
  • Page 12: Recommended Maintenance

    Recommended Maintenance The VetScan HM5 requires minimal maintenance, but does need attention to proper sample handling and cleanliness to ensure optimal results. Daily Run a blank. Weekly Wipe down the HM5 wash head with a damp, lintfree cloth. With each change of reagent pack Perform Vetscan HM5 bleach cleaning procedure.
  • Page 13: Shutting Down The Analyzer

    Shutting Down the Analyzer Proper shut downs rinse the analyzer to keep it in good condition. So it is important to perform shut downs when the analyzer will not be used for extended periods of time. It is preferable to leave the analyzer on overnight if samples will be run within a couple of days.
  • Page 14 Results Flags (Partial list) Capital letters indicate WBC related flags; lower-case letters indicate RBC or PLT-related flags. For more details, consult your HM5 operator’s manual. Flag Meaning Solution Empty sample Lower needle height Linearity error (WBC) Prime lyse and repeat blank. View - cell concentration manual smear to confirm high WBC too high...