Centricity™ RA600 Service Manual
RadWorks Release 6.0
FP10516
Revision 2
Published—August 2002
Do not duplicate.
© GE Medical Systems Information Technologies, 2002, All rights reserved
Table of Contents
loading

Summary of Contents for GE Centricity RA600

  • Page 1 GE Medical Systems Information Technologies Centricity™ RA600 Service Manual RadWorks Release 6.0 FP10516 Revision 2 Published—August 2002 Do not duplicate. © GE Medical Systems Information Technologies, 2002, All rights reserved...
  • Page 2 Regulatory Requirements This product complies with the regulatory requirements of the following: Medical Device Quality Systems Regulation, FDA (Food and Drug Administration, Department of Health and Human Services, USA). Underwriters’ Laboratories, Inc. (UL), an independent testing laboratory. Canadian Standards Association (CSA). International Electrotechnical Commission (IEC), international standards organization, when applicable.
  • Page 3 increase the separation between the equipment and the affected device power the equipment from a source different from that of the affected device consult the point of purchase or service representative for further suggestions The manufacturer is not responsible for any interference caused by using other than recommended interconnect cables or by unauthorized changes or modifications to this equipment.
  • Page 4 Rate gezogen und verstanden wurde. • Wird diese warnung nicht beachtet, so kann es zu Verletzungen des Kundendienst- technikers, des Bedieners oder des Patienten durch elektrische Schläge, mecha- nische oder sonstige Gefahren kommen. • Este Manual de Servicio sólo existe en inglés.
  • Page 5 • Este Manual de Assistência Técnica só se encontra disponível em Inglês. • Se qualquer outro serviço de assistência técnica, quE não a GEMS, solicitar estes ATENÇAO manuais noutro idioma, é da responsabilidade do cliente fornecer os serviços de tradução. •...
  • Page 6 Revision History Date Author Reason For Change 3/2002 Initial release for Centricity RA600 4/2002 SA/WB Add work group server, streaming video, Europe’s dongle replacement procedure, revised initial powerup config., Flat panel display coverage & other miscellaneous minor fixes 8/2002 DB/SA/ Add RA600 Identification &...
  • Page 7: Table Of Contents

    Table of Contents CHAPTER 1 INTRODUCTION ........11 CHAPTER 2 INSTALLATION STEERING GUIDE .
  • Page 8 Restoring RA600 Original System Configuration ....177 SW 002 Installing Centricity RA600 Software ......180 SW 003 (C, 2C) Matrox Video Board Driver Installation .
  • Page 9 TSG 006 Display Windows 2000 System Logs ......244 TSG 007 Display Centricity RA600 Application Logs ..... . . 245 TSG 008 Display IP/Netmask/Default Route Configuration Settings .
  • Page 10 TSG 013 Display Application Software Version ......251 TSG 014 Workstation Application Problem Solving ......252 TSG 015 Connect Frame Grabber to Monitor that has no pass-through .
  • Page 11: Chapter 1 Introduction

    Introduction Section 1: Introduction Throughout this manual the common term Workstation is used to refer to the Centricity RA600 Workstation which represents the installed configuration of monitor(s) and computer. Section 2: Chapter Descriptions Shown below is a summary of the chapters in this book and a description of each chapter.
  • Page 12 Chapter 10 Suggested Maintenance. Suggested maintenance procedures to be performed every six months. Section 3: Related Documentation Here is a list of documents related to the Centricity RA600 workstation. GE Document Number Document Name FP10515 Centricity RA600 Workstation Pre-Installation Manual OI10091 Centricity RA600 / RadWorks 6.0 User’s Guide...
  • Page 13 A VGA service monitor for system maintenance tasks. Illustration 1-1 shows a workstation configuration with two Barco CRT portrait monitors and the service monitor. Barco Monitors Computer Unit VGA Service Keyboard Mouse Monitor Illustration 1-1: Centricity RA600 showing BARCO monitors & service monitor Page 13...
  • Page 14 (See appropriate job card in Chapter 3 for cabling and the location of video board in Computer Unit.) Keyboard Standard 104 key Mouse 2-button with scrolling wheel. 4-2 Monitor Configurations These are the monitors used with the Centricity RA600. Refresh Video Model Type Type Resolution Rate Board...
  • Page 15 The service monitor is used to run diagnostic tests on the A+, A, B and B+ model workstations. Model Type Type Resolution Refresh Rate Video Board VGA Service Monitor Color 640 x 480 default onboard 4-4 Software Items Software Items Installed Item Description Windows 2000 (with SP2) Operating System Centricity RA600 Application Software Page 15...
  • Page 16 Introduction Illustration 1-2: Computer Unit, Front View Page 16...
  • Page 17 Introduction Illustration 1-3: Computer Unit, Rear View Section 5: List of Abbreviations The following abbreviations and terms are used throughout this manual. Abbreviation/Term Definition Refers to the Computer Unit, (Central Processor Unit) Internet Address, (Internet Protocol) Page 17...
  • Page 18 DICOM images, not as an internet access point. The machine will be most secure when set up with limited access and secure user account names and passwords. Only trained GE personnel and system administrators should add, modify, or delete pre-approved content on the workstation computer, minimizing virus vulnerability.
  • Page 19 Section 8: Dell Service Strategies 8-1 Service Strategy for Dell Equipment The local site FE (Field Engineer) will do the installation of all Centricity RA600 Workstation components (including Dell CPU/Server). When the customer calls into GE CARES, the ROC (Remote Online Center) will investigate the service call and determine if it’s related to the Dell equipment.
  • Page 20 Introduction not be valid outside the US. This transfer may take up to 3 weeks. If service needs to be done before the transfer, please contact the EMEA Program Manager at Dell for an immediate transfer. EMEA Program Manager Global Segment Tel: +44 (0) 1344 748931 Fax:...
  • Page 21 Customer Contact Phone Hardware Information Item Name Number CPU Type (Mfr., Model, Service Nol, etc.) RAM amount Hard Disk Size RAID Unit (Mfr., Model, Size) Software Information Item Name Number Centricity RA600 Software Version No. Centricity RA600 Dongle/License Serial No. Page 21...
  • Page 22 Subnet Mask AE Title Internet Connection Gateway Address Remote Connectivity Software Installed (VNC, PC Anywhere, etc.) VPN Connection Available for GE Service? (If yes, type and connection info) Dial -In 56KB Modem Phone No. RadWorks Installed Options Item Name Data / Value...
  • Page 23 Introduction Item Name Data / Value Archive Module Digitizer (Y/N) Digitizer ID (Mfr, Model, S/N) QC Module (Y/N) Dicom print option (Y/N) Dicom printer (Mfr, Model, S/N) Integration module (Y/N) Other Modalities Connected to the RA600/RadWorks System (Indicate direct/remote connection & if remote, type of wide-area network (WAN) used: POTS, DSL, cable, etc.) Modality Description...
  • Page 24 Introduction Modality Description IP Address S.M. Port AE Title Type Page 24...
  • Page 25: Chapter 2 Installation Steering Guide

    Installation steering guide Section 1: Introduction This chapter provides directions on how to install and update software in the Centricity RA600 Workstation. Section 2: Steering Flowchart Illustration 2-1 page 27 is an overall process steering diagram for installing the Centricity RA600 Workstation.
  • Page 26 Installation steering guide Physical Software Install/ Optional Item Installation Configure Video Board Driver for C monitor SW 003 Driver for A, A+, B, SW 004 Driver for B+ CNF 007 Frame Grabber board IST 008 FP10621 (kit document) (I-Color Static Frame Capture System) Single Media Archive CD-R IST 007 SW 006...
  • Page 27 (Not needed for new installation) SW001- Load from cold (Ghost) SW002- Load from warm Install any optional peripheral device(s). See Table 2-1 Configure Centricity RA600 system to site CNF001 Install & configure any peripheral software. See Table 2-1 Install any software patch(es)
  • Page 28: Chapter 3 Hardware Installation

    Hardware Installation Section 1: Summary of Job CardS Job Card No Hardware Installation Job Card Title IST 001 (A+) 2048 x 2506 Portrait Monitor Interconnections IST 002 (A) 1728 x 2304 Portrait Monitor Interconnections IST 003 (B) 1200 x 1600 Portrait Monitor Interconnections IST 004 (B+) 1536 x 2048 Flat Panel Portrait Monitor Connections IST 005...
  • Page 29: Ist 001 (A+) 2048 X 2506 Portrait Monitor Interconnections

    RA600 Workstation is 1 gauss to ensure proper operation. Section 2: Prerequisites The site must be prepared for installation of the workstation. Refer to the Centricity RA600 Workstation Pre-Installation Manual for detailed information. The appropriate line voltage, line frequency, and power supply line cords.
  • Page 30 Service Monitor Tool Table 3-1: Items used for installation of 2048 x 2506 workstation The Centricity RA600 Workstation should already be configured as per the factory settings. Also, the operating system software and applications software are installed at the factory. The person installing the workstation must physically interconnect workstation cables, and then configure the system using site-specific information (i.e.
  • Page 31 Hardware Installation IST 001 (A+) 2048 x 2506 Portrait Monitor Interconnections 3-5 Keyboard & Mouse Connections Connect the mouse and keyboard to the computer housing. 3-6 Monitor Connections Connect the monitors as shown in appropriate cabling illustration (two or four monitors). Interface interconnections to two Barco “A”...
  • Page 32 Hardware Installation IST 001 (A+) 2048 x 2506 Portrait Monitor Interconnections Interface interconnections to four Barco “A” 2048 x 2560 portrait monitors Video card connectors Monitor cable connector to Slot 1 Blank video interface Slot 2 Blank Slot 3 Blank Calibrator connector Slot 4 Video Card...
  • Page 33: Ist 002 (A) 1728 X 2304 Portrait Monitor Interconnections

    RA600 Workstation is 1 gauss to ensure proper operation. Section 2: Prerequisites The site must be prepared for installation of the workstation. Refer to the Centricity RA600 Workstation Pre-Installation Manual for detailed information. The appropriate line voltage, line frequency, and power supply line cords.
  • Page 34 Service Monitor Tool Table 3-2: Items used for installation of 1728 x 2304 workstation The Centricity RA600 Workstation should already be configured as per the factory settings. Also, the operating system software and applications software are installed at the factory. The person installing the workstation must physically interconnect workstation cables, and then configure the system using site-specific information (i.e.
  • Page 35 Hardware Installation IST 002 (A) 1728 x 2304 Portrait Monitor Interconnections 3-5 Keyboard & Mouse Connections Connect the mouse and keyboard to the computer housing. 3-6 Monitor Connections Connect the monitors as shown in appropriate cabling illustration (two or four monitors). Interface interconnections to two Barco “A”...
  • Page 36 Hardware Installation IST 002 (A) 1728 x 2304 Portrait Monitor Interconnections Interface interconnections to four Barco “A” 1728 x 2304 portrait monitors Video card connectors Monitor cable connector to Slot 1 video interface Blank Slot 2 Blank Slot 3 Blank Calibrator connector Slot 4 Video Card...
  • Page 37: Ist 003 (B) 1200 X 1600 Portrait Monitor Interconnections

    Centricity RA600 Workstation is 1 gauss to ensure proper operation. Section 2: Prerequisites The site must be prepared for installation of the workstation. Refer to the Centricity RA600 Workstation Pre-Installation Manual for detailed information. The appropriate line voltage, line frequency, and power supply line cords.
  • Page 38 Service Monitor Tool Table 3-3: tems used for installation of 1200 x 1600 workstation The Centricity RA600 Workstation should already be configured as per the factory settings. Also, the operating system software and applications software are installed at the factory. The person installing the workstation must physically interconnect workstation cables, and then configure the system using site-specific information (i.e.
  • Page 39 Hardware Installation IST 003 (B) 1200 x 1600 Portrait Monitor Interconnections 3-5 Keyboard & Mouse Connections Connect the mouse and keyboard to the computer housing. 3-6 Monitor Connections Connect the monitors as shown in appropriate cabling illustration (two or four monitors). Interface interconnections to two “B”...
  • Page 40 Hardware Installation IST 003 (B) 1200 x 1600 Portrait Monitor Interconnections Interface interconnections to four Barco “B” 1200 x 1600 portrait monitors Video board connectors Double-headed monitor cable connects to two interface connectors Slot 1 Blank Monitor 2 Monitor 1 Slot 2 Blank Slot 3...
  • Page 41: Ist 004 (B+) 1536 X 2048 Flat Panel Portrait Monitor Connections

    Centricity RA600 Workstation is 1 gauss to ensure proper operation. Section 2: Prerequisites The site must be prepared for installation of the workstation. Refer to the Centricity RA600 Workstation Pre-Installation Manual for detailed information. The appropriate line voltage, line frequency, and power supply line cords.
  • Page 42 Provided with monitor Table 3-4: Items used for installation of B+ workstation The Centricity RA600 Workstation should already be configured as per the factory settings. Also, the operating system software and applications software are installed at the factory. The person installing the workstation must physically interconnect workstation cables, and then configure the system using site-specific information (i.e.
  • Page 43 Hardware Installation IST 004 (B+) 1536 x 2048 Flat Panel Portrait Monitor Connections 3-5 Keyboard & Mouse Connections Connect the mouse and keyboard to the computer housing. 3-6 Monitor Connections The monitor video connections are behind the top rear panel of the B+ monitor. A faring in the panel provides outlet for the video and power cables.
  • Page 44 Hardware Installation IST 004 (B+) 1536 x 2048 Flat Panel Portrait Monitor Connections (primary monitor) Illustration 3-8: Connection of four "B+" monitors to workstation Page 44...
  • Page 45: Ist 005 (C) 1280 X 1024 Landscape Monitor Interconnections

    RA600 Workstation is 1 gauss to ensure proper operation. Section 2: Prerequisites The site must be prepared for installation of the workstation. Refer to the Centricity RA600 Workstation Technical Reference Manual for detailed information. The appropriate line voltage, line frequency, and power supply line cords.
  • Page 46 Service Monitor Tool Table 3-5: Items used for installation of 1280 x 1024 workstation The Centricity RA600 Workstation should already be configured as per the factory settings. Also, the operating system software and applications software are installed at the factory. The person installing the workstation must physically interconnect workstation cables, and then configure the system using site-specific information (i.e.
  • Page 47 Hardware Installation IST 005 (C) 1280 x 1024 Landscape Monitor Interconnections 3-5 Keyboard & Mouse Connections Connect the mouse and keyboard to the appropriate receptacles on the computer housing. 3-6 Monitor Connections Connect the monitors as shown in appropriate cabling illustration (one or two monitors). If the video card has two connectors, connect the primary monitor to connector 1.
  • Page 48 Hardware Installation IST 005 (C) 1280 x 1024 Landscape Monitor Interconnections Interface interconnections to two landscape monitors Slot 1 Blank Slot 2 Blank Slot 3 Blank Slot 4 Video Card Slot 5 Blank Slot 6 Video Card Landscape Landscape Monitor 2 Monitor 1 Illustration 3-10:connection oF two landscape monitors 3-7 Video Board Driver Installation...
  • Page 49: Ist 006 Site Configuration Forms

    Hardware Installation IST 006 Site Configuration Forms IST 006 Site Configuration Forms Section 1: Introduction This job card identifies what customer site specific data is necessary to configure the Centricity RA600 Workstation. Section 2: Procedure 2-1 Complete Data Sheet Information Complete the Workstation Installation Worksheet page 50...
  • Page 50 Hardware Installation IST 006 Site Configuration Forms Required Information Answers To OS Configuration Questions IP Address As supplied Normally supplied by the network administrator. Subnet mask As supplied (netmask) Normally supplied by the network administrator. Default Gateway As supplied Normally supplied by the network administrator.
  • Page 51 Hardware Installation IST 006 Site Configuration Forms Workstation Installation Worksheet Root Password: Hospital Name: User’s Interface Language: Application Support Telephone: Miscellaneous: Initial Software Revision: Software Upgrade Revision: Miscellaneous: Software Upgrade Revision: Miscellaneous: Software Upgrade Revision: Miscellaneous: Software Upgrade Revision: Miscellaneous: Software Upgrade Revision: Miscellaneous: Software Upgrade Revision:...
  • Page 52 Default route: a route (IP address) that is used to communicate to an unknown subset. This station Options. (Required on a multi-station FDO): Passwords Display Type A+,A,B,C Administrator CANal1 Password: (if assigned) Number Of 1,2,4 Centricity RA600 Radworks Monitors Password: Page 52...
  • Page 53: Ist 007 Adaptec Scsi Card Installation

    IST 007 Adaptec SCSI Card Installation IST 007 Adaptec SCSI Card Installation Section 1: Introduction This job card describes how to install an Adaptec SCSI controller card in the Centricity RA600 Workstation computer. Section 2: Tools REquired 5 mm cross-head screwdriver...
  • Page 54 Hardware Installation IST 007 Adaptec SCSI Card Installation Note: The components used in the workstation are static sensitive. Observe static precautions when performing this procedure. Power off the system. Remove the left-hand half of the workstation case. Slide the rear padlock hasp toward the chassis center and press the release button on the lower front of the bezel and remove the left side of the case.
  • Page 55: Ist 008 Installing Frame Grabber Board

    IST 008 Installing Frame Grabber Board IST 008 Installing Frame Grabber Board Section 1: Introduction This job card describes how to install the a frame grabber board into the Centricity RA600 Workstation. Section 2: Tools Required 5 mm cross-head screwdriver...
  • Page 56 Hardware Installation IST 008 Installing Frame Grabber Board Guide, Checking Inside the Compuuter. The Dell Online Documentation CD is shipped with each Dell 1400 CPU. Note: Be sure to observe proper ESD safety precautions when inserting the frame grabber board. Insert the frame grabber board into an available PCI slot.
  • Page 57: Ist 009 Optional Internal Hard Drive Connect & Configure

    IST 009 Optional Internal Hard Drive Connect & Configure IST 009 Optional Internal Hard Drive Connect & Configure Section 1: Introduction This job describes how to connect an additional 36GB internal hard drive to a Centricity RA600 Workstation. Section 2: Tools REquired...
  • Page 58 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure 1. Unpack optional hard drive and mounting screws and retain packing material until installation is validated. 2. Configure the hard drive as SCSI drive 2 by placing a jumper plug across pins 5 & 6 on the 28-pin connector at the front of the drive.
  • Page 59 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Open CPU case by removing side access panel. Release and remove the side access panel (use the release switch (padlock hasp) on lower rear panel and the push-button in front. Refer to Illustration 3-12 and on-line CPU documentation, as necessary.
  • Page 60 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Identify and disconnect ribbon data cable going to existing CPU hard drive. Disconnect power connector from existing hard drives. Illustration 3-13:Hard Drive Connector Page 60...
  • Page 61 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Remove Dell 1400 front panel by pressing down release button inside top of case and pulling front panel away and then up from chassis. Illustration 3-14:Front Panel Release Page 61...
  • Page 62 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure To access hard drive cage in CPU chassis, Pull out and down on the handle of the drive bay door on the front of the chassis until the cage front is about 1 inch out of the bay (Illustration 3-15), Illustration 3-15:Hard Drive Cage Access...
  • Page 63 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Then slide out the cage from the chassis to access the drive mounting screws on the top and bottom of the cage (Illustration 3-16). Illustration 3-16:Slide out hard drive cage Page 63...
  • Page 64 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Fit the new drive into a vacant area of the hard drive cage, oriented the same as the existing drive, with the connectors facing inward toward the center of the case. Illustration 3-17:New Drive Installed In Cage Mount the new drive to the drive cage with the four screws provided.
  • Page 65 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure If necessary, connect the system keyboard, mouse, and monitor to the appropriate connectors at the rear of the CPU chassis. Power up the CPU and note that the operating system recognizes the new drive. If the new drive is not listed during the power up display sequence, power down the CPU, reseat the cable connections and try again.
  • Page 66 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Illustration 3-18:Write signature to new disk Page 66...
  • Page 67 Note how the Computer Management window shows the new disk in the lower right pane. Illustration 3-19:New Disk 1 with written signature in Disk Management Right-click on the Centricity RA600 D: drive area and select Extend Volume... from the pop- up menu; then click Next on the Extend Volume Wizard Welcome window.
  • Page 68 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Illustration 3-20:Starting Extend Volume wizard Illustration 3-21:Can’t extend a volume originally created on basic (vis-a-vis dynamic) disk. Page 68...
  • Page 69 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Highlight Disk 1 in the All Available dynamic disks list and click Add>> to move this available space to the Selected dynamic disks list. Leave the default Sizes info set at Maximum and click Next.
  • Page 70 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure 4-3 Workaround for Extending a Volume Created on a Basic (non-dynamic) Disk. Windows 2000 does not allow extending a volume that was created on a Basic disk. The workaround is to create a standalone new dynamic disk, transfer all existing data from the old to the new disk, delete the old disk, and then extend the new disk onto the old disk space.
  • Page 71 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Illustration 3-24:Select Disk & Assign Drive Letter for Volume In the Select Disks window, high-light Disk 1 in the All available dynamic disks list and click Add>> to move it to the Selected dynamic disks list it for the volume. Leave the default values in the Size selection and click Next.
  • Page 72 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Illustration 3-25:Formatting & Finishing the Created Volume In the Format Volume window, select the Format this volume as follows radio button and select the following Formatting parameters: File system to use: Select NTFS Allocation unit size: Leave Default Volume label: New Volume Check the Perform a Quick Format check box.
  • Page 73 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Illustration 3-26:Computer Management view of new Disk 1 volume Copy the contents of the D: drive to the G: drive as follows: Right-click on the D: area of Disk 0 and select Open from the popup menu. Right-click on the G: area of Disk 1 and select Open from the popup menu.
  • Page 74 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Close both the D: and G: windows. Right-click on the D: area of Disk 0 and select Delete Volume... from the popup menu. Deleting the D: volume Illustration 3-27: When the Are you sure...
  • Page 75 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Illustration 3-28:Clicking through warnings Right-click on the G: drive area and select Extend Volume... Illustration 3-29:Extending the G: Drive Page 75...
  • Page 76 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Click Next in the Welcome to the Extend Volume Wizard window. Illustration 3-30:Finish G: Volume Extension In the Select Disks window, select Disk 0 In the All available dynamic disks field and then click Add>>...
  • Page 77 Right-click on the Disk 1 New Volume area and select Properties from the popup menu. On the General tab of the New Volume Properties window, enter Centricity RA600 in the Label field and click OK. Note the new name now appears on both areas of the new volume.
  • Page 78 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Illustration 3-32:Changing the drive letter of the new volume Change the drive letter of the new volume to D: as follows: Right click on the new volume area and select Change Drive Letter and Path... from the popup menu.
  • Page 79 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Illustration 3-33:Complete the drive letter change In the Edit Drive Letter.. window, select the Assign a drive letter radio button and highlight D:, then click OK. In the Confirm popup, click Yes. Note in the Disk Management console display, that the new drive is now designated D: Close out of the Disk Management window.
  • Page 80 Hardware Installation IST 009 Optional Internal Hard Drive Connect & Configure Illustration 3-34:The extended D: drive Use the Windows 2000 Explore function to verify that the Database folder and its subdirectories/files are present as before the installation. Page 80...
  • Page 81: Ist 010 Installing Additional Ram Modules In Cpu

    Adding 1, 2, or 3 additional 512MB RAM module(s) to a Dell 1400 CPU that has 512MB RAM. Minimum RAM for a Centricity RA600 CPU is 512MB. Section 2: Prerequisites: A properly installed and configured Dell PowerEdge 1400 CPU with Windows 2000 Pro installed and running.
  • Page 82 Hardware Installation IST 010 Installing Additional RAM Modules In CPU Unpack optional RAM module(s) and retain packing material until installation is validated. Power down CPU and disconnect power cord from the receptacle on the rear of the Dell 1400. Open CPU case by removing the side access panel. Remove the left-hand half of the workstation case.
  • Page 83 Hardware Installation IST 010 Installing Additional RAM Modules In CPU Identify the CPU power supply at the top rear of the chassis. Press the Release tab at the lower rear corner and rotate it up and away from the system board. The power supply mounting bracket has detents to hold it up while accessing the system board.
  • Page 84 Hardware Installation IST 010 Installing Additional RAM Modules In CPU Note: There are no system requirements for the placement or the order of the RAM modules. Rotate the power supply back down to its operating position and reconnect the CPU power cord.
  • Page 85 Hardware Installation IST 010 Installing Additional RAM Modules In CPU Illustration 3-37:Checking System RAM Recognition. If the system does not recognize the new RAM module(s), Power down the CPU, Disconnect the power cord Release and raise the power supply Reseat the RAM modules, and repeat steps through 11.
  • Page 86: Ist 011 Installing Global Modem

    Hardware Installation IST 011 Installing Global Modem IST 011 Installing Global Modem Section 1: Introduction This job card describes how to install the GE global modem. Section 2: Material Required GE Global Modem Kit, including DB24M-to-DB9F serial cable and Modem kit (Includes modem, universal power supply, phone cables, customer property label, phone company adapter kit, adhesive foot pads, and Velcro mounting tape strips).
  • Page 87 Hardware Installation IST 011 Installing Global Modem Unpack the Global Modem Kit and retain packing material until installation is validated. Power off the CPU. Connect the serial cable from the COM1 port of the CPU to the modem. Connect the power cord from the modem to the local power outlet. Connect the telephone cable from the modem to the local communication outlet.
  • Page 88: Ist 012 Installing Barco 3Mp2Fh Video Board

    Hardware Installation IST 012 Installing Barco 3MP2FH Video Board IST 012 Installing Barco 3MP2FH Video Board Section 1: Introduction This job card describes how to install the Barco 3MP2FH video controller board for the Barco 3M pixel flat panel monitor in the Dell 1400 computer. Section 2: Material Required Barco 3MP2FH video controller board Dell 1400 Computer...
  • Page 89 Hardware Installation IST 012 Installing Barco 3MP2FH Video Board Note: Prior to installing your 3MP2FH board(s) in your PC please take a few minutes to familiarize yourself with both the board and the PCI slots in your computer. Illustration 3-38:3MP2FH Switch Locations There are four switches located at SW1 on the 3MP2FH.
  • Page 90 Hardware Installation IST 012 Installing Barco 3MP2FH Video Board Prior to installing the BarcoMed 3MP2FH, check the setting of the VGA switches on the graphics board. By default, VGA should be enabled on the primary card. If installing 2 cards in the system, make sure that the VGA EN option on the 2nd card is set to OFF.
  • Page 91 Hardware Installation IST 012 Installing Barco 3MP2FH Video Board Turn off the power to your computer and disconnect the power cord. Remove the chassis cover according to the manufacturer’s instructions. Be sure to observe safety warnings. Install the first card in 3rd slot from bottom and the 2nd card in the bottom slot. Be sure that the board is seated firmly in the slot.
  • Page 92: Ist 013 Installing Raid Unit

    Hardware Installation IST 013 Installing RAID Unit IST 013 Installing RAID Unit Section 1: Introduction This job card outlines how to install the RAID storage unit (Dell PowerVault 221S) with Centricity RA600 (Dell 1400 platform) Work Group Server. The RAID unit is intended to be the primary short- term image storage repository for the RA600 with Workgroup Server option.
  • Page 93 Hardware Installation IST 013 Installing RAID Unit Connect the two SCSI type cables between the RAID unit and the SCSI controller card (in one of the expansion slots in the computer): Connect one cable between SCSI connector A on the RAID unit and SCSI connector 0 on the PERC 3 SCSI controller board in the computer.
  • Page 94 Hardware Installation IST 013 Installing RAID Unit Illustration 3-41:Dell PowerEdge1400 unit rear view - connectors 1 Bus config. switch - Set to middle pos. 2 Fan B fault 3 Fan A fault 4 Power supply status 5 Power supply fault 6 AC status 7 AC connector 8 Power supply on/off switch...
  • Page 95: Chapter 4 Software Configuration

    Software Configuration Section 1: SUMMARY OF job cardS This section of the service manual covers common procedures for the workstation. Some procedures are duplicated here from other sections for ease of use. Software Configuration Job Cards Card No. CNF 001 Initial Power Up and Software Configuration CNF 002 Configuring System IP Address, Subnet Mask or...
  • Page 96: Cnf 001 Initial Power Up And Software Configuration

    Software Configuration CNF 001 Initial Power Up and Software Configuration CNF 001 Initial Power Up and Software Configuration Section 1: Introduction This job card covers configuring the workstation OS software using customer site specific data. Refer to the forms in CNF 005 page 107 for the customer information.
  • Page 97 Software Configuration CNF 001 Initial Power Up and Software Configuration System must be connected to a live network. The RA600 license file requires an active network communication. If the network is not available, you must disable MediaSense before installation. A registry key setting has been provided for disabling MediaSense.
  • Page 98 Software Configuration CNF 001 Initial Power Up and Software Configuration Enter the computer name and click Next. Note: The preceding name entry opportunity does not appear for a Workgroup Server. Refer to page 228. , on Enter the current Date and Time information and click Next. The system proceeds with installation and then automatically reboots.
  • Page 99 3-2 Install Centricity RA600 License File Locate the Centricity RA600 Configuration floppy (license file diskette). Copy the RA600 license file into the d:\AMI_60\bin\ folder (where d: is the installation drive). Remove the blank license.dat file from the d:\AMI_60\bin folder, if present.
  • Page 100 (manually) unchanged between calibrations 3-5 Initialize Centricity RA600 Software Launch the RA600 software by double clicking the icon next to the Start button or by selecting Start > Programs > Centricity RA600 > Centricity RA600. Page 100...
  • Page 101: Cnf 002 Configuring System Ip Address, Subnet Mask Or Default Route/Gateway

    Software Configuration CNF 002 Configuring System IP Address, Subnet Mask or Default Route/Gateway CNF 002 Configuring System IP Address, Subnet Mask or Default Route/Gateway Section 1: Introduction This job card explains how to change the title items or add a subnet mask or default route from the local workstation to a gateway device.
  • Page 102 Software Configuration CNF 002 Configuring System IP Address, Subnet Mask or Default Route/Gateway In the popup window, choose whether you want a specific or automatically supplied IP address. If specific, you must enter the IP address, Subnet Mask, and Default Gateway addresses in the fields provided whether you want a specific or automatically supplied DNS server.
  • Page 103: Cnf 003 Changing Host/Computer Name

    Software Configuration CNF 003 Changing Host/Computer Name CNF 003 Changing Host/Computer Name Section 1: Introduction This job card provides instructions for properly changing the host / computer name. Section 2: Prerequisites All computer names must be unique. Having two workstations with the Caution same hostname can slow workstation response or cause workstations to be inoperable.
  • Page 104 Note: The computer name must be unique on the LAN. The preferred naming convention for the Centricity RA600 workstations is RWDxxx, where xxx is the last three numbers of the IP address. (RWD = RadWorks Display. Click OK to close the Identification Changes window and OK to close the System Properties window.
  • Page 105: Cnf 004 Adding/Configuring Tcp/Ip Printer

    Software Configuration CNF 004 Adding/Configuring TCP/IP Printer CNF 004 Adding/Configuring TCP/IP Printer Section 1: Introduction This job card provides instructions for properly configuring the use of a TCP/IP printer for use with the application software. TCP/IP printers are paper based printers used for such things as printing reports and work lists.
  • Page 106 Software Configuration CNF 004 Adding/Configuring TCP/IP Printer Click Start > Settings > Printers > Add Printer The Add Printer wizard initiates. Click Next in the Welcome window and follow the wizard instructions to add the printer. When prompted, Enter the name and/or IP address of the printer Select this printer as the default printer.
  • Page 107: Cnf 005 Vnc Server Installation And Configuration

    This job card Describes how to install and configure the VNC software to receive incoming support calls from the GEMS Remote Online Center. Section 2: Prerequisites VNC is installed on the workstation or available on Centricity RA600/RadWorks 6.0 Utility/Driver CD-ROM. Workstation user login name and password.
  • Page 108 Software Configuration CNF 005 VNC Server Installation and Configuration Click Next to choose the default install location. Click Next to select the default program folder. Click Finish to complete the installation. Select Start>Program>VNC>Administrative Tools>Install Default Registry Settings. Select Start>Program>VNC>Administrative Tools>Install WinVNC Services. Only if the workstation includes A and B monitors, ensure that the Palette Options is set for Dynamic Gray Palette as follows: Select Start>Settings>Control Panel>Display.
  • Page 109: Cnf 006 Monitor Calibration With Barco Tool (Medical Pro)

    Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) Section 1: Introduction This job card describes how to set up Barco’s MediCal Pro software and perform calibration on Barco and non-Barco monitors. The procedure here is an outline guide. For more detailed instructions, consult the Medical Pro User’s Guide provided with the software.
  • Page 110 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) RA600 video card(s) properly installed and configured. RA600 system monitors have been connected to the monitor controller card(s) per the applicable IST job card in Chapter 2. Section 3: System Setup Procedure Medical Pro software is pre-installed for your turnkey RA600 system.
  • Page 111 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) A message box may appear, asking you for a connection to MediCal Administrator. This is a separate Barco network administration program for monitor calibration. Select No if you do not use this program.
  • Page 112 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) If the number found does not match your configuration, click No and follow the troubleshooting directions of the wizard (check cable connections, etc.) until all the Barco displays are found. If the display is not Barco, the wizard reports 0 found and asks if this is correct.
  • Page 113 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) For the first non-Barco display found, the wizard terms it a “Remaining imaging board head” and asks if you want to control a display on this head. Click Yes and enter the requested information about the display connected, then click Next.
  • Page 114 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) When the Calibration screen appears, you can select to calibrate and do a Q/A check immediately. Select Next to continue. When the following screen appears, you can select to calibrate and do a Q/A check immediately.
  • Page 115 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) 16. For a B+ flat panel monitor, the process is entirely automated because it has a built-in screen sensor. For other monitor types, follow the on-screen instructions to connect the screen sensor to the workstation computer.
  • Page 116 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) For measuring the uniformity of your display, you can remove the shield of your sensor, place it on the highlighted patch and press the enter key on the following message. After finishing all Q/A tasks, you should see following message: 21.
  • Page 117 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) 22. Notice the graphic representation in the lower right-hand corner that shows the screen sensor used in the test. Close MediCal Pro now. Note: - For 2 megapixel grayscale displays, the white is set to 65 fL, the black to 0.2 fL (DICOM as display function with Ambient Light Correction).
  • Page 118 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) 4-2 Just do a Visual Check If you want to execute a visual check (ignoring if it is due or not), go to Start > Programs > MediCal Pro > MediCal Q/A Visual Check. Follow each step on the screen until the application closes.
  • Page 119 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) You now see an overview of the total history of the selected task for all your displays. The results are sorted according to their execution time. The top result is the most recent. Select a particular result and click on “Details”...
  • Page 120 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) Go to the history of the visual check task for a certain display as explained in the section “Look up histories of executed tasks”. When the detailed information for a particular result on a display is chosen, the task history shows the following view.
  • Page 121 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) Click on the unnamed button below Measurements to see the measurement values for a particular result. A window shows up with the measured values of the uniformity task. 4-7 Look up detailed results of a measurement task Go to the history of one of the measurement tasks (Measure Display White, Measure Display Black, Measure Quality Level) for a certain display as explained in the section “Look up histories of executed tasks”.
  • Page 122 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) Once you are in the task history details of that task for a certain display, you can view a graph of all the results of that task or you can view this information in a table format. See the illustrations below for a preview of what you can expect of these views.
  • Page 123 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) 5-1 Measure Display White / Measure Display Black / Measure Quality Level is ‘WARNING’ or ‘NOT OK’. In case one of the results of a measurement task has an insufficient result, it is recommended that you recalibrate the display for the task that had poor results.
  • Page 124 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) Finally, when the displays are calibrated again, retry the tasks with poor results. (See section “Run an individual task” for more information about executing a single task.) 5-2 Measure Uniformity Task is ‘WARNING’ or ‘NOT OK’ In case a result of the Measure Uniformity task is unsatisfying, find out which display is causing the problem.
  • Page 125 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) You need to be logged in as an advanced user. If you are not yet in advanced mode (your configuration is not visible at the bottom of the main window), you need to log in again. Go to Tools >...
  • Page 126 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) 5-3 Visual check is ‘NOT OK’ When there is a problem with the visual check, the first thing you should do is find out which part of the visual check is causing this problem. Follow the steps below to have a detailed analysis of the problem;...
  • Page 127 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) In the Q/A section, the visual check task will state its last result is “NOT OK”. Go to the history of the Visual Check task. (See the section “Look up histories of executed tasks”...
  • Page 128 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) it here. Select the display (you can only select one display at a time) where the result is “NOT OK” and click “Details”. The different questions of the visual check task are displayed with the corresponding answer next to it.
  • Page 129 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) 5-4 Visual check – SMPTE test pattern ‘squares not visible’ problem In case the squares of the SMPTE test pattern are not visible on a certain display, you should recalibrate that display.
  • Page 130 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) When the display is calibrated again, retry the visual check task and verify that the squares of the SMPTE test pattern are visible now (See section “Run an individual task” for more information about executing a single task).
  • Page 131 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) You need to be logged in as an advanced user. If you are not yet in advanced mode (your configuration is not visible at the bottom of the main window), you need to log in again. Go to Tools >...
  • Page 132 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) The display alignment tool shows on the display. Use the indicated tools for horizontal and vertical geometry settings to adjust the alignment of the display (see picture below). For further information on using the alignment tool, refer to the user manual for MediCal Pro.
  • Page 133 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) When the alignment of the display is adjusted, retry the visual check task and verify that the linearity has improved (See section “Run an individual task” for more information about executing a single task).
  • Page 134 Software Configuration CNF 006 Monitor Calibration with Barco Tool (MediCal Pro) configuration. Right-click on the test pattern for a popup menu with further actions on this pattern. Page 134...
  • Page 135: Cnf 007 Configuring Barco 3Mp2Fh Video Driver

    Software Configuration CNF 007 Configuring Barco 3MP2FH Video Driver CNF 007 Configuring Barco 3MP2FH Video Driver Section 1: Introduction This job card describes how to install and configure the Barco 3MP2FH video controller board driver for the Barco 3M pixel flat panel monitor. Section 2: Material Required Dell 1400 computer with 3MP2FH video board installed.
  • Page 136 Software Configuration CNF 007 Configuring Barco 3MP2FH Video Driver 4-1 Reinstalling or Updating your BarcoMed Drivers To reinstall your current BarcoMed Display Driver or install a new version of the BarcoMed Display Driver follow the steps below: Right click on the desktop and select “Properties ”. Select the “Settings” tab. Now select the rectangle that represents the BarcoMed 3MP2FH Imaging Board whose driver you wish to reinstall or update.
  • Page 137 Software Configuration CNF 007 Configuring Barco 3MP2FH Video Driver 4-2 Setting the Resolution of Your CORONIS Display Panel To set the resolution for your CORONIS display panel(s), right click on the desktop and select “Properties”. Select the “Settings” tab. Now select the rectangle that represents the BarcoMed 3MP2FH Imaging Board you are working with.
  • Page 138 Software Configuration CNF 007 Configuring Barco 3MP2FH Video Driver The system shown in the following screen capture has one BarcoMed 3MP2FH Imaging Boards installed and no 3 rd party VGA card. Your configuration may look different. This example shows Windows 2000 Display Properties Control Panel after the drivers have been installed and the resolution set.
  • Page 139 Software Configuration CNF 007 Configuring Barco 3MP2FH Video Driver In the system shown in the screen capture preceding, there is one BarcoMed 3MP2FH Imaging Board installed with Dual View disabled and no 3 rd party VGA card. Your configuration may look different.
  • Page 140 Software Configuration CNF 007 Configuring Barco 3MP2FH Video Driver In the system shown in the preceding screen capture, there is one BarcoMed 3MP2FH Imaging Board installed with Dual View with Dual View enabled. Rectangles “1” and “2” represent the two CORONIS displays of a single board/dual head installation.
  • Page 141 Software Configuration CNF 007 Configuring Barco 3MP2FH Video Driver When the drivers have finished copying, click “Finish” on the “Completing Found New Hardware Wizard” page to complete the installation. Windows may now prompt you to reboot your system. Click “Yes ”. Boot normally and log on using an account with administrator privileges.
  • Page 142 Software Configuration CNF 007 Configuring Barco 3MP2FH Video Driver Select the “Adapter” tab and then click on the “List All Modes…” button. Select the resolution and refresh rate that your CORONIS display panel supports from the dialog box and click “OK ”.
  • Page 143: Cnf 008 Raid System Recovery And Configuration

    Section 2: Prerequisites Windows 2000 Server registration ID number. Video cable is properly connected between the computer and the monitor. Centricity RA600 Workgroup Server Recovery CD-ROM set is available Centricity RA600 license file diskette is available Section 3: Procedure 3-1 Restoring Original System Configuration All data and programs on the system disk will be erased.
  • Page 144 Locate the computer name, if necessary, by right-clicking on the Network Neighborhood icon and selecting Properties. The Computer Name is shown on the Identification tab. Place the Centricity RA600 Recovery CD-ROM disk 1 into the CD-ROM drive. Restart the workstation.
  • Page 145 (F:) drive and select to share the AMI_60 folder. Copy RA600 license file to the Bin folder. Start the Centricity RA600 application and check that it is working. Use Windows Explorer to find and delete the following folder: c:\RW_BackUp.
  • Page 146 Software Configuration CNF 008 RAID System Recovery and Configuration From the Task menu, select Assign Global Hotspare. In right -hand pane, note that Array Disk 00 is now called Spare Array Disk in the Type column. Page 146...
  • Page 147 Software Configuration CNF 008 RAID System Recovery and Configuration High-light Logical Array > Array Group 0. From the Task menu, select Create Virtual Disk... In the resulting initial wizard window, click Next. In the following wizard window, click Next to accept defaults. In the following wizard window, leave all defaults except select RAID 5 from the Type pulldown menu and then click Next.
  • Page 148 Software Configuration CNF 008 RAID System Recovery and Configuration 3-3 Initializing the RAID Select Programs > Dell OpenManage Applications > Array Manager > Array Manager Console. Right click on Disk 1 in the left-hand pane, select Write Signature, and click Yes to confirm. Right click on Disk 1 and select Create Volume...
  • Page 149 Software Configuration CNF 008 RAID System Recovery and Configuration Click Next in the Welcome window of the Create Volume Wizard. Select Disk 1, Select Free Space, and then click Next. Page 149...
  • Page 150 Software Configuration CNF 008 RAID System Recovery and Configuration Select Primary Partition, accept the default value of Disk space to use (maximum), then click Next. Select Assign a drive letter, select F, and click Next Check the Format this volume checkbox, select NTFS file system, enter a Volume label of RA600 WGS, check the Perform Quick Format checkbox, then click Next.
  • Page 151: Cnf 009 Configuring Workgroup Server & Client System

    Software Configuration CNF 009 Configuring Workgroup Server & Client System CNF 009 Configuring Workgroup Server & Client System Section 1: Introduction This job card outlines how to configure the RA600 Workgroup server and client installation after installation per job cards 013, SW 011 012.
  • Page 152 Software Configuration CNF 009 Configuring Workgroup Server & Client System Start the Centricity RA600 program and wait for the operator console to display. Click on Data Selector tab in the Tooltab guide (see arrow in preceding screen capture). For each worklist name in the Worklist View pull-down list, Select the worklist name.
  • Page 153 Software Configuration CNF 009 Configuring Workgroup Server & Client System Uncheck the Allow Empty Query checkbox. Click OK. When the empty query option has been disallowed for all worklists, proceed to the next Configuration process. Page 153...
  • Page 154 CNF 009 Configuring Workgroup Server & Client System 3-2 Initial settings for the Workgroup Server and for each Client Machine In the Centricity RA600 console, select Configuration > Generic Properties > HIS-RIS Highlight Default HIS/RIS Implementation... and click on the Config... button.
  • Page 155 Software Configuration CNF 009 Configuring Workgroup Server & Client System Select Today Copy and click on Modify. Enter the new worklist name of Last 5 Days into the Description field. Page 155...
  • Page 156 Software Configuration CNF 009 Configuring Workgroup Server & Client System On the Active studies tab, Query group, select StudyDate in the Description column, modify the first limit by appending minus five (- 5) at the end of the limit as shown below: Page 156...
  • Page 157 Software Configuration CNF 009 Configuring Workgroup Server & Client System Close all open windows, then close and restart the Centricity RA600 program. Add a new worklist called "Last 5 Days" as follows: From the Database menu, select the Add View.
  • Page 158 Software Configuration CNF 009 Configuring Workgroup Server & Client System In the Add View window, make the following entry in the Description field: Last 5 Days [Do Not Delete] In the List pull-down list, select Last 5 Days. In the Copy Settings from field, ensure that the Local Studies... item is selected Click OK.
  • Page 159 In the Initial View window pulldown list, select the Last 5 Days [Do Not Delete] item and then click OK. Close and restart the Centricity RA600 program. In the Centricity RA600 console, ensure that the program has opened with the initial worklist display of Last 5 Days [Do Not Delete]. Page 159...
  • Page 160 3-3 Workgroup Server and ASP If this Workgroup server is to be used in an ASP (Application Service Provider) environment, configure Centricity RA600 Workgroup Server according to the ASP workflow environment. 3-4 Put Workgroup Server Machine Name in Client Hosts File...
  • Page 161: Chapter 5 Functional Check

    Functional Check Section 1: Summary of Job Cards Job Card Functional Check Job Card Title VF 001 Functional Checks / Setup Completion VF 002 Saving/Restoring System and User Settings VF 003 Customer Turnover Page 161...
  • Page 162: Functional Checks / Setup Completion

    Functional Check VF 001 Functional Checks / Setup Completion VF 001 Functional Checks / Setup Completion Section 1: Introduction This job card finishes the required system configurations before turnover to the customer. This procedure will: Test to verify valid network connections. Set the administrator password.
  • Page 163 Functional Check VF 001 Functional Checks / Setup Completion 3-1 Verify Ethernet Connection Click on Start > Programs > Accessories > Command Prompt Execute the ping command with the IP address of a workstation on the network. Example: ping 3.28.75.100 [Return] You should see a response from the ping if the Ethernet connection is good.
  • Page 164 3-4 Application Software Checkout Login to the system with the correct username and password. Launch Centricity RA600 and note that the Data Selector comes up in the control console. Select and double-click to display an image. 3-5 System and User Settings Backup Check (after final configuration) Ensure both system and user’s settings are backed up according to the...
  • Page 165: Saving/Restoring System And User Settings

    A formatted diskette (only if settings are to be stored offline) Section 3: Prerequisites Centricity RA600 must be installed and running.OS is loaded. System monitor(s), keyboard and mouse must be functional and monitor(s) powered on. All system and user settings have been completed Note: Only the correct drivers will successfully install with a particular video board.
  • Page 166 Functional Check VF 002 Saving/Restoring System and User Settings Locate and launch the RadWorksSettings.exe program. The program should be located at D: > AMI_60 > bin > RadWorksSettingsTool.exe. The Settings tool initial window appears. 4-2 Backing Up System Settings Log on to the system as administrator, start the Settings application and click the RadWorks System button.
  • Page 167 Functional Check VF 002 Saving/Restoring System and User Settings In the RadWorks System Settings window, click the Backup button. In the Save As window, enter a name for the backup file and click Save to save the available system files to the default Backup folder, else first navigate to the preferred backup device and/ or folder and then click Save.
  • Page 168 Functional Check VF 002 Saving/Restoring System and User Settings 4-4 Backing Up User Settings Log on to the system under your user ID and password (or as administrator), start the Settings application and click the RadWorks User button In the RadWorks User Settings window, highlight your assigned user profile name and click Save.
  • Page 169 Functional Check VF 002 Saving/Restoring System and User Settings Log on to the system under your user ID and password (or as administrator), start the Settings application and click the RadWorks User button In the RadWorks System Settings window, click the Load button. In the Open window, select the files you wish to restore and click Open to restore those system files from the default Backup folder, else first navigate to the preferred backup device and/or folder, select the desired files there and then click Open.
  • Page 170 Functional Check VF 002 Saving/Restoring System and User Settings Click OK to close the RadWorks User Settings window. 4-6 Clear User Settings Log on to the system under your user ID and password (or as administrator), start the Settings application and click the RadWorks User button In the RadWorks User Settings window, click the Clear button.
  • Page 171 This will remove only that settings file for the current user. Confirm the deletion in the popup confirmation dialog box. Click OK to close the RadWorks System Settings window. 4-8 Choose User Profiles Running Centricity RA600 Prerequisite: RA600 newly installed User starts the RA600 application for the first time.
  • Page 172 Functional Check VF 002 Saving/Restoring System and User Settings The user may select one of the listed profiles and click OK (this loads the selected profile) or click Cancel to reject the defaults and build his own custom profile from scratch (this loads the system defaults).
  • Page 173: Vf 003 Customer Turnover

    Functional Check VF 003 Customer Turnover VF 003 Customer Turnover Section 1: Introduction This procedure ensures that the installation of the workstation is complete and ready to turn over to the customer. Section 2: Prerequisites Workstation integration completed. Workstation operating system software installed (factory completed) and configured as appropriate with site specific data.
  • Page 174 Functional Check VF 003 Customer Turnover Check off Task Completed Ensure fans and other devices are clear of foreign bodies and functional. Confirm images can be displayed on the graphical user interface. Monitor(s) calibrated per procedure. If applicable. default route information is added. If applicable.
  • Page 175: Chapter 6 Software Procedures

    27 Job Card Software Procedure Job Card SW 001 Restoring RA600 Original System Configuration SW 002 Installing Centricity RA600 Software SW 003 (C, 2C) Matrox Video Board Driver Installation SW 004 (A+,A,B) Dome Video Board Driver Installation SW 005...
  • Page 176 SW 011 Installing Work Group Server on RA600 SW 012 Installing an RA600 Work Group Client SW 013 Install & Setup of Streaming Video Capture System (Refer to Kit document FP10621) SW 014 Installing Centricity RA600 Software Patch Page 176...
  • Page 177: Sw 001 Restoring Ra600 Original System Configuration

    Section 2: Items Required Centricity RA600 Recovery CD-ROM Windows 2000 registration ID number Centricity RA600 license file diskette available MediCal Pro registration ID number All data and programs on the system disc will be erased. Caution Section 3: Prerequisite A VGA service monitor (for any systems with A, A+ and B monitors) is required for all sites at this time.
  • Page 178 Software Procedures SW 001 Restoring RA600 Original System Configuration for an existing workstation are available, verify them against the workstation and then use them as a reference when configuring the workstation during the baselining process. If one exists, the configuration floppy may be used if the saved data is not suspect. Otherwise, simply note the workstation’s IP Address and Workstation Name, etc., for use when manually configuring the workstation.
  • Page 179 Software Procedures SW 001 Restoring RA600 Original System Configuration Insert RA600 Workstation Recovery CD-ROM (2of 2), when prompted. When the “Clone complete” message appears, remove (2of 2) CD from CD-ROM drive and reboot the computer. After reboot is complete, when the Windows 2000 Welcome window appears, perform job card 001, Initial Power Up and Software Configuration page...
  • Page 180: Sw 002 Installing Centricity Ra600 Software

    Software Procedures SW 002 Installing Centricity RA600 Software SW 002 Installing Centricity RA600 Software Section 1: Introduction This job card describes how to install the RA600 software on a new workstation hardware platform. There are several options that may be installed with the RA600 software. Please refer to the pertinent job cards in this manual or to the RadWorks Technical Reference Manual (FP10123) for installation instructions.
  • Page 181 Software Procedures SW 002 Installing Centricity RA600 Software 4-1 Typical Installation Place the RA600 CD-ROM into the CD-ROM drive. When the CD-ROM self-loads and the Welcome to ... Setup screen appears, click Next. Illustration 6-1: Install to folder D:\ami_60\ When the Destination Folder screen appears with a C:\ drive location specified, click Change.
  • Page 182 Software Procedures SW 002 Installing Centricity RA600 Software Note that the Destination Folder screen now shows the just-entered folder name in the Install RadWorks to location. Click Next. Illustration 6-2: Choose Installation Type and RadWorks Role RadWorks - Choose the RadWorks type and role:...
  • Page 183 Software Procedures SW 002 Installing Centricity RA600 Software RadWorks - Choose the licensing configuration and provide the licensing mechanism: Click on the license file radio button. For the license file choice only, type in the location of the license file in the field provided or click on Browse to navigate to the license file location.
  • Page 184 Software Procedures SW 002 Installing Centricity RA600 Software If the devices are not correct click Back on the next screen to make corrections and then proceed. If the list shows the correct choices, continue to the Transmit screen. Illustration 6-4: Specify Transmit Parameters and Miscellaneous Install Items Transmit: Enter the image/data transmission location data for this workstation in the local network.
  • Page 185 Software Procedures SW 002 Installing Centricity RA600 Software Check all the additional items you want to include with the installation on this workstation: Demo Data and/or Configure multimonitor. Click Next Illustration 6-5: Choose Demo Data (or not) and Complete Installation In the Ready to Install the Program screen, click Install.
  • Page 186 Software Procedures SW 002 Installing Centricity RA600 Software Note: Only if the Single Media Archive (SMA) is included with this installation, the VOB software install wizard will appear. Follow the wizard’s instructions and use defaults to install this option. Check for any RA600 software patch(es) included with the application. If any are present, install the patch(es) at this time according to the release notes included on the patch CD.
  • Page 187 Software Procedures SW 002 Installing Centricity RA600 Software Description: This parameter controls DHCP Media Sense behavior. If you set this value data to 1, DHCP, and even non-DHCP, clients ignore Media Sense events from the interface. By default, Media Sense events trigger the DHCP client to take an action, such as attempting to obtain a lease (when a connect event occurs), or invalidating the interface and routes (when a disconnect event occurs).
  • Page 188: Sw 003 (C, 2C) Matrox Video Board Driver Installation

    This job card describes how to install the driver for the Matrox G450 video board for the C, 2C, or 18” flat panel (1x or 2x) monitor. Section 2: Items Required Centricity RA600 Tools CD-ROM One or two C monitor(s) or 18” flat panel . Section 3: Prerequisites OS is loaded.
  • Page 189 Software Procedures SW 003 (C, 2C) Matrox Video Board Driver Installation Note: The Setup.exe program is also available in a Win2K > Drivers folder on the Utility Tools CD- ROM (P/N 2309786). In the first setup window, select the language desired for the installation instructions and click Next.
  • Page 190 Software Procedures SW 003 (C, 2C) Matrox Video Board Driver Installation Illustration 6-7: Final Matrox Setup Windows When the system reboots, a Matrox icon appears in the clock tray of the windows 2000 taskbar. Click on the Matrox icon and choose Display Properties from the popup menu. On the Display Properties window, click the Advanced...
  • Page 191 Software Procedures SW 003 (C, 2C) Matrox Video Board Driver Installation Illustration 6-8: Start Matrox properties selection. Select the Adapter tab and make the following selections: Click the List all modes button to bring up the List all modes window. For a 1C, single color monitor configuration, highlight 1280 x 1024, High Color (16 bit), 85 Hertz For an 18”...
  • Page 192 Software Procedures SW 003 (C, 2C) Matrox Video Board Driver Installation For an 18” 2x flat panel monitor configuration, highlight 2560 x 1024, High Color (16 bit), 60 Hertz Click OK to close the List all modes window. Illustration 6-9: Finish Monitor and Matrox properties selection On the Plug and Play...
  • Page 193: Sw 004 (A+,A,B) Dome Video Board Driver Installation

    This job card describes how to install the driver for the A+, A, and B monitor video board (Dome Md2, Md4, Md5). Section 2: Items Required Centricity RA600 Tools CD-ROM A VGA service monitor. Section 3: Prerequisites OS is loaded.
  • Page 194 Software Procedures SW 004 (A+,A,B) Dome Video Board Driver Installation If not already done, power on the workstation. Log in as user: Administrator. There may be no password set at this time. If a password has been set it could be: (CANal1). Navigate to C:\Drivers\ and choose the appropriate folder for your display(s).
  • Page 195 Software Procedures SW 004 (A+,A,B) Dome Video Board Driver Installation MD2 PCI 1200 x 1600 75.5 Hz See Note NOTE: All settings need to be set individually for each monitor. In addition to this resolution setting, monitor 1 should be set as Primary Monitor and Extend Windows Desktop box should be checked, monitor 2 &...
  • Page 196 Software Procedures SW 004 (A+,A,B) Dome Video Board Driver Installation Select the appropriate number of screens. Select the correct monitor orientation (side by side or left/right). A set of icons will display showing the number of monitors and their respective orientation. The monitor layout is self explanatory.
  • Page 197: Sw 005 Removing Applications Software

    Software Procedures SW 005 Removing Applications Software SW 005 Removing Applications Software Section 1: Introduction This job card describes how to remove the Workstation applications. Section 2: Prerequisites OS is loaded. Video board(s) switch settings are set correctly. Video board(s) are properly Installed in the workstation. Video cable(s) properly connected.
  • Page 198 Software Procedures SW 005 Removing Applications Software Confirm the deletion of the application. Ignore any warnings about shared programs. Page 198...
  • Page 199: Sw 006 Single Media Archive (Sma) Install, Configure & Test

    VOB software enables the SMA function Section 2: Items Required Centricity RA600 Application CD-ROM and any included software patch(es). Approved SMA Read-Write drive with power and signal cables. Writable CDs approved for use with the SMA drive. Section 3: Prerequisites Installation of SMA is part of the RA600 installation.
  • Page 200 Software Procedures SW 006 Single Media Archive (SMA) Install, Configure & Test Check for any RA600 software patch(es) included with the RA600 application. If any are present, install the patch(es) at this time according to the release notes included on the patch Start the RadWorks (RA600) application.
  • Page 201 Software Procedures SW 006 Single Media Archive (SMA) Install, Configure & Test In the Edit Column Properties dialog box make the new column properties 3109 102C - Selected for archiving and click OK. Click the Configuration menu in the RadWorks (RA600) application and choose Single Media Archive.
  • Page 202 Software Procedures SW 006 Single Media Archive (SMA) Install, Configure & Test For the Media Type, select Read Only from the drop down list. For the Device selection, click the Browse button and navigate to the CD write device. Highlight the device and click OK. For the Total Size field, enter the size of the CD capacity that will be used in the device, e.g.
  • Page 203 Software Procedures SW 006 Single Media Archive (SMA) Install, Configure & Test Click OK to complete the SMA configuration. To test the SMA installation first select some studies in the RadWorks Data Selector for writing to the SMA. Either Page 203...
  • Page 204 The approximate number of disks required to store the selected studies is shown in the Est. number of disks field on the Archive tab to the right of the Data Selctor. Refer to the Centricity RA600 / RadWorks 6.0 User Guide, OI10091, for information on using the Archive Index Tool.
  • Page 205 Software Procedures SW 006 Single Media Archive (SMA) Install, Configure & Test Start the write operation either - by clicking the Local Database > Create Volume button on the Archive tab to the right of the RadWorks Data Selector - by selecting Archiving > Create Volume with marked studies from the RadWorks menu bar.
  • Page 206 Software Procedures SW 006 Single Media Archive (SMA) Install, Configure & Test Page 206...
  • Page 207: Sw 007 Installing Foresight Static Frame Capture Software (Refer To Kit Document Fp10621)

    Software Procedures SW 007 Installing Foresight Static Frame Capture Software (Refer to Kit document FP10621) SW 007 Installing Foresight Static Frame Capture Software (Refer to Kit document FP10621) As of Revision 2, this job card is a placeholder for the kit document to ensure instruction sync with the kit.
  • Page 208: Sw 008 Common Windows 2000 Procedures

    Software Procedures SW 008 Common Windows 2000 Procedures SW 008 Common Windows 2000 Procedures Section 1: Introduction This job card describes commonly used Windows 2000 Pro procedures. In the following sections, “local” means “physically in front of” the workstation console, while “remote”...
  • Page 209 Software Procedures SW 008 Common Windows 2000 Procedures Section Topic Changing the Capacity of the Recycling Bin Renaming Desktop Icons Changing Mouse Sensitivity Changing OS Passwords 5-10 Changing Mouse Properties Right Handed / Left Handed Viewing Contents of a Floppy Disk Viewing Contents of Internal Hard Disk Kill Workstation Application Locally —...
  • Page 210 Software Procedures SW 008 Common Windows 2000 Procedures Section Topic Displaying Application Programs Starting Application Programs Changing Screen Saver Properties (wait value) Table 6-2: Common Windows 2000 Procedures List Section 2: Startup / Shutdown 2-1 Booting Workstation from a Power Off Condition Note: System will take about two minutes to boot, if using portrait monitors, no video will be displayed until the system has booted.
  • Page 211 Software Procedures SW 008 Common Windows 2000 Procedures Select Shut Down from the list provided and click OK. If the user needs to login with an alternate userid after rebooting the system, press CTRL + ALT + DEL to log onto the system. Section 3: Login / Logout 3-1 Changing to a Different Login Account Click on Start.
  • Page 212 Software Procedures SW 008 Common Windows 2000 Procedures Note: Hard blocked refers to a system which is totally locked up. The only recourse is a reset. Press the reset button located on the front of the CPU. 4-2 Shut Down System — from Command Prompt Select Start >...
  • Page 213 Software Procedures SW 008 Common Windows 2000 Procedures The clock appears in a popup dialog box. Set the Date /Time and Time Zone accordingly and click OK. 5-2 Remove Programs Using the Add / Remove Program Dialog Note: For instructions on removing applications software see SW 005 - Removing Applications Software.
  • Page 214 Software Procedures SW 008 Common Windows 2000 Procedures Uncheck Show small icons in Start menu. Check Show clock. 5-4 Setting Screen Saver Password With the right mouse button, click on the desktop. Select Properties. Click on the Screen Saver tab. Select a screen saver, if necessary and check the Password protected check box.
  • Page 215 Software Procedures SW 008 Common Windows 2000 Procedures Click on the Global or [C:] tab. Drag the slide bar to the appropriate level and click OK. 5-7 Renaming Desktop Icons Right click the mouse button on the icon to be renamed. Select Rename.
  • Page 216 Software Procedures SW 008 Common Windows 2000 Procedures Type the new password. Type the new password in the confirm box Click OK twice and close the control panel. 5-10 Changing Mouse Properties Right Handed / Left Handed Select Start > Settings > Control Panel. Double-click on the Mouse icon.
  • Page 217 Software Procedures SW 008 Common Windows 2000 Procedures Double-click on the My Computer icon on the desktop. Double-click on the (C:) or (D:) icon. 7-3 Kill Workstation Application Locally — Workstation Application Blocked, Windowing System OK. Application blocked refers to an application which is totally locked up. The only recourse is as follows: Press Ctrl - Alt - Delete at the same time.
  • Page 218 IP Routing Enabled..: Yes WINS Proxy Enabled..: No DNS Suffix Search List..: iis.med.ge.com Ethernet adapter Local Area Connection: Connection-specific DNS Suffix .
  • Page 219 Software Procedures SW 008 Common Windows 2000 Procedures 7-7 Locating Files Using The Search Utility Select Start > Search. Click on for Files or Folders. In the “Search for files or folders named” field, in the name that needs to be located. Click on Search Now.
  • Page 220 Software Procedures SW 008 Common Windows 2000 Procedures Double-click on the Recycle Bin icon on the desktop Select File. Select Empty Recycle Bin (items will be permanently be remove from system). Click Yes to removing the recycling bin contents 7-11 Displaying “Available Space” on Hard Drive Double-click on the My Computer icon on the desktop.
  • Page 221 Software Procedures SW 008 Common Windows 2000 Procedures Select the file or folder for which you want to change properties. Select appropriate attributes. Click OK. Section 8: Applications 8-1 Displaying Application Programs Click on Start. Click on Programs. 8-2 Starting Application Programs Click on Start.
  • Page 222 Software Procedures SW 008 Common Windows 2000 Procedures Select a screen saver, if none is currently selected, and then use the up/down arrow in the Wait field to change the time. The screen saver will be activated when there is no keyboard activity for the time indicated in the Wait field.
  • Page 223: Sw 009 Changing Dome Display Font Size

    SW 009 Changing DOME Display Font Size Section 1: Introduction This job card outlines how to change the size of the type font used with the DOME drivers supplied with some of the Centricity RA600 systems. Section 2: Prerequisites Latest DOME Video Driver must be installed Section 3: Changing DOME Display Font Size Right click on a blank screen area of the Dome display desktop and select Properties.
  • Page 224: Sw 010 Installing Film Digitizer Scsi Software

    Section 3: Prerequisites Ensure that the RA600 workstation has been installed with both hardware and software configured to GE specification. The RadWorks Acquisition module must be installed as part of the RA600 (RadWorks 6.0) application. Ensure that proper SCSI ID has been set on the device interface card, if necessary. Refer to manufacturer’s documentation for details.
  • Page 225 Software Procedures SW 010 Installing Film Digitizer SCSI Software In the Program Modification window, select the Modifiy radio button. In the Custom Setup window, double click the Acquisition module, select the desired digitizer (e.g., Lumiscan 75, HowteK, 450) and click Next. In the Ready to Modify the Program window, click Install.
  • Page 226 Software Procedures SW 010 Installing Film Digitizer SCSI Software When the setup sequence completes successfully, copy the driver software of the digitizer from the RA600 (RadWorks 6.0) CD (\3rdparty folder) into the manually created DataAcquisitionDrivers folder on the same root as the AMI software (ususally D:\). For an automated install, specify a typical install to the above-named folder for a SCSI-type scanner.
  • Page 227 Software Procedures SW 010 Installing Film Digitizer SCSI Software Start the RA600 (RadWorks 6.0) application and click Acquire > Device on the menu bar to open the Device Management dialog box Select the digitizer device and click Load to load the device. If you want to make this device a default device, select the device and click Activate.
  • Page 228: Sw 011 Installing Work Group Server On Ra600

    SW 011 Installing Work Group Server on RA600 Section 1: Introduction This job card outlines how to install & configure the Centricity RA600 Workgroup Server on the Dell 1400 workstation. The Workgroup Server allows sharing of image files (and other files with clients in a work group relationship.
  • Page 229 Section 4: Procedure Note: DO NOT CHANGE THE COMPUTER NAME OF THIS SYSTEM. By changing the Computer Name, Centricity RA600 Workgroup Server becomes un- operational. In any event, if the Computer Name must be changed, the ROC should be notified and the Registry Keys need to be edited.
  • Page 230: Sw 012 Installing An Ra600 Work Group Client

    SW 012 Installing an RA600 Work Group Client Section 1: Introduction This job card outlines how to install & configure the Centricity RA600 Workgroup client on the Dell 1400 workstation. The Workgroup Server allows sharing of image files (and other files with clients in a work group relationship.
  • Page 231 To provide the optional capability of installing a client (via the network) from the server, do the following (if not already done): If a Centricity RA600 folder is not present on the Workgroup server’s “C:” drive, create one and copy all the contents from the root of the Centricity RA600 CD to this folder.
  • Page 232: Sw 013 Install & Setup Of Streaming Video Capture System (Refer To Kit Document Fp10621)

    Software Procedures SW 013 Install & Setup of Streaming Video Capture System (Refer to Kit document FP10621) SW 013 Install & Setup of Streaming Video Capture System (Refer to Kit document FP10621) As of Revision 2, this job card is a placeholder for the kit document to ensure instruction sync with the kit.
  • Page 233: Sw 014 Installing Centricity Ra600 Software Patch

    Section 2: Prerequisites OS is loaded. Video monitor(s), board(s) and driver(s) are properly Installed & configured in the workstation. Centricity RA600 software application is installed and configured to the site. Software patch CD-ROM is at hand. Section 3: Procedure Login as user: Administrator. There may be no password set at this time. If a password has been set it could be: CANal1.
  • Page 234: Chapter 7 Troubleshooting

    Troubleshooting Section 1: Introduction The troubleshooting procedures are described in a series of job cards about the different steps, problems encountered with their corrective actions and useful procedures, when installing and using the workstation. Corrective action should be made by authorized and qualified personnel Caution only.
  • Page 235 Troubleshooting Job Card No. Troubleshooting Job Card TSG 007 Display Centricity RA600 Application Logs TSG 008 Display IP/Netmask/Default Route Configuration Settings TSG 009 Display Loaded Drivers TSG 010 Display System Services TSG 011 Display System Hardware & Software Configuration TSG 012...
  • Page 236: Tsg 001 Telneting To A Workstation

    Troubleshooting TSG 001 Telneting to a Workstation TSG 001 Telneting to a Workstation Section 1: Introduction This job card provides instructions to telenet into a workstation. Be aware that it is possible to render a system inoperable through Caution incorrect use of the telnet access software Section 2: Procedure Note: Telneting in from a Sun system sometimes works best when done from an xterm window.
  • Page 237 Troubleshooting TSG 001 Telneting to a Workstation Microsoft(R) Windows NT(TM) (C) Copyright 1985-1996 Microsoft Corp. c:\> Some “basic” keyboard functions (which may or may not be needed, de- pending on your terminal type): Delete the last character typed Erase the current line The next character will be sent as an “Alt”...
  • Page 238: Tsg 002 Workstation Directory Tree

    Troubleshooting TSG 002 Workstation Directory Tree TSG 002 Workstation Directory Tree Section 1: System Files This section shows an example directory structure for the RA600 (RadWorks) software. c:\Ami_60 +---Bin +---Connect +---Receive +---Work +---CDViewer +---CDViewer +---Colortables +---Database +---Archive +---Archive001 +---Archive002 +---Archive002a +---Archive003 +---Archive004 +---Archive005...
  • Page 239: Tsg 003 Troubleshooting Network Problems

    Troubleshooting TSG 003 Troubleshooting Network Problems TSG 003 Troubleshooting Network Problems Section 1: introduction This job card is used to verify connectivity of the ethernet network connection. Section 2: Monitoring Commands Login as administrator to use the following commands: Ping: connection test # ping 3.28.75.100 [Return] (IP address) 3.28.75.100 is alive...
  • Page 240 Troubleshooting TSG 003 Troubleshooting Network Problems ipconfig /all For troubleshooting isolation steps, run new CAT 5 cable to eliminate possible cable problems. Page 240...
  • Page 241: Tsg 004 Using The Vga Service Monitor

    Troubleshooting TSG 004 Using The VGA Service Monitor TSG 004 Using The VGA Service Monitor Section 1: Introduction This job card explains using the VGA service monitor to troubleshoot video driver problems. It is useful when debugging boot problems or performing any low level diagnostics on portrait monitor systems where the Video Drivers have not started.
  • Page 242 Troubleshooting TSG 004 Using The VGA Service Monitor Illustration 7-1: Service Monitor Connection Page 242...
  • Page 243: Tsg 005 Stop A Software Process

    Troubleshooting TSG 005 Stop a Software Process TSG 005 Stop a Software Process Section 1: Introduction This job card explains how to display and stop a software process. (If a particular process is thought to be problematic.) Section 2: GUI Based Procedure Right click the Taskbar and select Task Manager...
  • Page 244: Tsg 006 Display Windows 2000 System Logs

    Security—Security and Auditing logs. Application - application related logs, not to be confused with the RA600 (RadWorks) workstation applications log file that is separate. See TSG 007 Display Centricity RA600 Application Logs page 245. Section 2: Procedure Select Start > Programs > Administrative Tools > Event Viewer.
  • Page 245: Tsg 007 Display Centricity Ra600 Application Logs

    Troubleshooting TSG 007 Display Centricity RA600 Application Logs TSG 007 Display Centricity RA600 Application Logs Section 1: Introduction This job card explains how to display the RA600 (RadWorks) workstation application logs. The workstation application log file (radworks.log) is an ASCII based log file located in the c:\ami_60\log directory (assuming the RadWorks application is installed on the c:\ drive.
  • Page 246: Tsg 008 Display Ip/Netmask/Default Route Configuration Settings

    Troubleshooting TSG 008 Display IP/Netmask/Default Route Configuration Settings TSG 008 Display IP/Netmask/Default Route Configuration Settings Section 1: Introduction This job card explains how to display current network configuration settings. Section 2: Procedure Select Start > Settings > Control Panel > Network and Dial-up Connections.. Right click on Local area connections and select Properties.
  • Page 247: Tsg 009 Display Loaded Drivers

    Troubleshooting TSG 009 Display Loaded Drivers TSG 009 Display Loaded Drivers Section 1: Introduction This job card explains how to display the drivers currently loaded by the Operating System. Section 2: Procedure Select Start > Settings > Control Panel > Administrative Tools > Computer Management >...
  • Page 248: Tsg 010 Display System Services

    Troubleshooting TSG 010 Display System Services TSG 010 Display System Services Section 1: Introduction This job card explains how to display the system services being used by the Operating System. Section 2: Procedure Select Start > Settings > Control Panel > Administrative Tools > Services. Page 248...
  • Page 249: Tsg 011 Display System Hardware & Software Configuration

    Troubleshooting TSG 011 Display System Hardware & Software Configuration TSG 011 Display System Hardware & Software Configuration Section 1: Introduction This job card explains how to display the current hardware and software configuration of the workstation. Section 2: GUI Based Procedure Select Start >...
  • Page 250: Tsg 012 Display Os Version

    Troubleshooting TSG 012 Display OS Version TSG 012 Display OS Version Section 1: Introduction This job card explains how to display the current OS version. Section 2: GUI Based Procedure Select Start > Settings > Control Panel > System. Select the General tab. The Main Screen will show the OS and its build version, along with any service packs installed.
  • Page 251: Tsg 013 Display Application Software Version

    Troubleshooting TSG 013 Display Application Software Version TSG 013 Display Application Software Version Section 1: Introduction This job card explains how to display the current software version running on the workstation. Section 2: Procedure Select Start > Programs > RadWorks 6.0 > RadWorks 6.0, or click on the Workstation icon on the desktop.
  • Page 252: Tsg 014 Workstation Application Problem Solving

    TSG 014 Workstation Application Problem Solving TSG 014 Workstation Application Problem Solving Section 1: Introduction This job card presents possible solutions to problems that may be encountered in the use of the Centricity RA600 application software. Section 2: Problem Resolution Problem Possible Solution 1.
  • Page 253 6. Check that the imported hanging protocol name appears in the list in the Automated Hanging Protocols Manager (QC) dialog box. 5. Centricity RA600 won’t allow Check box must be selected for direct view. direct view Select Configuration>Viewing>General and check the Direct View box.
  • Page 254 Troubleshooting TSG 014 Workstation Application Problem Solving Problem Possible Solution 7. Centricity RA600 did not save Ensure that you have checked the Automatically save window settings my worklist check box. To find this check box select Configuration>Quality Control>General tab. 8. Can’t retrieve (import) from...
  • Page 255: Tsg 015 Connect Frame Grabber To Monitor That Has No Pass-Through

    Section 2: Procedure Turn on the Centricity RA600 workstation power. Remove the left-hand half of the Dell 1400 case to access the Hi Def Accura board: Slide the rear padlock hasp toward the chassis center and press the release button on the lower front of the bezel and remove the left side of the case.
  • Page 256 Video IN Video OUT Setting Replace the 1400 side cover and restart the Centricity RA600 system. Connect the video source to the CA channel you have selected and then connect the corresponding CT channel to the monitor. Follow job card SW 007 to install the Imagraph software and configure the channel.
  • Page 257: Chapter 8 Disassembly & Reassembly Procedures

    Disassembly & Reassembly Procedures Section 1: Introduction This chapter contains information on correct settings of the video boards inside the workstation computer and reference diagrams for rear panel connections to the workstation computer. The printed circuit boards and hard disk drives contain electronic components that are extremely sensitive to static electricity.
  • Page 258: Video Board Switch Settings

    Disassembly & Reassembly Procedures DR 001 Video Board Switch Settings DR 001 Video Board Switch Settings Section 1: Introduction This job card describes the various video board switch settings. These switches need to be set correctly or the system will not boot. Section 2: Tools Required 5 mm cross-head screwdriver 5 mm flat-head screwdriver...
  • Page 259 Disassembly & Reassembly Procedures DR 001 Video Board Switch Settings Section 3: Procedure Video boards should be installed in slots 4 & 6 (use slot 6 for the second board, if applicable) The following switch settings are used to disable the V settings of the Dome (MDxxxx) and Barco (3MP2FH) video boards and must be set correctly for the system to boot.
  • Page 260: Pc Port Connection Diagram

    Disassembly & Reassembly Procedures DR 002 PC Port Connection Diagram DR 002 PC Port Connection Diagram Section 1: Introduction This job card identifies the onboard ports of the PC. Section 2: Procedure Reference Illustration 8-1 below for connections. For more information on how slot positions are assigned to different monitors, refer to Hardware Installation page...
  • Page 261 Disassembly & Reassembly Procedures DR 002 PC Port Connection Diagram Illustration 8-1: PC Port Connections Page 261...
  • Page 262: Replace-Configure Perc 3/Dc Raid Controller Board

    Disassembly & Reassembly Procedures DR 003 Replace-Configure PERC 3/DC RAID controller board DR 003 Replace-Configure PERC 3/DC RAID controller board Section 1: Introduction This job card describes the replacement and configuration of the PERC 3/DC RAID controller circuit board in the Dell 1400 computer. Note: This procedure is necessary only when controller board is faulty.
  • Page 263 Disassembly & Reassembly Procedures DR 003 Replace-Configure PERC 3/DC RAID controller board Power up the computer. A mismatch error message will appear during system BIOS bootup. When the mismatch error message appears on the monitor screen, press the + keys to run the Configuration Utility. In the Management Menu box, use the arrow and Enter keys to select Configure >...
  • Page 264: Replacement Of 221S Raid Unit Components

    Disassembly & Reassembly Procedures DR 004 Replacement of 221S RAID Unit Components DR 004 Replacement of 221S RAID Unit Components Section 1: Introduction This job card contains information about replacement of RAID components. Section 2: Items Required The replacement FRU (Field Replaceable Unit) must be available (Refer to RAID FRU list in Chapter 9) Philips screwdrivers (Nos.
  • Page 265: Chapter 9 Renewal Parts

    Renewal Parts Section 1: INTRODUCTION This chapter contains the renewal parts lists and procedure for license renewal/replacement. Section 2: Workstation FRU List Note: If the Dell 1400 computer needs replacement, transfer all optional components (e.g. RAM upgrade, Hard drive upgrade, Video & other optiona boards, etc.) from the old to the new computer before returning the old computer for warranty purposes.
  • Page 266 Renewal Parts Part No. Description workstation display(s) 4362604 Dome Video Board MD5-PCX 2295365 Dome Video Board MD4(PCI) 4356812 Dome Video Board MD2(PCL) 2328679 Barco Coronis 3MP Controller Board - Dual Head 2334569 Matrox Millenium G450 Controller for C/FP Mon. replaces 4362476 2306872 Flat Panel Monitor - NEC 2328218...
  • Page 267 Renewal Parts Part No. Description Keyboards & mouse 2300612 104 Key, Keyboard (Dell P/N 310-4100) 2266207 French Keyboard 2266208 Spanish Keyboard 2266209 German Keyboard 2266210 Portugese Keyboard 2266212 Swedish Keyboard 2266215 European Power Cable 2266211 Italian Keyboard 2266216 Italian Power Cable 2303424 Mechanical Serial wheel mouse (Dell P/N 310-3777) Hi-Res service kit...
  • Page 268 Renewal Parts Part No. Description 2270474 Foresight Imaging HI*DEF Accura board 016030-200 2270476 Foresight Imaging 030063 8-channel video cable iColor Frame Grabber 2308031 Frame Grabber Board - Single frame, color 2324574 Frame Grabber Board - Streaming, color (TIMS) 2335175 S-Video Y/C Adapter for FRMGRB 2335180 25DB to 6BNC Jack 7ft Cable-GRMGRB Work Group Server...
  • Page 269 If the customer damages or loses this dongle, a replacement dongle can be attained through GE Cares. The dongle is set up as a FRU 1 and is in stock under part number 2253361.
  • Page 270 Renewal Parts Dongles are coded specifically to a serial number based on the modules and options purchased. A replacement dongle needs to be coded so it gives the customer the same capabilities as their original dongle. License files are used on some systems instead of a dongle. License files are coded to the MAC address of the primary network adapter in the computer and as such, can not be moved from machine to machine.
  • Page 271 Renewal Parts 3-3 Global Procedure for the FE to request the ROC to provide a license file or a dongle code. All requests must be handled by the ROC in the requesting pole. 3-3-1 Job to be performed by the FE during a new installation or an upgrade At installation time, the FE records the either dongle serial number or the MAC address in the Service Manual, At installation time, or for every new option or upgrade, the FE records the FDO number in the...
  • Page 272 3-4-4 PROCESS TO REPLACE A DONGLE BY A LICENSE FILE Radworks 5.1 Turnkey, damaged dongle GE FE contact the ROC with customer details (Name, Address, FDO), MAC address and registration information ( type of software and modules list) ROC generates a temporary license file (1 month) and send it to the FE, or upload it on the...
  • Page 273 On a quarterly basis, ROC send damaged dongles to manufacturing at Mt Prospect Radworks 5.1 Turnkey, lost / stolen dongle GE FE contacts the ROC with customer details Name, Address, FDO), MAC address and registration information (type of software and modules list)
  • Page 274 Renewal Parts [email protected] [email protected] Page 274...
  • Page 275: Chapter 10 Suggested Maintenance

    This chapter presents job cards related to suggested maintenance procedures for the Centricity RA600 Workstation. The Centricity RA600 Workstation does not require extensive routine hardware maintenance. Reasonable care of its components will help ensure continuous operation. Section 2: \Summary of Job Cards Job Card No.
  • Page 276: Mnt 001 Scheduled Maintenance

    Suggested Maintenance MNT 001 Scheduled Maintenance MNT 001 Scheduled Maintenance Section 1: INTRODUCTION This job card contains the list of annual maintenance tasks to perform on the Centricity RA600 Workstation. Pay special note to the electrostatic discharge precautions when working Caution inside the Computer chassis.
  • Page 277 Suggested Maintenance MNT 001 Scheduled Maintenance Carefully vacuum out the system, attending to heavy accumulation spots such as the processor area. Before replacing the computer cover visually verify that the cooling fans are working properly. Note: The use of compressed air is not recommended. Blowing dust out with a stream of air is not recommended because of the ESD (Electro-Static Discharge) hazards to components, and the dust hazard to the personnel in the area near the cleaning activity.
  • Page 278 GE Medical Systems – Information Technologies GE Medical Systems – Information Technologies: Telex 3787371 8200 W. Tower Ave., Milwaukee, Wisconsin, 53201 U.S.A. (Asia, Pacific, Latin America, North America) GE Medical Systems – Information Technologies GmbH Munzinger Strasse 3, D-79111, Frieburg, Germany...