HP MODEL8565A SPECTRUM ANAlYZER
All Serials
IMPROVED FREQUENCY RESPONSE PERFORMANCE TEST
The following is the recommended procedure for frequency response testing of all HP
Model 8565A Spectrum Analyzers. Replace test paragraph 4-20 in your Operation and
Service Manual with the service note.
A synchronizer and sweep oscillator are connected to make a tracking generator for the
8565A. The sweep oscillator is phase locked across each FREQUENCY BAND, and fre-
quency response tests are performed. The YTF is modulated with a 1 kHz sine wave to
eliminate amplitude variations due to small errors in YTF Tracking.