Siemens POLYMOBIL Plus Start-Up Instructions
Hide thumbs Also See for POLYMOBIL Plus:

Quick Links

POLYMOBIL Plus
Start-up Instructions
From Serial No. 12000
Print No.: SPR8-125.034.02.02.02
Replaces: SPR8-125.034.02.01.02
SP
© Siemens AG 2005
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 03.05
Table of Contents
loading

Summary of Contents for Siemens POLYMOBIL Plus

  • Page 1 POLYMOBIL Plus Start-up Instructions From Serial No. 12000 © Siemens AG 2005 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent...
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho- rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
  • Page 3: Table Of Contents

    5 _______Start-up report _________________________________________________ 5 - 1 POLYMOBIL Plus Start-up report ......5 - 1 6 _______Electrical safety / reports ________________________________________ 6 - 1 Protective conductor resistance/ report .
  • Page 4 0 - 4 Contents Page POLYMOBIL Plus SPR8-125.034.02 Page 4 of 4 Siemens AG Rev. 02 03.05 CS PS 24 Medical Solutions...
  • Page 5: Text Emphasis

    Checks marked with this symbol are to be entered into the test certificate at the end of these instructions. This symbol means "Attention, consult the documentation". This symbol indicates components sensitive to Electro-Static Discharge (ESD). Siemens AG SPR8-125.034.02 Page 1 of 10 POLYMOBIL Plus Medical Solutions Rev. 02 03.05 CS PS 24...
  • Page 6: Tools And Measurement Devices Required

    - Set radiation activity as low as possible. (low KV and mAs values) - Release radiation as short as necessary. Checks in which radiation has to be released are identified by the radiation protection warning symbol. POLYMOBIL Plus SPR8-125.034.02 Page 2 of 10 Siemens AG Rev. 02 03.05...
  • Page 7 This can cause life-threatening voltage to be present in the system even after a prolonged period of time. • Refer also to the POLYMOBIL Plus Service instructions RXR8-120.061... "Replacing the capacitor bank". Siemens AG SPR8-125.034.02 Page 3 of 10...
  • Page 8: Important Start-Up Information

    After completion of the start-up procedure, the POLYMOBIL Plus which has been completely tested and adjusted at the factory, is ready to operate. • The POLYMOBIL Plus can be connected to nominal line voltages from 110 V AC through ± 230 V AC ( 10%) and automatically self-adjusts to the existing power line voltage.
  • Page 9: Information On The Protective Conductor Resistance Test

    Repeat measurement In maintenance or repairs perform the protective conductor resistance measurement anew. Document and assess the values determined in the repeat measurement. Siemens AG SPR8-125.034.02 Page 5 of 10 POLYMOBIL Plus Medical Solutions Rev. 02 03.05 CS PS 24...
  • Page 10: Information On The Device Leakage Current Measurement

    However, observe the minimum resolution of the leakage current measuring instrument and additional manufacturer’s data which restrict the use of the measuring instrument. POLYMOBIL Plus SPR8-125.034.02 Page 6 of 10 Siemens AG Rev.
  • Page 11 DIN VDE 0751-1:2001-10, Fig. C5 for protective class 1 = System 2 = Application part (not available) 3 = Measuring arrangement (integrated in the measuring instrument) Siemens AG SPR8-125.034.02 Page 7 of 10 POLYMOBIL Plus Medical Solutions Rev. 02 03.05 CS PS 24...
  • Page 12 Document this in addition. Document the measuring method (differential measurement or direct measurement) and the used measuring equipment (designation and serial number). POLYMOBIL Plus SPR8-125.034.02 Page 8 of 10 Siemens AG Rev.
  • Page 13: Replacing Damaged Or Missing Screws

    Do not use abrasive cleaning agents or organic solvents such as benzine, alcohol or spot remover. • Do not spray water on the unit. • For additional information, refer to the Instructions for use, chapter "Cleaning/ Disinfecting". Siemens AG SPR8-125.034.02 Page 9 of 10 POLYMOBIL Plus Medical Solutions Rev. 02 03.05 CS PS 24...
  • Page 14 1 - 10 Requirements This page intentionally left blank. POLYMOBIL Plus SPR8-125.034.02 Page 10 of 10 Siemens AG Rev. 02 03.05 CS PS 24 Medical Solutions...
  • Page 15 Place the side panel with the integrated ramp (1/Fig. 1) down directly in front of the pallet and engage the retaining tabs. • Remove the plastic cover. Siemens AG SPR8-125.034.02 Page 1 of 2 POLYMOBIL Plus Medical Solutions Rev. 02 03.05 CS PS 24...
  • Page 16 Remove the 2 Allen screws and the bracket (7/Fig. 2, 17 mm spanner opening) from the stand trolley on the right and left side. • Push the POLYMOBIL Plus over the rail (8/Fig. 3) and move it off the pallet via the ramp (9/Fig. 3). POLYMOBIL Plus SPR8-125.034.02...
  • Page 17: Checking The Nominal Line Voltage

    Attaching the warning label • Attach the warning label printed in the appropriate language, item number 18 13 554, to the operating console (1/Fig. 1). NOTE The warning label is located in the POLYMOBIL Plus binder. µGy m MOB00675 Fig. 1 Siemens AG SPR8-125.034.02...
  • Page 18: Checking The Controls And Indicators

    Press the ON key (5/Fig. 5) on the control console. After switching the system on, it executes an initialization routine as follows: Initialization The POLYMOBIL Plus is ready – – – – – – for exposure: the kV display no...
  • Page 19: Checking The Counterbalance System

    Siemens AG SPR8-125.034.02 Page 3 of 6 POLYMOBIL Plus Medical Solutions Rev. 02 03.05 CS PS 24...
  • Page 20: Checking The Controls On The Double-Slot Diaphragm

    Close the double-slot diaphragm. Set 60 kV, 10 mAs. • Release an exposure. The radiation indicator (9/Fig. 5) lights during an exposure; at the same time, an acous- tic signal sounds. POLYMOBIL Plus SPR8-125.034.02 Page 4 of 6 Siemens AG Rev. 02 03.05...
  • Page 21: Checking The Exposure Control

    On the developed film, write the following data with a waterproof felt pen: - SID setting - film size - radiation field size Siemens AG SPR8-125.034.02 Page 5 of 6 POLYMOBIL Plus Medical Solutions Rev. 02 03.05 CS PS 24...
  • Page 22 NOTE In case of deviations greater than 1.6 cm, see Service Instructions Chapter "Checking coincidence of light and radiation fields". • Switch OFF the POLYMOBIL Plus and disconnect the line voltage cable. POLYMOBIL Plus SPR8-125.034.02 Page 6 of 6 Siemens AG Rev.
  • Page 23: Protective Conductor Test

    In the direct measurement of the device leakage current no hous- ing parts of the system may be touched during the measurement. Access to the system for third persons must be prevented. Siemens AG SPR8-125.034.02 Page 1 of 2 POLYMOBIL Plus Medical Solutions Rev. 02 03.05 CS PS 24...
  • Page 24: Reports

    Roll up the power line cable on the cable holder in back of the control console. • Move the support arm (Fig. 1) into the travel position and check the locking mechanism. POLYMOBIL Plus SPR8-125.034.02 Page 2 of 2 Siemens AG Rev.
  • Page 25: Polymobil Plus Start-Up Report

    Start-up report 5 - 1 POLYMOBIL Plus Start-up report The Start-up report follows the same format as the start-up instructions. Item number / Type number / Serial number: ....
  • Page 26 12. Device leakage current measurement (Measured values in separate report ’Device leakage current/ report’) Date: ....Service engineer....POLYMOBIL Plus SPR8-125.034.02 Page 2 of 2 Siemens AG Rev. 02 03.05...
  • Page 27: Protective Conductor Resistance/ Report

    Meas. instrument Type: Meas. instrument Ser-No.: Meas. instrument calibrated to: n.a. Assessment: Date: Name: Signature (*1) Measuring circuit: See Fig. 1, next page Siemens AG SPR8-125.034.02 Page 1 of 6 POLYMOBIL Plus Medical Solutions Rev. 02 03.05 CS PS 24...
  • Page 28 1 = System 2 = Application part (not available) 3 = Measuring arrangement (integrated in the measuring instrument) Remarks: Date Remark Name Signature POLYMOBIL Plus SPR8-125.034.02 Page 2 of 6 Siemens AG Rev. 02 03.05 CS PS 24 Medical Solutions...
  • Page 29: Device Leakage Current/ Report

    Meas. instrument Ser-No.: Meas. instrument calibrated up to: Assessment: Date: Name: Signature: (*1) Measuring circuit: See Fig. 2 to Fig. 3, next page Siemens AG SPR8-125.034.02 Page 3 of 6 POLYMOBIL Plus Medical Solutions Rev. 02 03.05 CS PS 24...
  • Page 30 VDE 0751-1:2001-10, Fig. C6 for protective class 1 = System 2 = Application part (if available) 3 = Measuring arrangement (integrated in the measuring instrument) POLYMOBIL Plus SPR8-125.034.02 Page 4 of 6 Siemens AG Rev. 02 03.05 CS PS 24...
  • Page 31 Electrical safety / reports 6 - 5 Remarks: Date Remark Name Signature Siemens AG SPR8-125.034.02 Page 5 of 6 POLYMOBIL Plus Medical Solutions Rev. 02 03.05 CS PS 24...
  • Page 32 6 - 6 Electrical safety / reports Date Remark Name Signature POLYMOBIL Plus SPR8-125.034.02 Page 6 of 6 Siemens AG Rev. 02 03.05 CS PS 24 Medical Solutions...
  • Page 33 New ’ Final work steps’ section inserted. Chapter 5 Start-up report adapted for protective conductor test and device leakage current measurement. Chapter 6 New ’Electrical safety/ reports’ chapter inserted. Siemens AG SPR8-125.034.02 Page 1 of 2 POLYMOBIL Plus Medical Solutions Rev. 02 03.05 CS PS 24...
  • Page 34 7 - 2 Changes to previous version This page intentionally left blank. POLYMOBIL Plus SPR8-125.034.02 Page 2 of 2 Siemens AG Rev. 02 03.05 CS PS 24 Medical Solutions...