Quick Links

Test Fixture
User's Guide
HP Part No. 16442-90000
Printed in Japan
August 1998
Edition 2
E0897
loading

Summary of Contents for HP 16442A

  • Page 1 Test Fixture User’s Guide HP Part No. 16442-90000 Printed in Japan August 1998 Edition 2 E0897...
  • Page 2: Product Warranty

    Product Warranty Hewlett-Packard warrants Hewlett-Packard hardware, accessories and supplies against defects in materials and workmanship for the period of one year from the warranty start date specified below. If Hewlett-Packard receives notice of such defects during the warranty period, Hewlett-Packard will, at its option, either repair or replace products which prove to be defective.
  • Page 3 Assistance Product maintenance agreements and other customer assistance agreements are available for Hewlett-Packard products. Certification Hewlett-Packard Company certifies that this product met its published specifications at the time of shipment from the factory. Hewlett-Packard further certifies that its calibration measurements are traceable to the National Institute of Standards and Technology (NIST), to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
  • Page 4: Ground The Instrument

    Safety Summary The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual may impair the protection provided by the equipment.
  • Page 5 Safety Symbols The general definitions of safety symbols used on equipment or in manuals are listed below. Instruction manual symbol: the product will be marked with this symbol when it is necessary for the user to refer to the instruction manual in order to protect against damage to the instrument.
  • Page 7: Installation

    Maintenance Specifications Installation The HP 16442A test fixture is designed for testing electronic devices. You mount the suitable socket module on the HP 16442A, which allows you to easily connect various devices (DUTs) to measurement units. HP 16442A has thirteen channels: 6 source/monitor unit (SMU) channels Up to 3 channels can be connected to SMUs by Kelvin connections.
  • Page 8 Inspecting HP 16442A Test Fixture upon Receiving When the HP 16442A arrives at your site, make sure that nothing is missing or damaged. Unpack the carton that contains the HP 16442A, then check the contents against the following figure and table. (For details about furnished and optional accessories, refer to ”Specifications”.)
  • Page 9 Installing HP 16442A Test Fixture You can make the HP 16442A stable by installing stabilizers as shown in the following figure. 1. Put a stabilizer on both sides of test fixture. 2. Screw a flathead screw into hole of each stabilizer.
  • Page 10: Operation

    You place the DUT into the provided socket modules in HP 16442A, then measure the dc characteristics by using the connected measurement units.
  • Page 11 HP 16442A is equipped with an interlock (Intlk) terminal, which prevents the user from getting an electric shock during operation. If more than 40 V will be forced to HP 16442A, be sure to connect Intlk terminal of HP 16442A to Interlock terminal of measurement instrument using an Interlock/LED cable (HP 16493J option 001 for 1.5 m, HP 16493J option 002 for 3.0 m).
  • Page 12 The following figure shows pin assignment of the Intlk terminal. 6 HP 16442A User’s Gude...
  • Page 13 Attaching a Socket Module onto HP 16442A Test Fixture After choosing correct socket module for DUT, install the socket module into HP 16442A. The following figure shows how to attach socket module to HP 16442A. Use the following procedure to attach the socket module: 1.
  • Page 14 Three universal socket modules and a blank Teflon board module for uniquely shaped devices. 1. Non-Kelvin Connections Use connection cables to connect the FORCE terminals on HP 16442A to the FORCE terminals on the socket module. 2. Kelvin Connections Socket Modules for Power Devices The three socket modules for power devices have separate FORCE and SENSE terminals for Kelvin connections.
  • Page 15 Figure 1-3, use miniature banana—pin plug connection cables to connect both FORCE and SENSE terminals of HP 16442A to the FORCE terminal on the socket module as follows: insert one pin plug into the FORCE terminal of the socket module, then insert the other pin plug into the tail hole of the inserted pin plug.
  • Page 16 Note that hydrochloric acid residue deposited by solder may cause electrical problems. Hewlett-Packard recommends that you solder the wiring on the universal board using low hydrochloric acid solder (HP part number: 8090-0433). Do not use the universal socket module for low-current measurements (less than 1 nA) that require low leakage current.
  • Page 17: Maintenance

    Performing test Servicing Cleaning HP 16442A Test Fixture To maintain high performance, HP 16442A must be kept clean. Oil, perspiration, hair, dust, and dirt degrade board insulation, which increases leakage current and decreases measurement accuracy. Hewlett-Packard recommends the following cleaning procedure: Chassis, lid, inner plate, and socket modules: 1.
  • Page 18 This section provides information for trained service personnel to repair the test fixture. ”Replaceable Parts” provides the HP part numbers of replaceable parts. When a part needs to be replaced, order the parts from the nearest HP Sales and Service Office.
  • Page 19 Figure 1-4. Removing the Bottom Cover When wiring the test fixture and socket modules, use low hydrochloric acid solder (HP part number: 8090-0433) to prevent the flux in the solder from spreading unnecessarily. When soldering, make sure the terminals do not make contact with each other. After soldering, make sure that there are no lint bridges so that leakage current is minimal.
  • Page 20 Replaceable Parts Figure 1-5. View of Cable Connections in the Test Fixture 14 HP 16442A User’s Gude...
  • Page 21 Wire 24 White/Green (12) 8150-0458 30cm×8 Wire 24 White/Brown (13) 8150-0462 32cm×1 Wire 24 White/Yellow (14) 8120-0122 16cm×1 Coaxial Cable (15) 8120-0122 16cm×1 Coaxial Cable (16) 8150-0456 5cm×4 Wire 24 White (17) 41422-24099 Soldering Pipe HP 16442A User’s Gude 15...
  • Page 22 Figure 1-6. How to Install Triaxial and BNC Connectors Table 1-2. Parts List for Connectors Reference Designation HP Part Number Quantity Description 1250-1906 Triaxial Connector 1250-0083 BNC Connector 2190-0016 Lock Washer 0360-1190 Terminal Washer 2950-0001 5000-4218 Terminal Washer 1252-1419 6-pin Connector...
  • Page 23 Figure 1-7. Connections between Lid and Body Table 1-3. Parts List for Connections between Lid and Body Reference Designation HP Part Number Quantity Description 16442-24001 Sleeve 0515-1973 Bolt M4 16442-61624 Cable Assy 0515-1550 Screw M3-L 8 P-H 16442-23001 HP 16442A User’s Gude 17...
  • Page 24 Figure 1-8. TO-3 or TO-66 Package Socket Module Figure 1-9. In-Line Package Socket Module Figure 1-10. Axial Lead Package Socket Module 18 HP 16442A User’s Gude...
  • Page 25 Figure 1-11. TO- (4-pin) Package Socket Module Figure 1-12. Dual-In-Line (28-pin) Package Socket Module Figure 1-13. Dual-In-Line (18-pin) Package Socket Module HP 16442A User’s Gude 19...
  • Page 26 Figure 1-14. TO- (8-, 10-, and 12-pin) Package Socket Module Figure 1-15. Universal Socket Module 20 HP 16442A User’s Gude...
  • Page 27: Specifications

    The following specifications data is specified at 23 ±5 C and 50 % relative humidity. Functions HP 16442A lets you test various shaped DUTs. HP 16442A has: 6 SMU channels (when using non-Kelvin connections). 3 SMU channels (when using Kelvin connections).
  • Page 28 16088-60009 axial lead package socket module 16088-60006 TO-package socket module (12-pin) 16442-90000 HP 16442A test fixture User’s Guide (this document) 1 Two f athead screws are furnished. Connection cables (furnished) Miniature Banana – Pin Plug 22 HP 16442A User’s Gude...
  • Page 29 Pin Plug – Pin Plug Miniature Banana – Miniature Clip Miniature Banana – Miniature Banana HP 16442A User’s Gude 23...
  • Page 30 Socket modules (furnished) Blank Teflon Board Dual-In-Line Packages (28-pin) 24 HP 16442A User’s Gude...
  • Page 31 Axial lead package TO-Package (12-pin) HP 16442A User’s Gude 25...
  • Page 32 TO-packages socket module (12-pin) 16088-60006 adds TO-packages socket module (TO-3 or TO-66) 16088-60007 adds in-line packages socket module (3-pin) 16088-60008 adds axial lead package socket module 16088-60009 adds carrying case for socket modules 16442-60101 26 HP 16442A User’s Gude...
  • Page 33 Connection pins (optional) Socket modules (optional) TO-package (TO-3 or TO-66) In-Line package (3-pin) HP 16442A User’s Gude 27...
  • Page 34 Dual-In-Line package (18-pin) TO-package (4-pin) 28 HP 16442A User’s Gude...
  • Page 35 TO-package (8-pin) TO-package (10-pin) HP 16442A User’s Gude 29...
  • Page 36 Universal socket module 30 HP 16442A User’s Gude...
  • Page 37: General Specifications

    5% to 90% relative humidity at 65 Weight Approximately 2.5 kg Dimensions (W × H × D) 260 mm × 140 mm × 260 mm (not including stabilizers) 260 mm × 140 mm × 320 mm (including stabilizers) HP 16442A User’s Gude 31...
  • Page 38 1000 V 20 mΩ 16088-60008 3pin Kelvin 1000 V 20 mΩ 16088-60009 axial Kelvin 1000 V 20 mΩ 16088-60010 Universal module 16442-60011 Universal module 16442-60012 Universal module Connection wire residual resistance : 20 mΩ typical 32 HP 16442A User’s Gude...