Quick Links

INSTRUCTION MANUAL
PFM-KCU11 ( Transducer )
PFM-M11P ( Sensor )
PFM-M11PT ( High-temperature type )
1.Outline・・・・・・・・・・・・・・・・・  2 
2.Specification・・・・・・・・・・・・・・  3 
3.Shape ・・・・・・・・・・・・・・・・・・ 4 
4.Mounting method ・・・・・・・・・・・・・ 5 
6.Part names and functions・・・・・・・・・ 8 
7.Start up      ・・・・・・・・・・・・・・ 9,10 
8.Setup example of Parameter・・・・・・・・ 11 
9.Maintenance ・・・・・・・・・・・・・・・ 11 
10.Trouble shooting     ・・・・・・・・・・12,13 
※The operator should read this Instruction Manual carefully and 
   handle the device correctly. 
1-8-18 Norimatsu-Higashi,Yahatanishi-ku,Kitakyushu 807-0837 Japan 
Phone No.(8193)691-3731  Fax No.(8193)691-3735 
http://www.matsushima-m-tech.com 
FOR
Dust Monitor
TYPE
Contents 
TPFM-007E 
Rev6:Jul.14,2017
loading

Summary of Contents for Panasonic PFM-KCU11

  • Page 1: Table Of Contents

    TPFM-007E  Rev6:Jul.14,2017 INSTRUCTION MANUAL Dust Monitor TYPE PFM-KCU11 ( Transducer ) PFM-M11P ( Sensor ) PFM-M11PT ( High-temperature type ) Contents  1.Outline・・・・・・・・・・・・・・・・・  2  2.Specification・・・・・・・・・・・・・・  3  3.Shape ・・・・・・・・・・・・・・・・・・ 4  4.Mounting method ・・・・・・・・・・・・・ 5  5.How to connect・・・・・・・・・・・・・・ 6,7  6.Part names and functions・・・・・・・・・ 8  7.Start up      ・・・・・・・・・・・・・・ 9,10  8.Setup example of Parameter・・・・・・・・ 11  9.Maintenance ・・・・・・・・・・・・・・・ 11  10.Trouble shooting     ・・・・・・・・・・12,13  ※The operator should read this Instruction Manual carefully and     handle the device correctly.  1-8-18 Norimatsu-Higashi,Yahatanishi-ku,Kitakyushu 807-0837 Japan  Phone No.(8193)691-3731  Fax No.(8193)691-3735  http://www.matsushima-m-tech.com  E-mail [email protected] ...
  • Page 2 Safety Precautions Be sure to thoroughly read the instruction manual before using the products. Keep the instruction manual in a safe, convenient location for future reference. All or part of the contents described in this manual may be changed without any notice. Due to our constant striving for further improvement of products, parts or products that differ from those described in this manual may be substituted.
  • Page 3: 1.Outline

    1.Outline       Electric charge will generate when a number of corpuscle contained in gas pass through and/or   get in touch with probe of dust monitor equipped with duct, plumbing pipe, etc.        The dust monitor amplifies this "Displacement of the charge" with the sensor, and sends the   amplified slight signal to the output unit.        This electric signal "Displacement of the charge" is both filtered and amplified and then output   instrumentation signal of 4〜20mA in proportion to dust amount and contact signal as well.  Important:View for density of dust monitor per Friction Electrification Mechanism  Output value of Friction Electric Charge Mechanic dust monitor will be variable due to  various  environmental  factors  such  as  「kind  of  particle」 「particle  diameter」 「specific  gravity 「current speed」through measure principle.         Under a regular environment, however, linear characteristic will not of course vary under that  environment.  As to considering density regarding to Dust Monitor of Friction Electric Charge Mechanism,  it requires you to consider influence due to variation of particle characteristics as afore  mentioned and it does not always meet density specified.        Density indicated on specification is for design development of machinery and is publicized  for you customers to use as a guideline in deciding the range.        When you customers decide the range, you please find the range and set it in which alarm level  is easy to set strictly from 0−100% signal and becomes pulse signal. ...
  • Page 4: 2.Specification

    2.Specification        2-1.Power supply              2-1-1.Power pressure         AC 110V or 220V -15%/+10%(specify on demand) 50/60Hz             2-1-2.Power consumed         7VA         2-2.Material to measure           Solid particle in gas        2-3.Quality of particle           All kind of solid particle         2-4.Particle size                 ≧ 0.3μm  3  3       2-5.Density of particle           0.1mg/m 〜 1000mg/m       2-6.Measure principle             Friction electric charge detection method     ...
  • Page 5 3.Shape(rough drawing)  Transducer:PFM-KCU11  Sensor:PFM-M11P  80  (30)  (85)  4-φ5  66  Lead outlet G1/2(※6)  Mtg.Holes  Housing:ADC  Earth terminal:M4 screw(※4)  DIN Rail  M5 Bolt,Nut(※5)  Mtg.Boss R1  Rail socket  60  1 -socket L=43mm  68  Signal cable  Insulator:PTFE   φ18  Probe:SUS316L(※3)  φ10  Transducer:PFM-KCU11  Sensor:PFM-M11PT   (30)   (85)  Lead outlet G1/2(※6)  80  4-φ5  66  Mtg.Holes  Housing:ADC  Earth terminal: M4 screw(※4)  Radiation fin:  Alminium  DIN Rail ...
  • Page 6: 4.Mounting Method

    4.Mounting method         To mount Dust Monitor, 1 -socket is provided with.          Weld the socket onto the pipe. Install the dust monitor so it is horizontal or vertical.  Welding should execute all around the pipe to shut off gas.      4-1.Precaution for mounting           Dust Monitor may mal-function in an environment as listed hereunder.           Mounting Monitor in such environment should be refrained from, please.  (1) Plumbing pipe to be extremely squeezed or close to a place bending.  (2) Mounting position within diameter of duct or 1.5 of vertical dimension.  (3) Near a fan  (4) Fitting plumbing pipe・short pipe・duct are nonconductive.  (5) Measure environment and conditions are out of specification.  (6) In case raising bulk up with short pipe in a place where dusts accumulate between short  pipe and probe.  (7) Place where tip of probe is within 20mm from plumbing pipe.  Important : Wiring of power and signal should be independent, please.          Cap is for transport protection and is not a raincoat.  Attention :・Exclusive cable is transmitting very weak signal. ...
  • Page 7: 5.How To Connect・・・・・・・・・・・・・・ 6,7

    5.How to connect  ②  Attaching Screw  Cover  There are Attaching Screw Covers at upper and lower  portions of a surface of a Transducer. Please remove the covers with a negative driver or the  like. Please keep the covers which you removed. Please attach the cover which you kept after wiring. ②  Attaching Screw  Cover  Upper:Main block part  ①  Lower:Terminal block part  Fig.6. Removing cover Mtg.Hole  Mtg.Hole  A terminal block part is joined to a main block part  with four points of screws and connectors. You can see a screw joining to the terminal block part  under holes for Transducer attaching when you remove  the attaching screw covers. Please loosen these joining screws of the  Joining  terminal block part with a Phillips driver. Screws  You can see the terminal block of the Transducer when  you remove the main block part from the terminal block  part after having loosened these Joining Screws. Mtg.Hole  Mtg.Hole  Fig.7. Joining screws of terminal and main block part When you wire the terminal block of the converter, you  Power terminal  must not make any false wiring. Board(Terminal)  After wiring, please attach the main block part with  the procedure the reverse of how to remove the main  DC4-20m terminal ...
  • Page 8 Transducer  White/Brown/Black/Green/Yellow ( 2 ) ( 3 ) ( 4 ) ( 1 ) ( 5 ) To cable grand  Exclusive line  Shilded cable (1)Power supply  (2)Alarm contact (SPDT)  (3)Upper limit 1 contact (SPDT)  Substrate fixing  screw  (4)Upper limit 2 contact (SPDT)  COM  NO       NC  (5)4-20mA (Analog output signal)  Sensor  ※Please earth the terminal FG.  Line-connecting terminal:        Max. 1.5mm Fig.9. Connection Important : EMC-cable gland is used for Sensor.                      Again, when wiring.  Figure.10. becomes points of wiring.    (Component part) Clamping member ...
  • Page 9: 6.Part Names And Functions

    6.Part names and functions  【for indication】  POWER  LED : Light on as power on  FAULT  LED : CPU on when abnormal  L1 ON  LED : Light on over limit1set  % L2 ON  LED : Light on over limit1set  4 beams indicate LED: detection density  will indicate per 0-100% complying with  setup.  【for button】  → :   use  when  changing  beam  of  number/value.  ↑ :  use when increasing number/value.  :  use when switching parameter set  MOD mode and measure mode.            Pushing after set change will  memorize  parameter and  return  to measure  mode.  : ...
  • Page 10 7.Start up    7-1.Warm up operation  Dust Monitor requires warm up operation for about 1 hour after switch on.          7-2.Measure range set  After finishing warm up operation, you please set most appropriate measure range.   You please set the range at 1 - 4 if the condition in which normal flow speed is regular,  filter is correct and measure environment is appropriate.  In case requiring observation of smaller change, you please set the range at 5〜9 so  that you can observe smaller change.  (In case density is high however, you may get over range.)                          4mA(0%)     20mA(100%)                    4mA(0%)       20mA(100%)  3         Measure range 1  :  0.1  〜  1000mg/m Measure range6 ...
  • Page 11       7-4.Limit set              Limit set can be done 2 different upper limits (alarm) in the range of 4〜20mA(signal   level 0〜100%)              In case signal change gets severe however, chattering might rise often.   In such case you may be recommended to use integral function or delay function at/from   higher position (Customers expertise) respectively.  【Limit 1,2 and action chart of trouble alarm】  Limit output(upper limit alarm) 20mA  Limit set value(ON)  Limit set value(OFF)  4mA  Time  COM-NO  open  COM-NO  close  COM-NO  open  LED Off  LED On  LED Off  Power OFF to open COM-NO  Fig.13. Operation chart(upper limit alarm)  Trouble alarm  Device normal  Trouble  Device normal ...
  • Page 12: 8.Setup Example Of Parameter

    8.Set example of parameter   8-1.Flow of parameter set              Parameter should set according to the flow as shown hereunder.  Measure  window  Parameter mode  Parameter select  Value set  SEL button  → Move  beam  by  MOD button  Push down  button  Press down  Respective  ↑ Each  button  (to    parameter  parameter select  push down change  mode)  value  SEL button push down  Push down MOD button   (after set parameter move  (after parameter set, to  measure  to other parameter item)  window) ...
  • Page 13 10.Trouble shooting       When your device has any problem, please refer to the following particulars and solve it by  yourselves first.   If the device does not recover to normal after those confirmations listed  on the right below have been done, please kindly contact Matsushima.  Table 1. Trouble shooting Contents  Cause thinkable  Items to confirm  Power LED and  ・Power is not supplied.  ・Confirm power line is wired  Other LEDs     to Terminal AC.  are off.    If not supplied, supply power.  ・Power specification is changed from the  ・Confirm  the  voltage  on  name  plate   ordered specification.  of Transducer is the same as the  supplied power.  If it is different, return us  Transducer,  then the  power spec.  will be changed at out site.  ・CPU abnormal.  ・Power OFF, then ON.  Output  ・Warm up operation after setup is not complete.  ・Make  warm  up  operation  for  1  hour Direction is  ...
  • Page 14 Output indication  ・Power is not supplied.  ・Confirm power line is wired  does not change.    to Terminal AC.    If not supplied, supply power.  ・The wiring between Sensor and Transducer is  ・Check wiring  of exclusive  cable.   either off or wrong.  (See Page 6, 5.How to connect.) If  wiring  is  wrong,  rewire  correctly.  ・Material sticking on Probe electrode.  ・Remove Sensor and check if any   sticking on Probe electrode.  If yes, power off and wipe them  off with moist cloth. (See NOTE) ・Probe electrode is in contact with pipe inside  ・Measure inside diameter and   (GND).   Probe electrode  length and  check  if they are not in contact.  If  yes,  change  electrode  length. (See NOTE)  ・Device is in trouble.  ・Consult with Sales Dept. to  check the soundness of Device.  ・CPU abnormal.  ・Power OFF, then ON. ...

This manual is also suitable for:

Pfm-m11pPfm-m11pt