GE Amersham ImageQuant 800 User Manual

GE Amersham ImageQuant 800 User Manual

Hide thumbs Also See for Amersham ImageQuant 800:
Amersham
ImageQuant 800
User Manual
Table of Contents
loading

Summary of Contents for GE Amersham ImageQuant 800

  • Page 1 Amersham ™ ImageQuant 800 User Manual...
  • Page 2: Table Of Contents

    The Amersham ImageQuant 800 instrument ................. System Preparation ..................... Amersham ImageQuant 800 control software ..............2.1.1 Introduction ............................2.1.2 Start the Amersham ImageQuant 800 control software ..........2.1.3 System settings ..........................2.1.4 Account management ......................... Amersham ImageQuant CONNECT ....................2.2.1 Introduction ............................
  • Page 3 Table of Contents Image problems ............................ Software problems ..........................Error messages ............................Index ........................Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 4: Introduction

    1 Introduction Introduction About this chapter This chapter contains important user information and a general description of the Amersham ImageQuant 800 and its intended use. In this chapter Section See page 1.1 About this manual 1.2 The Amersham ImageQuant 800 instrument...
  • Page 5: About This Manual

    1.1 About this manual About this manual About this manual The Amersham ImageQuant 800 User Manual provides instructions for setting up the Amersham ImageQuant 800 system, performing image capture, and handling image files using the Amersham ImageQuant 800 control software and Amersham ImageQuant CONNECT software.
  • Page 6 Notes and tips Note: A note is used to indicate information that is important for trouble-free and optimal use of the product. Tip: A tip contains useful information that can improve or optimize your procedures. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 7: The Amersham Imagequant 800 Instrument

    The Amersham ImageQuant 800 is available in several different configurations, which can be upgraded to improve functionality. Depending on the configuration of your Amersham ImageQuant 800 instrument, not all of the functions described in this manual will be available. For more information on Amersham ImageQuant 800 instrument con- figurations and upgrades, see the Amersham ImageQuant 800 Operating Instructions.
  • Page 8 660 nm IRlong Epi 775 nm White Epi 470 to 635 nm White Trans 470 to 635 nm A fully equipped filter turret in the Amersham ImageQuant 800 instrument contains the following emission filters: Emission filter Data 525BP20 Cy3(UV) 605BP40...
  • Page 9: System Preparation

    This chapter provides information on how to start and set up the Amersham ImageQuant 800 control software and Amersham ImageQuant CONNECT software. Section See page 2.1 Amersham ImageQuant 800 control software 2.2 Amersham ImageQuant CONNECT Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 10: Amersham Imagequant 800 Control Software

    2 System Preparation 2.1 Amersham ImageQuant 800 control software Amersham ImageQuant 800 control software About this section This section contains information on how to start the Amersham ImageQuant 800 control software, adjust settings, and manage user accounts. In this chapter Section See page 2.1.1 Introduction...
  • Page 11: Introduction

    The Amersham ImageQuant 800 is operated through the Amersham ImageQuant 800 control software either via the touchscreen panel or an external monitor. For information on the installation process and how to set up the Amersham ImageQuant 800 system, refer to the Amersham ImageQuant 800 Operating Instructions.
  • Page 12: Start The Amersham Imagequant 800 Control Software

    Introduction This section contains information on how to start the Amersham ImageQuant 800 system via Amersham ImageQuant 800 control software. For information on how to set up and install the system, see Amersham ImageQuant 800 Operating Instructions. Start the system Follow the steps below to start the Amersham ImageQuant 800.
  • Page 13 2 System Preparation 2.1 Amersham ImageQuant 800 control software 2.1.2 Start the Amersham ImageQuant 800 control software Step Action Enter your Username and Password. Note: To disable Log in, see the Turn off password protection, on page Tap Login. Wait for the CCD status to turn green, which shows that the CCD has been cooled to the target temperature.
  • Page 14 The self-diagnostic process will normally take a few minutes. Connection to a network may take longer. Turn on the computer and log into Windows. Double click on the Amersham ImageQuant 800 control software icon to connect to the instrument. If Login is turned On, enter your Username and Password.
  • Page 15 Home screen will display. Note: The default time until the instrument enters sleep mode is 30 minutes. To change this setting, refer to the Amersham ImageQuant 800 Operating Instructions. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 16: System Settings

    To adjust the focus for a specific imaging mode, follow the steps below. Step Action Open the imaging mode being used. Note: In this example, Chemiluminescence imaging mode is used. Open the Settings tab. Select Focus from the left side of the screen. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 17 Note: Focus changes will only affect the selected image capture mode. If the user changes to another mode, for example, Fluorescence, the focal distance will still be set to the default setting. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 18 2 System Preparation 2.1 Amersham ImageQuant 800 control software 2.1.3 System settings Screen display If an external monitor is being used to control the Amersham ImageQuant 800 instrument, the screen settings can be changed in the Settings tab. Part Function...
  • Page 19 Select whether the Focus is automatic, or manually set a fixed focal distance. Select whether the Correction is calculated automatically, or manually apply a set correction file. Note: For more information on correction file, see Image correction files, on page Tap Save. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 20 2.1.3 System settings Check filter information The Amersham ImageQuant 800 holds seven filters. These can be either system filters, or custom filters. The installed filters depend on the system configuration. A list of installed filters and their specifications can be seen by tapping the Filter specs button.
  • Page 21 Tap Save. To make the folder the default save location, tap the check box next to the folder name and tap Make default. Result: the default symbol will appear next to the selected folder. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 22: Account Management

    Turn off password protection By default, the Amersham ImageQuant 800 control software is password protected. It is possible to turn off the password requirement for the Amersham ImageQuant 800 by disabling the System login. To do this, use the following procedure.
  • Page 23 Slide the System user login ON/OFF toggle to OFF. Close the System user login ON/OFF window by tapping the cross in the top right corner. Result: the System login settings now shows as OFF. Tap Done. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 24 Tap cogwheel and tap Shut down to shut down the Amersham ImageQuant 800 system. Note: The new setting is available after restarting the Amersham ImageQuant 800 control software. Create a new user New users can be added in the Settings screen. To do this, use the following steps.
  • Page 25 Type the password again in the Confirm password space. Tap Add. Tap Done. Delete a user To delete a user account, use the following steps. Note: Account deletion can only be performed by an administrator with the master password. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 26 Create a new user, on page Change a password To change the password of the current user, follow the procedure below. Step Action In the Settings screen, tap User accounts on the left side of the screen. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 27 Result: the Change password screen shows. Enter the Current password for the current user. Enter a New password for the user. Type the password again in the Confirm password space. Tap Add. Tap Done. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 28 Tap the cogwheel icon in the top left corner and tap Settings. Tap the box next to the desired user to select that account. Tap Reset password. Enter the master password and tap OK. Result: the user's password is reset to 0123456789. Tap Save. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 29: Amersham Imagequant Connect

    This chapter contains information on how to connect to and manage Amersham ImageQuant 800 instruments via the Amersham ImageQuant CONNECT software package. In this chapter Section See page 2.2.1 Introduction 2.2.2 Connect to the instrument Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 30: Introduction

    Introduction Introduction It is is possible to access the Amersham ImageQuant 800 remotely via a computer elsewhere on the network using Amersham ImageQuant CONNECT. This allows users to remotely book experiments using the Scheduler function, and to manage image files.
  • Page 31: Connect To The Instrument

    To connect a new instrument to the Amersham ImageQuant CONNECT software, use the following steps. Step Action Click the options icon in the top right corner and click Settings. Click Add instrument. Result: a new line appears on the Instruments list. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 32 To log in to Amersham ImageQuant CONNECT and access the Amersham ImageQuant 800 from a remote computer, use the following procedure. Step Action Double click on the Amersham ImageQuant CONNECT icon. Carefully read the license agreement and click I Agree. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 33 Enter your User name and Password. Click OK. Result: the Amersham ImageQuant CONNECT Home screen shows. Note: To refresh the status of the connected instruments, click Sync. Select the instrument to connect. Result: the selected instrument Image Library screen shows. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 34 Click the Instrument drop-down menu at the top of the screen. Click the instrument to which you want to connect. Sync instruments To refresh the list of connected Amersham ImageQuant 800 instruments, click the refresh button at the top of the screen. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 35 Click the options icon in the top right corner and click Settings. Click the box next to the instrument to be deleted. Click Delete. Click Save. Note: To undo the deletion, click Cancel. This must be done before the Save button is clicked. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 36: Handle Methods In Imagequant 800 Control Software

    This chapter contains information on how to create and run methods using the Amersham ImageQuant 800 control software. In this chapter Section See page 3.1 Introduction 3.2 General settings 3.3 Advanced settings 3.4 Chemiluminescence image capture 3.5 Colorimetric image capture 3.6 Fluorescence image capture Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 37: Introduction

    Methods can be saved and opened at a later time. If an opened method is changed, it needs to be saved again. Image files contain information on the method and image settings used. See help text in the Amersham ImageQuant 800 control software for more information.
  • Page 38: General Settings

    Depending on the imaging method being used, samples can be placed in the upper (1) or lower (2) tray position. The position of the tray is detected automatically by the Amersham ImageQuant 800 control software and displayed at the bottom of the screen. The display is automatically updated if the tray position is changed.
  • Page 39 Multiple exposures are captured and merged by averaging the results into a single image. This improves the signal-to-noise ratio by removing artefacts and reducing false positive signals. This setting can be used for Chemiluminescence and Fluorescence imaging methods. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 40 Result: the image capture method screen is shown. Tap Start. Help function For information, tips, and help when using the ImageQuant 800 control software, tap the Help button in the top right corner. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 41: Advanced Settings

    Capture area The Amersham ImageQuant 800 has a maximum image capture area of 160 × 220 mm. This capture area is dependent on which tray position is being used and whether the NP lens is being used. The capture area under each setting is detailed in the table below.
  • Page 42 Changing the capture area is only possible in the Lower tray position. In the Upper tray position, it is possible to see a live view of the capture area to po- sition the sample within the default capture area. Tap Select. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 43 This allows for detection of a fainter signal in image analysis, but results in a loss of resolution compared to 2 × 2 binning. When performing image capture on the Amersham ImageQuant 800, it is possible to adjust the binning settings for either greater sensitivity or greater resolution. Increasing the sensitivity results in a decrease in the exposure time.
  • Page 44 Under Advanced settings, tap Select binning. Tap to drag the slider to move the slider towards High resolution, or High sensitivity. Tap Select. Tip: To avoid image saturation, either set a lower binning or reduce the exposure time. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 45: Chemiluminescence Image Capture

    This section contains information on how to create and run chemiluminescence imaging methods. In this section Section See page 3.4.1 Chemiluminescence imaging 3.4.2 Automatic chemiluminescence exposure 3.4.3 Manual chemiluminescence exposure 3.4.4 Time series chemiluminescence exposure 3.4.5 SNOW chemiluminescence exposure Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 46: Chemiluminescence Imaging

    The illustration below shows the main components of the Chemiluminescence screen. Part Function Part Function Save image capture settings as Toggle full dynamic range a method SNOW (noise reduction mode) Image library tab Time series exposure Scheduler tab Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 47 The option to add multiplex imaging is found in the Advanced settings panel when creating a method. To use the fluorescence multiplex imaging, use the following steps. Step Action Next to Fluorescence multiplex, tap Select. Tap Select dye. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 48 Enter the exposure time manually under Exposure time To add a second dye, tap Add dye. To remove a dye, tap the cross on that dye. Tap Select. Note: To cancel changes and return to the previous screen, tap Cancel. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 49 3.4.1 Chemiluminescence imaging Image a plate or petri dish The Amersham ImageQuant 800 is capable of artefact-free chemiluminescence imaging of multi-well plates (e.g., 96-well plates) and petri dishes with the use of the NP lens and NP Tray Guide accessories.
  • Page 50: Automatic Chemiluminescence Exposure

    Under Advanced settings, change the settings for Binning and Capture area as required. See Section 3.3 Advanced settings, on page 41 for more information. Enter the settings for Fluorescence multiplex if required, see Fluorescence multiplex, on page Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 51 Name of the method and tap Save. Tap Pre-capture. Note: To take an automatic exposure without predefining a region of interest, tap Start, without tapping Pre-capture. Tap the button to toggle between Point and Area measurement. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 52 Use an area as the frame of reference, toggle to Area. Tap and drag to select the desired reference area (1). Note: The area can be moved without redrawing it by using the arrow buttons (2). Tap Start. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 53: Manual Chemiluminescence Exposure

    + and - buttons, or the keypad to enter the desired value. Note: The maximum time that can be set is 10 hours. The minimum time that can be set is 0.1 sec. Tap OK to return to the create method window. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 54 Enter the settings for Fluorescence multiplex if required, see Fluorescence multiplex, on page To save the capture settings for future use, tap Save method. In the popup window, enter the Name of the method and tap Save. Tap Start. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 55: Time Series Chemiluminescence Exposure

    Under 1.Select Exposure, tap Time series. Under Number of images, use the + and - buttons to select the desired number of image to be taken. Under Exposure time per image, tap the white box. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 56 Section 3.3 Advanced settings, on page 41 for more information. To save the capture settings for future use, tap Save method. In the popup window, enter the Name of the method and tap Save. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 57 Tap Start. Result: the View image screen will show while all of the images are being taken. The current image being taken will show as Exposing and images to be taken afterwards as Pending. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 58: Snow Chemiluminescence Exposure

    To save the capture settings for future use, tap Save method. In the popup window, enter the Name of the method and tap Save. Tap Start to begin the image pre-capture Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 59 If it is not possible to select such a back- ground area, for example if a post-stained gel exhibits many dye speckles in the gel background, it is recommended to select a smaller area, or an area next to the gel.” Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 60 To manually enter a target improvement ratio during image capture, toggle off Auto stop, tap the box next to Target improvement ratio, and enter a new value. Result: the manual boundary will show in the graph. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 61: Colorimetric Image Capture

    Colorimetric image capture About this section This section contains information on how to create and run colorimetric imaging methods. In this section Section See page 3.5.1 Colorimetric imaging 3.5.2 Gel documentation 3.5.3 Optical Densitometry Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 62: Colorimetric Imaging

    To achieve the best possible accuracy, trans illumination is recommended. This method is only available in certain instrument configurations. For more information on instrument configurations and upgrading the Amersham ImageQuant 800 instrument, see Amersham ImageQuant 800 Operating Instructions. The Colorimetric tab The illustration below shows the main components of the Colorimetric screen.
  • Page 63 3 Handle methods in ImageQuant 800 control software 3.5 Colorimetric image capture 3.5.1 Colorimetric imaging Part Function Part Function OD measurement Show or hide advanced settings Gel documentation Start capture Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 64: Gel Documentation

    Section 3.3 Advanced settings, on page 41 for more information. To save the capture settings for future use, tap Save method. In the popup window, enter the Name of the method and tap Save. Tap Start. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 65: Optical Densitometry

    To save the capture settings for future use, tap Save method. In the popup window, enter the Name of the method and tap Save. Tap Start. Note: Both color and greyscale images for analysis are saved in the image file folder. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 66: Fluorescence Image Capture

    This section contains information on how to create and run fluorescence imaging methods. In this section Section See page 3.6.1 Fluorescence imaging 3.6.2 Automatic fluorescence exposure 3.6.3 Manual fluorescence exposure 3.6.4 SNOW fluorescence exposure Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 67: Fluorescence Imaging

    Part Function Save image capture settings as Select binning a method SNOW (noise reduction mode) Select dyes Manual exposure Select capture area Auto exposure Toggle colorimetric marker Open method Toggle full dynamic range Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 68 To select which dye or dyes are being used in the sample, use the following steps. A maximum of three dyes can be analyzed at one time. Step Action To select the dyes used, under 2.Basic settings, tap Select. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 69 (i.e., parallel to the short sides of the tray). This will minimize any potential gradient effect when measuring bands of a similar size in different lanes. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 70: Automatic Fluorescence Exposure

    If shorter exposure times are required, tap the toggle button next to Full dynamic range to the off position. Note: Turning off Full dynamic range will reduce the range of intensity levels used in the image. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 71 Name of the method and tap Save. Tap Pre-capture. Note: To take an automatic exposure without predefining a region of interest, tap Start, without tapping Pre-capture. Tap the button to toggle between Point and Area measurement. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 72 Use an area as the frame of reference, toggle to Area. Tap and drag to select the desired reference area (1). Note: The area can be moved without redrawing it by using the arrow buttons (2). Tap Start. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 73: Manual Fluorescence Exposure

    Under 1.Select Exposure, tap Manual. To select the dyes used, tap Select. In the Fluorescence dyes window, tap the drop-down arrow and select the dye being used. Under Exposure time, tap the white box. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 74 Enter an Exposure time for the new dye. If needed, add a third dye. If an overlay image of the sample and marker is required, tap the toggle button next to Colorimetric marker. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 75 Section 3.3 Advanced settings, on page 41 for more information. To save the capture settings for future use, tap Save method. In the popup window, enter the Name of the method and tap Save. Tap Start. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 76: Snow Fluorescence Exposure

    To save the capture settings for future use, tap Save method. In the popup window, enter the Name of the method and tap Save. Tap Start to begin the image pre-capture Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 77 If it is not possible to select such a back- ground area, for example if a post-stained gel exhibits many dye speckles in the gel background, it is recommended to select a smaller area, or an area next to the gel.” Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 78 To manually enter a target improvement ratio during image capture, toggle off Auto stop, tap the box next to Target improvement ratio, and enter a new value. Result: the manual boundary will show in the graph. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 79: Manage Images In Imagequant 800 Control Software

    This chapter contains information on how to view and manage images in ImageQuant 800 control software. In this chapter Section See page 4.1 View the image after capture 4.2 The Image library screen Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 80: View The Image After Capture

    Zoom in and out of image Editing tools Image overlays Depending on the imaging type used, the image overlays tabs will be different. Imaging method Display options Chemiluminescence with colorimetric Toggle Marker on and off marker Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 81 Crop. Set image crops if multiple samples have been used. Note: Using Half crop and Quarter crop will separate the image into 2 or 4 image files when saved. Rotate image. Exposure time for each image type. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 82 To measure the signal intensity, use the following procedure. Note: Signal intensity measurements are not reliable in saturated images. Saturated parts of an image are highlighted in purple. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 83 In the Image view screen, tap Intensity. Tap and drag to select the area of interest. Read the intensity signals in the right table. Note: If multiple images are being viewed, the intensities of all dyes will be displayed. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 84 4.1 View the image after capture Step Action Use the arrow buttons (1) to move the area selection without resizing. Time-series images When taking a Time series exposure, different options are available on the View image screen. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 85 Select which individual image to Display selected image on full save screen Toggle colorimetric Marker on Zoom in all images and off Selected image for full screen (in- Zoom out all images dicated by blue border) Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 86: The Image Library Screen

    Toggle between list and grid Selected image information (thumbnail) view Image thumbnails from select- Open selected image ed folder Select image folder Copy/move image Select imaging method Delete image Sort images Select all images Search images Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 87 If multiple images have been selected, the option to Open is not available. The option to Copy/Move this images is available for multiple images. Move image Image files can be copied and moved in Amersham ImageQuant 800 control software. Step Action In the Image library screen, tap to select the image or images to be moved.
  • Page 88 Tap Move to move the image to the new location and remove it from the current location. Print image Amersham ImageQuant 800 control software can print images directly to a printer connected on the network. To do this, use the following steps. Step Action In the View image screen, tap Print.
  • Page 89: Imagequant Connect Operation

    This chapter describes the ImageQuant CONNECT software and provides information on how to use the different functions. In this chapter Section See page 5.1 Introduction 5.2 Use the scheduler tool 5.3 Manage images Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 90: Introduction

    The Amersham ImageQuant CONNECT software enables the user to connect remotely to any Amersham ImageQuant 800 instrument on the network. Using this package, it is possible to monitor runs, book a time slot for using the machine, as well as view, and download image files from the instrument.
  • Page 91: Use The Scheduler Tool

    5 ImageQuant CONNECT operation 5.2 Use the scheduler tool Use the scheduler tool Introduction The ImageQuant CONNECT scheduler tool can be used to remotely book or edit a time slot on the Amersham ImageQuant 800 instrument remotely. The Scheduler screen Part Function Part...
  • Page 92 By default, time slots are set to 15 minutes. Click Save. Edit a booking To edit a booking, use the following steps. Step Action In the Scheduler screen, select the booking to be edited. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 93 5 ImageQuant CONNECT operation 5.2 Use the scheduler tool Step Action Click Edit booking. In the Edit booking window, enter the new information. Click Save. To delete a booking, select the booking slot and click Delete. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 94: Manage Images

    Image information Select folder Open image Filter by method Copy image Sort images by date or alphabeti- Select all images in the folder cally Search images Toggle between list and thumb- nail image view Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 95 Click on the desired image to display the image information in the right panel. Note: If more than one image is selected, no information will be displayed. Click Open. Result: the image is opened in the image editing screen. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 96 The Intensity function can be used to measure the signal intensity of either a single point or a defined area of the image. to use this function, follow the instructions in Measure the signal intensity, on page Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 97: Troubleshooting

    If the suggested actions in this guide do not solve the problem, or if the problem is not covered by this guide, contact your GE Healthcare representative for advice. In this chapter...
  • Page 98: Instrument Problems

    Action Make a note of the error code and error message on the monitor. Shut down the Amersham ImageQuant 800 control software using the usual procedure, see the Amersham ImageQuant 800 Operating Instructions. If the shut down procedure does not work, use the task manager to force the Amersham ImageQuant 800 control software to exit.
  • Page 99 The instrument cannot be The ambient air tempera- Lower the room temperature to 28°C or cooled down properly ture is too high lower Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 100 The instrument is not con- Unplug the instrument and check that all appear to respond to soft- nected to the external connections between the instrument and ware control computer external computer are secure Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 101: Image Problems

    White Epi source relative to the gel gel so that the lanes of the gel are parallel imaging in the upper tray lanes to the long side of the tray position Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 102 If there is no such flat background stopped area, select a background area next to the membrane or gel Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 103: Software Problems

    Check the that instrument is connected to CONNECT is not showing the network. See Amersham ImageQuant on host instrument 800 Operating Instructions Unable to log in Forgotten password Reset the user's password, see Reset a password, on page 28 Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 104: Error Messages

    Tap OK. Make sure that system doors are interrupted firmly shut and take the image again. IP address al- Tap OK. change the IP address of the new ready exists instrument to a unique identifier Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 105 The images captured in this session have not been saved. To return to the software and save the image, tap No. To exit the software without saving and new images or methods, tap Yes. Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 106: Index

    41 Tray guide, 49 copy, 87 Typographical conventions, 6 correction files, 20 editing, 81 histogram, 82 User account Image view screen, 80 create user, 24 Measure signal intensity, 82 delete user, 25 Amersham ImageQuant 800 User Manual 29383813 AA...
  • Page 107 All goods and services are sold subject to the terms and conditions of sale of Björkgatan 30 the company within GE Healthcare which supplies them. A copy of these terms 751 84 Uppsala and conditions is available on request. Contact your local GE Healthcare repre- sentative for the most current information.

Table of Contents