Firmware Version FW 2.1
Hardware Version HW 2.1
Service Manual
2001589-094
Revision D
Table of Contents
loading

Summary of Contents for GE TONOPORT V

  • Page 1 GE Healthcare TONOPORT V Ambulatory Blood Pressure System Firmware Version FW 2.1 Hardware Version HW 2.1 Service Manual 2001589-094 Revision D...
  • Page 2 NOTE: The information in this manual only applies to the product TONOPORT V, Firmware Version FW2.1. Due to continuing product innovation, specifications in this manual are subject to change without notice. © 2009-2016 General Electric Company. All rights reserved TONOPORT V Revision D...
  • Page 3: Table Of Contents

    Firmware Loading Procedure ..... . 6-1 Settings for TONOPORT V Update Program V3.0 for Windows ..6-1 Loading Procedure .
  • Page 4 Exchange Units ..........12-3 TONOPORT V...
  • Page 5: Introduction

    Initial release 2012-01 Chapter 4: TONOPORT V, V2.0 corrected to TONOPORT V, V2.1 Chapter 5: 2 pictures for TONOPORT V, version HW2.1 Rev. B added and difference between Rev. A and Rev. B explained Chapter 6: Required Tools: Windows 7 added Chapter 8.2: 1 picture for TONOPORT V, V2.1 Rev.
  • Page 6: Safety Information

    Introduction: Safety Information Safety Information 1.2.1 Responsibility GE Healthcare is responsible for the effects of safety, reliability, and performance only if: Assembly operations, extensions, readjustments, modifications, or repairs are  carried out by persons authorized by GE Healthcare. The electrical installation of the relevant room complies with the requirements ...
  • Page 7: Service Information

    Any unauthorized attempt to repair equipment under warranty voids that  warranty. It is the user’s responsibility to report the need for service to GE Healthcare or  to one of their authorized agents. Failure on the part of the responsible individual, hospital, or institution using ...
  • Page 8 Introduction: Service Information For your notes TONOPORT V Revision D 2001589-094...
  • Page 9: Block Diagram

    Block Diagram Tubing system Revision D TONOPORT V 2-1 2001589-094...
  • Page 10 For your notes TONOPORT V Revision D 2001589-094...
  • Page 11: Functional Description

    OFF Info Start mmHg unit is mmHg Info unit is kPa Measurement limits - Max. cuff inflation time = 130 seconds - Max. measurement time after start of measurement = 180 seconds Revision D TONOPORT V 3-1 2001589-094...
  • Page 12 For your notes TONOPORT V Revision D 2001589-094...
  • Page 13: Pin Connections

    9600 Baud / 8 Bit / 1 Stopbit none Parity Cable USB TONOPORT V - PC (2001589-040) - not shown This cable connects the USB port of TONOPORT V, V2.1 (USB B-mini, male, 5-pin) to the USB port (USB A, male) of the PC. Revision D TONOPORT V 4-1...
  • Page 14 For your notes TONOPORT V Revision D 2001589-094...
  • Page 15: Pictures

    Pictures Pictures of TONOPORT V for version HW2.1, Rev. A   (GE part number: 2001589-092, Rev. A) Revision D TONOPORT V 5-1 2001589-094...
  • Page 16  (GE part number: 2001589-092 Rev. B) NOTE TONOPORT V, HW2.1, Rev. B uses a main board with another type of deflation valve than Rev. A. Difference between Rev. A and Rev. B: The main boards of Rev. A and Rev. B differ in the deflation valves used. The defla- tion valves are tightly connected with the main boards in Rev.
  • Page 17: Firmware Loading Procedure

    TONOPORT V version and loads the corresponding firmware for TONOPORT V, HW2.1: firmware FW2.1 or higher version. NOTE If the update program detects an older TONOPORT V version, it loads for TONOPORT V, HW2.0: firmware FW2.0 and for TONOPORT V, HW1.4 and earlier: firmware version FW1.4...
  • Page 18: Loading Procedure

    Firmware Loading Procedure: Loading Procedure Loading Procedure Connect inactive TONOPORT V to USB port or RS232 port via the USB or  RS232 cable. Turn on TONOPORT V.  NOTE After TONOPORT V has been switched on, updating can only be carried out within the first 10 minutes.
  • Page 19 Firmware Loading Procedure: Loading Procedure Check the correct firmware version on TONOPORT V:  Step Action Diplay Switch on TONOPORT HH:MM (time) Press “INFO” 5 times Press “START” 21 : 00 (21 = Version FW 2.1 00 = no direct printout) Press "START"...
  • Page 20 Firmware Loading Procedure: Loading Procedure For your notes TONOPORT V Revision D 2001589-094...
  • Page 21: Troubleshooting Guide / Error Codes

    E 14 Diastolic reading below 40 mmHg. Code appears when the cuff pressure has dropped to 40 mmHg and no diastolic pressure could be identified (TONOPORT V does not measure diastolic pressures below 40 mmHg). E 15 Motion artifact during diastole detection.
  • Page 22 At the same time the device switches to a deflation rate of 4 mmHg/s. When it detects more than 13 oscillations later on, the rate is changed to 6 mmHg/s. TONOPORT V Revision D 2001589-094...
  • Page 23: Adjustment Instructions

    Fixed volume of 0.5 L as an air container  T-adapter  Step-by-Step Instructions Only the pressure amplifier needs to be calibrated. This is done after switching  on the TONOPORT V and waiting until the time is displayed on the LCD. Revision D TONOPORT V 8-1 2001589-094...
  • Page 24 Use a screw driver to turn the trimming potentiometer R53 until the LCD  indicates a value of 50 ±20. Connect a reference pressure gauge, a rubber bulb and air container (0.5 L) to  the TONOPORT pressure connector and press "START". TONOPORT V Revision D 2001589-094...
  • Page 25 Using a screwdriver, turn the trimming potentiometer R40 until the LCD  indicates a value of 100.  Also check the values indicated on the LCD at the following pressures   (max. permitted deviation for each measurement: ±2 mm Hg): 50 mmHg  Revision D TONOPORT V 8-3 2001589-094...
  • Page 26 Adjustment Instructions: Step-by-Step Instructions 150 mmHg 200 mmHg Switch the TONOPORT off.  TONOPORT V Revision D 2001589-094...
  • Page 27: Main Board - View Of The Components

    Adjustment Instructions: Main Board - View of the Components Main Board - View of the Components Main Board, Rev. A (GE part number 2002866-005, Rev. A)   for TONOPORT V, HW2.1, Rev. A TP9 TP10 Revision D TONOPORT V 8-5...
  • Page 28 Adjustment Instructions: Main Board - View of the Components Main Board, Rev. B (GE part number 2002866-005, Rev. B)   for TONOPORT V, HW2.1, Rev. B TP9 TP10 TONOPORT V Revision D 2001589-094...
  • Page 29: Technical Specifications

    Transport and storage  Temperature -20...+ 70 °C Relative humidity 10...90 %, no condensation Atmospheric pressure 500 hPa (altitude +4500 m relative to sea level) to 1060 hPa  (-400 m relative to sea level) Revision D TONOPORT V 9-1 2001589-094...
  • Page 30 Technical Specifications: Dimensions  Height 27 mm Width 80 mm Depth 100 mm Weight  199 g with batteries TONOPORT V Revision D 2001589-094...
  • Page 31: Technical Inspections

     Measure the time between start running the pump and stopping the pump.  The time must be 130 sec max. If the time is over this value, the TONOPORT V  must be sent in for inspection. Revision D TONOPORT V 10-1...
  • Page 32: Technical Inspections Of The Measuring System

    Press START: display indicates internal value, which must be 50 ±20. If the  value lies outside this range, the TONOPORT V must be sent in for inspection. Press START again: display indicates "0" (the device is now in the calibration ...
  • Page 33: Checkout Procedure

    Checkout Procedure After repair or maintenance of the TONOPORT V, a checkout procedure is recom- mended. This checkout procedure consists of the adjustment of the device, see Chapter 8, “Adjustment Instructions”  the technical inspection of the measuring system, see Chapter 10, “Technical ...
  • Page 34 Overpressure test passed? o Yes o No Time limit test passed? o Yes o No Visual inspection passed? o Yes o No Checkout OK? o Yes o No Date:......Name and signature of test person: ................11-2 TONOPORT V Revision D 2001589-094...
  • Page 35: Parts Lists

    TONOPORT V Operator Manual Serbian 2001589-111 TONOPORT V Operator Manual Estonian 2001589-112 TONOPORT V Operator Manual Rumanian 2001589-113 TONOPORT V Operator Manual English US 2001589-127 TONOPORT V Operator Manual Indonesian 2001589-128 TONOPORT V Operator Manual Croatian Revision D TONOPORT V 12-1...
  • Page 36: Accessories

    TONOPORT V Operator Manual Latvian 2001589-131 TONOPORT V Operator Manual Lithuanian 12.2 Accessories Part No. Description 2001589-011 Cable RS232 TONOPORT V ———> PC (with CardioSoft V4.14 or higher) 2001589-040 Cable USB TONOPORT V ———> PC (approx. 1.5 m) 2001589-015 Carrying Bag 2001589-016...
  • Page 37: Exchange Units

    PCB HW2.1 (with Motorola Pressure Sensor and LCD) (see pictures in chapter 5 and chapter 8.2 of this Service Manual) 2032506-001 Speaker TONOPORT V 12.4 Exchange Units Part No. Description 2003386-003 Exchange Recorder ABP TONOPORT V, V2.1 Revision D TONOPORT V 12-3 2001589-094...
  • Page 38 Parts Lists: Exchange Units For your notes 12-4 TONOPORT V Revision D 2001589-094...
  • Page 40 Fax: +49 30 2138542 Asia Headquarters GE Medical Systems GE Medical Systems Information Technologies, GmbH Information Technologies, Inc. GE (China) Co., Ltd. No1 Huatuo Road, Munzinger Straße 5 8200 West Tower Avenue Zhangjiang Hi-Tech Park Pudong, Milwaukee, WI 53223 USA 79111 Freiburg Shanghai, P.R.China 201203...