Table of Contents

Quick Links

FUJI-Reader
Service Manual
Preventive Maintenance
Print No.: SPB7-420.840.55.01.02
Replaces: n.a.
SP
© Siemens AG 2003
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 01.03
Table of Contents
loading

Summary of Contents for Siemens FUJI-Reader

  • Page 1 FUJI-Reader Service Manual Preventive Maintenance © Siemens AG 2003 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent...
  • Page 2 CR-IR347/CR-IR347P Service Manual Preventive Maintenance (PM)
  • Page 3: Table Of Contents

    Contents Preventive Maintenance CR-IR347 Service Manual – Contents Preventive Maintenance (PM) Preventive Maintenance Program ................. PM-2 Preventive Maintenance Program List ..............PM-2 Preventive Maintenance Flow ................PM-4 Checking the Intervals ..................PM-8 Precautions for Preventive Maintenance ............PM-9 Checks Before Maintenance ..................PM-10 Checking the Error Log ..................
  • Page 4 Contents Preventive Maintenance Erasure Conveyor (Cleaning/Replacing the Shock Absorber-Attached Guide Plates) ....................... PM-54 Erasure Conveyor (Cleaning the Rubber Rollers) ........... PM-55 Erasure Conveyor (Installing the Lamp House Assembly/Shock Absorber- Attached Reflection Plate) ................. PM-56 Erasure Conveyor (Installing the Cleaning Rollers) ........PM-57 8.10 Installing the Erasure Conveyor ................
  • Page 5 PM - 1 Preventive Maintenance (PM) Control Sheet Control Sheet Issue date Revision number Reason Pages affected 10/20/2000 New release (FM2732) All pages 05/15/2001 Corrections (FM3052) PM - 94, 97 08/30/2001 Design changes; measures against noise PM - 35, 38, 43, 55, 57-59, 62, (FM3142) 68, 69, 72, 74, 82, 90, 91, 94 08/30/2002...
  • Page 6: Preventive Maintenance Program

    PM - 2 Preventive Maintenance Program Preventive Maintenance Program List Work time Tools/jigs, measuring Periodically Refer to Program Intervals (min) apparatus replaced parts Checking the Error Log 0.5Y 5 min Checking the Image 0.5Y 20 min Cleaning the FD Drive and Magnetic FD drive cleaner 0.5Y 5 min...
  • Page 7 PM - 3 Work time Tools/jigs, measuring Periodically Refer to Program Intervals (min) apparatus replaced parts Subscanning Unit (Cleaning the SUS Belt/ 10 min ¤ Ethanol Rubber Belt/Kapton Belt) Pulling Out the Scanner Unit 5 min Removing the Side-Positioning Conveyor 5 min Side-Positioning Conveyor (Cleaning/ Shock absorber-...
  • Page 8: Preventive Maintenance Flow

    PM - 4 Preventive Maintenance Flow START 2.1 Checking the Error Log 2.2 Checking the Image Cleaning the FD Drive and Magnetic Card Reader Clean 0.5Y <> Turning OFF the High-Voltage Switch Cleaning/Replacing the Air Filters Replace Clean...
  • Page 9 PM - 5 Removing the Erasure Conveyor Erasure Conveyor (Cleaning/Replacing the Erasure Lamps) Replace Clean Erasure lamps Erasure Conveyor (Cleaning/Replacing the Cleaning Replace Clean Rollers) Cleaning rollers Replace Erasure Conveyor (Cleaning/Replacing the Shock Clean Shock absorber- Absorber-Attached Reflection Plate) attached reflection plate Replace Erasure Conveyor (Cleaning/Replacing the Shock Clean...
  • Page 10 PM - 6 Side-Positioning Conveyor (Cleaning/Replacing the Shock Replace Absorber-Attached Guide Plates) Shock absorber- Clean attached guide plates Side-Positioning Conveyor (Cleaning the Rubber Rollers) Clean Subscanning Unit (Removing the Light-Collecting Guide Assemblies) Subscanning Unit (Cleaning the Light-Collecting Guide Clean Assemblies) Subscanning Unit (Removing the Scanning Optics Unit) Subscanning Unit (Cleaning the Light-Collecting Mirror) Clean...
  • Page 11 PM - 7 Turning ON the High-Voltage Switch Installing and Cleaning the Covers Checking the Date and Time Deleting the Error Log 0.5Y Checking the Image FR7H6189.EPS 009-058-00 CR-IR347 Service Manual PM - 7 10.20.2000 FM2732 (2)
  • Page 12: Checking The Intervals

    PM - 8 Checking the Intervals To facilitate the understanding of the inspection intervals of each preventive maintenance item, the item is indicated by one of the following icons in this volume. Indicates that the procedures should be performed at intervals of half year. 0.5Y Indicates that the procedures should be performed at intervals of one year.
  • Page 13: Precautions For Preventive Maintenance

    PM - 9 Precautions for Preventive Maintenance CAUTION The red-painted screws should never be removed. INSTRUCTION The yellow-painted screws require adjustments when components are reinstalled. When reinstalling the components, follow the check/adjustment procedures. REFERENCE With this machine, the mounting screws for the shock absorber-attached guide plates to be replaced are blue-painted.
  • Page 14 PM - 10 Checks Before Maintenance At intervals of half year Check. 0.5Y Checking the Error Log (1) Start the M-Utility. (2) Open the error log. REFERENCE Below is shown an example where errors and warning messages are displayed. To check the number of occurrences for each error code, select “2.
  • Page 15 PM - 11 CODE DATE 081D 1998.04.16 14:36 11.E738.000002.8A0AC1.000000.000 C19.8 000001.000009.00001B 1722 1998.04.16 09:58 85AA1319.05972166 0.END 1.NEXT 2.BEFORE (DEFAULT=1) : FR7H6102.EPS (3) Check the error log displayed. CHECK Check if error logs are generated. If error logs are generated, see [Troubleshooting Volume]. 009-058-00 CR-IR347 Service Manual PM - 11...
  • Page 16: Checking For Image Problems

    PM - 12 Checking for Image Problems I Check for Nonuniformity, Sensitivity, and Density (1) Expose two IPs of all sizes found at the institution (both ST and HR types). Expose the first in the normal direction, and the second in a 180-degree rotated orientation. The IP exposure conditions are as follows.
  • Page 17 PM - 13 (4) Using the output film, check the sensitivity. N CHECK N Make sure that the S value on the film is held within the following range. ST: 140 ≤ S value ≤ 260 HR: 84 ≤ S value ≤ 156 (5) Make sure that there is no density unbalance in the image of the output film.
  • Page 18 PM - 14 I Check for Jitter and HR (1) Place steel rules and a washer on IPs of all sizes found at the institution, and expose them. (Both ST and HR types. For the layout, see the illustration below.) The IP exposure conditions are as follows.
  • Page 19 PM - 15 (2) Using the test menu, record and output an image. ❍ ST “ ” (for Japanese version)/“IMAGE FORMAT” (for versions other than Japanese) ❍ HR “ ” (for Japanese version)/“BREAST, L (DSR)” or “BREAST, R (DSR)” (for U.S. version)/“IMAGE FORMAT2” (for European version) (3) Make sure that there is no jitter in the image of the border and steel rules on the output film.
  • Page 20 PM - 15.1 I Check for Format Make sure that the white blank portion, as measured from the outermost side of the image border of the output film, is 2 mm or less in actual size on the IP. N CHECK N Make sure that the white blank portion is 2 mm or less.
  • Page 21 PM - 15.2 I Check for Output Characters Make sure that the output characters are indicated correctly on the output film. N CHECKS N • Make sure that the output characters are indicated correctly. If anything abnormal is found, refer to “4. [2] CONFIGURATION SETTING” in the “Maintenance Utility”...
  • Page 22 PM - 15.3 BLANK PAGE 009-058-03 CR-IR347 Service Manual PM - 15.3 08.30.2002 FM3476...
  • Page 23 PM - 15.4 BLANK PAGE 009-058-03 CR-IR347 Service Manual PM - 15.4 08.30.2002 FM3476...
  • Page 24: Cleaning The Fd Drive And Magnetic Card Reader

    PM - 16 Cleaning the FD Drive and Magnetic Card Reader At intervals of half year Clean. 0.5Y Cleaning the FD Drive (1) Start the M-Utility. (2) Select the FD reading. 0. QUIT 1. ERROR LOG UTILITY 2. CONFIGURATION SETTING 3.
  • Page 25 PM - 17 (3) Insert the FD drive cleaner into the FD drive. (4) Execute the FD reading. REFERENCE The number to be selected for the FD reading at step #1 may be either 1 through 6. “5. CSL MENU DATA” is selected by way of example. 0.QUIT 1.SCANNER DATA 2.CONFIGURATION DATA...
  • Page 26: Removing The Covers

    PM - 18 Removing the Covers At intervals of one year, or when 60,000 IPs have been processed Remove. WARNING In order to avoid electric shock hazards, power OFF the machine before performing any procedures. [Remove] T4x12 (4) Upper left-hand side cover [Remove] T4x6 (x4) [Remove] BR4x8 (x2) Upper front cover...
  • Page 27 PM - 19 Upper rear cover [Remove] T4x12 (x6) #10 [Remove] T4x6 (x3) T4x12 (x4) Lower left-hand side cover #11 [Remove] T4x12 (x8) Lower rear cover #6 [Remove] T4x6 (x7) [Remove] BR4x8 (x6) #5 [Loosen] T4x12 (x2) Inner left-hand side cover [Remove] BR4x8 (x5) Inner right-hand side cover...
  • Page 28: Turning Off The High-Voltage Switch

    PM - 20 Turning OFF the High-Voltage Switch At intervals of one year, or when 60,000 IPs have been processed Turn OFF. CAUTION Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is powered ON with any of the covers removed, the photomultiplier will be damaged. Turn OFF the high-voltage switch (S1) of the SCN08D board.
  • Page 29 PM - 21 BLANK PAGE 009-058-00 CR-IR347 Service Manual PM - 21 10.20.2000 FM2732 (2)
  • Page 30: Cleaning/Replacing The Air Filters

    PM - 22 Cleaning/Replacing the Air Filters At intervals of one year, or when 60,000 IPs have been processed Clean. At intervals of two years, or when 120,000 IPs have been processed Replace. Upper Left-Hand Side Cover and Lower Left-Hand Side Cover Upper left-hand side cover Air filter [Clean] Vacuum cleaner...
  • Page 31 PM - 23 Lower Right-Hand Side Cover Velcro tape Air filter [Clean] Vacuum cleaner [Replace] Lower right-hand side cover FR7H6021.EPS 009-058-00 CR-IR347 Service Manual PM - 23 10.20.2000 FM2732 (2)
  • Page 32: Cassette Set Unit And Ip Removal Unit

    PM - 24 Cassette Set Unit and IP Removal Unit Removing the Cassette Set Unit At intervals of one year, or when 60,000 IPs have been processed Remove. (1) Disconnect the connectors. [Disconnect] Connectors IP removal unit CNL2 CNL3 CNL1 [Disconnect] Connectors Cassette set unit CNU1...
  • Page 33 PM - 25 (2) Pull out the tray. #1 [Loosen] BR4x8 (x2) Bracket #2 [Remove] BR4x8 (x2), W6 [Remove] BR4x8 (x2) [Pull-out] Tray FR7H6928.EPS 009-058-00 CR-IR347 Service Manual PM - 25 10.20.2000 FM2732 (2)
  • Page 34 PM - 26 (3) Remove the front shelf cover. [Remove] Screw cover [Remove] T4x6 (x2) Stack table [Remove] BR4x8 (x2) Front shelf cover [Remove] BR4x8 (x4) FR7H6093.EPS 009-058-00 CR-IR347 Service Manual PM - 26 10.20.2000 FM2732 (2)
  • Page 35 PM - 27 (4) Remove the cassette set unit. NOTE To remove the cassette set unit, move the IP removal arm to the position shown in the illustration below. [Rotate] Timing belt IP removal unit IP removal arm [Remove] BR4x8 (x3) Cassette set unit Cassette FR7H3315.E...
  • Page 36: Ip Removal Unit (Cleaning The Suction Cups

    PM - 28 IP Removal Unit (Cleaning the Suction Cups) At intervals of one year, or when 60,000 IPs have been processed Clean. NOTE To clean the suction cups, move the IP removal arm to the position shown in the illustration below.
  • Page 37 PM - 29 (2) Clean the suction cups. Suction face view [Clean] Moistened cloth FR7H6156.EPS (3) Put the IP removal arm back toward the rear. 009-058-00 CR-IR347 Service Manual PM - 29 10.20.2000 FM2732 (2)
  • Page 38: Ip Removal Unit (Cleaning/Replacing The Rubber Rollers Located Under The Ip Removal Arm And Shock Absorber-Attached Guide Plates

    PM - 30 IP Removal Unit (Cleaning/Replacing the Rubber Rollers Located Under the IP Removal Arm and Shock Absorber- Attached Guide Plates) At intervals of one year, or when 60,000 IPs have been processed Clean the rubber rollers and shock absorber-attached guide plates. At intervals of three years, or when 180,000 IPs have been processed Replace the shock absorber-attached guide plates.
  • Page 39 PM - 31 (3) Clean or replace the shock absorber-attached guide plates. [Remove] BR4x8 Shock absorber-attached [Remove] BR4x16 guide plate [Clean] Vacuum cleaner [Replace] [Remove] BR4x8 Stay [Loosen] B3x6 (x4) FR7H6031.EPS 009-058-00 CR-IR347 Service Manual PM - 31 10.20.2000 FM2732 (2)
  • Page 40 PM - 32 (4) By reversing the procedures of step (1), reinstall the accepting roller assemblies (small and large). CHECKS • There should be no gap between the shock absorber-attached guide plates and the stay. Shock absorber-attached guide plate Stay FR7H6929.EPS •...
  • Page 41: Ip Removal Unit (Cleaning/Replacing The Shock Absorber-Attached Guide Plates Located Under The Ip Removal Unit

    PM - 33 IP Removal Unit (Cleaning/Replacing the Shock Absorber- Attached Guide Plates Located Under the IP Removal Unit) At intervals of one year, or when 60,000 IPs have been processed Clean. At intervals of three years, or when 180,000 IPs have been processed Replace.
  • Page 42: Ip Removal Unit (Cleaning/Replacing The Shock Absorber-Attached Guide Plates In The Conveyance Path Of The Ip Removal Unit

    PM - 34 IP Removal Unit (Cleaning/Replacing the Shock Absorber- Attached Guide Plates in the Conveyance Path of the IP Removal Unit) At intervals of one year, or when 60,000 IPs have been processed Clean the rubber rollers and shock absorber-attached guide plates. At intervals of three years, or when 180,000 IPs have been processed Replace the shock absorber-attached guide plates.
  • Page 43 PM - 35 Cleaning/Replacement Procedures (1) Remove the shock absorber-attached guide plate A. [Remove] B3x6 (x6) [Remove] BR4x8 [Disconnect] Connector [Remove] Bracket A3x6 (x2) Anti-static member Shock absorber-attached guide plate A [Clean] Vacuum cleaner [Replace] FR7H6097.EPS 009-058-00 009-058-02 CR-IR347 Service Manual PM - 35 10.20.2000 FM2732 (2) 08.30.2001 FM3142...
  • Page 44 PM - 36 (2) Remove the shock absorber-attached guide plates B and C. [Remove] B3x6 (x4) [Remove] B3x6 (x4) Shock absorber-attached Shock absorber-attached guide plate C guide plate B [Clean] [Clean] Vacuum cleaner Vacuum cleaner [Replace] [Replace] FR7H6098.EPS 009-058-00 CR-IR347 Service Manual PM - 36 10.20.2000 FM2732 (2)
  • Page 45 PM - 37 (3) Remove the shock-absorber-attached guide plate D. [Remove] B3x6 (x4) Shock absorber-attached guide plate D [Clean] Vacuum cleaner [Replace] FR7H6150.EPS 009-058-00 CR-IR347 Service Manual PM - 37 10.20.2000 FM2732 (2)
  • Page 46: Ip Removal Unit (Cleaning The Rubber Rollers

    PM - 38 IP Removal Unit (Cleaning the Rubber Rollers) At intervals of one year, or when 60,000 IPs have been processed Clean. (1) Clean the rubber rollers. (2) By reversing the procedures described in “7.5 IP Removal Unit (Cleaning/Replacing the Shock Absorber-Attached Guide Plates in the Conveyance Path of the IP Removal Unit),”...
  • Page 47: Cassette Set Unit (Cleaning The Rubber Roller

    PM - 39 Cassette Set Unit (Cleaning the Rubber Roller) At intervals of one year, or when 60,000 IPs have been processed Clean. Rubber Roller [Clean] Moistened cloth FR7H6152.EPS 009-058-00 CR-IR347 Service Manual PM - 39 10.20.2000 FM2732 (2)
  • Page 48: Installing The Cassette Set Unit

    PM - 40 Installing the Cassette Set Unit At intervals of one year, or when 60,000 IPs have been processed Install. NOTE To install the cassette set unit, move the IP removal arm to the position shown in the illustration below.
  • Page 49 PM - 41 (2) Push in the tray. [Attach] BR4x8 [Push-in] Tray #3 [Connect] Connectors #4 [Connect] Connectors IP removal unit Cassette set unit CNL2 CNL3 CNK1 CNSOLK1 CNL1 BCR08A CN1 CNK2 FR7H6099.EPS 009-058-00 CR-IR347 Service Manual PM - 41 10.20.2000 FM2732 (2)
  • Page 50 PM - 42 (3) Install the front shelf cover. [Tighten] BR4x8 (x2) Bracket [Attach] BR4x8 (x4), W6 Bracket [Attach] BR4x8 (x5) Front shelf cover [Connect] CNU1 [Attach] Screw cover [Attach] T4x6 (x2) Stack table FR7H6153.EPS 009-058-00 CR-IR347 Service Manual PM - 42 10.20.2000 FM2732 (2)
  • Page 51 PM - 43 NOTE When installing the front shelf cover, align the cassette insertion face of the front shelf cover to the guide protrusion of the cassette set unit. If they are not aligned, a cassette hold failure may result. CHECK DETAIL A #2 [Align]...
  • Page 52: Erasure Conveyor

    PM - 44 Erasure Conveyor Removing the Erasure Conveyor At intervals of one year, or when 60,000 IPs have been processed Remove. NOTE The erasure lamp should be handled carefully. If it is hit, the erasure lamp may be broken, so you may get injured.
  • Page 53 PM - 45 (2) Remove the erasure conveyor. #1 [Disconnect] Connectors CN-M2 CN-M1 Erasure conveyor [Remove] BR4x8 (x2) FR7H6041.EPS 009-058-00 CR-IR347 Service Manual PM - 45 10.20.2000 FM2732 (2)
  • Page 54 PM - 46 Erasure Conveyor (Cleaning/Replacing the Erasure Lamps) At intervals of one year, or when 60,000 IPs have been processed ¡ Clean. At intervals of two years, or when 120,000 IPs have been processed ¡ Replace. CAUTION Since the erasure lamp may be heated to a high temperature, exercise care not to suffer burns. (1) Remove the five erasure lamps.
  • Page 55 PM - 47 (2) Clean the reflection face within the erasure lamp assembly and the five erasure lamps, or replace the five erasure lamps. Erasure lamp (x5) [Clean] Moistened cloth Socket [Replace] Reflection face [Clean] Moistened cloth FR7H6154.EPS (3) By reversing the procedures of step (1), reinstall the erasure lamps. 009-058-00 CR-IR347 Service Manual PM - 47...
  • Page 56: Erasure Conveyor (Cleaning/Replacing The Cleaning Rollers

    PM - 48 Erasure Conveyor (Cleaning/Replacing the Cleaning Rollers) At intervals of one year, or when 60,000 IPs have been processed Clean. At intervals of two years, or when 120,000 IPs have been processed Replace. (1) Remove the cleaning roller assembly. Cleaning roller assembly [Remove] BR4x8 FR7H6045.EPS...
  • Page 57 PM - 49 (2) Clean the cleaning rollers and inside the cleaning roller assembly, or replace the cleaning rollers. Cleaning roller [Clean] Vacuum cleaner #1 [Loosen] B3x6 (x4) Cleaning roller assembly [Clean] Vacuum cleaner [Remove] B3x8...
  • Page 58: Erasure Conveyor

    PM - 50 Erasure Conveyor (Cleaning/Replacing the Shock Absorber-Attached Reflection Plate) At intervals of one year, or when 60,000 IPs have been processed Clean. At intervals of three years, or when 180,000 IPs have been processed Replace. #2 Shock absorber-attached reflection plate [Clean] Moistened cloth...
  • Page 59 PM - 51 BLANK PAGE 009-058-00 CR-IR347 Service Manual PM - 51 10.20.2000 FM2732 (2)
  • Page 60: Filter

    PM - 52 Erasure Conveyor (Cleaning/Replacing the Shock Absorber-Attached Filter) At intervals of one year, or when 60,000 IPs have been processed Clean At intervals of three years, or when 180,000 IPs have been processed Replace NOTE When cleaning the shock absorber-attached filter with a moistened cloth, use care not to touch the shock absorber.
  • Page 61 PM - 53 (2) Clean or replace the shock absorber-attached filter. Lamp house assembly [Remove] BR4x8 (x5) Spacer Shock absorber-attached filter [Clean] Moistened cloth [Remove] BR4x6 [Replace] FR7H6105.EPS 009-058-00 CR-IR347 Service Manual PM - 53 10.20.2000 FM2732 (2)
  • Page 62: Erasure Conveyor (Cleaning/Replacing The Shock Absorber-Attached Guide Plates

    PM - 54 Erasure Conveyor (Cleaning/Replacing the Shock Absorber-Attached Guide Plates) At intervals of one year, or when 60,000 IPs have been processed Clean At intervals of three years, or when 180,000 IPs have been processed Replace (1) Remove the shock absorber-attached guide plates. Erasure conveyor Shock absorber-attached guide plate...
  • Page 63: Erasure Conveyor (Cleaning The Rubber Rollers

    PM - 55 Erasure Conveyor (Cleaning the Rubber Rollers) At intervals of one year, or when 60,000 IPs have been processed Clean. Rubber roller CHECK Anti-static member CHECK CHECK Rubber roller Anti-static [Clean] member Moistened cloth CHECK Erasure conveyor FR7H6159.EPS CHECK Make sure that the anti-static members are brought into contact with the rollers.
  • Page 64: Erasure Conveyor (Installing The Lamp House Assembly/Shock Absorber-Attached Reflection Plate

    PM - 56 Erasure Conveyor (Installing the Lamp House Assembly/ Shock Absorber-Attached Reflection Plate) At intervals of one year, or when 60,000 IPs have been processed Install. (1) Install the lamp house assembly. Lamp house assembly Erasure conveyor [Attach] BR4x8 (x4) FR7H6160.EPS (2) Install the shock absorber-attached reflection plate.
  • Page 65: Erasure Conveyor (Installing The Cleaning Rollers

    PM - 57 Erasure Conveyor (Installing the Cleaning Rollers) At intervals of one year, or when 60,000 IPs have been processed Install. DETAIL A Cleaning roller Anti-static assembly member CHECK [Attach] BR4x8 FR7H6169.EPS CHECK Make sure that the anti-static members are brought into contact with the rollers. 009-058-02 009-058-00 CR-IR347 Service Manual...
  • Page 66: Installing The Erasure Conveyor

    PM - 58 8.10 Installing the Erasure Conveyor At intervals of one year, or when 60,000 IPs have been processed Install. CAUTION The erasure lamp should be handled carefully. If it is hit, the erasure lamp may be broken, so you may get injured.
  • Page 67 PM - 59 (2) Install the erasure conveyor. [Attach] BR4x8 (x2) [Connect] [Connect] CN-M1 CN-M2 FR7H6938.EPS (3) Insert the erasure lamp assembly into the erasure conveyor. Erasure conveyor [Connect] Connectors Positioning pin DCOUT5 DCOUT4 ACOUT1 ACOUT2 [Insert] Erasure lamp assembly [Attach] BR4x8 FR7H6051.EPS 009-058-02...
  • Page 68: Subscanning Unit And Side-Positioning Conveyor

    PM - 60 Subscanning Unit and Side-Positioning Conveyor Subscanning Unit (Cleaning the SUS Belt/Rubber Belt/ ® Kapton Belt) At intervals of two years, or when 120,000 IPs have been processed Clean. CAUTION Wear gloves when cleaning the belts. REFERENCE Remove/install the rubber belt while rotating the driving shaft. SUS and Rubber Belts Rubber belt [Clean]...
  • Page 69: Subscanning Unit (Cleaning The Sus Belt/Rubber Belt/Kapton ® Belt

    PM - 61 ® Kapton Belt ® Kapton belt [Clean] Cloth (ethanol) DETAIL A Transparent cover [Remove] Extension [Remove] BR4x8 (x2) spring FR7H6920.EPS INSTRUCTIONS ® • When installing the Kapton belt, ensure that the arrow mark printed on it faces toward the subscanning unit.
  • Page 70: Pulling Out The Scanner Unit

    PM - 62 Pulling Out the Scanner Unit At intervals of one year, or when 60,000 IPs have been processed Pull out. CAUTION When servicing any printed circuit board, be sure to wear an anti-static wristband to ground your body. If your body is not grounded, static electricity built on your body may cause damage to electric parts on the board.
  • Page 71 PM - 63 HHS label #2 [Disconnect] Connectors CN-Z12 CN12 SCR08D board SCN08D board FR7H6163.EPS 009-058-00 CR-IR347 Service Manual PM - 63 10.20.2000 FM2732 (2)
  • Page 72 PM - 64 (2) Pull out the scanner unit. HHS label #2 Ground [Remove] CS4x8 [Remove] BR4x8 (x2) [Pull-out] Scanner unit FR7H6164.EP 009-058-00 CR-IR347 Service Manual PM - 64 10.20.2000 FM2732 (2)
  • Page 73 PM - 65 BLANK PAGE 009-058-00 CR-IR347 Service Manual PM - 65 10.20.2000 FM2732 (2)
  • Page 74: Removing The Side-Positioning Conveyor

    PM - 66 Removing the Side-Positioning Conveyor At intervals of one year, or when 60,000 IPs have been processed Remove. (1) Disconnect the connectors. DETAIL A #1 [Disconnect] Connectors #2 [Disconnect] Connectors CN-MN4 CN-MN1 CN-MN2 CN-MN3 #3 [Disconnect] Connectors Scanner unit BCR08A CN-N1 FR7H6165.EPS...
  • Page 75 PM - 67 (2) Remove the side-positioning conveyor. [Remove] BR4x8 (x2) Side-positioning conveyor Scanner unit [Remove] BR4x8 FRONT Motor [Remove] BR4x8 FR7H6166.EPS 009-058-00 CR-IR347 Service Manual PM - 67 10.20.2000 FM2732 (2)
  • Page 76: Side-Positioning Conveyor (Cleaning/Replacing The Shock Absorber-Attached Guide Plates

    PM - 68 Side-Positioning Conveyor (Cleaning/Replacing the Shock Absorber-Attached Guide Plates) At intervals of one year, or when 60,000 IPs have been processed Clean. At intervals of three years, or when 180,000 IPs have been processed Replace. (1) Remove the cleaning guide assembly. Guide plate E Guide plate D FRONT...
  • Page 77 PM - 69 (2) Clean or replace the shock absorber-attached guide plates A and B. NOTE The cleaning guide assembly should not be installed on the side-positioning conveyor. DETAIL A Threaded Threaded hole hole GOOD Shock absorber-attached guide plates A and B [Clean] [Remove] BR4x8 (x2) Vacuum cleaner...
  • Page 78 PM - 70 (3) Clean or replace the shock absorber-attached guide plates C, D, and E. NOTE The shock absorber-attached guide plates C, D, and E should not be installed on the side-positioning conveyor. Shock absorber attached guide plates C, D, E [Remove] B3x6 (x2) [Clean] Vacuum cleaner...
  • Page 79 PM - 71 (4) Remove the rubber roller assembly (upper). [Remove] Tension coil spring (x8) (L = 116 mm) (L=116 mm) (L=121 mm) [Remove] Rubber roller assembly (x2) Side-positioning conveyor (L=121 mm) FR7H6114.EPS 009-058-00 CR-IR347 Service Manual PM - 71 10.20.2000 FM2732 (2)
  • Page 80 PM - 72 (5) Clean the shock absorber-attached guide plates F, G, and H, or replace them. NOTE Do not install the shock absorber-attached guide plates F, G, and H to the side-posi- tioning conveyor. Shock absorber- attached guide plate G FRONT Shock absorber-attached guide plate G...
  • Page 81: Side-Positioning Conveyor (Cleaning The Rubber Rollers

    PM - 73 Side-Positioning Conveyor (Cleaning the Rubber Rollers) At intervals of one year, or when 60,000 IPs have been processed Clean. (1) Clean the rubber rollers. FRONT Rubber roller Rubber roller [Clean] Moistened cloth Side-positioning conveyor FR7H6174.EPS 009-058-00 CR-IR347 Service Manual PM - 73 10.20.2000 FM2732 (2)
  • Page 82 PM - 74 (2) By reversing the procedures described in “9.4 Side-Positioning Conveyor (Cleaning/ Replacing the Shock Absorber-Attached Guide Plates),” reinstall the shock absorber- attached guide plates A through H and the cleaning roller assembly. CHECKS • Make sure that the right and left rubber roller assemblies are not installed incorrectly. •...
  • Page 83 PM - 75 BLANK PAGE 009-058-00 CR-IR347 Service Manual PM - 75 10.20.2000 FM2732 (2)
  • Page 84: Subscanning Unit (Removing The Light-Collecting Guide Assemblies

    PM - 76 Subscanning Unit (Removing the Light-Collecting Guide Assemblies) At intervals of one year, or when 60,000 IPs have been processed Remove. CAUTIONS • Never touch the incidence surface of the light-collecting guide even when you wear gloves. • When servicing a PC board, be sure to wear a wristband to ground your body. If you do not ground your body, the buildup of static electricity on your body may damage electronic components on the PC board.
  • Page 85 PM - 77 (2) Rotate the lever counterclockwise. (3) Remove the light-collecting guide assembly (for the front). Light-collecting guide assembly FRONT Roller bearing Support (3) [Remove] Light-collecting guide assembly (for the front) (2) #2 [Rotate] Lever (2) #1 [Remove] BR4x8 FR7H6122.EPS 009-058-00 CR-IR347 Service Manual...
  • Page 86 PM - 78 Light-Collecting Guide Assembly (for the Back) (1) Disconnect the PMR08C board connectors. [Disconnect] [Disconnect] CN1 PMR08C board [Disconnect] CN2 FR7H6123.EPS (2) Remove the bracket. [Remove] BR4x8 [Remove] Bracket FR7H6903.EPS 009-058-00 CR-IR347 Service Manual PM - 78 10.20.2000 FM2732 (2)
  • Page 87 PM - 79 (3) Rotate the lever clockwise. (4) Remove the light-collecting guide assembly (for the back). [Rotate] Lever [Remove] Light-collecting guide assembly (for the back) DETAIL A Rubber roller (driving side) Guide Latch Light-collecting guide assembly Guide Latch FR7H6124.EPS 009-058-00 CR-IR347 Service Manual PM - 79...
  • Page 88: Subscanning Unit (Cleaning The Light-Collecting Guide Assemblies

    PM - 80 Subscanning Unit (Cleaning the Light-Collecting Guide Assemblies) At intervals of one year, or when 60,000 IPs have been processed Clean. NOTES • Wipe the light-collecting guide assembly in one direction at a constant rate. • If the lens cleaning paper becomes soiled, replace it with a new one. •...
  • Page 89 PM - 81 BLANK PAGE 009-058-00 CR-IR347 Service Manual PM - 81 10.20.2000 FM2732 (2)
  • Page 90: Subscanning Unit (Removing The Scanning Optics Unit

    PM - 82 Subscanning Unit (Removing the Scanning Optics Unit) At intervals of one year, or when 60,000 IPs have been processed Remove. CAUTIONS • Carefully handle the scanning optics unit, using care not to shock it. • Never put your finger into the emission port located in the bottom of the scanning optics unit. •...
  • Page 91 PM - 83 (2) Remove the scanning optics unit. Scanning optics unit [Remove] BR4x20 (x4) FR7H6904.EPS 009-058-00 CR-IR347 Service Manual PM - 83 10.20.2000 FM2732 (2)
  • Page 92: Subscanning Unit (Cleaning The Light-Collecting Mirror

    PM - 84 Subscanning Unit (Cleaning the Light-Collecting Mirror) At intervals of one year, or when 60,000 IPs have been processed Clean. CAUTIONS • Wear gloves when cleaning the light-collecting mirror. • Never touch the reflection surface of the light-collecting mirror even when you are wearing the gloves.
  • Page 93 PM - 85 (2) Remove the mirror mechanism assembly. REFERENCE To remove the screw (CS3x6) of the mirror assembly, insert an Allen wrench or the like into the shaft of the mirror mechanism. [Remove] Tension coil spring [Disconnect] Connector (CNZ11) Allen wrench [Remove] CS3x6 [Remove]...
  • Page 94 PM - 86 (3) Clean the reflection surface of the light-collecting mirror. Mirror mechanism assembly Light-collecting mirror Reflection surface [Clean] Lens cleaning paper (ethanol) FR7H6066.EPS 009-058-00 CR-IR347 Service Manual PM - 86 10.20.2000 FM2732 (2)
  • Page 95: Subscanning Unit (Cleaning The Rubber Rollers

    PM - 87 9.10 Subscanning Unit (Cleaning the Rubber Rollers) At intervals of one year, or when 60,000 IPs have been processed Clean. FRONT Subscanning unit Center guide plate Rubber roller (x4) [Clean] Moistened cloth FR7H6906.EPS 009-058-00 CR-IR347 Service Manual PM - 87 10.20.2000 FM2732 (2)
  • Page 96: Subscanning Unit (Cleaning The Shock Absorber-Attached Center Guide Plate/Shock Absorber-Attached Guide Plate

    PM - 88 9.11 Subscanning Unit (Cleaning the Shock Absorber-Attached Center Guide Plate/Shock Absorber-Attached Guide Plate) At intervals of one year, or when 60,000 IPs have been processed Clean. Shock absorber [Clean] Vacuum cleaner FRONT Subscanning unit FR7H6907.EPS 009-058-00 CR-IR347 Service Manual PM - 88 10.20.2000 FM2732 (2)
  • Page 97: Subscanning Unit (Cleaning The Glass Guide

    PM - 89 9.12 Subscanning Unit (Cleaning the Glass Guide) At intervals of one year, or when 60,000 IPs have been processed Clean. CAUTIONS • Wear gloves when cleaning the glass guide. • Carefully handle the glass guide to avoid damage to it due to contact with surrounding objects.
  • Page 98: Subscanning Unit (Replacing The Shock Absorber-Attached Center Guide Plate/Shock Absorber-Attached Guide Plate

    PM - 90 9.13 Subscanning Unit (Replacing the Shock Absorber- Attached Center Guide Plate/Shock Absorber-Attached Guide Plate) At intervals of three years, or when 180,000 IPs have been processed Replace. CAUTION Carefully handle the glass guide to avoid damage to it due to contact with surrounding objects. Shock Absorber-Attached Center Guide Plate (1) Remove the upper rubber roller (driving side).
  • Page 99 PM - 91 (2) Remove the grip arm (driving side) Anti-static member Ground cable CHECK [Remove] Grip arm (driving side) [Remove] BR4x8 [Release] Clamp [Loosen] B3x6 [Loosen] B3x6 [Push] Bracket Grip arm [Push] Bracket Grip arm FR7H6130.EPS 009-058-02 009-058-00 CR-IR347 Service Manual PM - 91 08.30.2001 FM3142 10.20.2000 FM2732 (2)
  • Page 100 PM - 92 (3) Release the grip of the upper rubber roller (driven side). Transparent cover FRONT Grip motor (driven side) [Rotate] Allen wrench [Attach] Allen wrench FRONT Upper rubber roller (driven side) Shock absorber-attached center guide plate FR7H6131.EPS 009-058-00 CR-IR347 Service Manual PM - 92 10.20.2000 FM2732 (2)
  • Page 101 PM - 93 (4) Remove the shock absorber-attached center guide plate. [Remove] S3x6 (x6) Shock absorber-attached center guide plate FR7H6132.EPS 009-058-00 CR-IR347 Service Manual PM - 93 10.20.2000 FM2732 (2)
  • Page 102 PM - 94 (5) Replace the shock absorber-attached center guide plate. [Remove] B3x6 (x2) Glass guide [Remove] B3x6 (x2) Leaf spring Narrow Shock absorber-attached center guide plate [Replace] Wide FR7H6927.EPS (6) Reversing the procedures of steps (1) through (5), reinstall the shock absorber-at- tached center guide plate.
  • Page 103 PM - 95 Shock Absorber-Attached Guide Plate Replace the shock absorber-attached guide plate. Shock absorber-attached guide plate [Replace] [Remove] B3x6 (x2) FR7H6909.EPS 009-058-00 CR-IR347 Service Manual PM - 95 10.20.2000 FM2732 (2)
  • Page 104: Subscanning Unit (Installing The Scanning Optics Unit

    PM - 96 9.14 Subscanning Unit (Installing the Scanning Optics Unit) At intervals of one year, or when 60,000 IPs have been processed Install (1) Reversing the procedures described in “9.9 Subscanning Unit (Cleaning the Light- Collecting Mirror),” install the mirror mechanism assembly. NOTE Make sure that the slack of the SZ1 connector cable is not sticking out on the reflection surface side of the light-collecting mirror.
  • Page 105 PM - 97 (2) Install the duct assembly. Duct assembly [Tighten] T4x8 (x2) FR7H6910.EPS (3) Reversing the procedures described in “9.8 Subscanning Unit (Removing the Scanning Optics Unit),” install the scanning optics unit. CHECK Make sure that the scanning optical unit is brought into contact with the positioning pins.
  • Page 106: Subscanning Unit (Installing The Light-Collecting Guide Assemblies

    PM - 98 9.15 Subscanning Unit (Installing the Light-Collecting Guide Assemblies) At intervals of one year, or when 60,000 IPs have been processed Install Light-Collecting Guide Assembly (for the Front) (1) Rotate the lever counterclockwise. [Rotate] Lever FR7H6912.EPS 009-058-00 CR-IR347 Service Manual PM - 98 10.20.2000 FM2732 (2)
  • Page 107 PM - 99 (2) Install the light-collecting guide assembly (for the front) INSTRUCTION When installing the light-collecting guide assembly, properly position it by pressing it against the mirror assembly and positioning pin. Mirror assembly Positioning pin CHECK Light-collecting guide assembly FR7H6932.EPS [Install] Light-collecting guide assembly Light-collecting...
  • Page 108 PM - 100 (3) Rotate the lever clockwise. [Rotate] Lever FR7H6914.EPS 009-058-00 CR-IR347 Service Manual PM - 100 10.20.2000 FM2732 (2)
  • Page 109 PM - 101 (4) Connect the PMT08D board connectors. [Connect] [Connect] [Connect] CN1 PMT08D board FR7H6915.EPS 009-058-00 CR-IR347 Service Manual PM - 101 10.20.2000 FM2732 (2)
  • Page 110 PM - 102 BLANK PAGE 009-058-00 CR-IR347 Service Manual PM - 102 10.20.2000 FM2732 (2)
  • Page 111 PM - 103 Light-Collecting Guide Assembly (for the Back) (1) Rotate the lever clockwise. [Rotate] Lever FR7H6916.EPS 009-058-00 CR-IR347 Service Manual PM - 103 10.20.2000 FM2732 (2)
  • Page 112 PM - 104 (2) Install the light-collecting guide assembly (for back surface). NOTE When installing the light-collecting guide assembly, ensure that its claws are positioned above the guide pins. Light-collecting guide Claw assembly [Attach] Claw [Attach] FR7H6933.EPS 009-058-00 CR-IR347 Service Manual PM - 104 10.20.2000 FM2732 (2)
  • Page 113 PM - 105 Rubber roller (driving-side) Guide [Attach] Light-collecting guide assembly (for back surface) Guide Light-collecting guide assembly Claw Light-collecting guide assembly FR7H3934.EPS 009-058-00 CR-IR347 Service Manual PM - 105 10.20.2000 FM2732 (2)
  • Page 114 PM - 106 (3) Move the light-collecting guide assembly (for the rear) toward the reference plane. (4) Rotate the lever in the direction of the arrow in the figure below to secure the light- collecting guide assembly (for back surface). [Move] Light-collecting guide assem (for the back) (4) #1...
  • Page 115 PM - 107 (5) Attach the bracket. INSTRUCTION When installing the bracket, secure it while pressing its positioning screw against the side plate. CHECK [Attach] BR4x8 [Press] Positioning screw Bracket Side plate Positioning screw FR7H6935.EPS Bracket 009-058-00 CR-IR347 Service Manual PM - 107 10.20.2000 FM2732 (2)
  • Page 116 PM - 108 (6) Connect the PMR08C board connectors. [Connect] CN4 [Connect] CN1 PMR08C board [Connect] CN2 FR7H6917.EPS CHECKS • Check that the PMT08DPMR08C board socket section is parallel to the light- collecting guide connector section. Parallel Not parallel Connector section Socket section PMT08D/PMR08C board GOOD...
  • Page 117: Installing The Side-Positioning Conveyor

    PM - 109 9.16 Installing the Side-Positioning Conveyor At intervals of one year, or when 60,000 IPs have been processed Install NOTES • Press the side-positioning conveyor against the positioning screws. • When tightening the screws, begin by tightening the screws on the right-hand side plate for the subscanning unit.
  • Page 118: Pushing In The Scanner Unit

    PM - 110 9.17 Pushing In the Scanner Unit At intervals of one year, or when 60,000 IPs have been processed Push in. NOTES • When pushing in the scanner unit, exercise care so that the disconnected connectors and cables are not caught. •...
  • Page 119 PM - 111 BLANK PAGE 009-058-00 CR-IR347 Service Manual PM - 111 10.20.2000 FM2732 (2)
  • Page 120: Turning On The High-Voltage Switch

    PM - 112 10. Turning ON the High-Voltage Switch At intervals of one year, or when 60,000 IPs have been processed Turn ON. Turn ON the high-voltage switch (S1) of the SCN08D board. HHS label #2 ON position High-voltage switch SCN08D board FR7H6925.EPS 11.
  • Page 121: Checking The Date And Time

    PM - 113 12. Checking the Date and Time At intervals of one year, or when 60,000 IPs have been processed Check. (1) Power ON the machine. (2) Start the M-Utility. (3) [Date/Time Setting] (4) Check the current date and time. (5) [Return] 009-058-00 CR-IR347 Service Manual...
  • Page 122: Deleting The Error Log

    PM - 114 13. Deleting the Error Log At intervals of half year Delete. (1) Start the M-Utility. (2) Delete the error log. 0. QUIT 1. ERROR LOG UTILITY 2. CONFIGURATION SETTING 3. TEST MODE 4. ELECTRICAL UTILITY 5. SCANNER UTILITY 6.
  • Page 123: Checking The Image

    PM - 115 14. Checking the Image At intervals of half year Check. “2.2 Checking the Image” 009-058-00 CR-IR347 Service Manual PM - 115 10.20.2000 FM2732 (2)