Quick Links

A B B M E A S U R E M E N T & A N A LY T I C S | U S E R G U I D E | I M/ D O S P M P R E V. B
AAO100
Dosing pump
Introduction
AAO100
Dosing pump
The AAO100 dosing pump is used to introduce a
cleaning solution to the jetwash of an AP200 pH
system at preset intervals. The timing, duration and
amount of cleaning is controlled by an AX400 pH
analyzer.
Measurement made easy
Table of Contents
loading

Summary of Contents for ABB AAO100

  • Page 1 Introduction — AAO100 Dosing pump The AAO100 dosing pump is used to introduce a cleaning solution to the jetwash of an AP200 pH system at preset intervals. The timing, duration and amount of cleaning is controlled by an AX400 pH...
  • Page 2 For more information Further publications are available for free download from: www.abb.com/analytical or by scanning this code: Search for or click on AP200 Rugged pH/Redox (ORP) sensor systems DS/AP200-EN Data Sheet AP200 series Rugged pH/Redox (ORP) sensor systems IM/AP200 User Guide...
  • Page 3: Table Of Contents

    Dosing Pump AAO100 Contents 1 Introduction ..............2 2 Installation ...............2 Fitting the Pump to the Reservoir ......2 Electrical Connections ..........3 2.2.1 Dosing Pump Mains Connections ....3 2.2.2 Analyzer Connections ........3 Dosing Pump Settings ..........3 Pump Tubing Connections ........4 2.4.1 AP200 Connection ........4 2.4.2 Connecting the Tubing to the Pump .....5...
  • Page 4: Introduction

    1 Introduction 1 Introduction 2 Installation The AAO100 dosing pump is used to introduce a cleaning Note. Installation must be carried out in accordance with solution to the jetwash of an AP200 pH system at preset local water company and council by-laws.
  • Page 5: Electrical Connections

    Dosing Pump AAO100 2 Installation 2.2 Electrical Connections 2.2.1 Dosing Pump Mains Connections 2.2.2 Analyzer Connections The power supply range is 100 to 230 V AC. Referring to Fig. 2.3: 1. Plug the controller cable into the dosing pump Caution. Before making any electrical connections, ensure 2.
  • Page 6: Pump Tubing Connections

    Dosing Pump AAO100 2 Installation 2.4 Pump Tubing Connections 2.4.1 AP200 Connection Note. Wash tubing is not supplied. Use 6mm OD semi-rigid polyethylene tube (part no. 0212035). Referring to Fig. 2.4: 1. Remove jetwash port sealing nut and retain for further use.
  • Page 7: Connecting The Tubing To The Pump

    Dosing Pump AAO100 2 Installation 2.4.2 Connecting the Tubing to the Pump Referring to Fig. 2.5: 1. Cut the end of the tubing straight across. 2. Slide union nut and circlip over the tubing. 3. Push the end of the tubing over nozzle 4.
  • Page 8: Analyzer Settings

    Dosing Pump AAO100 3 Analyzer Settings 3 Analyzer Settings Access the Wash Cycle Configuration menu – see Section 5 of the AX400 pH analyzer’s User Guide (IM/AX4PH). Wash Mode Set to Cont. (continuous) for use with the dosing pump. Cont.
  • Page 9: Cleaning Solutions

    Dosing Pump AAO100 4 Cleaning Solutions 4 Cleaning Solutions 5 Dimensions The AP200 spray jet tube is constructed in 316 stainless steel. Dimensions in mm (in.) Possible contaminants together with recommended cleaning solutions are shown in Table 4.1. Contaminant Cleaning Solution...
  • Page 10: Notes

    Dosing Pump AAO100 Notes Notes IM/DOSPMP Rev. B...
  • Page 11 Sales Service...
  • Page 12 We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document.

Table of Contents