GE B40 Reference Manual

GE B40 Reference Manual

Patient monitor
Hide thumbs Also See for B40:
GE Healthcare
B40/B20 Patient Monitor
HL7 Reference Manual
B40/B20 Patient Monitor
English
2062665-001 G (paper)
© 2014 General Electric Company.
All Rights Reserved.
Table of Contents
loading

Summary of Contents for GE B40

  • Page 1 GE Healthcare B40/B20 Patient Monitor HL7 Reference Manual B40/B20 Patient Monitor English 2062665-001 G (paper) © 2014 General Electric Company. All Rights Reserved.
  • Page 2 NOTE: The information in this manual applies to GE B40/B20 Patient Monitor’s software VSP-B, VSP-C and later. It does not apply to earlier software versions. Due to continuing product innovation, specifications in this manual are subject to change without notice.
  • Page 3: Table Of Contents

    Contents Introduction ........1 Manual Information ..........2 Revision History .
  • Page 4 PCD-01 ........12 PCD-01 description .
  • Page 5 Abbreviations ........45 Abbreviations ........... . 46 2062665-001 G...
  • Page 6 2062665-001 G...
  • Page 7: Introduction

    Introduction: Introduction 2062665-001 G...
  • Page 8: Manual Information

    The device can be connected to hospital information systems (HIS) by the Health Level Seven standard interface (HL7). This document: Intended to be a technical reference to GE customers implementing the HL7  outbound interface. Not intended to provide instructions for implementing and using the HL7 ...
  • Page 9: Manual Conventions

    Any unauthorized attempt to repair equipment under warranty voids the  warranty. For GE’s products, it is the user’s responsibility to report the need for service  to GE or to one of their authorized agents. Failure on the part of the responsible individual, hospital, or institution using ...
  • Page 10: Safety Message Signal Words

    Introduction: Service Information Safety message signal words Safety message signal words designate the severity of a potential hazard. WARNING indicates a hazardous situation that, if not avoided, could result in death or serious injury. CAUTION indicates a hazardous situation that, if not avoided, could result in minor or moderate injury.
  • Page 11: Hl7 Interface

    HL7 Interface: Service Information HL7 Interface 2062665-001 G...
  • Page 12: Hl7 Standard Background

    HL7 Interface: HL7 Standard Background HL7 Standard Background Overview The Health Level Seven (HL7) standard is used to exchange data between computer systems. It does not require a specific computer operating system, programming language, or communication protocol for its implementation. The goal of the HL7 standard is to harmonize message content and usage, while allowing user-specific variations within the standard.
  • Page 13: Hl7 Communication

     data handling  Communication between HIS personnel and GE interface personnel is essential to determine the customer-specific use of the HL7 standard. NOTE: This document is not intended to provide instructions for implementing and using the HL7 standard. Details of the HL7 standard can be found in Health Level Seven Version 2.6.
  • Page 14: Observation Results Interface

    Observation results interface: HL7 Standard Background Observation results interface 2062665-001 G...
  • Page 15: General Description

    Observation results interface: General Description General Description Introduction The HL7 standard interface is used to connect a hospital’s information system (HIS) to the monitor for the transfer of data. Using the HL7 outbound interface, the monitor can return result messages to the HIS. The Patient Monitor Outbound implementation of the HL7 outbound interface protocol provides result messages.
  • Page 16: Low Level Communications

    The hospital or the third party can choose GE offered switch or other types.  If you need to use GE offered switch, please contact to GE ND&I Team to setup CARESCAPE Network Infrastucture version 2. If hospital or the third party design networks by themselves, the hospital or ...
  • Page 17: Supported Numeric Data Categories

    If the device or the HIS server connection is interrupted from the network (e.g., disconnected, in transport, etc.), the trended data for that time period will be lost. NOTE: GE recommends that you enabling the HL7 acknowledgment option in monitor’s service menu. Disabling this option may miss some data, if the HL7 receiver system has the feature of sending acknowledge message.
  • Page 18: Pcd-01

    PCD-01: HIS Disconnection or Network Outage PCD-01 2062665-001 G...
  • Page 19: Pcd-01 Description

     The value of this identifier is composed of two components. Component #1 is an EUI-64 identifier that starts with the GE Healthcare company code, plus the EUI-64 separator, followed by the last 3 octects of the system's MAC address in hex upper-case with no separators., plus a string that identifies the monitoring equipment.
  • Page 20: Obx-4

    PCD-01: PCD-01 description OBX-4 Condition predicate: This field is used to distinguish between multiple OBX segments by providing an unambiguous mapping from observation contained in the OBX segment to the IEEE 11073 containment tree for the Medical Device System sourcing of the observation.
  • Page 21 PCD-01: PCD-01 description This is a logical grouping of semantically related observations and numeric categories. The OBXs that follow are the actual results for the VMD. OBX|11|NM|147842^MDC_ECG_HEART_RATE^MDC|1.5.1.1|80|26 4864^MDC_DIM_BEAT_PER_MIN^MDC|||||R|||||||080019FFFE0 B4020^B40_GE OBX|12|NM|148066^MDC_ECG_V_P_C_RATE^MDC|1.5.1.2|0|264 864^MDC_DIM_BEAT_PER_MIN^MDC|||||R|||||||080019FFFE0B 4020^B40_GE Invasive/Non Invasive Blood Pressure also have an explicit channel OBX as shown in the example below.
  • Page 22: Vmd Reference Ids, Mdc Codes, And Containment Trees

    PCD-01: PCD-01 description VMD Reference IDs, MDC Codes, and Containment Trees The following table lists all of the defined VMD and Channel related to each numeric category. Parameter VMD Reference ID VMD MDC Containment Code MDC_DEV_ECG_VMD 69798 1.5.0.0 ECG RESP MDC_DEV_ECG_RESP_VMD 70666 1.6.0.0...
  • Page 23: Sample Messages

    Sample Messages: PCD-01 description Sample Messages 2062665-001 G...
  • Page 24: Hl7 Minimal Lower Layer Protocol

    Sample Messages: HL7 minimal lower layer protocol HL7 minimal lower layer protocol HL7 Outbound messages are framed using the HL7 Minimal Lower Layer Protocol. This protocol is documented in the HL7 2.6 specification. All HL7 messages begin with one start block character. The framing is completed by an end character and a carriage return character at the end of the message.
  • Page 25: Result Message Examples

    This example has one OBR segment followed by the corresponding OBX segments collected at a user-configured time interval. The OBX segments contain periodic vital sign data for heart rate, PVC, temperature, and invasive blood pressures. MSH|^~\&|VSP^080019FFFE134535^EUI-64|GE Healthcare|||20130122134100||ORU^R01^ORU_R01|00409713 4535|P|2.6|||NE|AL||UNICODE UTF-8|||PCD_DEC_001^IHE PCD^1.3.6.1.4.1.19376.1.6.1.1.1^ISO...
  • Page 26 864^MDC_DIM_BEAT_PER_MIN^MDC|||||R|||||||080019FFFE13 4535^B40_GE OBX|13||69902^MDC_DEV_METER_TEMP_VMD^MDC|1.26.0.0|||| |||X OBX|14|NM|150344^MDC_TEMP^MDC|1.26.1.1|37.1|268192^MD C_DIM_DEGC^MDC|||||R|||||||080019FFFE134535^B40_GE Example 2 This example shows an OBR segment followed by the corresponding OBX segments containing various periodic vital sign data collected at a user configured time interval. MSH|^~\&|VSP^080019FFFE0B4020^EUI-64|GE Healthcare|||20121109160900||ORU^R01^ORU_R01|000C290B 4020|P|2.6|||NE|AL||UNICODE|||PCD_DEC_001^IHE PCD^1.3.6.1.4.1.19376.1.6.1.1.1^ISO PID|||999999999^^^PID^MR||^^^^^^L||| 2062665-001 G...
  • Page 27 Sample Messages: Transactions from the monitor HL7 interface PV1||E|X^^B4020 OBR|1|080019FFFE0B402020121109160900^VSP^080019FFFE0B 4020^EUI- 64|080019FFFE0B402020121109160900^VSP^080019FFFE0B402 0^EUI-64|182777000^monitoring of patient^SCT|||20121109160900 OBX|1||69965^MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS^MDC|1 .0.0.0|||||||X OBX|2||69854^MDC_DEV_METER_PRESS_BLD_VMD^MDC|1.13.0.0 |||||||X OBX|3||69855^MDC_DEV_METER_PRESS_BLD_CHAN^MDC|1.13.1. 0|||||||X OBX|4|NM|150033^MDC_PRESS_BLD_ART_SYS^MDC|1.13.1.1|11 2|266016^MDC_DIM_MMHG^MDC|||||R|||||||080019FFFE0B402 0^B40_GE OBX|5|NM|150034^MDC_PRESS_BLD_ART_DIA^MDC|1.13.1.2|76 |266016^MDC_DIM_MMHG^MDC|||||R|||||||080019FFFE0B4020 ^B40_GE OBX|6|NM|150035^MDC_PRESS_BLD_ART_MEAN^MDC|1.13.1.3|9 5|266016^MDC_DIM_MMHG^MDC|||||R|||||||080019FFFE0B402 0^B40_GE OBX|7|NM|149522^MDC_BLD_PULS_RATE_INV^MDC|1.13.1.4|80 |264864^MDC_DIM_BEAT_PER_MIN^MDC|||||R|||||||080019FF FE0B4020^B40_GE OBX|8||69855^MDC_DEV_METER_PRESS_BLD_CHAN^MDC|1.13.2. 0|||||||X OBX|9|NM|150087^MDC_PRESS_BLD_VEN_CENT_MEAN^MDC|1.13. 2.1|9|266016^MDC_DIM_MMHG^MDC|||||R|||||||080019FFFE0 B4020^B40_GE OBX|10||69798^MDC_DEV_ECG_VMD^MDC|1.5.0.0|||||||X OBX|11|NM|147842^MDC_ECG_HEART_RATE^MDC|1.5.1.1|80|26 4864^MDC_DIM_BEAT_PER_MIN^MDC|||||R|||||||080019FFFE0 B4020^B40_GE OBX|12|NM|148066^MDC_ECG_V_P_C_RATE^MDC|1.5.1.2|0|264...
  • Page 28 Sample Messages: Transactions from the monitor HL7 interface OBX|14|NM|131842^MDC_ECG_AMPL_ST_II^MDC|1.5.1.4|0.04| 266419^MDC_DIM_MICRO_VOLT^MDC|||||R|||||||080019FFFE0 B4020^B40_GE OBX|15|NM|131901^MDC_ECG_AMPL_ST_III^MDC|1.5.1.5|0.04 |266419^MDC_DIM_MICRO_VOLT^MDC|||||R|||||||080019FFFE 0B4020^B40_GE OBX|16|NM|131843^MDC_ECG_AMPL_ST_V1^MDC|1.5.1.6|0.04| 266419^MDC_DIM_MICRO_VOLT^MDC|||||R|||||||080019FFFE0 B4020^B40_GE OBX|17|NM|131844^MDC_ECG_AMPL_ST_V2^MDC|1.5.1.7|0.00| 266419^MDC_DIM_MICRO_VOLT^MDC|||||R|||||||080019FFFE0 B4020^B40_GE OBX|18|NM|131845^MDC_ECG_AMPL_ST_V3^MDC|1.5.1.8|0.00| 266419^MDC_DIM_MICRO_VOLT^MDC|||||R|||||||080019FFFE0 B4020^B40_GE OBX|19|NM|131846^MDC_ECG_AMPL_ST_V4^MDC|1.5.1.9|0.00| 266419^MDC_DIM_MICRO_VOLT^MDC|||||R|||||||080019FFFE0 B4020^B40_GE OBX|20|NM|131847^MDC_ECG_AMPL_ST_V5^MDC|1.5.1.10|0.00 |266419^MDC_DIM_MICRO_VOLT^MDC|||||R|||||||080019FFFE 0B4020^B40_GE OBX|21|NM|131848^MDC_ECG_AMPL_ST_V6^MDC|1.5.1.11|0.00 |266419^MDC_DIM_MICRO_VOLT^MDC|||||R|||||||080019FFFE 0B4020^B40_GE OBX|22|NM|131902^MDC_ECG_AMPL_ST_AVR^MDC|1.5.1.12|- 0.04|266419^MDC_DIM_MICRO_VOLT^MDC|||||R|||||||080019 FFFE0B4020^B40_GE OBX|23|NM|131903^MDC_ECG_AMPL_ST_AVL^MDC|1.5.1.13|0.0 0|266419^MDC_DIM_MICRO_VOLT^MDC|||||R|||||||080019FFF E0B4020^B40_GE OBX|24|NM|131904^MDC_ECG_AMPL_ST_AVF^MDC|1.5.1.14|0.0 4|266419^MDC_DIM_MICRO_VOLT^MDC|||||R|||||||080019FFF E0B4020^B40_GE OBX|25||69766^MDC_DEV_GEN_CONC_AWAY_VMD^MDC|1.11.0.0| ||||||X...
  • Page 29 Sample Messages: Transactions from the monitor HL7 interface OBX|28|NM|151594^MDC_CO2_RESP_RATE^MDC|1.11.1.3|18|26 4928^MDC_DIM_RESP_PER_MIN^MDC|||||R|||||||080019FFFE0 B4020^B40_GE OBX|29|NM|151748^MDC_AWAY_O2_EXP^MDC|1.11.1.4|16|2626 88^MDC_DIM_PERCENT^MDC|||||R|||||||080019FFFE0B4020^B 40_GE OBX|30|NM|152196^MDC_CONC_AWAY_O2_INSP^MDC|1.11.1.5|2 1|262688^MDC_DIM_PERCENT^MDC|||||R|||||||080019FFFE0B 4020^B40_GE OBX|31|NM|152320^MDC_AWAY_N2O_EXP^MDC|1.11.1.6|5|2626 88^MDC_DIM_PERCENT^MDC|||||R|||||||080019FFFE0B4020^B 40_GE OBX|32|NM|152192^MDC_CONC_AWAY_N2O_INSP^MDC|1.11.1.7| 8|262688^MDC_DIM_PERCENT^MDC|||||R|||||||080019FFFE0B 4020^B40_GE OBX|33|NM|0^MDC_CONC_MAC_SUM^MDC|1.11.1.8|0|262656^MD C_DIM_DIMLESS^MDC|||||R|||||||080019FFFE0B4020^B40_GE OBX|34|NM|0^MDC_CONC_MAC_SUM_AGE_CORR^MDC|1.11.1.9|0| 262656^MDC_DIM_DIMLESS^MDC|||||R|||||||080019FFFE0B40 20^B40_GE OBX|35|NM|152312^MDC_AWAY_ISOFL_EXP^MDC|1.11.1.10|0|2 62688^MDC_DIM_PERCENT^MDC|||||R|||||||080019FFFE0B402 0^B40_GE OBX|36|NM|152184^MDC_CONC_AWAY_ISOFL_INSP^MDC|1.11.1. 11|1|262688^MDC_DIM_PERCENT^MDC|||||R|||||||080019FFF E0B4020^B40_GE OBX|37||0^MDC_DEV_EEG_ENTROPY_VMD^MDC|1.8.0.0|||||||X OBX|38|NM|155024^MDC_EEG_PAROX_CRTX_BURST_SUPPRN^MDC| 1.8.1.1|0|262688^MDC_DIM_PERCENT^MDC|||||R|||||||0800 19FFFE0B4020^B40_GE OBX|39|NM|0^MDC_EEG_ENTROPY_RESPONSE^MDC|1.8.1.2|81|2 62656^MDC_DIM_DIMLESS^MDC|||||R|||||||080019FFFE0B402 0^B40_GE OBX|40|NM|0^MDC_EEG_ENTROPY_STATE^MDC|1.8.1.3|75|2626 56^MDC_DIM_DIMLESS^MDC|||||R|||||||080019FFFE0B4020^B...
  • Page 30: Application High Level Acknowledgment Messages

    HL7 receiver system. The default acknowledgment configuration is Enable in monitor HL7 service menu. NOTE: GE recommends enabling the HL7 acknowledgement option in the monitor's service menu. Disabling this option may miss some data, if the HL7 receiver system has the feature of sending acknowledge message.
  • Page 31: Acknowledgment Message Examples

    Sample Messages: Transactions from the monitor HL7 interface MSA — Message segment acknowledgment  HL7 acknowledgment messages may contain one of statuses from the receiving system: Application Accept (AA)  Commit Accept (CA)  Application Error (AE)  Application Reject (AR) ...
  • Page 32 Sample Messages: Transactions from the monitor HL7 interface MSA|AR|17f1be29-aa2a-11d2-a655- 00105a1e9b67|Unsupported message type Acknowledgment Error MSH|^~\&|NUR^^|.^^|MMS HL7|.^^|199812021002||ACK^R01|17f1be29-aa2a-11d2- a655-00105a1e9b67|P^|2.3 MSA|AE|17f1be29-aa2a-11d2-a655-00105a1e9b67|Unable to find patient 2062665-001 G...
  • Page 33: Hl7 Data Segment Definitions

    HL7 Data Segment Definitions: Transactions from the monitor HL7 interface HL7 Data Segment Definitions 2062665-001 G...
  • Page 34: Introduction

    The following tables define the various data segments and are shown with the standard HL7 message layout. The lengths given for fields are the maximum lengths supported by the GE system, not the HL7 definitions for field lengths. When the GE HL7 interface support column indicates N, the data may be optimally sent in the message but it will be ignored by the GE HL7 interface.
  • Page 35 HL7 Data Segment Definitions: Introduction Abbreviations Description Not used with this trigger event. Data Types Description Coded element Structured numeric Composite ID w/check digit String Composite quantity w/units Timing quantity Extended composite ID w/check digit Time stamp Driver's license number Text data Date Extended composite ID number...
  • Page 36: Data Segments

    HL7 Data Segment Definitions: Data Segments Data Segments MSH — Message Header Segment REQ/OPT TABLE# ELEMENT NAME Field Separator Encoding Characters 0361 Sending Application 0362 Sending Facility 0361 Receiving Application 0362 Receiving Facility Date/Time of Message Security Message Type Message Control ID Processing ID Version ID Sequence Number...
  • Page 37: Msa - Message Acknowledgment

    HL7 Data Segment Definitions: Data Segments MSA — Message Acknowledgment REQ/OPT TABLE# ELEMENT NAME 0008 Acknowledgement code Message Control Id Text Message Delayed Acknowledgment Type 0357 Error Condition PID — Patient Identification REQ/OPT TBL# ITEM# ELEMENT NAME 00104 Set ID - PID 00105 Patient ID 00106...
  • Page 38 HL7 Data Segment Definitions: Data Segments REQ/OPT TBL# ITEM# ELEMENT NAME 00124 Mother's Identifier 0189 00125 Ethnic Group 00126 Birth Place 0136 00127 Multiple Birth Indicator 00128 Birth Order 0171 00129 Citizenship 0172 00130 Veterans Military Status 0212 00739 Nationality 00740 Patient Death Data and Time 0136...
  • Page 39: Pv1 - Patient Visit

    HL7 Data Segment Definitions: Data Segments PV1 — Patient Visit REQ/OPT TABLE# ELEMENT NAME Set ID - PV1 0004 Patient Class Patient Location 0007 Admission Type Preadmit Number Prior Patient Location 0010 Attending Doctor 0010 Referring Doctor 0010 Consulting Doctor 0069 Hospital Service Temporary Location...
  • Page 40 HL7 Data Segment Definitions: Data Segments REQ/OPT TABLE# ELEMENT NAME Transfer to Bad Debt Date 0021 Bad Debt Agency Code Bad Debt Transfer Amount Bad Debt Recovery Amount 0111 Delete Account Indicator Delete Account Date 0112 Discharge Disposition 0113 Discharged to Location 0114 Diet Type 0115...
  • Page 41: Obr - Observation Request (To Host)

    HL7 Data Segment Definitions: Data Segments OBR — Observation Request (To Host) USAGE ELEMENT NAME Set ID Placer Order Number Filler Order Number Universal Service Identifier Priority Requested Date/Time Observation Date/Time OBX — Result USAGE TABLE# ELEMENT NAME Set Id 0125 Value Type Observation Identifier...
  • Page 42 HL7 Data Segment Definitions: Data Segments USAGE TABLE# ELEMENT NAME Date/Time of the Analysis 0163 Observation Site 2062665-001 G...
  • Page 43: Supported Parameters

    HL7 Data Segment Definitions: Data Segments Supported parameters 2062665-001 G...
  • Page 44: Supported Parameters

    HL7 Data Segment Definitions: Supported parameters Supported parameters NOTE: Some parameters in this table are only available from certain versions of the acquisition source. NOTE: This list is subject to change without notice. HL7 Supported Parameters Invasive Pulmonary PA-S mm[Hg] MDC_PRESS_BLD_ART_ 150045 MDC_DIM_MMH...
  • Page 45 HL7 Data Segment Definitions: Supported parameters Arterial ART-M mm[Hg] MDC_PRESS_BLD_ART_ 150035 MDC_DIM_MMH 266016 Pressure ABP_MEAN Mean Abdominal ABP-S mm[Hg] MDC_PRESS_BLD_ART_ 150037 MDC_DIM_MMH 266016 Pressure ABP_SYS Systolic Abdominal ABP-D mm[Hg] MDC_PRESS_BLD_ART_ 150038 MDC_DIM_MMH 266016 Pressure ABP_DIA Diastolic Abdominal ABP-M mm[Hg] MDC_PRESS_BLD_ART_ 150039 MDC_DIM_MMH 266016...
  • Page 46 HL7 Data Segment Definitions: Supported parameters IBP1 - Mean P1 - mm[Hg] MDC_PRESS_BLD_MEA 150019 MDC_DIM_MMH 266016 Mean IBP2 - P2 - mm[Hg] MDC_PRESS_BLD_SYS 150017 MDC_DIM_MMH 266016 Systolic Systolic IBP2 - P2 - mm[Hg] MDC_PRESS_BLD_DIA 150018 MDC_DIM_MMH 266016 Diastolic Diastolic IBP2 - Mean P2 - mm[Hg] MDC_PRESS_BLD_MEA...
  • Page 47 HL7 Data Segment Definitions: Supported parameters Premature {beat}/ MDC_V_P_C_RATE 148065 MDC_DIM_BEAT 264864 Ventricular _PER_MIN Count Respiration IMPR-RR {breath} MDC_RESP_RATE 151562 MDC_DIM_RESP 264928 Rate - ECG /min _PER_MIN Expired CO2 CO2-EX mm[Hg] MDC_CONC_AWAY_CO 151712 MDC_DIM_MMH 266016 2_EXP Inspired CO2 CO2-IN mm[Hg] MDC_CONC_AWAY_CO 151716 MDC_DIM_MMH...
  • Page 48 HL7 Data Segment Definitions: Supported parameters Inspired ENF-IN MDC_CONC_AWAY_EN 152172 MDC_DIM_PERC 262688 Enflurane FL_INSP Expired DES-EX MDC_AWAY_DESFL_EX 152296 MDC_DIM_PERC 262688 Desflurane Inspired DES-IN MDC_CONC_AWAY_DE 152168 MDC_DIM_PERC 262688 Desflurane SFL_INSP Expired SEV-EX MDC_AWAY_SEVOFL_E 152308 MDC_DIM_PERC 262688 Sevoflurane Inspired SEV-IN MDC_CONC_AWAY_SEV 152180 MDC_DIM_PERC 262688 Sevoflurane...
  • Page 49 HL7 Data Segment Definitions: Supported parameters ST Segment ST-AVF MDC_ECG_AMPL_ST_A 131904 MDC_DIM_MICR 266419 O_VOLT Non- Noninvasive NBP-S mm[Hg] MDC_PRESS_BLD_NONI 150021 MDC_DIM_MMH 266016 Invasive BP Systolic NV_SYS Blood Pressure (NIBP) Noninvasive NBP-D mm[Hg] MDC_PRESS_BLD_NONI 150022 MDC_DIM_MMH 266016 BP Diastolic NV_DIA Noninvasive NBP-M mm[Hg] MDC_PRESS_BLD_NONI...
  • Page 50 HL7 Data Segment Definitions: Supported parameters Skin TP-Skin MDC_TEMP_SKIN 150388 MDC_DIM_DEGC 268192 Temperature Airway TP-AirW MDC_TEMP_AWAY 150356 MDC_DIM_DEGC 268192 Temperature Room MDC_TEMP_ROOM MDC_DIM_DEGC 268192 Temperature Room Myocardial TP-Myo MDC_TEMP_MYO MDC_DIM_DEGC 268192 Temperature Core TP-Core MDC_TEMP_CORE 150368 MDC_DIM_DEGC 268192 Temperature Surface TP-Surf MDC_TEMP_BODY 150364...
  • Page 51 HL7 Data Segment Definitions: Supported parameters Abbreviations 2062665-001 G...
  • Page 52 HL7 Data Segment Definitions: Abbreviations Abbreviations Abbreviations and symbols which you may encounter while reading this manual or using the monitor HL7 application are listed below with their meanings. Abbreviation Definition acknowledgment code Addr, AD address Admit, discharge, transfer ASCII American Standard Code for Information Interchange Addendum...
  • Page 53 HL7 Data Segment Definitions: Abbreviations Merge Segment Message Segment Header Message Segment Acknowledgment NACK negative acknowledgment code numeric number observation request observation observation result optional Patient Data Model part number, person name Patient Identification Patient Visit Query Definition request Req. Field required field revision sequence...
  • Page 54 HL7 Data Segment Definitions: Abbreviations 2062665-001 G...
  • Page 56 Headquarters Asian Headquarters GE Medical Systems GE Medical Systems GE Healthcare Finland Oy Information Technologies Asia Information Technologies, Inc. Kuortaneenkatu 2 1 Huatuo Road 8200 West Tower Avenue Zhangjiang Hi-tech Park Pudong FI-00510 Milwaukee, WI 53223 USA Shanghai, P.R. China, 201203...