KERN CKE 6k0.02 Operating Manual

KERN CKE 6k0.02 Operating Manual

Counting balances
Hide thumbs Also See for CKE 6k0.02:
Operating Manual
Counting balances
KERN CKE
Version 1.3
01/2009
GB
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
Phone: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
CKE-BA-e-0913
Table of Contents
loading

Summary of Contents for KERN CKE 6k0.02

  • Page 1 KERN & Sohn GmbH Ziegelei 1 Phone: +49-[0]7433- 9933-0 D-72336 Balingen Fax: +49-[0]7433-9933-149 email: [email protected] Internet: www.kern-sohn.com Operating Manual Counting balances KERN CKE Version 1.3 01/2009 CKE-BA-e-0913...
  • Page 2: Table Of Contents

    Adjusting (see chapter 7.4.1)..................10 Verification ........................10 Operation........................11 Control panel CKE ......................11 Operation.......................... 12 7.2.1 Volume assessment - KERN QKE only !!!................... 14 7.2.1.1 Volume assessment without manual input ................14 7.2.1.2 Volume assessment with manual tare input ................ 14 7.2.1.3 Volume assessment with manual tare and gross input ............
  • Page 3 Interface RS 232C......................16 7.4.1 There are 4 kinds of data output via RS 232C ................16 7.4.2 Explanation of the data transfer....................16 7.4.3 Numerator ............................. 16 Printer..........................17 Underfloor weighing ......................17 Service, maintenance, disposal ................18 Cleaning ..........................18 Service, maintenance.......................18 Disposal ..........................18 Instant help .........................
  • Page 4: Mode - Menü

    1 MODE – MENÜ How to invoke the mode menu: Turn on balance, press and hold the tare key and press the ON/OFF key. Let go of the tare key. Factory settings: 9600bd: Fast: Keyboard overview / Function CKE-BA-e-0913...
  • Page 5: Technical Data

    2 Technical Data 2.1 KERN CKE KERN CKE 6k0.02 CKE 8k0.05 CKE 16k0.05 CKE 16k0.1 Readability (d) 0,02 g 0,05 g 0,05 g 0,1 g Weighing range (max) 6.100 g 8.100 g 16.100 g 16.100 g Taring range (subtractive) 6.100 g 8.100 g...
  • Page 6 KERN CKE 36k0.1 CKE 36k0.2 CKE 65k0.2 CKE 65k0.5 Readability (d) 0,1 g 0,2 g 0,2 g 0,5 g Weighing range (max) 36.100 g 36.100 g 65.100 65.100 Taring range (subtractive) 36.100 g 36.100 g 65.100 65.100 Reproducibility 0,1 g...
  • Page 7: Basic Information (General)

    The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com) with regard to the monitoring of balance test substances and the test weights required for this. Our accredited DKD calibration laboratory offers fast and inexpensive adjustment for test weights and weighing balances (reset to national normal weight).
  • Page 8: Basic Safety Precautions

    4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances. 4.2 Personnel training The appliance may only be operated and maintained by trained personnel.
  • Page 9: Unpacking, Setup And Commissioning

    Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the balance from the power supply. With your balance, only use accessories and peripheral devices by KERN, as they are ideally tuned to your balance.
  • Page 10: Adjustment

    6.6 Adjustment As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out during the initial start-up, after change in location and variation of surrounding temperature.
  • Page 11: Operation

    7 Operation 7.1 Control panel CKE ON/OFF Taring; Invoke the mode menu by pressing the ON/OFF key. Generate reference using 5 parts; In mode menu: NO function Generate reference using 10 parts Generate reference using 20 parts; In mode menu: YES function Switchover pcs ↔...
  • Page 12: Operation

    7.2 Operation CKE-BA-e-0913...
  • Page 13 CKE-BA-e-0913...
  • Page 14: Volume Assessment - Kern Qke Only

    7.2.1 Volume assessment - KERN QKE only !!! 7.2.1.1 Volume assessment without manual input Switch on balance Place an empty vessel and tare Place a vessel containing ideal filling (100%) and save the value by pressing the MAX key It is now possible to qualify additional filled vessels by using the percentage display (%) 7.2.1.2 Volume assessment with manual tare input...
  • Page 15: Data Output Rs 232 C

    Baud rate selectable from 2400, 4800, 9600 Baud (factory setting) and 19200 Baud. Sub-D plug 9-channel required For operation with interface faultless operation is only ensured with the correct KERN – interface cable (max. 2m) Pin allocation of the balance output socket (front view)
  • Page 16: Interface Rs 232C

    7.4 Interface RS 232C Data output via interface RS 232C General Information The previous condition for the data transfer between balance and a peripherical device (e.g. printer, PC ...) is that the appliances are set to the same interface parameters (e.g. baud rate, parity ...). 7.4.1 There are 4 kinds of data output via RS 232C Data output using the PRINT key The printing process can be triggered by pressing the PRINT key.
  • Page 17: Printer

    Printer The serial interface RS 232 facilitates the connection of a printer. The printout shows the weight in grams. In counting mode either the piece number or the weight details will be printed out. In percentage mode the percentage proportion or the weight details will be printed out. Printout will take place after pressing the PRINT key.
  • Page 18: Service, Maintenance, Disposal

    Spilled weighing goods must be removed immediately. 8.2 Service, maintenance The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply. 8.3 Disposal Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
  • Page 19: Declaration Of Conformity

    EN 55011:1998+A1:1999+A2 :2002 EN 61000-3-2 :2006-04 EN61000-3-3 :1995-01+A1:2001-06+A2 :2005-11 EN45501 :1992-10+AC :1993-08 OIML R 76-1 :2006 2006/95/EC EN60950 Date: 26.11.2008 Signature: Gottl. KERN & Sohn GmbH Management Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149 CKE-BA-e-0913...

Table of Contents